ForceTriad FT6009Recalls & reported problems
Covidien LP · Bipolar Resection Footswitch · Class 2
All models of ForceTriad (7)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 10884524002903), received between Jun 2, 2023 and Apr 23, 2025 — classified as 1 injury, 5 malfunctions.
The most frequently reported specific device problem is “Break” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem4
- Break1
- Energy Output Problem1
- Inaccurate Delivery1
- Intermittent Energy Output1
- No Apparent Adverse Event1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- Ascites1
- Burn(s)1
- Hemorrhage/Blood Loss/Bleeding1
- Pleural Effusion1
- Superficial (First Degree) Burn1
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
InjuryApr 23, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
ACCORDING TO THE LITERATURE, A RETROSPECTIVE STUDY ANALYZED THE OUTCOMES OF PATIENTS WHO UNDERWENT ROBOTIC DISTAL GASTRECTOMY WITH A SINGLE-PORT ROBOTIC PLATFORM FOR THE TREATMENT OF GASTRIC CANCER BETWEEN MARCH 2023 AND APRIL 2024. IT WAS NOTED THAT A LIGASURE DEVICE OR COMPETITOR PRODUCT WAS UTILIZED INTRA-OPERATIVELY TO DIVIDE THICK TISSUE AND VESSELS. THERE WERE 20 PATIENTS INCLUDED IN THE STUDY AND COMPLICATIONS INCLUDED PLEURAL EFFUSION AND CHYLE ASCITES. NO RE-OPERATION OR MORTALITY WAS REPORTED; HOWEVER, INTERVENTIONS AND PATIENT OUTCOMES ARE UNKNOWN.
Patient problems: Ascites; Pleural Effusion
MAUDE report 21893010 ↗MalfunctionApr 17, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT AFTER THE PROCEDURE, PATIENT HAD A FIRST-DEGREE BURN ON THE BUTTOCKS WHICH ALSO SHOWED IN THE PHOTO.
Patient problems: Burn(s)
MAUDE report 21852930 ↗MalfunctionJun 3, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
ACCORDING TO THE REPORTER, POST OPERATIVELY, THE PATIENT HAD A FIRST-DEGREE BURN AROUND THE RETURN ELECTRODE MONITORING PLATE (REM) ON THE RIGHT SIDE OF THE PATIENT AND IT WAS TREATED.
Patient problems: Superficial (First Degree) Burn
MAUDE report 19445161 ↗MalfunctionJan 29, 2024Energy Output Problem; No Apparent Adverse Event; Energy Output Problem; No Apparent Adverse Event
ACCORDING TO THE REPORTER, DURING PROCEDURE, ACTIVATION AND END TONE WERE HEARD BUT SEALING WAS INADEQUATE PARTIAL. BLEEDING WAS OBSERVED DIRECTLY FROM THE LIGASURE SEAL.
Patient problems: Hemorrhage/Blood Loss/Bleeding
MAUDE report 18602682 ↗MalfunctionDec 14, 2023Break; Inaccurate Delivery; Break; Inaccurate Delivery
ACCORDING TO THE REPORTER, DURING SERVICING, THE UNIT WAS CLEANED AND DISINFECTED ACCORDING TO MANUFACTURER¿S GUIDELINE. THE UNIT WAS RECEIVED IN GOOD VISUAL CONDITION. THE REM FUNCTION FAILED ON 1 OHM TEST, DRY SOLDER ON REM CONNECTOR AT STEERING RELAY PCBA (PRINTED CIRCUIT BOARD ASSEMBLY) SIDE, CURRENT SOFTWARE VERSION: 4.00.03. THE REM CONNECTOR SOLDERING WAS REFLOWED ON STEERING RELAY PCBA. OUTPUTS TESTED IN ALL MODES OF OPERATION TEST: PASSED. REM-SYSTEM TEST: PASSED. HIGH FREQUENCY LEAKAGE TEST: PASSED. LOW FREQUENCY ELECTRICAL SAFETY TEST: PASSED. CROSS COUPLING TEST: PASSED. FUNCTIONAL TEST OF ALL HAND SWITCH PORTS, FOOTSWITCH CONNECTIONS AND TOUCHSCREEN TEST: PASSED. THE UNIT HAS BEEN TESTED SUCCESSFULLY AND WAS FULLY CHECKED FOR FUNCTIONALITY AND PATIENT ELECTRICAL SAFETY ACCORDING TO MANUFACTURER¿S GUIDELINES AND NORM. DEVICE WAS TECHNICALLY EFFICIENT AND SAFE TO USE. THERE WAS NO PATIENT INVOLVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18327915 ↗MalfunctionJun 2, 2023Adverse Event Without Identified Device or Use Problem; Intermittent Energy Output; Adverse Event Without Identified Device or Use Problem; Intermittent Energy Output
ACCORDING TO THE REPORTER, DURING BIPOLAR HYSTEROSCOPY RESECTION, THE FOOTSWITCH STOPPED WORKING AND DOES NOT POWER UP, IT STOPPED THE PROCEDURE AND POSTPONED PATIENT WAS UNDER ANESTHESIA ALL THE TIME. THEY PERFORMED AROUND 10 TO 12 PROCEDURES WITH THE FOOTSWITCH. USED THE SAME GENERATOR WITH ANOTHER FOOTSWITCH AND IT WAS WORKING. THE PROCEDURE WAS RESCHEDULED. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17054731 ↗
Other ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES devices
- Vasoview Hemopro 2 VH-4000Maquet Cardiovascular LLC · Electrosurgical, Cutting & Coagulation & Accessories6 recalls1 death3,232 reports
- LigaSure LF1937Covidien LP · Electrosurgical, cutting & coagulation & accessories3 deaths2,090 reports
- Olympus WB91051WOLYMPUS Winter & Ibe GmbH · Electrosurgical, Cutting & Coagulation & Accessories1 death1,951 reports
- Valleylab FT10 VLFT10GENCovidien LP · Electrosurgical, cutting & coagulation & accessories1 recall5 deaths1,062 reports
- LigaSure LF1837Covidien LP · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES1 recall967 reports
- LigaSure LF2019Covidien LP · Electrosurgical, cutting & coagulation & accessories948 reports
- VAPR 225028Depuy Mitek, LLC · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES895 reports
- LigaSure LF1923Covidien LP · Electrosurgical, cutting & coagulation & accessories855 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page