VAPR COOLPULSE 228147Recalls & reported problems

Depuy Mitek, LLC · VAPR COOLPULSE 90 Electrode with Hand Controls 90 DEGREES Suction with Integrated Handpiece · Class 2

All models of VAPR COOLPULSE (2)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Quick summary

This DePuy Mitek VAPR CoolPulse is an electrosurgical tool with suction, used in arthroscopic surgery. Many of its 300 reports describe fluid flow problems or suction failures, and the tool jamming or not responding, with another device used to finish the surgery. The reports include 13 injuries and no deaths. Reports rose from 47 in the 12 months before to 69 in the latest 12 months. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
298
Deaths reported
0
Injuries reported
13
Recalls naming it
0

Summary

The FDA's MAUDE database contains 298 adverse event reports that list this exact device (DI 10886705010097), received between Aug 28, 2014 and Aug 5, 2026 — classified as 13 injuries, 285 malfunctions.

The most frequently reported specific device problem is “Infusion or Flow Problem” (79 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 1 reportsJan 2024: 7 reportsFeb 2024: 4 reportsMar 2024: 1 reportsApr 2024: 2 reportsMay 2024: 3 reportsJun 2024: 3 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 1 reportsOct 2024: 4 reportsNov 2024: 4 reportsDec 2024: 5 reportsJan 2025: 3 reportsFeb 2025: 1 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 4 reportsJul 2025: 7 reportsAug 2025: 4 reportsSep 2025: 12 reportsOct 2025: 5 reportsNov 2025: 6 reportsDec 2025: 12 reportsJan 2026: 14 reportsFeb 2026: 11 reportsMar 2026: 8 reportsApr 2026: 0 reportsMay 2026: 8 reportsJun 2026: 6 reportsJul 2026: 0 reportsAug 2026: 2 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports20242025202614
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Infusion or Flow Problem79
  • Suction Failure40
  • Detachment of Device or Device Component30
  • Use of Device Problem28
  • Electrical /Electronic Property Problem26
  • Break24
  • Output Problem22
  • Device Contaminated During Manufacture or Shipping13

Patient problems

  • No Clinical Signs, Symptoms or Conditions192
  • No Consequences Or Impact To Patient84
  • Not Applicable15
  • Foreign Body In Patient7
  • No Known Impact Or Consequence To Patient4
  • Partial thickness (Second Degree) Burn3
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
  • Burn(s)1

Recent adverse event reports

The 25 most recent of 298 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 5, 2026Output Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50139283 ↗
  • MalfunctionAug 5, 2026Output Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50139281 ↗
  • MalfunctionJun 9, 2026Infusion or Flow Problem

    IT WAS REPORTED THAT DURING AN ARTHROSCOPIC ROTATOR CUFF REPAIR PROCEDURE THE VAPR CLPLSE90 ELECTRODE W HAND CNTRLS DEVICE JAMMED AND COULD NOT SUCTION. ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO DELAY IN THE PROCEDURE REPORTED. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25447506 ↗
  • MalfunctionJun 5, 2026Infusion or Flow Problem

    IT WAS REPORTED THAT DURING A ROTATOR CUFF REPAIR PROCEDURE, THE APR CLPLSE90 ELECTRODE W HAND CNTRLS DEVICE BECAME JAMMED. IT WAS NOT REPORTED IF THERE WERE ANY DELAYS TO THE SURGICAL PROCEDURE. IT WAS REPORTED THAT A SPARE DEVICE WAS USED TO COMPLETE THE SURGERY. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25414582 ↗
  • MalfunctionJun 4, 2026Electrical /Electronic Property Problem

    IT WAS REPORTED THAT DURING SHOULDER SLEEVE REPAIR, THE VAPR COOLPULSE HAD SPARKS AND DID NOT RESPOND. ANOTHER DEVICE WAS USED TO COMPLETE THE SURGERY. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25390074 ↗
  • MalfunctionJun 4, 2026Use of Device Problem; Infusion or Flow Problem

    IT WAS REPORTED THAT DURING SHOULDER CUFF REPAIR BLADE BLOCKAGE, THE VAPR COOLPULSE GOT JAMMED AND DID NOT WORK. ANOTHER DEVICE WAS USED TO COMPLETE THE SURGERY. THERE WERE NO ADVERSE CONSEQUENCES RELATED TO THE PATIENT. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25390002 ↗
  • MalfunctionJun 4, 2026Infusion or Flow Problem

    IT WAS REPORTED AN ANTERIOR CRUCIATE LIGAMENT RECONSTRUCTION, IT WAS FOUND THAT THE VAPR COOLPULSE DID NOT SUCTION. ANOTHER DEVICE WAS USED TO COMPLETE THE SURGERY. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25389784 ↗
  • MalfunctionJun 1, 2026Electrical /Electronic Property Problem; Material Fragmentation; Use of Device Problem

