Skip to content

VAPR 225304Recalls & reported problems

Depuy Mitek, LLC · VAPR Angled End Effect Electrode 3.5 mm x 160 mm · Class 2

All models of VAPR (30)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
6
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 10886705009282), received between Apr 3, 2018 and Jun 10, 2026 — classified as 1 injury, 5 malfunctions.

The most frequently reported specific device problem is “Blocked Connection” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 1 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Blocked Connection2
  • Break2
  • Detachment of Device or Device Component1
  • Electrical /Electronic Property Problem1
  • Output Problem1
  • Torn Material1
  • Use of Device Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions4
  • Foreign Body In Patient1
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 10, 2026Output Problem

    IT WAS REPORTED THAT DURING ANTERIOR CRUCIATE LIGAMENT RECONSTRUCTION SURGERY, EVEN IF PUSHING THE FOOT SWITCH, THE VAPR SOMETIMES DID NOT OUTPUT. THE SURGERY WAS COMPLETED WITHOUT ANY ISSUES USING A REPLACEMENT PROBE. THERE WAS NO SURGICAL DELAY AND NO HARM TO THE PATIENT. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25455389 ↗
  • MalfunctionFeb 24, 2025Blocked Connection; Blocked Connection

    IT WAS REPORTED THAT DURING AN ARTHROSCOPIC BANKART REPAIR (ABR) SHOULDER PROCEDURE, WHEN THE VAPR 3.5MM END 21 DEG ELECTRODE DEVICE WAS CONNECTED TO A VAPR HANDPIECE, THE MAIN UNIT DID NOT ACHIEVE THE CONNECTION COMPLETION STATE. IT WAS DETERMINED THAT THERE WERE NO PROBLEMS WITH THE HANDPIECE DEVICE BECAUSE OTHER PROBES HAD BEEN USED ALREADY WITH THE HANDPIECE. A NEW PRODUCT WAS OPENED AND CONNECTED WITHOUT ANY PROBLEMS. THERE WAS NO SURGICAL DELAY AND NO HARM TO THE PATIENT. NO ADDITIONAL INFORMATION COULD BE PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21453193 ↗
  • MalfunctionFeb 22, 2023Blocked Connection; Blocked Connection

    IT WAS REPORTED BY THE AFFILIATE IN JAPAN THAT DURING AN UNKNOWN PROCEDURE OF THE SHOULDER ON (B)(6) 2023, IT WAS OBSERVED THAT THE ANGLED END EFFECT ELECTRODE, 21º DEVICE BY A HANDPIECE WHEN IT WAS CONNECTED TO THE HANDPIECE. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE PATIENT CONSEQUENCES NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16419063 ↗
  • MalfunctionFeb 21, 2023Electrical /Electronic Property Problem; Use of Device Problem; Detachment of Device or Device Component; Electrical /Electronic Property Problem; Use of Device Problem; Detachment of Device or Device Component

    IT WAS REPORTED BY THE AFFILIATE IN JAPAN THAT DURING AN UNKNOWN PROCEDURE ON (B)(6) 2022, IT WAS OBSERVED THAT THE ANGLED END EFFECT ELECTRODE, 21º DEVICE DID NOT RUN. DURING IN-HOUSE ENGINEERING EVALUATION, IT WAS DETERMINED THAT THE RETURN WIRE WAS NO LONGER SOLDERED TO THE RETURN TUBE. ACCORDING TO THE INVESTIGATION, THE WIRE HAD BECOME DETACHED RESULTING IN NO CONTINUITY ON THE RETURN CIRCUIT; AND THEREFORE, NO ACTIVATION DURING THE ABLATE AND COAGULATE MODE. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE PATIENT CONSEQUENCES NOR SURGICAL DELAY REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16412285 ↗
  • InjurySep 19, 2018Break; Break

    ON (B)(6) 2018, THE REPORTED DEVICE WAS USED FOR THE SYNOVIAL RESECTION DURING SCREW REMOVAL SURGERY AFTER HTO. AFTER THE SURGERY, THE SURGEON RECOGNIZED THAT THE PROBE TIP WAS BROKEN AND REMAINED IN THE PATIENT¿S BODY. THEN, THE BROKEN FRAGMENT WAS REMOVED FROM THE PATIENT¿S BODY WHEN THE SURGEON PERFORMED ADDITIONAL REMOVAL SURGERY LATER. IT WAS BRAND NEW AND THE FIRST USE WHEN THE ISSUE OCCURRED. NO FURTHER INFORMATION WAS PROVIDED BY THE HOSPITAL.

    Patient problems: Foreign Body In Patient

    MAUDE report 7890772 ↗
  • MalfunctionApr 3, 2018Break; Torn Material; Break; Torn Material

    IT WAS REPORTED THAT THE REPORTED DEVICE WAS USED DURING THE SURGERY ON (B)(6) 2018. IT WAS NOTICED THAT THE BLACK COATING ON THE ELECTRODE SHAFT CAME OFF DURING THE PROCEDURE. THEREFORE, THE SURGEON STOPPED USING THE REPORTED DEVICE. IT WAS BRAND NEW AND THE FIRST USE WHEN THE ISSUE OCCURRED. THERE WAS NO DELAY IN SURGERY AND NO HARM TO THE PATIENT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7394036 ↗

Other ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES devices

All ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES devices →

Sources and updates