VAPR 228504Recalls & reported problems
Depuy Mitek, LLC · VAPR Premiere50 Electrode with Hand Controls 3.0mm, 50 DEGREES Suction with Integrated Handpiece · Class 2
All models of VAPR (30)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
Quick summary
This DePuy Mitek VAPR Premiere50 electrode is a hand-controlled electrosurgical and suction device used in arthroscopic surgery. The 50 reports mostly describe it failing mid-procedure: buttons not working, sparks, loss of cutting or coagulation, and suction failures sometimes traced to tissue debris or a cut suction tube. Reports describe no deaths and no injuries. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 50
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 50 adverse event reports that list this exact device (DI 10886705020942), received between Sep 3, 2018 and Jun 25, 2026 — classified as 50 malfunctions.
The most frequently reported specific device problem is “Loss of or Failure to Bond” (22 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Loss of or Failure to Bond22
- Electrical /Electronic Property Problem8
- Suction Failure5
- Infusion or Flow Problem3
- Output Problem3
- Break2
- Device Contaminated During Manufacture or Shipping2
- Disconnection2
Patient problems
- No Consequences Or Impact To Patient34
- No Clinical Signs, Symptoms or Conditions14
- Not Applicable2
Recent adverse event reports
The 25 most recent of 50 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 25, 2026Output Problem
IT WAS REPORTED THAT DURING AN UNSPECIFIED SURGERY, THE VAPR DEVICE WORKED INTERMITTENTLY. CHANGED TO ANOTHER VAPR DEVICE TO CONTINUE THE SURGERY BUT THE SAME PROBLEM HAPPENED AGAIN. ANOTHER DEVICE WAS USED TO COMPLETE THE SURGERY. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION WAS PROVIDED. THIS IS REPORT 1 OF 2 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25622396 ↗MalfunctionJun 3, 2026Infusion or Flow Problem; Device Contaminated During Manufacture or Shipping; Output Problem
IT WAS REPORTED THAT DURING A MENISCAL REPAIR SURGICAL PROCEDURE, WHILE USING THE VAPR P50 W/ HAND CONTROLS DEVICE, IT WAS OBSERVED THAT AN ABNORMALITY ALARM WAS FOUND. DURING IN-HOUSE ENGINEERING EVALUATION IT WAS DETERMINED THAT TISSUE DEBRIS WAS IN THE ACTIVE TIP, SALINE CRYSTALLIZATION ON THE TIP AND ESCHAR BUILD UP ON MANIFOLD. IT WAS FURTHER DETERMINED THAT THERE WAS SUCTION FAILURE. ANOTHER DEVICE WAS USED TO COMPLETE THE SURGERY. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION COULD BE PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25379840 ↗MalfunctionJun 2, 2026Electrical /Electronic Property Problem; Melted; Device Contaminated During Manufacture or Shipping; Physical Resistance/Sticking
IT WAS REPORTED THAT DURING AN ARTHROSCOPY HIP GLUTEAL MUSCLE CONTRACTURE SURGICAL PROCEDURE, WHEN USING THE VAPR P50 W/ HAND CONTROLS DEVICE, IT WAS OBSERVED THAT THE BUTTONS FAILED, AND THE DEVICE DID NOT OPERATE WHEN THE BUTTONS WERE DEPRESSED. IT WAS REPORTED THAT ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE. DURING IN-HOUSE ENGINEERING EVALUATION IT WAS DETERMINED THAT THE COAGULATION TEST DID NOT WORK AS INTENDED, AND THE TIP COMPONENTS WERE IN USED CONDITION, WITH SOME RESIDUE AROUND THE MANIFOLD AND THE ACTIVE TIP WAS CHARRED. THERE WERE NO ADVERSE CONSEQUENCES RELATED TO THE PATIENT. NO ADDITIONAL INFORMATION COULD BE PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25363867 ↗MalfunctionJun 30, 2025Device Remains Activated; Physical Resistance/Sticking; Device Remains Activated; Physical Resistance/Sticking