    IT WAS REPORTED THAT DURING IN-HOUSE ENGINEERING EVALUATION, IT WAS DETERMINED THAT THE VAPR CLPLSE90 ELECTRODE W HAND CNTRLS DEVICE HAD FOREIGN SUBSTANCE/DEBRIS/CLEANING/STERILIZATION AND WOULD NOT COAGULATE. IT WAS INITIALLY REPORTED THAT DURING AN ARTHROSCOPIC LIGAMENT RECONSTRUCTION PROCEDURE, IT WAS DISCOVERED THAT THE DEVICE DID NOT WORK. THERE WERE NO DELAYS IN THE SURGICAL PROCEDURE. IT WAS REPORTED THAT A SPARE DEVICE WAS USED TO COMPLETE THE SURGERY. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25348136 ↗
  • MalfunctionMay 28, 2026Infusion or Flow Problem

    IT WAS REPORTED THAT DURING A ROTATOR CUFF REPAIR SURGICAL PROCEDURE, WHEN USING THE VAPR CLPLSE90 ELECTRODE W HAND CNTRLS DEVICE, IT WAS OBSERVED THAT THE DEVICE BECAME JAMMED AND COULD NOT SUCTION. ANOTHER DEVICE WAS USED TO COMPLETE THE SURGERY. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION COULD BE PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25322979 ↗
  • MalfunctionMay 27, 2026Infusion or Flow Problem

    IT WAS REPORTED THAT DURING A ROTATOR CUFF REPAIR SURGERY, THE VAPR COOLPULSE DEVICE COULD NOT SUCTION. THE ELECTRODE CONNECTED TO THE ASPIRATOR AND PUT IT INTO THE ARTICULAR CAVITY. IT WAS FOUND THAT THERE WAS NO SUCTION. ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO REPORTS OF A DELAY IN THE PROCEDURE REPORTED. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25305998 ↗
  • MalfunctionMay 27, 2026Electrical /Electronic Property Problem; Output Problem

    IT WAS REPORTED THAT DURING A GLUTEAL MUSCLE CONTRACTURE SURGICAL PROCEDURE IT WAS OBSERVED THAT THE VAPR CLPLSE90 ELECTRODE W HAND CNTRLS DEVICE SUDDENLY HAD NO ABILITY TO PERFORM ABLATION DURING SURGERY. IT WAS REPORTED THAT ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION COULD BE PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25305774 ↗
  • MalfunctionMay 27, 2026Infusion or Flow Problem

    IT WAS REPORTED THAT DURING A ROTATOR CUFF REPAIR SURGICAL PROCEDURE, WHILE USING A 4.5 HEALIX ADVANCE BR W/OCORD ANCHOR DEVICE IT WAS OBSERVED THAT THE ANCHOR WAS BROKEN OFF. IT WAS REPORTED THAT ALL BROKEN PARTS WERE REMOVED. IT WAS REPORTED THAT ANOTHER DEVICE WAS USED TO COMPLETE THE SURGERY. IT WAS REPORTED THAT IN THE SAME PROCEDURE, WHILE USING A VAPR CLPLSE90 ELECTRODE W HAND CNTRLS DEVICE IT WAS OBSERVED THAT WHEN THE ELECTRODE WAS CONNECTED TO THE ASPIRATOR, AND PUT IT INTO THE ARTICULAR CAVITY, IT IS FOUND THAT THERE WAS NO SUCTION. ANOTHER DEVICE WAS USED TO COMPLETE THE SURGERY. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION COULD BE PROVIDED. THIS IS REPORT 2 OF 2 FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25304392 ↗
  • MalfunctionMay 11, 2026Degraded; Device Contaminated During Manufacture or Shipping

    IT WAS REPORTED THAT BEFORE THE MENISCAL REPAIR SURGERY WITH A VAPR CLPLSE90 ELECTRODE W HAND CNTRLS DEVICE, UPON OPENING THE DEVICE PACKAGE, IT WAS OBSERVED THAT THE DEVICE TIP HAD RUST. ADDITIONALLY, IT WAS REPORTED THAT \"EMBROIDERY ECCHYMOSIS\" AT THE TIP OF THE DEVICE. ANOTHER DEVICE WAS USED TO COMPLETE THE SURGERY. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION COULD BE PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25142754 ↗
  • MalfunctionMay 11, 2026Infusion or Flow Problem

    IT WAS REPORTED THAT DURING MENISCAL REPAIR SURGERY, THE VAPR COOLPULSE GOT JAMMED AND DID NOT WORK. ANOTHER DEVICE WAS USED TO COMPLETE THE SURGERY. THERE WERE NO ADVERSE CONSEQUENCES RELATED TO THE PATIENT. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25137732 ↗
  • MalfunctionMay 4, 2026Device Remains Activated; Physical Resistance/Sticking

