Skip to content

Olympus WB990310Recalls & reported problems

OLYMPUS Winter & Ibe GmbH · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
6
Deaths reported
0
Injuries reported
6
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 14042761083529), received between Sep 13, 2022 and Sep 13, 2022 — classified as 6 injuries.

The most frequently reported specific device problem is “Output above Specifications” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Output above Specifications3
  • Use of Device Problem3

Patient problems

  • Necrosis6

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • InjurySep 13, 2022Use of Device Problem; Use of Device Problem

    IT WAS REPORTED BY AN OLYMPUS TERRITORY MANAGER, THE PATIENT'S TURBINATE WAS BEING OVERTREATED DURING A SUBMUCOSAL RESECTION (SMR) PROCEDURE USING THE ELECTROSURGICAL GENERATOR. IT WAS REPORTED THE OUTPUT FROM THE HANDPIECE WAS HIGH AND THE TISSUE EXPERIENCED NECROSIS. THE CUSTOMER WAS ADVISED TO TURN DOWN THE OUTPUT OF THE DEVICE. THE PROCEDURE WAS COMPLETED WITHOUT DELAY, AND THERE WERE NO REPORTS OF A DEVICE MALFUNCTION. IT WAS REPORTED THE DEVICE WAS INSPECTED PRIOR TO THE PROCEDURE, AND NO ANOMALIES WERE FOUND. FOLLOW-UP WITH THE PHYSICIAN INDICATED THE NECROSIS ISSUE WAS NOTED AT THE FIRST FOLLOW-UP APPOINTMENT AFTER THE PROCEDURE. THE PHYSICIAN REPORTED THREE (3) PATIENTS EXPERIENCED NECROSIS OF THE HEAD AND MID-BODY OF THE INFERIOR TURBINATE. THERE WERE ALLEGEDLY NO CLINICAL SYMPTOMS, BUT THE PHYSICIAN HAD A CONCERN FOR LONG TERM EMPTY NOSE SYNDROME. THE TISSUE NECROSIS WAS TREATED WITH DEBRIDEMENT AND OBSERVATION. THE PHYSICIAN REPORTED 2 DEVICES WERE INVOLVED WITH THE EVENT: WA90002A, SERIAL (B)(4), AND WB990310, LOT 1000084723. THE 3 PATIENTS INVOLVED AND THE CORRESPONDING DEVICES ARE AS FOLLOWS: CASE WITH PATIENT IDENTIFIER (B)(6) REPORTS 75 YEAR OLD ASIAN MALE, WA90002

    Patient problems: Necrosis

    MAUDE report 15414554 ↗
  • InjurySep 13, 2022Output above Specifications

    IT WAS REPORTED BY AN OLYMPUS TERRITORY MANAGER, THE PATIENT'S TURBINATE WAS BEING OVERTREATED DURING A SUBMUCOSAL RESECTION (SMR) PROCEDURE USING THE ELECTROSURGICAL GENERATOR. IT WAS REPORTED THE OUTPUT FROM THE HANDPIECE WAS HIGH AND THE TISSUE EXPERIENCED NECROSIS. THE CUSTOMER WAS ADVISED TO TURN DOWN THE OUTPUT OF THE DEVICE. THE PROCEDURE WAS COMPLETED WITHOUT DELAY, AND THERE WERE NO REPORTS OF A DEVICE MALFUNCTION. IT WAS REPORTED THE DEVICE WAS INSPECTED PRIOR TO THE PROCEDURE, AND NO ANOMALIES WERE FOUND. FOLLOW-UP WITH THE PHYSICIAN INDICATED THE NECROSIS ISSUE WAS NOTED AT THE FIRST FOLLOW-UP APPOINTMENT AFTER THE PROCEDURE. THE PHYSICIAN REPORTED THREE (3) PATIENTS EXPERIENCED NECROSIS OF THE HEAD AND MID-BODY OF THE INFERIOR TURBINATE. THERE WERE ALLEGEDLY NO CLINICAL SYMPTOMS, BUT THE PHYSICIAN HAD A CONCERN FOR LONG TERM EMPTY NOSE SYNDROME. THE TISSUE NECROSIS WAS TREATED WITH DEBRIDEMENT AND OBSERVATION. THE PHYSICIAN REPORTED 2 DEVICES WERE INVOLVED WITH THE EVENT: WA90002A, SERIAL (B)(4), LOT 1000084723. THE 3 PATIENTS INVOLVED AND THE CORRESPONDING DEVICES ARE AS FOLLOWS: CASE WITH PATIENT IDENTIFIER (B)(6) REPORTS 75 YEAR OLD ASIAN MALE, (B)(4), CASE WITH PAT