IT WAS REPORTED THAT DURING AN UNKNOWN SURGERY, IT WAS OBSERVED THAT MULTIPLE BUTTON OF THE VAPR P50 W/ HAND CONTROLS DEVICE WERE PRESSED AND THE RF WAND WAS STILL WORKING AND COULD NOT CLOSE. ANOTHER DEVICE WAS USED TO COMPLETE THE SURGERY. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION COULD BE PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22356844 ↗MalfunctionJan 2, 2025Electrical /Electronic Property Problem; Electrical /Electronic Property Problem
REPORT 1 OF 3 FOR (B)(4). IT WAS REPORTED THAT DURING AN UNSPECIFIED SURGICAL PROCEDURE, THE VAPR P50 W/ HAND CONTROLS DEVICE GENERATED SPARKS. ACCORDING TO THE REPORTER, CHANGED TO ANOTHER TWO DEVICES TO CONTINUE THE SURGERY AND THE SAME PROBLEM HAPPENED AGAIN. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO DELAY IN THE PROCEDURE REPORTED. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21053903 ↗MalfunctionOct 22, 2024Break; Electrical /Electronic Property Problem; Use of Device Problem; Break; Electrical /Electronic Property Problem; Use of Device Problem
IT WAS REPORTED THAT DURING A MENISCAL REPAIR PROCEDURE, IT WAS OBSERVED THAT THE VAPR P50 W/ HAND CONTROLS DEVICE DID NOT WORK; AND AN OUTPUT SHORT ERROR APPEARED. DURING IN-HOUSE ENGINEERING EVALUATION, IT WAS DETERMINED THAT THE SUCTION TUBE ON THE DEVICE WAS CUT OFF. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS PATIENT INVOLVEMENT. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20506776 ↗MalfunctionAug 15, 2024Electrical /Electronic Property Problem; Electrical /Electronic Property Problem
IT WAS REPORTED FROM CHINA THAT DURING AN ANTERIOR CRUCIATE LIGAMENT RECONSTRUCTION PROCEDURE, IT WAS OBSERVED THAT THE VAPR PREMIERE50 ELECTRODE WITH HAND CONTROLS 3.0MM, 50 DEGREES SUCTION WITH INTEGRATED HANDPIECE DEVICE COULD NOT CUT (ABLATE). DURING IN-HOUSE ENGINEERING EVALUATION, IT WAS DETERMINED THAT THE DEVICE COULD NOT COAGULATE AS IT FAILED THAT HIPOT ACTIVATE-RETURN ELECTRICAL TESTING AND THE ACTIVATION FUNCTIONAL TEST. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS PATIENT INVOLVEMENT. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19991892 ↗MalfunctionJun 5, 2024Electrical /Electronic Property Problem; Electrical /Electronic Property Problem
THIS IS REPORT 2 OF 4 FOR (B)(4). IT WAS REPORTED FROM CHINA THAT DURING A ROTATOR CUFF REPAIR PROCEDURE, IT WAS OBSERVED THAT THE VAPR P50 W/ HAND CONTROLS DEVICE GENERATED SPARKS; AND THEREFORE, WAS NOT USED ON THE PATIENT. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19467776 ↗MalfunctionJan 12, 2024Device Reprocessing Problem; Output Problem; Device Reprocessing Problem; Output Problem
IT WAS REPORTED BY THE SALES REP IN CHINA THAT DURING AN UNSPECIFIED SURGICAL PROCEDURE, THE VAPR P50 W/ HAND CONTROLS DEVICE AFTER CONNECT WITH GENERATOR, AN ALARM SOUND APPEARS. DURING IN-HOUSE ENGINEERING EVALUATION, IT WAS DETERMINED THAT THE DEVICE HAD INTERMITTENT OPERATION AND FOREIGN SUBSTANCE/DEBRIS/CLEANING. THERE WERE NO ADVERSE PATIENT CONSEQUENCES NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18505885 ↗MalfunctionOct 19, 2023Infusion or Flow Problem; Infusion or Flow Problem