    IT WAS REPORTED THAT DURING AN ARTHROSCOPIC SHOULDER STABILIZATION PROCEDURE THE VAPR CLPLSE90 ELECTRODE W HAND CNTRLS DEVICE BUTTON STUCK REMAINING IN ABLATION MODE. ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS A 5 MINUTE DELAY IN THE PROCEDURE REPORTED. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25072418 ↗
  • MalfunctionMay 4, 2026Infusion or Flow Problem

    IT WAS REPORTED THAT DURING AN UNSPECIFIED ARTHROSCOPIC PROCEDURE THE VAPR CLPLSE90 ELECTRODE W HAND CNTRLS DEVICE WAS CLOGGED CAUSING SUCTION FAILURE. ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO DELAY IN THE PROCEDURE REPORTED. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25071925 ↗
  • MalfunctionMar 31, 2026Detachment of Device or Device Component

    IT WAS REPORTED THAT DURING AN UNSPECIFIED SURGICAL PROCEDURE IT WAS OBSERVED THAT A METAL COMPONENT OF THE VAPR CLPLSE90 ELECTRODE W HAND CNTRLS ELECTRODE DEVICE DETACHED AND FELL INTO THE PATIENT¿S KNEE. IT WAS NOT REPORTED IF THERE WERE ANY ADVERSE CONSEQUENCES TO THE PATIENT, OR IF THERE WAS A SURGICAL DELAY. THE EXACT DATE OF THE EVENT WAS NOT REPORTED; HOWEVER, IT WAS REPORTED THAT THE EVENT OCCURRED IN MARCH OF 2026. NO ADDITIONAL INFORMATION COULD BE PROVIDED.

    Patient problems: Foreign Body In Patient

    MAUDE report 24742284 ↗
  • MalfunctionMar 26, 2026Product Quality Problem; Detachment of Device or Device Component; Device Contaminated During Manufacture or Shipping

    IT WAS REPORTED THAT DURING IN-HOUSE ENGINEERING EVALUATION, IT WAS DETERMINED THAT THE VAPR CLPLSE90 ELECTRODE W HAND CNTRLS DEVICE FELL APART AND HAD FOREIGN SUBSTANCE/DEBRIS/CLEANING/STERILIZATION. IT WAS INITIALLY REPORTED THAT THE COOL PULSE 90 VAPR WAND WAS OPENED AND THE FACEPLATE WAS MISSING. IT WAS NOT REPORTED IF THE DEVICE WAS USED IN SURGERY. IT WAS NOT REPORTED IF THERE WERE ANY DELAYS IN A SURGICAL PROCEDURE OR IF A SPARE DEVICE WAS AVAILABLE. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24699156 ↗
  • MalfunctionMar 20, 2026Use of Device Problem; Infusion or Flow Problem; Output Problem

    IT WAS REPORTED THAT DURING AN ARTHROSCOPIC SHOULDER SURGICAL PROCEDURE IT WAS OBSERVED THAT WHEN THE VAPR CLPLSE90 ELECTRODE W HAND CNTRLS DEVICE WAS CONNECTED AS USUAL IT FUNCTIONED CORRECTLY DURING THE INITIAL PART OF THE PROCEDURES. AFTER SOME TIME, WITHOUT ANY CHANGES TO THE SETUP, THE SUCTION FUNCTION ON THE INSTRUMENT STOPPED WORKING. THE USER TROUBLESHOOTED THE SUCTION SOURCE, WHICH WAS FUNCTIONAL. WE CHECKED THE TUBING FOR CLOGS, BUT NONE WERE FOUND. WE ALSO INSPECTED THE TIP OF THE INSTRUMENT FOR BLOCKAGES, BUT IT WAS CLEAR. THE SURGEON HAD TO USE THE SUCTION ON THE ARTHROSCOPY TOWER INSTEAD, WHICH WAS LESS CONVENIENT BUT ALLOWED THE PROCEDURES TO CONTINUE. THERE WERE NO REPORTS OF DELAY IN THE PROCEDURE. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24649355 ↗
  • MalfunctionMar 18, 2026Use of Device Problem; Infusion or Flow Problem; Output Problem