    Patient problems: Necrosis

    MAUDE report 15414485 ↗
  • InjurySep 13, 2022Use of Device Problem; Use of Device Problem

    IT WAS REPORTED BY AN OLYMPUS TERRITORY MANAGER, THE PATIENT'S TURBINATE WAS BEING OVERTREATED DURING A SUBMUCOSAL RESECTION (SMR) PROCEDURE USING THE ELECTROSURGICAL GENERATOR. IT WAS REPORTED THE OUTPUT FROM THE HANDPIECE WAS HIGH AND THE TISSUE EXPERIENCED NECROSIS. THE CUSTOMER WAS ADVISED TO TURN DOWN THE OUTPUT OF THE DEVICE. THE PROCEDURE WAS COMPLETED WITHOUT DELAY, AND THERE WERE NO REPORTS OF A DEVICE MALFUNCTION. IT WAS REPORTED THE DEVICE WAS INSPECTED PRIOR TO THE PROCEDURE, AND NO ANOMALIES WERE FOUND. FOLLOW-UP WITH THE PHYSICIAN INDICATED THE NECROSIS ISSUE WAS NOTED AT THE FIRST FOLLOW-UP APPOINTMENT AFTER THE PROCEDURE. THE PHYSICIAN REPORTED THREE (3) PATIENTS EXPERIENCED NECROSIS OF THE HEAD AND MID-BODY OF THE INFERIOR TURBINATE. THERE WERE ALLEGEDLY NO CLINICAL SYMPTOMS, BUT THE PHYSICIAN HAD A CONCERN FOR LONG TERM EMPTY NOSE SYNDROME. THE TISSUE NECROSIS WAS TREATED WITH DEBRIDEMENT AND OBSERVATION. THE PHYSICIAN REPORTED 2 DEVICES WERE INVOLVED WITH THE EVENT: WA90002A, SERIAL (B)(4), AND WB990310, LOT 1000084723. THE 3 PATIENTS INVOLVED AND THE CORRESPONDING DEVICES ARE AS FOLLOWS: CASE WITH PATIENT IDENTIFIER (B)(6) REPORTS 75 YEAR OLD ASIAN MALE, WA90002

    Patient problems: Necrosis

    MAUDE report 15414474 ↗
  • InjurySep 13, 2022Output above Specifications

    IT WAS REPORTED BY AN OLYMPUS TERRITORY MANAGER, THE PATIENT'S TURBINATE WAS BEING OVERTREATED DURING A SUBMUCOSAL RESECTION (SMR) PROCEDURE USING THE ELECTROSURGICAL GENERATOR. IT WAS REPORTED THE OUTPUT FROM THE HANDPIECE WAS HIGH AND THE TISSUE EXPERIENCED NECROSIS. THE CUSTOMER WAS ADVISED TO TURN DOWN THE OUTPUT OF THE DEVICE. THE PROCEDURE WAS COMPLETED WITHOUT DELAY, AND THERE WERE NO REPORTS OF A DEVICE MALFUNCTION. IT WAS REPORTED THE DEVICE WAS INSPECTED PRIOR TO THE PROCEDURE, AND NO ANOMALIES WERE FOUND. FOLLOW-UP WITH THE PHYSICIAN INDICATED THE NECROSIS ISSUE WAS NOTED AT THE FIRST FOLLOW-UP APPOINTMENT AFTER THE PROCEDURE. THE PHYSICIAN REPORTED THREE (3) PATIENTS EXPERIENCED NECROSIS OF THE HEAD AND MID-BODY OF THE INFERIOR TURBINATE. THERE WERE ALLEGEDLY NO CLINICAL SYMPTOMS, BUT THE PHYSICIAN HAD A CONCERN FOR LONG TERM EMPTY NOSE SYNDROME. THE TISSUE NECROSIS WAS TREATED WITH DEBRIDEMENT AND OBSERVATION. THE PHYSICIAN REPORTED 2 DEVICES WERE INVOLVED WITH THE EVENT: WA90002A, SERIAL (B)(4), LOT 1000084723. THE 3 PATIENTS INVOLVED AND THE CORRESPONDING DEVICES ARE AS FOLLOWS: CASE WITH PATIENT IDENTIFIER (B)(6) REPORTS 75 YEAR OLD ASIAN MALE, WA90002A, CASE WITH P

