Olympus WA22306DRecalls & reported problems
OLYMPUS Winter & Ibe GmbH · Class 2
Data updated Nov 27, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
Quick summary
This Olympus electrosurgical resection electrode has 117 adverse event reports on file. Many describe the loop wire breaking, bending or fragmenting during set-up or use, and a few mention pieces left in the patient. The reports include no deaths and 21 injuries. There is 1 recall, still open since June 2023, over the loop wire breaking after contact with metal objects such as other endoscopic equipment, implants or stents, which can let fragments fall inside the patient.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 117
- Deaths reported
- 0
- Injuries reported
- 21
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 117 adverse event reports that list this exact device (DI 14042761051689), received between May 15, 2018 and May 14, 2026 — classified as 21 injuries, 96 malfunctions.
The most frequently reported specific device problem is “Break” (55 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Recalls naming this device
- Jun 8, 2023Class IIOpen, Classified
Additional Caution statements in IFU: (1) Due to complaints where the loop wire at the distal end of the HF resection electrode broke after getting in contact with metal objects, such as other endoscopic equipment, implants or stents. As a result, fragments can fall inside the patient and must be retrieved. (2) The second Caution refers to the risk of sparkover when getting in contact with metal parts as this can cause uncontrolled heating of the HF-resection electrode and may result in damage and breakage to the HF-resection electrode tip
Z-2413-2023 · Root cause: Process design · Recall details · FDA record ↗
Most reported problems
Device problems
- Break55
- Material Twisted/Bent22
- Fracture19
- Material Fragmentation19
- Mechanical Problem18
- Material Deformation12
- Detachment of Device or Device Component8
- Melted8
Patient problems
- No Clinical Signs, Symptoms or Conditions94
- No Known Impact Or Consequence To Patient9
- Foreign Body In Patient6
- Device Embedded In Tissue or Plaque4
- Internal Organ Perforation3
- Asystole1
- Burn(s)1
- Ecchymosis1
Recent adverse event reports
The 25 most recent of 117 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 14, 2026Device Damaged Prior to Receipt by User
IT WAS REPORTED THAT THE RESECTION ELECTRODE WAS BENT. THE ISSUE WAS FOUND DURING PREPARATION FOR USE. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25166956 ↗MalfunctionMay 4, 2026Material Deformation
IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE RESECTION ELECTRODE EXHIBITED THE DISTAL END OF THE PRODUCT WAS BENT. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25075900 ↗MalfunctionMay 1, 2026Material Deformation
IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE RESECTION ELECTRODE EXHIBITED THE THE DISTAL END OF THE PRODUCT WAS BENT. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25059644 ↗MalfunctionApr 24, 2026Material Deformation
IT WAS REPORTED THE HF-RESECTION ELECTRODE LOOP WAS DEFORMED. THE DISTAL END WAS BENT. THE ISSUE OCCURRED DURING SET UP FOR A THERAPEUTIC TUR PROCEDURE. THE PROCEDURE WAS COMPLETED USING A SIMILAR DEVICE. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24984014 ↗MalfunctionApr 23, 2026Material Deformation
IT WAS REPORTED THE HF-RESECTION ELECTRODE HAD A DISTORTED DISTAL END OF THE ELECTRODE BEFORE USE. THIS MALFUNCTION OCCURRED DURING INSPECTION FOR USE. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24969070 ↗MalfunctionApr 22, 2026Material Twisted/Bent
IT WAS REPORTED THE HF-RESECTION ELECTRODE LOOP WAS BENT. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24960972 ↗MalfunctionApr 22, 2026Fracture
IT WAS REPORTED THE HF-RESECTION ELECTRODE INSERTION SECTION WAS BROKEN WHEN UNPACKED. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24960920 ↗MalfunctionApr 21, 2026Material Twisted/Bent
IT WAS REPORTED THE RESECTION ELECTRODE APPEARED BENT. THE ISSUE WAS FOUND DURING A THERAPEUTIC PROCEDURE. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24947468 ↗MalfunctionApr 20, 2026Material Twisted/Bent
IT WAS REPORTED THE HF-RESECTION DISTAL END ELECTRODE WAS BENT. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24931704 ↗MalfunctionApr 13, 2026Break
IT WAS REPORTED THE RESECTION ELECTRODE ELECTRIC CUTTING LOOP HAD BROKE OFF. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24865121 ↗MalfunctionApr 7, 2026Material Twisted/Bent
IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE RESECTION ELECTRODE HAD A BENT DISTAL END. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24809152 ↗MalfunctionApr 3, 2026Material Deformation
NO ADDITIONAL INFORMATION PROVIDED BY THE CUSTOMER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24780652 ↗MalfunctionMar 16, 2026Degraded
IT WAS REPORTED THE HF-RESECTION ELECTRODE LOOP'S WIDTH OF THE DISTAL END BECAME NARROW. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24606276 ↗MalfunctionMar 12, 2026Break
