PlumePen PLPUL2020Recalls & reported problems

Conmed Corporation · PlumePen Ultra Surgical Smoke Evacuation Pencil, 7/8" (22 mm), 10 ft. (3 m) · Class 2

All models of PlumePen (8)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Quick summary

This Conmed PlumePen Ultra is a surgical smoke evacuation pencil that surgeons hold to suction smoke and coagulate tissue during operations. Most of the 53 reports describe the plastic body cracking or breaking apart during surgery, sometimes leaving fragments in the surgical site that had to be removed with forceps, and a few describe it activating without the surgeon pressing the button or small burns on the patient's skin. Seven reports involved injuries and there were no deaths. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
50
Deaths reported
0
Injuries reported
6
Recalls naming it
0

Summary

The FDA's MAUDE database contains 50 adverse event reports that list this exact device (DI 20845854061210), received between Sep 9, 2022 and Jun 4, 2026 — classified as 6 injuries, 44 malfunctions.

The most frequently reported specific device problem is “Material Fragmentation” (30 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 1 reportsMar 2024: 1 reportsApr 2024: 2 reportsMay 2024: 5 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 4 reportsSep 2024: 1 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 3 reportsMay 2025: 2 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 2 reportsDec 2025: 3 reportsJan 2026: 5 reportsFeb 2026: 1 reportsMar 2026: 1 reportsApr 2026: 1 reportsMay 2026: 2 reportsJun 2026: 1 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520265
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Fragmentation30
  • Detachment of Device or Device Component4
  • Adverse Event Without Identified Device or Use Problem3
  • Break3
  • Self-Activation or Keying3
  • Crack2
  • Overheating of Device2
  • Arcing1

Patient problems

  • No Clinical Signs, Symptoms or Conditions40
  • Burn(s)4
  • Foreign Body In Patient4
  • Insufficient Information3
  • Superficial (First Degree) Burn2

Recent adverse event reports

The 25 most recent of 50 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 4, 2026Material Fragmentation

    THE DISTRIBUTOR REPORTED ON BEHALF OF THEIR CUSTOMER THAT THE PLPUL2020 ULTRA SURGICAL SMOKE EVACUATION PENCIL, DEVICE WAS BEING USED ON (B)(6) 2026 DURING A WHIPPLE SURGERY PROCEDURE WHEN IT WAS REPORTED ¿THE BLADE HOLDER COLLAPSE AND TELESCOPIC CAPTURE PORT CRACKED DURING PROCEDURE.¿. FURTHER ASSESSMENT QUESTIONING FOUND THAT THE DEVICE DID FRAGMENT INTO THE SURGICAL SITE AND WAS RETRIEVED BY A FORCEP. THE PROCEDURE WAS COMPLETED AND THERE WAS NO REPORT OF INJURY, MEDICAL INTERVENTION OR EXTENDED HOSPITALIZATION FOR THE PATIENT OR USER. THIS REPORT IS BEING RAISED DUE TO THE REPORTED MALFUNCTION WITH POTENTIAL FOR INJURY UPON REOCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25391515 ↗
  • MalfunctionMay 15, 2026Material Fragmentation

    THE CUSTOMER REPORTED THAT THE PLPUL2020 ULTRA SURGICAL SMOKE EVACUATION PENCIL, DEVICE WAS BEING USED ON (B)(6) 2026 DURING AN UNKNOWN PROCEDURE WHEN IT WAS REPORTED ¿ELECTRODE INSERTION PORT ON PENCIL CRACKED AND ELECTRODE FELL IN WOUND.¿ FURTHER ASSESSMENT QUESTIONING FOUND THAT THE FRAGMENT WAS RETRIEVED FROM THE SURGICAL SITE. THE PROCEDURE WAS COMPLETED WITH THE USE OF THE SAME ALTERNATE DEVICE AND A 5-MINUTE DELAY WAS REPORTED. THERE WAS NO REPORT OF INJURY, MEDICAL INTERVENTION OR EXTENDED STAY FOR THE PATIENT. THIS REPORT IS BEING RAISED DUE TO THE REPORTED MALFUNCTION WITH POTENTIAL FOR INJURY UPON REOCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25182616 ↗
  • MalfunctionMay 7, 2026Material Fragmentation

