InMode RF hand piece HP101306ARecalls & reported problems
Inmode M.D Ltd · HPL10D13 InModeRF HP · Class 2
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 4
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 17290016633211), received between Jul 16, 2021 and May 18, 2023 — classified as 4 injuries.
The most frequently reported specific device problem is “Use of Incorrect Control/Treatment Settings” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Use of Incorrect Control/Treatment Settings2
- Adverse Event Without Identified Device or Use Problem1
- Improper or Incorrect Procedure or Method1
- Use of Device Problem1
Patient problems
- Full thickness (Third Degree) Burn1
- Movement Disorder1
- Neuralgia1
- Neuropathy1
- Skin Inflammation/ Irritation1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
InjuryMay 18, 2023Use of Incorrect Control/Treatment Settings; Use of Incorrect Control/Treatment Settings
3RD DEGREE BURN DEVELOPED FOLLOWING TREATMENT WITH FACETITE, LOCATED IN THE CENTRAL ANTERIOR AREA OF THE NECK. THE ABLATIVE LESION, ROUNDED WITH A DIAMETER OF ABOUT 2 CM, APPEARS DECORTICATED OF THE DERMIS. PROPHYLACTIC MEDICAL INTERVENTION REQUIRED TO AVOID SECONDARY INFECTION AND FACILITATE WOUND HEALING. SCAR FORMATION CANNOT BE RULED OUT AT THE MOMENT.
Patient problems: Full thickness (Third Degree) Burn
MAUDE report 16953031 ↗InjuryFeb 17, 2023Use of Device Problem; Use of Device Problem
EVIDENT NEURAPRAXIA IN THE RIGHT LOWER LIP AFTER A FACETITE TREATMENT ON (B)(6) 2021.
Patient problems: Movement Disorder
MAUDE report 16395171 ↗InjuryFeb 17, 2023Use of Incorrect Control/Treatment Settings; Adverse Event Without Identified Device or Use Problem; Use of Incorrect Control/Treatment Settings; Adverse Event Without Identified Device or Use Problem
A TISSUE NECROSIS IN THE SUPRAPUBIC REGION (MONS) FOLLOWING A REDUCTION TREATMENT WITH BODYTITE. SUSPICIOUS FOR ERYSIPELAS AS THE UNDERLYING TISSUE BLANCHED FEW DAYS AFTER. A WEEK LATER, THE WOUNDS EVOLVED AND PSEUDOESCHAR WAS SEEN. IN THE LAST PHOTO THERE IS SIGNIFICANT STRAW-COLORED DRAINAGE, POSSIBLY A SMALL SEROMA DRAINING FROM A FULL THICKNESS WOUND.
Patient problems: Skin Inflammation/ Irritation
MAUDE report 16395050 ↗InjuryJul 16, 2021Improper or Incorrect Procedure or Method
PATIENT RIGHT SIDE OF MOUTH SAGGING POST FACETITE TREATMENT. EVIDENT NEURAPRAXIA IN THE RIGHT LOWER LIP AFTER A FACETITE TREATMENT ON FEBRUARY 11TH. BASED ON OUR PAST EXPERIENCE, IN SUCH CASES, THE NATURE OF THE INJURY IS ALMOST ALWAYS TRANSIENT. DOCTOR (B)(6) REFUSES TO COLLABORATE WITH OUR REPRESENTATIVES AND WOULD NOT FILL IN THE CCQ, NEITHER ACCEPT A TECHNICAL INSPECTION OF THE SYSTEM. MANUFACTURER CASE INVESTIGATION: FURTHERMORE, HE HAS ANNOUNCED THAT HE NO LONGER INTENDS TO USE THIS TECHNOLOGY. THE NEURAPRAXIA THAT OCCURRED IN THIS EVENT MAY HAVE ARISEN FROM AN INVOLUNTARY IMPACT OF THE INSTRUMENT IN THE MANDIBULAR MARGINAL NERVE WHICH IN A CERTAIN POINT ALONG THE MANDIBULAR LINE PROTRUDES IN A SUPERFICIAL POSITION. THIS RESULTS IN A TEMPORARY NEUROPATHY WHICH USUALLY RESOLVES SPONTANEOUSLY WITHIN 2 - 3 MONTHS. IT IS THEREFORE TO BE CONSIDERED A MECHANICAL USER ERROR RELATED TO AN INCORRECT MANEUVER OF THE CANNULA IN THE AREA OF THE PATH OF THE MARGINAL MANDIBULAR NERVE. FDA SAFETY REPORT ID # (B)(4).
Patient problems: Neuralgia; Neuropathy
MAUDE report 12188979 ↗
Other Electrosurgical, Cutting & Coagulation & Accessories devices
- Vasoview Hemopro 2 VH-4000Maquet Cardiovascular LLC · Electrosurgical, Cutting & Coagulation & Accessories6 recalls1 death3,232 reports
- LigaSure LF1937Covidien LP · Electrosurgical, cutting & coagulation & accessories3 deaths2,090 reports
- Olympus WB91051WOLYMPUS Winter & Ibe GmbH · Electrosurgical, Cutting & Coagulation & Accessories1 death1,951 reports
- Valleylab FT10 VLFT10GENCovidien LP · Electrosurgical, cutting & coagulation & accessories1 recall5 deaths1,062 reports
- LigaSure LF1837Covidien LP · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES1 recall967 reports
- LigaSure LF2019Covidien LP · Electrosurgical, cutting & coagulation & accessories948 reports
- VAPR 225028Depuy Mitek, LLC · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES895 reports
- LigaSure LF1923Covidien LP · Electrosurgical, cutting & coagulation & accessories855 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page