IT WAS REPORTED BY THE SALES REP THAT DURING AN UNSPECIFIED SURGICAL PROCEDURE ON (B)(6)2023 THE VAPR P50 W/ HAND CONTROLS DEVICE HAD NO SUCTION. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO DELAY IN THE PROCEDURE REPORTED. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17966599 ↗MalfunctionJul 6, 2023Electrical /Electronic Property Problem; Electrical /Electronic Property Problem
IT WAS REPORTED BY A HEALTHCARE PROFESSIONAL IN CHINA THAT DURING A KNEE ARTHROSCOPY PROCEDURE ON (B)(6) 2023, IT WAS OBSERVED THAT THE VAPR PREMIERE50 ELECTRODE W/HAND CONTROLS, 3.0MM, 50° SUCTION W/INTEGRATED HANDPIECE DEVICE HAD AN ELECTRICAL SHORT. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE PATIENT CONSEQUENCES NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17270940 ↗MalfunctionApr 19, 2023Infusion or Flow Problem; Infusion or Flow Problem
IT WAS REPORTED BY A HEALTHCARE PROFESSIONAL IN CHINA THAT DURING AN UNKNOWN PROCEDURE ON (B)(6) 2023, IT WAS OBSERVED THAT THE VAPR PREMIERE50 ELECTRODE W/HAND CONTROLS, 3.0MM, 50° SUCTION W/INTEGRATED HANDPIECE DEVICE HAD NO SUCTION. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE PATIENT CONSEQUENCES NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16771379 ↗MalfunctionFeb 22, 2023Electrical /Electronic Property Problem; Electrical /Electronic Property Problem
IT WAS REPORTED BY THE SALES REP IN CHILE THAT DURING AN ANTERIOR CRUCIATE LIGAMENT RECONSTRUCTION PROCEDURE ON AN UNKNOWN DATE, IT WAS OBSERVED THAT THE VAPR PREMIERE50 ELECTRODE W/HAND CONTROLS, 3.0MM, 50° SUCTION W/INTEGRATED HANDPIECE DEVICE WAS DEFICIENT IN COAGULATION. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE WITH A DELAY OF 15 MINUTES. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16420734 ↗MalfunctionFeb 3, 2021Electrical /Electronic Property Problem; Power Problem; Electrical /Electronic Property Problem; Power Problem
IT WAS REPORTED BY THE AFFILIATE IN CHILE THAT DURING A KNEE ARTHROSCOPY PROCEDURE ON (B)(6) 2021, IT WAS OBSERVED THAT THE COAGULATION FUNCTIONALITY ON THE VAPR P50 W/ HAND CONTROLS DEVICE DID NOT WORK EVEN AT MAXIMUM POWER. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE WITH A 10 MINUTE DELAY. THERE WERE NO ADVERSE PATIENT CONSEQUENCES REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11273445 ↗MalfunctionOct 22, 2020Device Damaged Prior to Use; Device Damaged Prior to Use
IT WAS REPORTED BY THE SALES REP THAT ONE VAPR PREMIERE50 ELECTRODE W/HAND CONTROLS WAS DAMAGED UPON RECEIPT. THE DEVICES WERE NOT USED. NO PROCEDURE, OR PATIENT INVOLVEMENT. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10717517 ↗MalfunctionOct 16, 2020Loss of or Failure to Bond; Loss of or Failure to Bond
IT WAS REPORTED THAT DURING THE ACL OPERATION, AFTER CONNECT WITH GENERATOR, DOES NOT FUNCTION AT ALL, THE SCREEN DID NOW SHOW ANY ALERT CODE. ANOTHER DEVICE WAS USED TO COMPLETE THE SURGERY. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION COULD BE PROVIDED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10693421 ↗MalfunctionApr 22, 2020Blocked Connection; Blocked Connection
IT WAS REPORTED BY AFFILIATE VIA EMAIL THAT DURING AN ARTHRO PROCEDURE PERFORMED, THE VAPR P50 W/ HAND CONTROLS WASN'T RECOGNIZED. NO PATIENT CONSEQUENCE. THE DEVICE IS AVAILABLE TO BE RETURNED FOR EVALUATION.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9988444 ↗MalfunctionDec 12, 2019Loss of or Failure to Bond; Loss of or Failure to Bond