    A SALES REPRESENTATIVE REPORTED ON BEHALF OF A CUSTOMER THAT THE PLPUL2020, ULTRA SURGICAL SMOKE EVACUATION PENCIL DEVICE WAS USED IN AN UNKNOWN PROCEDURE ON (B)(6) 2026, AND ¿THE ACTIVE ELECTRODE CONNECTION BROKE OFF DURING THE PROCEDURE RENDERING THE PRODUCT UNUSABLE, AND ANOTHER DEVICE HAS TO BE OPENED TO COMPLETE THE PROCEDURE. NO PATIENT WAS INJURED IN THE PROCEDURE, AND THE PROCEDURE WAS COMPLETED WITH ONLY A MINOR DELAY WITH OPENING A NEW DEVICE. THE BROKEN OFF PIECE WAS RECOVERED AND REMOVED FROM THE SURGICAL FIELD.¿. FURTHER ASSESSMENT FOUND "THE FRAGMENT FELL ONTO A SUPERFICIAL INCISION OF THE PATIENT AND WAS EASILY RETRIEVED. IT DID NOT FALL INTO A CAVITY OR INSIDE THE PATIENT. THE HOUSING THAT HOLDS THE PENCIL BLADE/TIP AND THE BLADE ITSELF IS WHAT FELL ONTO THE PATIENT.". THERE WAS NO MEDICAL/SURGICAL INTERVENTION, OR EXTENDED HOSPITALIZATION FOR THE PATIENT. THIS REPORT IS BEING RAISED ON THE BASIS OF MALFUNCTION WITH POTENTIAL FOR INJURY UPON REOCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25110517 ↗
  • MalfunctionApr 3, 2026Material Fragmentation

    THE SALES REPRESENTATIVE REPORTED ON BEHALF OF THE CUSTOMER THAT THE DEVICE, PLPUL2020, ULTRA SURGICAL SMOKE EVACUATION PENCIL, WAS BEING USED DURING AN UNKNOWN PROCEDURE ON (B)(6) 2026 WHEN IT WAS REPORTED, ¿PLUMEPEN ULTRA CRACKED IN THE MIDDLE OF THE PROCEDURE. THE PRODUCT IS AVAILABLE FOR RETURN. ACCOUNT REPORTED THIS HAPPENING TWO OTHER TIMES REPORTING THAT THE CRACK RESULTED IN PLASTIC PIECES FALLING INTO THE PATIENT.¿. THERE WAS NO REPORT OF INJURY, MEDICAL INTERVENTION, OR HOSPITALIZATION FOR THE PATIENT. THE PROCEDURE WAS COMPLETED AS PLANNED. FURTHER ASSESSMENT FOUND THAT ALL FRAGMENTATION WAS RETRIEVED FROM THE PATIENT. THIS REPORT IS BEING RAISED DUE TO THE REPORTED MALFUNCTION WITH POTENTIAL FOR INJURY UPON REOCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24774836 ↗
  • MalfunctionMar 4, 2026Material Fragmentation

    THE SALES REPRESENTATIVE REPORTED ON BEHALF OF THE CUSTOMER THAT THE PLPUL2020 ULTRA SURGICAL SMOKE EVACUATION PENCIL DEVICE WAS BEING USED DURING A LAPAROTOMY PROCEDURE ON (B)(6) 2026 WHEN IT WAS REPORTED ¿A PIECE OF PLASTIC BROKE OFF THE INSIDE OF THE DIATHERMY TIP WHEN CHANGING THE TIP. THE CLEAR PLASTIC WAS FOUND ON THE DRAPES NEAR THE SURGICAL INCISION SITE.¿. THE PROCEDURE WAS COMPLETED WITH THE USE OF AN ALTERNATE COVIDIEN DEVICE AND A 2-MINUTE DELAY. THERE WAS NO REPORT OF INJURY, MEDICAL INTERVENTION, OR EXTENDED HOSPITALIZATION FOR THE PATIENT OR USER. THIS REPORT IS BEING RAISED DUE TO THE REPORTED MALFUNCTION WITH POTENTIAL FOR INJURY UPON REOCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24505213 ↗
  • MalfunctionFeb 6, 2026Material Fragmentation

