Valleylab Force Argon II-20Recalls & reported problems
Covidien LP · Argon Gas Delivery Unit II,120 V · Class 2
All models of Valleylab (97)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10884524002552), received between Apr 26, 2019 and Apr 26, 2019 — classified as 1 injury.
The most frequently reported specific device problem is “Energy Output Problem” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem1
- Energy Output Problem1
Patient problems
- Blood Loss1
- No Code Available1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
InjuryApr 26, 2019Energy Output Problem; Adverse Event Without Identified Device or Use Problem; Energy Output Problem; Adverse Event Without Identified Device or Use Problem
ACCORDING TO THE REPORTER, DURING THE LIVER MASS BIOPSY, THE UNIT'S HAND PIECES WOULD NOT FIRE PROPERLY - IT WAS VERY WEAK AND NEVER ARCED PROPERLY. HAD IT TURNED ALL THE WAY UP AND BOTH TANKS WERE FULL. HAD TO OPEN THE PATIENT (CONVERTED TO LAPAROTOMY) TO CONTROL BLEEDING. THREE (3) LAPAROSCOPIC TIPS AND TWO (2) HAND PIECES WERE OPENED ¿ COULD NOT GET ANY OF THEM TO WORK PROPERLY. ANOTHER GENERATOR WAS USED WITH A REGULAR HAND PENCIL CAUTERY BECAUSE HAD TO OPEN THE PATIENT. USED CAUTERY AND ENDO TIES TO COMPLETE THE SURGERY. THERE WAS A 6000 ML OF BLEEDING AND HAD TO TRANSFUSE PLASMA X5, PLATELETS X 2, RBC X 15. PATIENT WAS DISCHARGED.
Patient problems: Blood Loss; No Code Available
MAUDE report 8556110 ↗
Other ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES devices
- Vasoview Hemopro 2 VH-4000Maquet Cardiovascular LLC · Electrosurgical, Cutting & Coagulation & Accessories6 recalls1 death3,232 reports
- LigaSure LF1937Covidien LP · Electrosurgical, cutting & coagulation & accessories3 deaths2,090 reports
- Olympus WB91051WOLYMPUS Winter & Ibe GmbH · Electrosurgical, Cutting & Coagulation & Accessories1 death1,951 reports
- Valleylab FT10 VLFT10GENCovidien LP · Electrosurgical, cutting & coagulation & accessories1 recall5 deaths1,062 reports
- LigaSure LF1837Covidien LP · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES1 recall967 reports
- LigaSure LF2019Covidien LP · Electrosurgical, cutting & coagulation & accessories948 reports
- VAPR 225028Depuy Mitek, LLC · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES895 reports
- LigaSure LF1923Covidien LP · Electrosurgical, cutting & coagulation & accessories855 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page