Olympus WA94002ARecalls & reported problems
OLYMPUS Winter & Ibe GmbH · Single Use Plasma Coagulation Probe Axial Argon PA-211U 2.3 x 2200 mm · Class 2
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 14042761083420), received between Apr 24, 2023 and Sep 17, 2024 — classified as 2 injuries, 1 malfunction.
The most frequently reported specific device problem is “Output Problem” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Output Problem2
- No Device Output1
Patient problems
- Bowel Perforation2
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 17, 2024No Device Output; No Device Output
IT WAS REPORTED, THE PLASMA COAGULATION PROBE OUTPUT DID NOT START WHEN ATTEMPTING TO STOP BLEEDING. THE ISSUE OCCURRED DURING AN UNSPECIFIED PROCEDURE. THE PROCEDURE WAS COMPLETED AFTER SWITCHING TO THE SAME PRODUCT. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20236966 ↗InjuryApr 24, 2023Output Problem; Output Problem
THE CUSTOMER REPORTED TO OLYMPUS THAT DURING AN ATTEMPT TO COAGULATE AN ANGIODYSPLASIA IN THE COECUM USING THIS SINGLE USE PLASMA COAGULATION PROBE, THE PATIENT'S MUSCLES CONTRACTED WHEN THE FOOT SWITCH WAS ACTUATED. THE ARGON PROBE DID NOT IGNITE, AND NO COAGULATION WAS ACHIEVED ON THE SITE. THEN A CHANGE WAS MADE TO A NEW PROBE, AND THE SAME PROBLEMS OCCURRED AT THE SAME SPOT. IT WAS ONLY WHEN A SPOT ABOUT 5 CM AWAY FROM THE ORIGINAL SPOT WAS ACCESSED THAT IGNITION AND COAGULATION OCCURRED. IN THE EVENING, THE PATIENT WAS THEN FOUND TO HAVE A PERFORATION IN THE COECUM AT THE FIRST LOCATION. ADDITIONAL INFORMATION RECEIVED THAT THE PATIENT HAD TO BE OPERATED BECAUSE OF THE PERFORATION BUT IS NOW SYMPTOM-FREE. THE PROBE WAS CHECKED FOR INTACTNESS AND RINSED 2 TIMES BEFORE THE EXAMINATION. NO ABNORMALITIES WERE FOUND. THERE WERE NO REPORTS OF FURTHER PATIENT OR USER HARM ASSOCIATED WITH THIS EVENT. PATIENT IDENTIFIERS: (B)(6) 1ST OF 2 PROBES USED. (B)(6) 2ND OF 2 PROBES USED. THIS REPORT IS FOR (B)(6).
Patient problems: Bowel Perforation
MAUDE report 16802846 ↗InjuryApr 24, 2023Output Problem; Output Problem
THE CUSTOMER REPORTED TO OLYMPUS THAT DURING AN ATTEMPT TO COAGULATE AN ANGIODYSPLASIA IN THE COECUM USING THIS SINGLE USE PLASMA COAGULATION PROBE, THE PATIENT'S MUSCLES CONTRACTED WHEN THE FOOT SWITCH WAS ACTUATED. THE ARGON PROBE DID NOT IGNITE, AND NO COAGULATION WAS ACHIEVED ON THE SITE. THEN A CHANGE WAS MADE TO A NEW PROBE, AND THE SAME PROBLEMS OCCURRED AT THE SAME SPOT. IT WAS ONLY WHEN A SPOT ABOUT 5 CM AWAY FROM THE ORIGINAL SPOT WAS ACCESSED THAT IGNITION AND COAGULATION OCCURRED. IN THE EVENING, THE PATIENT WAS THEN FOUND TO HAVE A PERFORATION IN THE COECUM AT THE FIRST LOCATION. ADDITIONAL INFORMATION RECEIVED THAT THE PATIENT HAD TO BE OPERATED BECAUSE OF THE PERFORATION BUT IS NOW SYMPTOM-FREE. THE PROBE WAS CHECKED FOR INTACTNESS AND RINSED 2 TIMES BEFORE THE EXAMINATION. NO ABNORMALITIES WERE FOUND. THERE WERE NO REPORTS OF FURTHER PATIENT OR USER HARM ASSOCIATED WITH THIS EVENT. PATIENT IDENTIFIERS: (B)(6) 1ST OF 2 PROBES USED. (B)(6) 2ND OF 2 PROBES USED. THIS REPORT IS FOR (B)(6).
Patient problems: Bowel Perforation
MAUDE report 16802656 ↗
Other Electrosurgical, Cutting & Coagulation & Accessories devices
- Vasoview Hemopro 2 VH-4000Maquet Cardiovascular LLC · Electrosurgical, Cutting & Coagulation & Accessories6 recalls1 death3,232 reports
- LigaSure LF1937Covidien LP · Electrosurgical, cutting & coagulation & accessories3 deaths2,090 reports
- Olympus WB91051WOLYMPUS Winter & Ibe GmbH · Electrosurgical, Cutting & Coagulation & Accessories1 death1,951 reports
- Valleylab FT10 VLFT10GENCovidien LP · Electrosurgical, cutting & coagulation & accessories1 recall5 deaths1,062 reports
- LigaSure LF1837Covidien LP · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES1 recall967 reports
- LigaSure LF2019Covidien LP · Electrosurgical, cutting & coagulation & accessories948 reports
- VAPR 225028Depuy Mitek, LLC · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES895 reports
- LigaSure LF1923Covidien LP · Electrosurgical, cutting & coagulation & accessories855 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page