Medline 0118RRecalls & reported problems
Medline Industries, Inc. · 0118 E-Z CLEAN,MICROFINE NDL ELTD BLU · Class 2
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10888277377950), received between Jul 18, 2025 and Jul 18, 2025 — classified as 1 malfunction.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Detachment of Device or Device Component1
- Failure to Disconnect1
- Human-Device Interface Problem1
Patient problems
- Superficial (First Degree) Burn1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 18, 2025Failure to Disconnect; Detachment of Device or Device Component; Human-Device Interface Problem; Failure to Disconnect; Detachment of Device or Device Component; Human-Device Interface Problem
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE; PATIENT RECEIVED BURNS ON INNER AND OUTER MOUTH FROM CAUTERY. A NEEDLE TIP BOVIE WAS USED AND PATIENT WAS GROUNDED ON LEFT LATERAL THIGH WITH BOVIE PAD. SURGEON NOTICED THAT BOVIE TIP WASN'T PROPERLY INSULATED AND THAT A PART OF THE INSULATED PORTION WAS MISSING, HOWEVER THE BURNS SEEM TO COME FROM THE WHITE INSULATED PART OF THE BOVIE PENCIL THAT WAS RESTING ON THE PATIENT LIP IN BETWEEN USES. IT WOULD SEEM THAT THERE WAS A MALFUNCTION WITH THE BOVIE PENCIL ITSELF BECAUSE UPON CHANGING THE BOVIE TIP THE SAME ISSUE REOCCURRED WITH THE PATIENT BEING BURNED AROUND THE MOUTH. AS A RESULT, A NEW BOVIE PENCIL WAS INSTEAD OPENED AND THE ISSUE WAS RESOLVED. AT THE END OF SURGERY NOTICEABLE BURNS WERE NOTED ON INSIDE AND OUTSIDE MOUTH AREAS. AND SURGEON REQUESTED BACITRACIN OINTMENT TO BE APPLIED BEFORE GOING TO PACU. SURGEON WAS INFORMED THAT THE PATIENT HAD SUSTAINED BURNS ON INNER AND OUTER MOUTH FROM ELECTROCAUTERY PENCIL. THE SURGEON NOTICED THAT THE NEEDLE TIP BOVIE WAS NOT PROPERLY INSULATED AND SOME PARTS WERE MISSING. A NEW PENCIL WAS PROVIDED AND THE ISSUE WAS RESOLVED. THE MALFUNCTIONING PIECES WERE SAVED AND PICTURES WERE TAKEN.
Patient problems: Superficial (First Degree) Burn
MAUDE report 22545936 ↗
Other Electrosurgical, cutting & coagulation & accessories devices
- Vasoview Hemopro 2 VH-4000Maquet Cardiovascular LLC · Electrosurgical, Cutting & Coagulation & Accessories6 recalls1 death3,232 reports
- LigaSure LF1937Covidien LP · Electrosurgical, cutting & coagulation & accessories3 deaths2,090 reports
- Olympus WB91051WOLYMPUS Winter & Ibe GmbH · Electrosurgical, Cutting & Coagulation & Accessories1 death1,951 reports
- Valleylab FT10 VLFT10GENCovidien LP · Electrosurgical, cutting & coagulation & accessories1 recall5 deaths1,062 reports
- LigaSure LF1837Covidien LP · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES1 recall967 reports
- LigaSure LF2019Covidien LP · Electrosurgical, cutting & coagulation & accessories948 reports
- VAPR 225028Depuy Mitek, LLC · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES895 reports
- LigaSure LF1923Covidien LP · Electrosurgical, cutting & coagulation & accessories855 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page