IT WAS REPORTED BY AFFILIATE VIA EMAIL THAT DURING A KNEE ARTHROSCOPY THE COAGULATION BUTTON OF THE VAPR P50 W/ HAND CONTROLS DIDN¿T WORKED, IT WAS ALSO TESTED WITH THE PEDAL AND IT DIDN¿T COAGULATE, THE CONSOLE WAS SWITCHED AND THE FAILURE CONTINUED. THE PROCEDURE WAS COMPLETED USING ANOTHER DEVICE. NO PATIENT CONSEQUENCE AND NO SURGICAL DELAY WAS REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9460753 ↗MalfunctionDec 4, 2019Loss of or Failure to Bond; Loss of or Failure to Bond
IT WAS REPORTED BY THE AFFILIATE VIA EMAIL THAT DURING A MENISCECTOMY THE VAPR P50 W/ HAND CONTROLS DIDN¿T RESPOND WHEN COAGULATING, AND THE RADIO FREQUENCY CUT FUNCTION. SURGEON OPENED A NEW VAPR PREMIERE50 ELECTRODE -EA WITH NEW CODE BUT SAME SIZE AND AGAIN IT DIDN¿T COAGULATE. THE PROCEDURE WAS COMPLETED USING A DEVICE FROM A DIFFERENT COMPANY. NO PATIENT CONSEQUENCE AND NO SURGICAL DELAY WAS REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9413439 ↗MalfunctionDec 4, 2019Loss of or Failure to Bond; Loss of or Failure to Bond
IT WAS REPORTED BY THE AFFILIATE IN CHILE THAT DURING A KNEE ARTHROSCOPY, IT WAS OBSERVED THAT THE VAPR P50 W/ HAND CONTROLS DEVICE FAILED TO COAGULATE. AS PER THE AFFILIATE, THE SURGEON TRIED RESOLVING THE ISSUE BY CONNECTING THE PEDAL BUT THE ISSUE PERSISTED. THE SURGEON COMPLETED THE PROCEDURE WITH ANOTHER DEVICE THAT PRESENTED VERY WEEK COAGULATION. ONCE THE PROCEDURE WAS COMPLETED BOTH OF THE RADIO FREQUENCIES WERE TESTED USING ANOTHER CONSOLE BUT THE FAILURE CONTINUED. NO PATIENT CONSEQUENCE AND NO SURGICAL DELAY WAS REPORTED. THERE WAS NO DELAY IN THE SURGICAL PROCEDURE. THERE WAS A SPARE DEVICE AVAILABLE FOR USE TO COMPLETE THE SURGERY. THERE WAS PATIENT INVOLVEMENT. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9413046 ↗MalfunctionNov 25, 2019Loss of or Failure to Bond; Loss of or Failure to Bond
IT WAS REPORTED BY THE AFFILIATE IN (B)(6) THAT DURING A KNEE ARTHROSCOPY, IT WAS OBSERVED THAT THE VAPR P50 W/ HAND CONTROLS DEVICE FAILED TO COAGULATE. AS PER THE AFFILIATE, THE SURGEON TRIED RESOLVING THE ISSUE BY CONNECTING THE PEDAL BUT THE ISSUE PERSISTED. THE SURGEON COMPLETED THE PROCEDURE WITH ANOTHER DEVICE THAT PRESENTED VERY WEEK COAGULATION. ONCE THE PROCEDURE WAS COMPLETED BOTH OF THE RADIO FREQUENCIES WERE TESTED USING ANOTHER CONSOLE BUT THE FAILURE CONTINUED. NO PATIENT CONSEQUENCE AND NO SURGICAL DELAY WAS REPORTED. THERE WAS NO DELAY IN THE SURGICAL PROCEDURE. THERE WAS A SPARE DEVICE AVAILABLE FOR USE TO COMPLETE THE SURGERY. THERE WAS PATIENT INVOLVEMENT. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9370920 ↗MalfunctionNov 19, 2019Loss of or Failure to Bond; Loss of or Failure to Bond
IT WAS REPORTED BY THE AFFILIATE VIA EMAIL THAT DURING AN UNKNOWN PROCEDURE THE VAPR P50 W/ HAND CONTROLS DIDN¿T WORKED. ANOTHER UNIT HAD TO BE USED TO COMPLETE THE PROCEDURE. NO PATIENT CONSEQUENCE AND NO SURGICAL DELAY WAS REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED. ADDITIONAL INFORMATION PROVIDED BY THE AFFILIATE REPORTED THE EXACT FAILURE OF THE ELECTRODE WAS THAT THE HAND PIECE DID NOT FUNCTION AT ALL; WHEN A BUTTON WAS DEPRESSED NOTHING OCCURRED. IT WAS ALSO REPORTED THAT A SMALL SURGICAL DELAY OCCURRED AND THERE WERE NO KNOWN CONSEQUENCES TO THE PATIENT OR USER. THE AFFILIATE REPORTED THE KNEE ARTHROSCOPY PROCEDURE WAS COMPLETED WITH AN ALTERNATIVE HAND PIECE AND ADDITIONAL SURGICAL INTERVENTION WAS NOT REQUIRED. IN HOUSE EVALUATION OF THE RETURNED DEVICE FOUND THAT THE ELECTRODE EXHIBITED NO OUTPUT WHEN THE COAGULATION OR ABLATION FUNCTION WERE ACTIVATED. THE ELECTRODE WAS SENT TO THE SUPPLIER FOR AN IN-DEPTH INVESTIGATION.