IT WAS REPORTED THE RESECTION ELECTRODE LOOP PORTION OF THE LOOP ELECTRODE WAS DAMAGED. THE ISSUE WAS FOUND DURING A TRANSURETHRAL RESECTION, THERAPEUTIC PROCEDURE. THE INTENDED PROCEDURE WAS COMPLETED WITH AN OLYMPUS BACK-UP DEVICE. THERE WERE NO REPORTS OF PATIENT INJURY OR HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24583896 ↗MalfunctionFeb 18, 2026Material Deformation
NO ADDITIONAL INFORMATION RECEIVED FROM CUSTOMER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24378797 ↗MalfunctionFeb 13, 2026Material Twisted/Bent
NO ADDITIONAL INFORMATION RECEIVED FROM CUSTOMER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24339343 ↗MalfunctionFeb 11, 2026Break
IT WAS REPORTED THE HF-RESECTION ELECTRODE, LOOP BROKE DURING THE RESECTION. THE ISSUE WAS FOUND DURING A THERAPEUTIC PROCEDURE. THE PROCEDURE WAS COMPLETED USING A SIMILAR DEVICE. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24318211 ↗MalfunctionFeb 4, 2026Material Twisted/Bent
IT WAS REPORTED THE HF-RESECTION ELECTRODE HAD A BENT ELECTRODE AT THE DISTAL END. THE MALFUNCTION WAS NOTICED DURING INSPECTION FOR USE AND THE PROCEDURE WAS COMPLETED WITH A SIMILAR DEVICE. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24261677 ↗MalfunctionJan 5, 2026Material Integrity Problem
IT WAS REPORTED THAT THE RESECTION ELECTRODE LOOP WAS ATTACHED BACKWARDS. THE ISSUE WAS FOUND DURING INSPECTION FOR USE. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23970198 ↗MalfunctionDec 16, 2025Melted; Melted
NO ADDITIONAL INFORMATION PROVIDED BY THE CUSTOMER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23820552 ↗MalfunctionNov 14, 2025Material Twisted/Bent; Material Twisted/Bent
NO ADDITIONAL INFORMATION RECEIVED FROM THE CUSTOMER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23555076 ↗MalfunctionNov 7, 2025Material Twisted/Bent; Material Twisted/Bent
IT WAS REPORTED THE RESECTION ELECTRODE WAS BENT DURING UNPACKING. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23503628 ↗MalfunctionNov 7, 2025Material Twisted/Bent; Material Twisted/Bent
IT WAS REPORTED THAT THE LOOP WAS BENT ON THE RESECTION ELECTRODE. THE UNSPECIFIC PROCEDURE WAS COMPLETED BY REPLACING IT WITH A SIMILAR DEVICE. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23496848 ↗MalfunctionNov 5, 2025Material Twisted/Bent; Material Twisted/Bent
NO ADDITIONAL INFORMATION PROVIDED BY THE CUSTOMER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23479979 ↗MalfunctionNov 4, 2025Material Twisted/Bent; Material Twisted/Bent
NO ADDITIONAL INFORMATION PROVIDED BY THE INFORMATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23468426 ↗
FDA clearance and classification
The FDA regulates product code GEI (electrosurgical, cutting & coagulation & accessories) under 21 CFR 878.4400 as a Class II device. 21 CFR 878.4400 ↗
Devices of this type normally need 510(k) clearance before sale.
Common questions about the Olympus WA22306D
Has the Olympus WA22306D been recalled?
Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-2413-2023 (Class II), began on Jun 8, 2023: Additional Caution statements in IFU: (1) Due to complaints where the loop wire at the distal end of the HF resection electrode broke after getting in contact with metal objects, such as other endoscopic equipment,… If you use this device, follow the manufacturer's notice and your healthcare provider's advice.
What problems have been reported with the Olympus WA22306D?
The FDA's MAUDE database lists 117 adverse event reports for this device, received between May 15, 2018 and May 14, 2026. In those reports, the most common specific device problems are “Break” (55), “Material Twisted/Bent” (22) and “Fracture” (19); the most common patient problems are “Foreign Body In Patient” (6), “Device Embedded In Tissue or Plaque” (4) and “Internal Organ Perforation” (3). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the Olympus WA22306D?
None of the reports are classified as deaths, 21 as injuries and 96 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Who makes the Olympus WA22306D, and is it still sold?
The FDA's device database lists OLYMPUS Winter & Ibe GmbH as the company responsible for it (brand Olympus, model WA22306D). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Electrosurgical, Cutting & Coagulation & Accessories devices
- Vasoview Hemopro 2 VH-4000Maquet Cardiovascular LLC · Electrosurgical, Cutting & Coagulation & Accessories6 recalls1 death3,137 reports
- LigaSure LF1937Covidien LP · Electrosurgical, cutting & coagulation & accessories3 deaths2,038 reports
- Olympus WB91051WOLYMPUS Winter & Ibe GmbH · Electrosurgical, Cutting & Coagulation & Accessories1 death1,891 reports
- Valleylab FT10 VLFT10GENCovidien LP · Electrosurgical, cutting & coagulation & accessories1 recall5 deaths1,046 reports
- LigaSure LF1837Covidien LP · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES1 recall957 reports
- LigaSure LF2019Covidien LP · Electrosurgical, cutting & coagulation & accessories929 reports
- VAPR 225028Depuy Mitek, LLC · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES874 reports
- LigaSure LF1923Covidien LP · Electrosurgical, cutting & coagulation & accessories831 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026 · each recall links to the FDA recall notice
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page