SYSTEM 5000 60-8018-SYSRecalls & reported problems
Conmed Corporation · THE SYSTEM 5000 - FULL SYSTEM 100 - 230 VAC 50/60 Hz Electrosurgical Unit · Class 2
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 20653405040979), received between Jan 19, 2024 and Feb 12, 2024 — classified as 1 injury, 1 malfunction.
The most frequently reported specific device problem is “Overheating of Device” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Overheating of Device1
- Unintended Electrical Shock1
Patient problems
- Burn(s)1
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 12, 2024Unintended Electrical Shock; Unintended Electrical Shock
THE CUSTOMER REPORTED THAT THE DEVICE, 60-8018-SYS, THE SYS 5000-FULL, ENG FP, INT, DURING POST-OP ON (B)(6) 2024 WAS REPORTED AS, ¿AFTER CASE WAS FINISHED AND MACHINE WAS UNPLUGGED, SCRUB NURSE TOUCHED THE PRONGS OF THE POWER CABLE AND RECEIVED MILD SHOCK.¿ THERE WAS NO REPORT OF INJURY, MEDICAL INTERVENTION, OR HOSPITALIZATION FOR THE PATIENT. THIS REPORT IS BEING RAISED DUE TO THE REPORTED MALFUNCTION WITH POTENTIAL FOR INJURY UPON REOCCURRENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18683998 ↗InjuryJan 19, 2024Overheating of Device; Overheating of Device
THE SALES REPRESENTATIVE REPORTED ON BEHALF OF THE CUSTOMER THAT 60-8018-SYS, THE SYS 5000-FULL, ENG FP, INT WAS BEING USED ON 24NOV23 AND ¿CAUTERY DEVICE WOULD NOT START CAUTERIZING FOR A LEG VEIN HARVEST. THE BUTTONS REPEATEDLY WOULD NOT FUNCTION. AFTER A TIME TRYING TO USE THE PENCIL, SUDDENLY IT STARTED DELIVERING ENERGY INTENSELY AND CAUSED A .5CM X1CM BURN TO THE PATIENT'S SHIN NEAR THE OPERATIVE SITE. SURGEON AWARE.¿. A GOOD FAITH EFFORT WAS MADE TO OBTAIN FURTHER INFORMATION FROM THE REPORTER; HOWEVER THE REPORTER HAS NOT RESPONDED TO OUR ATTEMPTS FOR INFORMATION. THIS REPORT IS BEING RAISED ON REPORTED INJURY DUE TO A BURN OF UNKNOWN DEGREE TO THE PATIENT.
Patient problems: Burn(s)
MAUDE report 18544162 ↗
Other ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES devices
- Vasoview Hemopro 2 VH-4000Maquet Cardiovascular LLC · Electrosurgical, Cutting & Coagulation & Accessories6 recalls1 death3,232 reports
- LigaSure LF1937Covidien LP · Electrosurgical, cutting & coagulation & accessories3 deaths2,090 reports
- Olympus WB91051WOLYMPUS Winter & Ibe GmbH · Electrosurgical, Cutting & Coagulation & Accessories1 death1,951 reports
- Valleylab FT10 VLFT10GENCovidien LP · Electrosurgical, cutting & coagulation & accessories1 recall5 deaths1,062 reports
- LigaSure LF1837Covidien LP · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES1 recall967 reports
- LigaSure LF2019Covidien LP · Electrosurgical, cutting & coagulation & accessories948 reports
- VAPR 225028Depuy Mitek, LLC · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES895 reports
- LigaSure LF1923Covidien LP · Electrosurgical, cutting & coagulation & accessories855 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page