    IT WAS REPORTED THAT DURING AN ARTHROSCOPIC SHOULDER SURGERY, THE VAPR CLPLSE90 ELECTRODE WITH HAND CONTROLS WAS CONNECTED AS USUAL AND FUNCTIONED CORRECTLY DURING THE INITIAL PORTION OF THE PROCEDURE. AFTER SOME TIME, WITHOUT ANY CHANGES TO THE SETUP, THE INSTRUMENT'S SUCTION FUNCTION STOPPED WORKING. THE REPORTER INDICATED THAT THE SUCTION SOURCE WAS FUNCTIONAL. THE TUBING WAS CHECKED AND SHOWED NO CLOGS. THE INSTRUMENT TIP WAS INSPECTED AND WAS CLEAR. IT WAS REPORTED THAT THE SURGEON USED SUCTION FROM THE ARTHROSCOPY TOWER INSTEAD. THIS WAS LESS CONVENIENT BUT ALLOWED THE PROCEDURES TO CONTINUE. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24627960 ↗
  • MalfunctionMar 17, 2026Electrical /Electronic Property Problem; Infusion or Flow Problem

    IT WAS REPORTED THAT DURING AN ARTHROSCOPIC ROTATOR CUFF REPAIR PROCEDURE THAT THE VAPR CLPLSE90 ELECTRODE W HAND CNTRLS DEVICE THE COAGULATION FUNCTION AND SUCTION WERE NOT WORKING. ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO DELAY IN THE PROCEDURE REPORTED. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24622054 ↗
  • MalfunctionMar 11, 2026Output Problem

    IT WAS REPORTED THAT DURING SERVICE AND EVALUATION, IT WAS DETERMINED THAT THE TIP OF THE VAPR CLPLSE90 ELECTRODE W HAND CNTRLS DEVICE HAD FOREIGN MATTER, PRESUMABLY BIOLOGICAL MATTER. IT WAS REPORTED IN THE SERVICE ORDER THAT THE DEVICE DISPLAYED AN ALERT WHEN CONNECTED. IT WAS UNKNOWN WHEN THE EVENT OCCURRED. THERE WERE NO REPORTS OF DELAYS TO A SURGICAL PROCEDURE. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24575534 ↗
  • MalfunctionMar 11, 2026Product Quality Problem

    IT WAS REPORTED THAT THE VAPR CLPLSE90 ELECTRODE W HAND CNTRLS DEVICE METAL PLATE WAS MISSING. IT WAS NOT REPORTED IF THE DEVICE WAS USED IN SURGERY. IT WAS NOT REPORTED IF THERE WERE ANY DELAYS IN A SURGICAL PROCEDURE, OR IF A SPARE DEVICE WAS AVAILABLE FOR USE. IT WAS NOT REPORTED IF THERE WERE ANY INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24567932 ↗
  • MalfunctionMar 10, 2026Use of Device Problem; Contamination /Decontamination Problem; Electrical Power Problem

    IT WAS REPORTED THAT DURING A ROTATOR CUFF REPAIR PROCEDURE, THE VAPR CLPLSE90 ELECTRODE W HAND CNTRLS DEVICE OPENED THE PACKING (NOT USED ON PATIENT), THE DEVICE AFTER CONNECT WITH GENERATOR, DOES NOT FUNCTION AT ALL, THE SCREEN DID NOT SHOW ANY ALERT CODE. ANOTHER DEVICE WAS USED TO COMPLETE THE SURGERY. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24560635 ↗
  • MalfunctionFeb 20, 2026Infusion or Flow Problem

    IT WAS REPORTED THAT DURING A ROTATOR CUFF REPAIR SURGICAL PROCEDURE, IT WAS OBSERVED THAT THE VAPR CLPLSE90 ELECTRODE W HAND CNTRLS COULD NOT SUCTION. ANOTHER DEVICE WAS USED TO COMPLETE THE SURGERY. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION COULD BE PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24405260 ↗

FDA clearance and classification

The FDA regulates product code GEI (electrosurgical, cutting & coagulation & accessories) under 21 CFR 878.4400 as a Class II device. 21 CFR 878.4400 ↗

Devices of this type normally need 510(k) clearance before sale.

Common questions about the VAPR COOLPULSE 228147

Has the VAPR COOLPULSE 228147 been recalled?

No FDA recall record in our data names this exact device identifier (DI 10886705010097), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the VAPR COOLPULSE 228147?

The FDA's MAUDE database lists 298 adverse event reports for this device, received between Aug 28, 2014 and Aug 5, 2026. In those reports, the most common specific device problems are “Infusion or Flow Problem” (79), “Suction Failure” (40) and “Detachment of Device or Device Component” (30); the most common patient problems are “Not Applicable” (15), “Foreign Body In Patient” (7) and “Partial thickness (Second Degree) Burn” (3). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the VAPR COOLPULSE 228147?

None of the reports are classified as deaths, 13 as injuries and 285 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Who makes the VAPR COOLPULSE 228147, and is it still sold?

The FDA's device database lists Depuy Mitek, LLC as the company responsible for it (brand VAPR COOLPULSE, model 228147). Its record shows it as in commercial distribution.