Patient problems: Not Applicable
MAUDE report 9344122 ↗MalfunctionOct 31, 2019Loss of or Failure to Bond; Loss of or Failure to Bond
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THE AFFILIATE VIA EMAIL DURING AN UNKNOWN PROCEDURE ON OCTOBER 8, 2019, THE VAPR P50 W/ HAND CONTROLS DIDN¿T COAGULATE AND DIDN¿T ABLATE. THE PROCEDURE WAS COMPLETED USING A REPLACEMENT. THERE WAS SURGICAL DELAY OF APPROXIMATELY FIFTEEN (15) MINUTES. NO PATIENT CONSEQUENCE WAS REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED. THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9263192 ↗MalfunctionOct 28, 2019Loss of or Failure to Bond; Loss of or Failure to Bond
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2019 DURING KNEE ARTHROSCOPY, THE PRODUCT DID NOT CONDUCT ENERGY AT THE TIME OF CUTTING AND COAGULATION. IT WAS REPLACED WITH ANOTHER ONE WITH ANOTHER CODE AND DIFFERENT LOT, WHICH WORKED WITHOUT PROBLEM AND SURGERY WAS BEING ABLE TO PERFORM WITHOUT INCONVENIENCE FOR THE PATIENT AND FOR THE DOCTOR. THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9246717 ↗MalfunctionOct 16, 2019Electrical Power Problem; Electrical Power Problem
DEVICE REPORT FROM SYNTHES MITEK REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED BY THE AFFILIATE VIA PHONE THAT DURING A KNEE ARTHROSCOPY CLEANING THE VAPR P50 W/ HAND CONTROLS DOES NOT FUNCTION. THE PACKING OF THE DEVICE WAS OPENED, CONNECT THE GENERATOR, DOES NOT SHOW ANY ERROR CODE, BUT DID NOT WORK. CHANGED ANOTHER TWO DEVICES, THE EVENT RE-HAPPENED. ANOTHER DEVICE WAS USED TO COMPLETE THE SURGERY. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. THE SURGERY TIME WAS NOT DELAYED. NO ADDITIONAL INFORMATION COULD BE PROVIDED. THIS COMPLAINT IS FOR ONE (1) VAPR PREMIERE 50 ELECTRODE W/HAND CONTROLS, 3.0MM, 50° SUCTION W/INTEGRATED HANDPIECE. THIS REPORT IS 2 OF 2 FOR (B)(4).
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9200557 ↗
Other ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES devices
- Vasoview Hemopro 2 VH-4000Maquet Cardiovascular LLC · Electrosurgical, Cutting & Coagulation & Accessories6 recalls1 death3,232 reports
- LigaSure LF1937Covidien LP · Electrosurgical, cutting & coagulation & accessories3 deaths2,090 reports
- Olympus WB91051WOLYMPUS Winter & Ibe GmbH · Electrosurgical, Cutting & Coagulation & Accessories1 death1,951 reports
- Valleylab FT10 VLFT10GENCovidien LP · Electrosurgical, cutting & coagulation & accessories1 recall5 deaths1,062 reports
- LigaSure LF1837Covidien LP · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES1 recall967 reports
- LigaSure LF2019Covidien LP · Electrosurgical, cutting & coagulation & accessories948 reports
- VAPR 225028Depuy Mitek, LLC · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES895 reports
- LigaSure LF1923Covidien LP · Electrosurgical, cutting & coagulation & accessories855 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page