VAPR 225312Recalls & reported problems
Depuy Mitek, LLC · VAPR flexible Side Effect Electrode 3.5 mm x 160 mm · Class 2
All models of VAPR (30)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 10886705009312), received between Sep 18, 2018 and May 4, 2022 — classified as 5 malfunctions.
The most frequently reported specific device problem is “Electrical /Electronic Property Problem” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Most reported problems
Device problems
- Electrical /Electronic Property Problem2
- Appropriate Device Problem Term/Code Not Available1
- Break1
- Device Damaged Prior to Use1
- Electromagnetic Interference1
- Use of Device Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
- No Consequences Or Impact To Patient3
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 4, 2022Electrical /Electronic Property Problem; Use of Device Problem; Electrical /Electronic Property Problem; Use of Device Problem
IT WAS REPORTED BY THE CUSTOMER IN (B)(6) THAT DURING AN ARTHROSCOPIC SHOULDER STABILIZATION SURGERY ON (B)(6) 2021, IT WAS OBSERVED THAT THE VAPR 3.5MM FLEX SIDE DEVICE DID NOT WORK AND EVEN AFTER THE CABLE WAS CHANGED, IT STILL DID NOT WORK. DURING IN-HOUSE ENGINEERING EVALUATION, IT WAS DETERMINED THAT THERE WAS POOR ACTIVE SOLDER CONNECTION AND GAP BETWEEN SOLDER AND PCB ON THE DEVICE WHICH FURTHER SHOWED THAT THE DEVICE HAD AN ELECTRICAL PROBLEM; AND THAT IT DID NOT COAGULATE NOR ABLATE. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE PATIENT CONSEQUENCES NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14278750 ↗MalfunctionMay 4, 2022Electrical /Electronic Property Problem; Appropriate Device Problem Term/Code Not Available; Electrical /Electronic Property Problem; Appropriate Device Problem Term/Code Not Available
IT WAS REPORTED BY THE CUSTOMER IN (B)(6) THAT DURING AN ARTHROSCOPIC SHOULDER STABILIZATION SURGERY ON (B)(6) 2021, IT WAS OBSERVED THAT THE VAPR 3.5MM FLEX SIDE DEVICE DID NOT WORK AND EVEN AFTER THE CABLE WAS CHANGED, IT STILL DID NOT WORK. DURING IN-HOUSE ENGINEERING EVALUATION, IT WAS DETERMINED THAT THERE WAS POOR ACTIVE SOLDER CONNECTION AND GAP BETWEEN SOLDER AND PCB ON THE DEVICE WHICH FURTHER SHOWED THAT THE DEVICE HAD AN ELECTRICAL PROBLEM; AND THAT IT DID NOT COAGULATE NOR ABLATE. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE PATIENT CONSEQUENCES NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14278702 ↗MalfunctionJul 14, 2020Device Damaged Prior to Use; Device Damaged Prior to Use
IT WAS REPORTED BY THE AFFILIATE IN (B)(6) THAT DURING AN UNKNOWN SURGERY, IT WAS OBSERVED THAT THE ELECTRODE DEVICE WAS DAMAGED UPON OPENING ITS PACKAGE. IT WAS NOT REPORTED IF THERE WAS A DELAY IN THE SURGERY OR IF A SPARE DEVICE WAS AVAILABLE FOR USE TO COMPLETE THE SURGERY. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Clinical Signs, Symptoms or Conditions; No Consequences Or Impact To Patient
MAUDE report 10270096 ↗MalfunctionMar 23, 2020Break; Break
IT WAS REPORTED BY AFFILIATE VIA EMAIL THAT DURING SHOULDER REPAIR THE VAPR 3.5MM FLEX SIDE -EA THE TIP OF THE ELECTRODE WAS BROKEN IN THE MIDDLE OF SURGERY. NO PATIENT CONSEQUENCES OR SURGICAL DELAY REPORTED. THE DEVICE IS AVAILABLE TO BE RETURNED FOR EVALUATION. THE AFFILIATE ALSO REPORTED ALL FRAGMENTS WERE REMOVED FROM THE PATIENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9865939 ↗MalfunctionSep 18, 2018Electromagnetic Interference; Electromagnetic Interference
IT WAS REPORTED THE ELECTRODE CAUSED DISTURBANCES IN THE VIO 3- ERBE DEVICE WHEN USED NORMALLY. THE VIO 3- ERBE STOOD ON THE TOWER BUT WAS NOT IN USE. ONLY THE NEUTRAL ELECTRODE WAS ATTACHED TO THE PATIENT. DURING THE USE OF THE VAPR THE VIO 3 JUMPED THROUGH ITS PROGRAM, INDICATING A FAILURE BY THE VAPR. CUSTOMER HAD PURCHASED 2 OTHER ELECTRODES WITH THE SAME BATCH, WHICH DID NOT WORK WHICH IS CAPTURED IN COMPLAINT 15126MT. THE CUSTOMER HAS ELEVEN (11) MORE UNOPENED ELECTRODES FROM THE SAME BATCH WHICH WILL BE RETURNED. NEW ELECTRODES WILL BE PURCHASED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7885009 ↗
FDA clearance and classification
Cleared through 510(k) K974022 on Jan 12, 1998 (decision: Substantially Equivalent). Submitted by Mitek Products as “Vapr”. The FDA review took 90 days. FDA record ↗
The FDA regulates product code GEI (electrosurgical, cutting & coagulation & accessories) under 21 CFR 878.4400 as a Class II device. 21 CFR 878.4400 ↗
Common questions about the VAPR 225312
Has the VAPR 225312 been recalled?
No FDA recall record in our data names this exact device identifier (DI 10886705009312), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the VAPR 225312?
The FDA's MAUDE database lists 5 adverse event reports for this device, received between Sep 18, 2018 and May 4, 2022. In those reports, the most common specific device problems are “Electrical /Electronic Property Problem” (2), “Break” (1) and “Device Damaged Prior to Use” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the VAPR 225312?
None of the reports are classified as deaths, 0 as injuries and 5 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the VAPR 225312 FDA approved?
It was cleared through 510(k) K974022 on Jan 12, 1998. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Who makes the VAPR 225312, and is it still sold?
The FDA's device database lists Depuy Mitek, LLC as the company responsible for it (brand VAPR, model 225312). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES devices
- Vasoview Hemopro 2 VH-4000Maquet Cardiovascular LLC · Electrosurgical, Cutting & Coagulation & Accessories6 recalls1 death3,137 reports
- LigaSure LF1937Covidien LP · Electrosurgical, cutting & coagulation & accessories3 deaths2,038 reports
- Olympus WB91051WOLYMPUS Winter & Ibe GmbH · Electrosurgical, Cutting & Coagulation & Accessories1 death1,891 reports
- Valleylab FT10 VLFT10GENCovidien LP · Electrosurgical, cutting & coagulation & accessories1 recall5 deaths1,046 reports
- LigaSure LF1837Covidien LP · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES1 recall957 reports
- LigaSure LF2019Covidien LP · Electrosurgical, cutting & coagulation & accessories929 reports
- VAPR 225028Depuy Mitek, LLC · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES874 reports
- LigaSure LF1923Covidien LP · Electrosurgical, cutting & coagulation & accessories831 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.