Snowden-Pencer SP8302Recalls & reported problems
Steris Corporation · Snowden-Pencer MIS SCISSORS INSERT, MINI METZ 36CM, 10/BOX · Class 2
All models of Snowden-Pencer (62)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 10885403184154), received between Oct 22, 2021 and Jan 11, 2024 — classified as 4 malfunctions.
The most frequently reported specific device problem is “Break” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Break2
- Detachment of Device or Device Component2
Patient problems
- Foreign Body In Patient3
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 11, 2024Detachment of Device or Device Component; Detachment of Device or Device Component
THE USER FACILITY REPORTED THAT DURING A PATIENT PROCEDURE ONE OF THE HALVES OF THEIR SNOWDEN-PENCER MINI METZ SCISSORS INSERT BROKE OFF INTO THE PATIENT'S ABDOMEN. THE DETACHED COMPONENT WAS SUCCESSFULLY RETRIEVED WITH A GRASPER. THE PROCEDURE WAS COMPLETED SUCCESSFULLY. NO REPORT OF INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18498111 ↗MalfunctionSep 6, 2023Break; Break
MAT# SP8302, LOT # 76722. MINI SHEAR BROKE DURING A CASE TODAY, THEY WERE ABLE TO RETRIEVE THE BROKEN PIECE AND IS TAPED TO THE MINI SHEAR TO BE SENT IN. NO FURTHER INFORMATION IS NOT AVAILABLE.
Patient problems: Foreign Body In Patient
MAUDE report 17692769 ↗MalfunctionSep 6, 2023Break; Break
THE COMPLAINT IS REGARDING BD ITEM # SP8302, MINI METZ SCISSOR DISPOSABLE INSERT, 5MM, 36CM IN WHICH ONE OF THE BLADES BROKE OFF INTO THE ABDOMEN OF A PATIENT DURING A LAPAROSCOPIC CHOLECYSTECTOMY PROCEDURE. THE BROKEN BLADE WAS RETRIEVED FROM THE PATIENT¿S ABDOMEN. ADDITIONAL INFORMATION PROVIDED WAS THE LOT # 77278, EXPIRATION DATE 6.1.2025." NO FURTHER INFORMATION IS NOT AVAILABLE.
Patient problems: Foreign Body In Patient
MAUDE report 17692033 ↗MalfunctionOct 22, 2021Detachment of Device or Device Component; Detachment of Device or Device Component
PER COMPLAINT DETAILS RECEIVED: IT WAS REPORTED THAT TIP OF SCISSOR BROKE WITHIN THE PATIENT. PATIENT IS FINE. IT TOOK NEARLY TWO HOURS TO RETRIEVE BROKEN PIECE. INCIDENT WAS REPORTED TO FDA BY THE CUSTOMER. 22OCT2021 ADDITIONAL INFORMATION: WHAT WAS THE PROCEDURE THAT WAS BEING PERFORMED? LAPAROSCOPIC APPENDECTOMY DID ANY PART THE INSTRUMENT FALL INTO THE PATIENT¿S BODY, AND IF SO, HOW WAS IT RETRIEVED? YES, THE BOTTOM BLADE BROKE OFF AND FELL INTO THE PATIENT AND WAS RETRIEVED WITH LAPAROSCOPY. WAS THERE A MEDICAL PROCEDURE PERFORMED TO VERIFY IF THE INSTRUMENT WAS IN THE PATIENT¿S BODY, SUCH AS AN X-RAY? YES, THREE FLAT PLATE ABDOMINAL X-RAYS WERE TAKEN WITH THE BLADE IDENTIFIED IN EACH VIEW. WHAT WAS THE PATIENT¿S OUTCOME? NO OBVIOUS ENTEROTOMY OR INJURY FROM THE FRACTURED PIECE WAS NOTED. THE PATIENT WAS AWAKENED AND RETURNED TO THE RECOVERY ROOM IN STABLE CONDITION. WAS THE PROCEDURE COMPLETED AS PLANNED? INTRAABDOMINAL FRACTURE OF THE LAPAROSCOPIC SCISSORS EXTENDED THE PROCEDURE TIME 1 1/2 TO 2 HOURS. AFTER THE FRACTURED PIECE WAS LOCATED AND REMOVED, THE PROCEDURE WAS COMPLETED AS PLANNED. CAN YOU PLEASE SEND ALL PARTS OF THE INSTRUMENT FOR EVALUATION? YES. DO YOU HAVE THE
Patient problems: Foreign Body In Patient
MAUDE report 12684846 ↗
Other ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES devices
- Vasoview Hemopro 2 VH-4000Maquet Cardiovascular LLC · Electrosurgical, Cutting & Coagulation & Accessories6 recalls1 death3,232 reports
- LigaSure LF1937Covidien LP · Electrosurgical, cutting & coagulation & accessories3 deaths2,090 reports
- Olympus WB91051WOLYMPUS Winter & Ibe GmbH · Electrosurgical, Cutting & Coagulation & Accessories1 death1,951 reports
- Valleylab FT10 VLFT10GENCovidien LP · Electrosurgical, cutting & coagulation & accessories1 recall5 deaths1,062 reports
- LigaSure LF1837Covidien LP · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES1 recall967 reports
- LigaSure LF2019Covidien LP · Electrosurgical, cutting & coagulation & accessories948 reports
- VAPR 225028Depuy Mitek, LLC · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES895 reports
- LigaSure LF1923Covidien LP · Electrosurgical, cutting & coagulation & accessories855 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page