    A CUSTOMER REPORTED THAT THE PLPUL2020, ULTRA SURGICAL SMOKE EVACUATION PENCIL DEVICE WAS USED IN AN ABDOMINOPLASTY-FLEUR DE LIS PROCEDURE ON (B)(6) 2026, AND ¿THE PLUME PENS ARE STILL BREAKING. THE ONE THEY ARE USING IN THE CURRENT CASE DID THE SAME THING SHE DESCRIBED TO YOU WHEN THE PORTION OF THE PEN INSIDE THAT HOLDS THE TIP INTACT FAILS CAUSING THE TIP TO BE LOOSE AND WIGGLE AROUND INSIDE THE SHEATH¿ IT'S NOT THE SMOKE CAPTURE PORT (SHEATH), IT'S THE PORTION INSIDE OF THAT HOLDS THE TIP THAT BREAKS AND DOESN'T 'HOLD' THE TIP WELL, OR BECOMES LOOSE ALTOGETHER.... IT'S LIKE A PORTION OF WHATEVER HOLDS THE TIP IN BREAKS AND THEN IT CAN'T HOLD THE TIP WELL OR AT ALL. THEY WERE DOING AN ABDOMINOPLASTY-FLEUR DE LIS. SHE SAID IT DID FALL OUT ONTO THE PATIENT." FURTHER ASSESSMENT FOUND "IT WAS ONLY THE TIP THAT FELL OUT ONTO THE PATIENT, NOT THE METAL COMPONENT THAT HOLD THE TIP." THE TIP FELL ONTO - NOT INTO - THE PATIENT, AND WAS REMOVED FROM THE FIELD. THIS REPORT IS BEING RAISED ON THE BASIS OF MALFUNCTION WITH POTENTIAL FOR INJURY UPON REOCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24278276 ↗
  • MalfunctionJan 28, 2026Material Fragmentation

    A CUSTOMER REPORTED THAT THE PLPUL2020, ULTRA SURGICAL SMOKE EVACUATION PENCIL DEVICE WAS USED IN AN UNKNOWN PROCEDURE ON (B)(6) 2025, AND ¿THE PART THAT HOLDS THE ELECTRODE BECOMES WEAK AND BREAKS. SO FAR, THIS HAS HAPPENED TWICE, AND ALL PIECES WERE RECOVERED, BUT IT IS VERY DANGEROUS IF ANY REMAIN INSIDE THE PATIENT (TRANSPARENT PLASTIC MATERIAL).¿. THERE WAS NO REPORT OF INJURY, MEDICAL/SURGICAL INTERVENTION, OR EXTENDED HOSPITALIZATION FOR THE PATIENT. ATTEMPTS WERE MADE TO OBTAIN FURTHER INFORMATION FROM THE CUSTOMER, BUT NO RESPONSE HAS BEEN RECEIVED TO DATE. THIS REPORT IS BEING RAISED ON THE BASIS OF MALFUNCTION WITH POTENTIAL FOR INJURY UPON REOCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24198430 ↗
  • MalfunctionJan 28, 2026Material Fragmentation

    A CUSTOMER REPORTED THAT THE PLPUL2020, ULTRA SURGICAL SMOKE EVACUATION PENCIL DEVICE WAS USED IN AN UNKNOWN PROCEDURE ON AN UKNOWN DATE, AND ¿THE PART THAT HOLDS THE ELECTRODE BECOMES WEAK AND BREAKS. SO FAR, THIS HAS HAPPENED TWICE, AND ALL PIECES WERE RECOVERED, BUT IT IS VERY DANGEROUS IF ANY REMAIN INSIDE THE PATIENT (TRANSPARENT PLASTIC MATERIAL).¿. THERE WAS NO REPORT OF INJURY, MEDICAL/SURGICAL INTERVENTION, OR EXTENDED HOSPITALIZATION FOR THE PATIENT. ATTEMPTS WERE MADE TO OBTAIN FURTHER INFORMATION FROM THE CUSTOMER, BUT NO RESPONSE HAS BEEN RECEIVED TO DATE. THIS REPORT IS BEING RAISED ON THE BASIS OF MALFUNCTION WITH POTENTIAL FOR INJURY UPON REOCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24198429 ↗
  • MalfunctionJan 14, 2026Material Fragmentation

    A SALES REPRESENTATIVE REPORTED ON BEHALF OF A CUSTOMER THAT THE PLPUL2020, ULTRA SURGICAL SMOKE EVACUATION PENCIL DEVICE WAS USED IN AN UNKNOWN PROCEDURE ON (B)(6) 2025, AND ¿A PLASTIC PART OF THE CANNULA BROKE DURING THE PROCEDURE. ALL PARTS WERE RETRIEVED BY SURGEONS AND NURSES. THE PROCEDURE AND HOW THE CANNULA BROKE ARE UNKNOWN, UNAVAILABLE.¿ THE PROCEDURE WAS COMPLETED WITH THE USE OF AN ALTERNATE DEVICE AND A DELAY OF 2 MINUTES. FURTHER ASSESSMENT FOUND THE DEVICE HAD FRAGMENTED AND FALLEN INTO THE PATIENT AND WAS EASILY FOUND AND RETRIEVED. THERE WAS NO REPORT OF INJURY, MEDICAL/SURGICAL INTERVENTION, OR EXTENDED HOSPITALIZATION FOR THE PATIENT. THIS REPORT IS BEING RAISED ON THE BASIS OF MALFUNCTION WITH POTENTIAL FOR INJURY UPON REOCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24066387 ↗
  • MalfunctionJan 13, 2026Material Fragmentation

