Valleylab E2515HGNSBRecalls & reported problems
Covidien LP · Rocker Switch Pencil Holster; Caution: For manufacturing, processing, or repacking. · Class 2
All models of Valleylab (97)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 8
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 10884524003634), received between May 24, 2018 and Sep 18, 2023 — classified as 1 injury, 7 malfunctions.
The most frequently reported specific device problem is “Fire” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Fire4
- Detachment of Device or Device Component2
- Adverse Event Without Identified Device or Use Problem1
- Failure to Fire1
- Inappropriate/Inadequate Shock/Stimulation1
- Peeled/Delaminated1
- Self-Activation or Keying1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
- Burn(s)2
- No Consequences Or Impact To Patient2
- No Information2
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 18, 2023Peeled/Delaminated; Detachment of Device or Device Component; Peeled/Delaminated; Detachment of Device or Device Component
ACCORDING TO THE REPORTER, DURING HIP PROCEDURE, THE DEVICE'S TEFLON TIP WAS LOSING COATING AND HAD DIFFERENT COLOR. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17766434 ↗MalfunctionDec 2, 2022Detachment of Device or Device Component; Detachment of Device or Device Component
ACCORDING TO THE REPORTER, PRIOR TO USE, THE CLIP BROKE OFF. THERE WAS NO PATIENT INVOLVEMENT. NONE FELL INTO THE PATIENT CAVITY. THE INCIDENT DID NOT CAUSE THE PROCEDURE TO EXTEND. THERE WAS NO OTHER MEDICAL INTERVENTION DONE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15903836 ↗MalfunctionSep 27, 2022Inappropriate/Inadequate Shock/Stimulation; Adverse Event Without Identified Device or Use Problem; Inappropriate/Inadequate Shock/Stimulation; Adverse Event Without Identified Device or Use Problem
ACCORDING TO THE REPORTER, DURING A PROCEDURE, THE SURGEON FELT A SHOCK FROM HIS LEFT MIDDLE FINGER TIP UP TO HIS ELBOW AND GOT A BURN DUE TO THE CAUTERY PENCIL.
Patient problems: Burn(s)
MAUDE report 15490982 ↗MalfunctionFeb 24, 2022Self-Activation or Keying; Failure to Fire; Self-Activation or Keying; Failure to Fire
ACCORDING TO THE REPORTER, DURING BREAST REDUCTION, THE PENCIL ACTIVATED WITHOUT THE SURGEON PRESSING THE BUTTON. IT WAS UNKNOWN IF THE FOOT PEDAL WAS ACTIVATED. ERROR CODES POPPED UP THAT STATED ACTIVATION SHOULD BE STOPPED. AFTER IT STOPPED, THE PENCIL WOULD NOT ACTIVATE WHEN SURGEON PRESSED THE ACTIVATION BUTTONS. THE ACCOUNT SAID THEY PLUGGED THE PENCIL INTO MONOPOLAR 2. SERIAL NUMBER OF THE FT10 USED WAS (B)(4). SAME PENCIL WAS PLUGGED INTO A SECOND FT10 AND IT STILL WOULD NOT ACTIVATE. A NEW PENCIL WAS OPENED FOR THE CASE AND IT WORKED WITHOUT ISSUE. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13590769 ↗MalfunctionOct 25, 2019Fire; Fire
ACCORDING TO THE REPORTER, THE TIP OF THE DEVICE CAUGHT FIRE SIZE OF A BIRTHDAY CANDLE FLAME. THERE WAS NO PATIENT INJURY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9241499 ↗MalfunctionAug 3, 2018Fire
ACCORDING TO THE REPORTER, DURING USE, THE CAUTERY TIP IGNITED.
Patient problems: No Consequences Or Impact To Patient; No Information; No Known Impact Or Consequence To Patient
MAUDE report 7747299 ↗InjuryJul 6, 2018Fire; Fire
ACCORDING TO THE REPORTER, DURING A PLANNED CLOSURE OF MOHS DEFECT TO NASAL TIP AND UPPER CHEST WITH FULL THICKNESS SKIN GRAFT TO NOSE FROM RIGHT PRE-AURICULAR AREA AND PRIMARY CLOSURE OF THE UPPER MEDIAL CHECK DEFECT UNDER ANESTHESIA MONITORED INTRAVENOUS SEDATION. THEY INITIALLY RECEIVED THE PATIENT WITH OXYGEN VIA NASAL CANNULA, BUT WERE UNABLE TO MAINTAIN THE OXYGEN SATURATION ABOVE 90. THEY SWITCHED TO ORAL CAVITY AND A NON-REBREATHER MASK. THE HAIR AND BEARD WERE CLIPPED AND SKIN PREPPED WITH ANTISEPTIC SCRUB. ONE-HALF STRENGTH ANTISEPTIC PAINT WAS USED ON THE FACE AND FULL STRENGTH ANTISEPTIC PAINT WAS USED ON THE CHEST. MORE THAN THREE MINUTES LAPSED PRIOR TO DRAPING. PART OF DRAPE PLACED OVER BOTTOM PORTION OF THE NON-REBREATHER OXYGEN RESERVOIR OVER THE LEFT CHEST AND SHOULDER. BIPOLAR ELECTROCAUTERY WAS USED ON THE NASAL AREA AND THE RIGHT PRE-AURICULAR SKIN DONOR SITE AREA. WHEN ACTIVATING A MONOPOLAR ELECTROCAUTERY DEVICE WITH A NEEDLE TIP ON THE CHEST DEFECT, THERE WAS AN ARC AND FLASH OF FIRE. DRAPES WERE IMMEDIATELY REMOVED AND THE AREA WAS IMMEDIATELY FLUSHED WITH SALINE. THE ORAL AIRWAY WAS REMOVED AND INSPECTED AND A NASAL AIRWAY PLACED. THE PATIENT WAS INTUBATED
Patient problems: Burn(s)
MAUDE report 7667690 ↗MalfunctionMay 24, 2018Fire; Fire
ACCORDING TO THE REPORTER, THE CAUTERY TIP CAUSED A FIRE IN THE OPERATING ROOM.
Patient problems: No Information
MAUDE report 7541633 ↗
Other ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES devices
- Vasoview Hemopro 2 VH-4000Maquet Cardiovascular LLC · Electrosurgical, Cutting & Coagulation & Accessories6 recalls1 death3,232 reports
- LigaSure LF1937Covidien LP · Electrosurgical, cutting & coagulation & accessories3 deaths2,090 reports
- Olympus WB91051WOLYMPUS Winter & Ibe GmbH · Electrosurgical, Cutting & Coagulation & Accessories1 death1,951 reports
- Valleylab FT10 VLFT10GENCovidien LP · Electrosurgical, cutting & coagulation & accessories1 recall5 deaths1,062 reports
- LigaSure LF1837Covidien LP · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES1 recall967 reports
- LigaSure LF2019Covidien LP · Electrosurgical, cutting & coagulation & accessories948 reports
- VAPR 225028Depuy Mitek, LLC · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES895 reports
- LigaSure LF1923Covidien LP · Electrosurgical, cutting & coagulation & accessories855 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page