Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 10884524002781), received between Aug 21, 2017 and Jun 19, 2025 — classified as 1 injury, 3 malfunctions.

The most frequently reported specific device problem is “Temperature Problem” (2 reports).

1 other recall cover devices cleared under the same FDA submission (K973297); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem2
  • Temperature Problem2
  • Break1
  • Overheating of Device1
  • Smoking1

Patient problems

  • No Patient Involvement2
  • Biliary Leak1
  • Fistula1
  • Liver Damage/Dysfunction1
  • No Code Available1
  • No Consequences Or Impact To Patient1
  • Pain1
  • Pneumothorax1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • InjuryJun 19, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    ACCORDING TO LITERATURE SOURCE OF RETROSPECTIVE STUDY PERFORMED BETWEEN FEBRUARY 2013 AND MAY 2021, THE LONG-TERM OUTCOMES OF RFA USING COOL-TIP RFA FOR PERIBILIARY AND NON-PERIBILIARY HCC USING PROPENSITY SCORE MATCHING (PSM), 282 PATIENTS WERE INCLUDED IN THIS RETROSPECTIVE STUDY. COMPLICATIONS INCLUDED INTRAHEPATIC BILE DUCT DILATATION (18), BILOMA FORMATION (1), BILIARY FISTULA (1), LIVER DECOMPENSATION (55), THROMBUS IN PERITUMORAL VESSEL (10), PAIN REQUIRING TREATMENT (19), INFECTION (14), HYDROPNEUMOTHORAX REQUIRING DRAINAGE (3). BILOMA FORMATION AND BILIARY FISTULA WERE TREATED WITH PERCUTANEOUS BILIARY DRAINAGE.

    Patient problems: Biliary Leak; Fistula; Liver Damage/Dysfunction; Pain; Pneumothorax; Thrombosis/Thrombus; Unspecified Infection

    MAUDE report 22293714 ↗
  • MalfunctionOct 2, 2019Break; Overheating of Device; Smoking; Adverse Event Without Identified Device or Use Problem; Break; Overheating of Device; Smoking; Adverse Event Without Identified Device or Use Problem

    ACCORDING TO THE REPORTER, DURING PROCEDURE, AFTER THE PHYSICIAN PLACE THE CLUSTER ELECTRODE IN THE PATIENT BODY AND ONCE THE MACHINE WAS SWITCHED ON, THE MACHINE STARTED TO HEAT UP SUDDENLY AND THERE WERE BURNING SMOKE SENSATION OCCURRED. THE PROCEDURE WAS STOPPED AND DID NOT PROCEED ABLATING USING RADIO FREQUENCY ABLATION. SURGEON SWITCHED TO EMPRINT ANTENNA TO COMPLETE THE PROCEDURE WITH NO POST-PROCEDURAL ADVERSE EVENT HAPPENED TO THE PATIENT. WHOLE SET OF MACHINES AND CLUSTER ELECTRODE WERE RETURN BACK TO THE OFFICE. SERVICE ENGINEER AT THE FACILITY FOUND OUT THE PROBLEM THAT CAUSED THE SMOKE FROM THE UNIT WAS A FAULTY SYSTEM MAIN CIRCUIT BOARD.

    Patient problems: No Code Available; No Consequences Or Impact To Patient

    MAUDE report 9145582 ↗
  • MalfunctionApr 15, 2019Temperature Problem; Temperature Problem

    ACCORDING TO THE REPORTER, PRIOR TO USE, THE DEVICE HAD A TEMPERATURE TEST FAILURE. THERE WAS NO PATIENT INVOLVED.

    Patient problems: No Patient Involvement

    MAUDE report 8515246 ↗
  • MalfunctionAug 21, 2017Temperature Problem; Temperature Problem

    THE CUSTOMER REPORTED THAT PRIOR TO USE, THE TEMPERATURE TEST FAILED. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Patient Involvement

    MAUDE report 6807528 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K973297. One submission often covers many models, so check the product and lot details in each record.

  • Z-2301-2012 · Aug 1, 2012 · Terminated
    DGPHP RFA High-Power Single Use Grounding Pads and Cool-Tip RFA Electrode Kits (containing DGPHP RFA High-Power single use grounding pads) The product is used during ablation surg

Other ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES devices

All ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES devices →