    THE CUSTOMER REPORTED THAT THE THAT THE PLPUL2020 ULTRA SURGICAL SMOKE EVACUATION PENCIL, DEVICE WAS BEING USED ON 02DEC25 DURING A WHIPPLE PROCEDURE WHEN IT WAS REPORTED ¿THE BLADE HOLDER COLLAPSE AND TELESCOPIC CAPTURE PORT MELTED DURING PROCEDURE.¿. FURTHER ASSESSMENT QUESTIONING FOUND THE DEVICE DID FRAGMENT INTO THE SURGICAL SITE AND WAS REMOVED WITH THE USE OF FORCEPS. THE PROCEDURE WAS COMPLETED WITH THE USE OF THE SAME ALTERNATE DEVICE AND A 15-MINUTE DELAY. THERE WAS NO REPORT OF INJURY, MEDICAL INTERVENTION OR EXTENDED HOSPITALIZATION FOR THE PATIENT OR USER. THIS REPORT IS BEING RAISED DUE TO THE REPORTED MALFUNCTION WITH POTENTIAL FOR INJURY UPON REOCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24055485 ↗
  • MalfunctionJan 8, 2026Material Fragmentation

    THE SALES REPRESENTATIVE REPORTED ON BEHALF OF THE CUSTOMER THAT THE PLPUL2020 ULTRA SURGICAL SMOKE EVACUATION PENCIL DEVICE WAS FOUND DURING A NON-SURGICAL EVENT ON 16DEC25 WHEN IT WAS REPORTED ¿PIN LOOSE IN PACKAGE.¿. FURTHER ASSESSMENT QUESTIONING FOUND THAT THE DEVICE¿S ELECTRODE WAS LOOSE IN THE PACKAGE. THERE WAS NO REPORT OF INJURY, MEDICAL INTERVENTION, OR EXTENDED HOSPITALIZATION FOR THE PATIENT OR USER. THIS REPORT IS BEING RAISED DUE TO THE REPORTED MALFUNCTION WITH POTENTIAL FOR INJURY UPON REOCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24002325 ↗
  • MalfunctionDec 19, 2025Material Fragmentation; Material Fragmentation

    THE CUSTOMER REPORTED THAT THE DEVICE, PLPUL2020, ULTRA SURGICAL SMOKE EVACUATION PENCIL, WAS BEING USED DURING A GYN PROCEDURE ON (B)(6) 2025 WHEN IT WAS REPORTED, ¿DURING A CASE WITH DR. VIAU IN GYNECOLOGY, THE PLUMEPEN ULTRA BROKE WHEN HE TRIED TO UNCLIP IT TO LENGTHEN IT. A PIECE FELL INTO THE PATIENT, AND WAS LATER FOUND.". THERE WAS NO REPORT OF INJURY, MEDICAL INTERVENTION, OR HOSPITALIZATION FOR THE PATIENT. THE PROCEDURE WAS REPORTED TO HAVE BEEN COMPLETED USING AN ALTERNATE DEVICE WHICH CAUSED A 5-MINUTE DELAY. THIS REPORT IS BEING RAISED DUE TO THE REPORTED MALFUNCTION WITH POTENTIAL FOR INJURY UPON REOCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23848202 ↗
  • MalfunctionDec 16, 2025Material Fragmentation; Material Fragmentation

