Valleylab E2516HNSBRecalls & reported problems
Covidien LP · Button Switch Pencil Holster; Caution: For manufacturing, processing, or repacking. · Class 2
All models of Valleylab (97)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10884524002996), received between Aug 13, 2018 and Jun 12, 2023 — classified as 1 injury, 1 malfunction.
The most frequently reported specific device problem is “Connection Problem” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem1
- Connection Problem1
- Fire1
- Self-Activation or Keying1
Patient problems
- Burn(s)1
- No Patient Involvement1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryJun 12, 2023Self-Activation or Keying; Connection Problem; Adverse Event Without Identified Device or Use Problem; Self-Activation or Keying; Connection Problem; Adverse Event Without Identified Device or Use Problem
ACCORDING TO THE REPORTER, DURING A PROCEDURE, THE PENCIL WAS PLUGGED INTO THE NON MEDTRONIC GENERATOR, THE PENCIL WAS LAID ON THE PATIENT AND THE STAFF HEARD A BUZZING SOUND. THE DEVICE ACTIVATED WITHOUT THE ACTIVATION BUTTON OR SWITCH BEING PUSHED. NO ONE WAS NEAR THE PENCIL DURING THIS TIME AND NURSE RAN OVER TO UNPLUG THE CORD. PATIENT HAD A BURN IN THE STOMACH ABOUT 1 INCH IN SIZE. THE BURN WAS CUT OUT AND STITCHED UP. THERE WAS AN ERROR CODE ON THE BOVIE WHEN TRIED TO PLUGGED INTO THE VLFT10GEN GENERATOR-E320. A NEW PENCIL WAS USED. FT10 WAS NOT INVOLVED IN THE BURN INCIDENT.
Patient problems: Burn(s)
MAUDE report 17112785 ↗MalfunctionAug 13, 2018Fire; Fire
ACCORDING TO THE REPORTER, THE DEVICE PRODUCED FLAME AND DOCTOR QUIT USING IT. THERE WAS NO PATIENT INVOLVEMENT. A PHOTO SUBMITTED SHOWING THAT THE TIP OF THE PENCIL HAD BURNED.
Patient problems: No Patient Involvement
MAUDE report 7778618 ↗
Other ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES devices
- Vasoview Hemopro 2 VH-4000Maquet Cardiovascular LLC · Electrosurgical, Cutting & Coagulation & Accessories6 recalls1 death3,232 reports
- LigaSure LF1937Covidien LP · Electrosurgical, cutting & coagulation & accessories3 deaths2,090 reports
- Olympus WB91051WOLYMPUS Winter & Ibe GmbH · Electrosurgical, Cutting & Coagulation & Accessories1 death1,951 reports
- Valleylab FT10 VLFT10GENCovidien LP · Electrosurgical, cutting & coagulation & accessories1 recall5 deaths1,062 reports
- LigaSure LF1837Covidien LP · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES1 recall967 reports
- LigaSure LF2019Covidien LP · Electrosurgical, cutting & coagulation & accessories948 reports
- VAPR 225028Depuy Mitek, LLC · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES895 reports
- LigaSure LF1923Covidien LP · Electrosurgical, cutting & coagulation & accessories855 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page