    Patient problems: Necrosis

    MAUDE report 15414445 ↗
  • InjurySep 13, 2022Use of Device Problem; Use of Device Problem

    IT WAS REPORTED BY AN OLYMPUS TERRITORY MANAGER, THE PATIENT'S TURBINATE WAS BEING OVERTREATED DURING A SUBMUCOSAL RESECTION (SMR) PROCEDURE USING THE ELECTROSURGICAL GENERATOR. IT WAS REPORTED THE OUTPUT FROM THE HANDPIECE WAS HIGH AND THE TISSUE EXPERIENCED NECROSIS. THE CUSTOMER WAS ADVISED TO TURN DOWN THE OUTPUT OF THE DEVICE. THE PROCEDURE WAS COMPLETED WITHOUT DELAY, AND THERE WERE NO REPORTS OF A DEVICE MALFUNCTION. IT WAS REPORTED THE DEVICE WAS INSPECTED PRIOR TO THE PROCEDURE, AND NO ANOMALIES WERE FOUND. FOLLOW-UP WITH THE PHYSICIAN INDICATED THE NECROSIS ISSUE WAS NOTED AT THE FIRST FOLLOW-UP APPOINTMENT AFTER THE SMR PROCEDURE. THE PHYSICIAN REPORTED THREE (3) PATIENTS EXPERIENCED NECROSIS OF THE HEAD AND MID-BODY OF THE INFERIOR TURBINATE. THERE WERE ALLEGEDLY NO CLINICAL SYMPTOMS, BUT THE PHYSICIAN HAD A CONCERN FOR LONG TERM EMPTY NOSE SYNDROME. THE TISSUE NECROSIS WAS TREATED WITH DEBRIDEMENT AND OBSERVATION. THE PHYSICIAN REPORTED 2 DEVICES WERE INVOLVED WITH THE EVENT: WA90002A, SERIAL (B)(4), AND WB990310, LOT 1000084723. THE 3 PATIENTS INVOLVED AND THE CORRESPONDING DEVICES ARE AS FOLLOWS: CASE WITH PATIENT IDENTIFIER (B)(6) REPORTS 75 YEAR OLD ASIAN MALE, WA9

    Patient problems: Necrosis

    MAUDE report 15414418 ↗
  • InjurySep 13, 2022Output above Specifications

    IT WAS REPORTED BY AN OLYMPUS TERRITORY MANAGER, THE PATIENT'S TURBINATE WAS BEING OVERTREATED DURING A SUBMUCOSAL RESECTION (SMR) PROCEDURE USING THE ELECTROSURGICAL GENERATOR. IT WAS REPORTED THE OUTPUT FROM THE HANDPIECE WAS HIGH AND THE TISSUE EXPERIENCED NECROSIS. THE CUSTOMER WAS ADVISED TO TURN DOWN THE OUTPUT OF THE DEVICE. THE PROCEDURE WAS COMPLETED WITHOUT DELAY, AND THERE WERE NO REPORTS OF A DEVICE MALFUNCTION. IT WAS REPORTED THE DEVICE WAS INSPECTED PRIOR TO THE PROCEDURE, AND NO ANOMALIES WERE FOUND. FOLLOW-UP WITH THE PHYSICIAN INDICATED THE NECROSIS ISSUE WAS NOTED AT THE FIRST FOLLOW-UP APPOINTMENT AFTER THE PROCEDURE. THE PHYSICIAN REPORTED THREE (3) PATIENTS EXPERIENCED NECROSIS OF THE HEAD AND MID-BODY OF THE INFERIOR TURBINATE. THERE WERE ALLEGEDLY NO CLINICAL SYMPTOMS, BUT THE PHYSICIAN HAD A CONCERN FOR LONG TERM EMPTY NOSE SYNDROME. THE TISSUE NECROSIS WAS TREATED WITH DEBRIDEMENT AND OBSERVATION. THE PHYSICIAN REPORTED 2 DEVICES WERE INVOLVED WITH THE EVENT: WA90002A, SERIAL (B)(4), AND WB990310, LOT 1000084723. THE 3 PATIENTS INVOLVED AND THE CORRESPONDING DEVICES ARE AS FOLLOWS: CASE WITH PATIENT IDENTIFIER (B)(6) REPORTS 75 YEAR OLD ASIAN MALE, WA90002

    Patient problems: Necrosis

    MAUDE report 15414366 ↗

Other Electrosurgical, Cutting & Coagulation & Accessories devices

All Electrosurgical, Cutting & Coagulation & Accessories devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page