    THE SALES REPRESENTATIVE REPORTED ON BEHALF OF THE CUSTOMER THAT THE PLPUL2020 ULTRA SURGICAL SMOKE EVACUATION PENCIL DEVICE WAS BEING USED DURING A BREAST AUGMENTATION PROCEDURE ON (B)(6) 2025 WHEN IT WAS REPORTED ¿PLPUL2020 PLASTIC ELECTRODE HOLDER SNAPPED DURING USE.¿ FURTHER ASSESSMENT QUESTIONING FOUND THAT ¿NO FRAGMENTS WERE REPORTED TO HAVE FALLEN INTO SURGICAL SITE. THE ISSUE OCCURRED WHEN CLEANING THE TIP OF AN ELECTRODE, THIS ELECTRODE WAS NOT THE ELECTRODE THAT THE PLPUL2020 HAS FITTED WHEN PACKAGED. THE TIP WAS CHANGED DURING THE PROCEDURE. THE PLASTIC HOUSING THAT SECURES THE TIP WAS REPORTED TO HAVE BEEN "BROKEN.¿ THE PROCEDURE WAS COMPLETED WITHOUT THE USE OF AN ALTERNATE DEVICE AND THERE WAS A 5-MINUTE DELAY REPORTED. THERE WAS NO REPORT OF INJURY, MEDICAL INTERVENTION, OR EXTENDED HOSPITALIZATION FOR THE PATIENT OR USER. THIS REPORT IS BEING RAISED DUE TO THE REPORTED MALFUNCTION WITH POTENTIAL FOR INJURY UPON REOCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23815323 ↗
  • InjuryDec 12, 2025Material Fragmentation; Material Fragmentation

    THE SALES REPRESENTATIVE REPORTED ON BEHALF OF THE CUSTOMER THAT THE DEVICE, PLPUL2020, ULTRA SURGICAL SMOKE EVACUATION PENCIL, WAS BEING USED DURING AN OPEN LIVER PROCEDURE ON (B)(6) 2025 WHEN IT WAS REPORTED, ¿THE INNER PLASTIC OF THE PENCIL (WHERE THE DIATHERMY TIP ATTACHES TO THE HANDPIECE) SNAPPED WHILE ATTACHING THE TIP.¿. FURTHER ASSESSMENT WAS REQUESTED FROM THE REPORTER. THE STATEMENT RECEIVED READ, ¿I KNOW FOR CERTAIN THE GUARDED ELECTRODE TIP FELL INTO THE ABDOMEN AND WAS RETRIEVED PROMPTLY BY HAND. I CANNOT SAY FOR CERTAIN IF ANY OTHER PIECE OF PLASTIC FROM THE PLUME PEN FELL INTO THE CAVITY OR NOT. IF IT DID, THIS WAS NOT RETRIEVED.¿ THERE WAS NO REPORT OF MEDICAL INTERVENTION OR HOSPITALIZATION FOR THE PATIENT. THE PROCEDURE WAS REPORTED AS HAVING BEEN COMPLETED AS PLANNED. THIS REPORT IS BEING RAISED DUE TO THE REPORTED INJURY OF POSSIBLE FOREIGN BODY LEFT IN PATIENT.

    Patient problems: Foreign Body In Patient

    MAUDE report 23788098 ↗
  • MalfunctionNov 18, 2025Material Fragmentation; Material Fragmentation

    THE SALES REPRESENTATIVE REPORTED ON BEHALF OF THE CUSTOMER THAT THE PLPUL2020, ULTRA SURGICAL SMOKE EVACUATION PENCIL WAS BEING USED ON (B)(6) 2025 AND ¿¿USING THE ULTRA AND THE PIECE INSIDE THE ULTRA BROKE OFF WHEN SWAPPING OUT THE ELECTRODE¿" THERE WAS NO REPORT OF IMPACT OR INJURY TO THE PATIENT. THIS REPORT IS BEING RAISED DUE TO THE REPORTED MALFUNCTION WITH POTENTIAL FOR INJURY UPON REOCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23580551 ↗
  • MalfunctionNov 18, 2025Material Fragmentation; Material Fragmentation

    THE SALES REPRESENTATIVE REPORTED ON BEHALF OF THE CUSTOMER THAT THE PLPUL2020, ULTRA SURGICAL SMOKE EVACUATION PENCIL WAS BEING USED ON (B)(6) 2025 AND ¿¿USING THE ULTRA AND THE PIECE INSIDE THE ULTRA BROKE OFF WHEN SWAPPING OUT THE ELECTRODE¿" THERE WAS NO REPORT OF IMPACT OR INJURY TO THE PATIENT. THIS REPORT IS BEING RAISED DUE TO THE REPORTED MALFUNCTION WITH POTENTIAL FOR INJURY UPON REOCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23580546 ↗
  • InjuryJun 11, 2025Material Fragmentation; Material Fragmentation

    THE SALES REPRESENTATIVE REPORTED ON BEHALF OF THE CUSTOMER THAT THE DEVICE, PLPUL2020, ULTRA SURGICAL SMOKE EVACUATION PENCIL, WAS BEING USED DURING AN UNKNOWN PROCEDURE ON (B)(6) 2025 WHEN IT WAS REPORTED, ¿THE INNER PIECE CRACKED AND PIECES BROKE OFF AND HAD TO BE LOCATED INSIDE THE PATIENT.¿. AN ALTERNATE SAME DEVICE WAS USED TO COMPLETE THE PROCEDURE CAUSING A DELAY OF UNKNOWN MINUTES. THERE WAS NO REPORT OF INJURY, MEDICAL INTERVENTION, OR HOSPITALIZATION FOR THE PATIENT. THE PATIENT¿S STATUS WAS REPORTED AS ¿GOOD¿. A GOOD FAITH EFFORT WAS CONDUCTED BY CONMED TO DETERMINE IF ALL FRAGMENTATION HAD BEEN RETRIEVED; HOWEVER, TO DATE WE HAVE NOT RECEIVED A RESPONSE. THIS REPORT IS BEING RAISED DUE TO THE REPORTED INJURY OF POSSIBLE FOREIGN BODY LEFT IN PATIENT.

    Patient problems: Foreign Body In Patient

    MAUDE report 22182856 ↗
  • MalfunctionMay 6, 2025Material Fragmentation; Material Fragmentation

    THE CUSTOMER REPORTED THAT THE PLPUL2020, ULTRA SURGICAL SMOKE EVACUATION PENCIL, DEVICE WAS BEING USED ON (B)(6) 2025 DURING A BBR PROCEDURE WHEN IT WAS REPORTED ¿TOWARDS THE END OF A LONG CASE, THE PLASTIC COMPONENT THAT HOLDS THE ELECTRODE HAS CRACKED AND BROKEN, WITH SHARDS OF PLASTIC FOUND WITHIN THE WOUND. THE ELECTRODE IS PLACED IN THE HOLDER AT THE BEGINNING OF THE CASE AND IS NOT CHANGED OR REMOVED.¿. FURTHER ASSESSMENT FOUND THAT THE FRAGMENT WAS REMOVED FROM THE WOUND. THE PROCEDURE WAS COMPLETED WITH THE USE OF THE SAME ALTERNATE DEVICE AND A 10-MINUTE DELAY. THERE WAS NO REPORT OF INJURY, MEDICAL INTERVENTION, OR EXTENDED HOSPITALIZATION FOR THE PATIENT OR USER. THIS REPORT IS BEING RAISED ON THE REPORTED MALFUNCTION WITH POTENTIAL FOR INJURY UPON REOCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21969250 ↗
  • MalfunctionMay 6, 2025Material Fragmentation; Material Fragmentation

    THE CUSTOMER REPORTED THAT THE PLPUL2020, ULTRA SURGICAL SMOKE EVACUATION PENCIL, DEVICE WAS BEING USED ON (B)(6) 2025 DURING AN ABDOMINOPLASTY & MASTOPEXY PROCEDURE WHEN IT WAS REPORTED ¿TOWARDS THE END OF A LONG CASE, THE PLASTIC COMPONENT THAT HOLDS THE ELECTRODE HAS CRACKED AND BROKEN, WITH SHARDS OF PLASTIC FOUND WITHIN THE WOUND. THE ELECTRODE IS PLACED IN THE HOLDER AT THE BEGINNING OF THE CASE AND IS NOT CHANGED OR REMOVED.¿ FURTHER ASSESSMENT FOUND THAT THE FRAGMENT WAS REMOVED FROM THE WOUND. THE PROCEDURE WAS COMPLETED WITH THE USE OF THE SAME ALTERNATE DEVICE AND A 10-MINUTE DELAY. THERE WAS NO REPORT OF INJURY, MEDICAL INTERVENTION, OR EXTENDED HOSPITALIZATION FOR THE PATIENT OR USER. THIS REPORT IS BEING RAISED ON THE REPORTED MALFUNCTION WITH POTENTIAL FOR INJURY UPON REOCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21969248 ↗
  • MalfunctionApr 24, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    A SALES REPRESENTATIVE REPORTED ON BEHALF OF A CUSTOMER THAT THE PLPUL2020, ULTRA SURGICAL SMOKE EVACUATION PENCIL DEVICE WAS USED IN PEDIATRIC AND ADULT HERNIA PROCEDURES ON (B)(6) 2025, AND "ON TWO SEPARATE DAYS, WITH TWO DIFFERENT GENERAL SURGERY SURGEONS, THEY BURNED THEMSELVES USING THE CAUTERY. THE SURGEONS CHECKED FOR ANY HOLES IN THE GLOVES". FURTHER ASSESSMENT FOUND THE ELECTRODE CAUSED A "LIGHT BURN...ON THE FINGERS", AND "THE SURGEONS CONFIRMED THERE WERE NO HOLES IN THE GLOVES". THE PROCEDURE WAS COMPLETED "WITHOUT PROBLEM" AND WITH NO DELAY, USING "ANOTHER OF THE SAME CAUTERY". THERE WAS NO MEDICAL OR SURGICAL INTERVENTION REQUIRED FOR THE BURN, AND NO IMPACT TO THE PATIENT OR OTHERS. THERE WAS NO ALLEGATION OF A DEVICE MALFUNCTION. THIS REPORT IS BEING RAISED ON THE BASIS OF MALFUNCTION WITH POTENTIAL FOR INJURY UPON REOCCURRENCE.

    Patient problems: Superficial (First Degree) Burn

    MAUDE report 21899110 ↗
  • MalfunctionApr 24, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    A SALES REPRESENTATIVE REPORTED ON BEHALF OF A CUSTOMER THAT THE PLPUL2020, ULTRA SURGICAL SMOKE EVACUATION PENCIL DEVICE WAS USED IN PEDIATRIC AND ADULT HERNIA PROCEDURES ON APPROXIMATELY ON (B)(6) 2025, AND "ON TWO SEPARATE DAYS, WITH TWO DIFFERENT GENERAL SURGERY SURGEONS, THEY BURNED THEMSELVES USING THE CAUTERY. THE SURGEONS CHECKED FOR ANY HOLES IN THE GLOVES¿. FURTHER ASSESSMENT FOUND THE ELECTRODE CAUSED A "LIGHT BURN...ON THE FINGERS", AND "THE SURGEONS CONFIRMED THERE WERE NO HOLES IN THE GLOVES". THE PROCEDURE WAS COMPLETED "WITHOUT PROBLEM" AND WITH NO DELAY, USING "ANOTHER OF THE SAME CAUTERY". THERE WAS NO MEDICAL OR SURGICAL INTERVENTION REQUIRED FOR THE BURN, AND NO IMPACT TO THE PATIENT OR OTHERS. THERE WAS NO ALLEGATION OF A DEVICE MALFUNCTION. THIS REPORT IS BEING RAISED ON THE BASIS OF MALFUNCTION WITH POTENTIAL FOR INJURY UPON REOCCURRENCE.

    Patient problems: Superficial (First Degree) Burn

    MAUDE report 21899054 ↗
  • InjuryApr 1, 2025Self-Activation or Keying; Self-Activation or Keying

    THE SALES REPRESENTATIVE REPORTED ON BEHALF OF THE CUSTOMER THAT THE PLPUL2020, ULTRA SURGICAL SMOKE EVACUATION PENCIL WAS BEING USED ON (B)(6) 2025 AND THERE WAS ¿SIGNIFICANT INTERFERENCE GENERATED BY THE ELECTROCAUTERY, PREVENTING THE EMG (EVOKED POTENTIAL) DEVICE FROM FUNCTIONING PROPERLY. CONSEQUENCE: LOSS OF CONFIDENCE IN THE EMG DEVICE, PATIENT HAD TO BE AWAKENED QUICKLY TO CHECK FOR ANY IMPAIRMENTS. RESULT: SLIGHT IMPAIRMENTS, WHICH MAY BE TEMPORARY OR PERMANENT. OCCURRED AT THE END OF A SCOLIOSIS CASE, DURING CLOSURE. GENERATOR USED FT10. GENERATOR FT10 WAS ACTIVATED FOR 5 MINUTES WITHOUT NOTICING WHICH INTERFERED WITH ECG READING, ACCOUNT IS WONDERING IF THERE WAS A MALFUNCTION WITH ELECTROSURGICAL PEN AS THAT ACTIVATED ITSELF OR IF THE STAFF ACTIVATED IT WITHOUT NOTICING BECAUSE THE ENERGY ALARM SOUND WAS TOO LOW ON THE GENERATOR.¿. THE PROCEDURE WAS COMPLETED WITHOUT THE USE OF AN ALTERNATE DEVICE. THERE WAS A DELAY OF ¿30+ MINUTES¿. FURTHER INFORMATION HAS BEEN REQUESTED; HOWEVER, TO DATE, A REPLY HAS NOT BEEN RECEIVED. THIS REPORT IS BEING RAISED DUE TO THE REPORTED INJURY OF SLIGHT IMPAIRMENTS.

    Patient problems: Insufficient Information

    MAUDE report 21737687 ↗
  • MalfunctionOct 28, 2024Material Fragmentation; Material Fragmentation

    THE SALES REPRESENTATIVE REPORTED ON BEHALF OF THE CUSTOMER THAT THE DEVICE, PLPUL2020, ULTRA SURGICAL SMOKE EVACUATION PENCIL, WAS BEING USED DURING A TOTAL KNEE PROCEDURE ON (B)(6) 2024 WHEN IT WAS REPORTED, ¿SMOKE PENCIL PLASTIC ENCAPSULATION TIP BROKE DURING THE CASE. SHARDS ENDED UP IN PATIENT. ALL REMOVED.". THERE WAS NO REPORT OF INJURY, MEDICAL INTERVENTION, OR HOSPITALIZATION FOR THE PATIENT. THE PROCEDURE WAS COMPLETED WITH AN ALTERNATE DEVICE CAUSING A 10-MINUTE DELAY. THE PATIENT RECEIVED ADDITIONAL WASHINGS. THIS REPORT IS BEING RAISED DUE TO THE REPORTED MALFUNCTION WITH POTENTIAL FOR INJURY UPON REOCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20545258 ↗
  • MalfunctionSep 12, 2024Detachment of Device or Device Component; Detachment of Device or Device Component

    THE SALES REPRESENTATIVE REPORTED ON BEHALF OF THE CUSTOMER THAT THE PLPUL2020, ULTRA SURGICAL SMOKE EVACUATION PENCIL WAS BEING USED DURING A RIGHT KNEE ARTHROPLASTY ON (B)(6) 2024 AND ¿DURING A RIGHT TOTAL KNEE ARTHROPLASTY THE BOVIE TIP ON AN ULTRA FELL INTO THE PATIENT. THE SURGEON HAD TO STOP THE CASE AND WAS ABLE TO FIND THE BOVIE TIP. A NEW PEN WAS BROUGHT IN TO FINISH THE CASE¿".THERE WAS NO IMPACT OR INJURY TO THE PATIENT OR USER. THE PROCEDURE WAS COMPLETED WITH A NEW PLUMEPEN ULTRA AND A 10 MINUTE DELAY. THE PATIENT WAS REPORTED AS ¿STABLE¿. THIS REPORT IS BEING RAISED DUE TO THE REPORTED MALFUNCTION WITH POTENTIAL FOR INJURY UPON REOCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20201757 ↗
  • InjuryAug 26, 2024Device Remains Activated; Device Remains Activated

    A DISTRIBUTOR REPORTED ON BEHALF OF A CUSTOMER THAT THE PLPUL2020, 20/CA ULTRA NONSTICK 10FT DEVICE WAS USED DURING A MINOR LAP PROCEDURE ON AN UNKNOWN DATE AND ¿COMPONENT STAYED ON AFTER SURGEON RELEASED THE ACTIVATION BUTTON AND CAUSED A BURN TO THE PATIENT¿S SKIN¿. FOLLOW-UP ASSESSMENT QUESTIONS WERE ASKED, AND THE DISTRIBUTOR STATED THERE WAS ¿NO FURTHER INFORMATION TO PROVIDE¿. THIS REPORT IS BEING RAISED DUE TO THE REPORTED INJURY OF A BURN OF UNKNOWN DEGREE TO THE PATIENT.

    Patient problems: Burn(s)

    MAUDE report 20070508 ↗

FDA clearance and classification

The FDA regulates product code GEI (electrosurgical, cutting & coagulation & accessories) under 21 CFR 878.4400 as a Class II device. 21 CFR 878.4400 ↗

Devices of this type normally need 510(k) clearance before sale.

Common questions about the PlumePen PLPUL2020

Has the PlumePen PLPUL2020 been recalled?

No FDA recall record in our data names this exact device identifier (DI 20845854061210), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the PlumePen PLPUL2020?

The FDA's MAUDE database lists 50 adverse event reports for this device, received between Sep 9, 2022 and Jun 4, 2026. In those reports, the most common specific device problems are “Material Fragmentation” (30), “Detachment of Device or Device Component” (4) and “Break” (3); the most common patient problems are “Burn(s)” (4), “Foreign Body In Patient” (4) and “Superficial (First Degree) Burn” (2). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the PlumePen PLPUL2020?

None of the reports are classified as deaths, 6 as injuries and 44 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Who makes the PlumePen PLPUL2020, and is it still sold?

The FDA's device database lists Conmed Corporation as the company responsible for it (brand PlumePen, model PLPUL2020). Its record shows it as in commercial distribution.