PlumePen PLPUL2520Recalls & reported problems

Conmed Corporation · PlumePen Ultra Surgical Smoke Evacuation Pencil, 7/8" (22 mm), 15 ft. (4.6 m) · Class 2

All models of PlumePen (8)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 20845854061227), received between Oct 26, 2022 and Sep 15, 2025 — classified as 1 injury, 3 malfunctions.

The most frequently reported specific device problem is “Material Fragmentation” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Fragmentation4

Patient problems

  • No Clinical Signs, Symptoms or Conditions3
  • Foreign Body In Patient1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 15, 2025Material Fragmentation; Material Fragmentation

    THE CUSTOMER REPORTED THAT THE DEVICE, PLPUL2520, ULTRA SURGICAL SMOKE EVACUATION PENCIL WAS BEING USED ON (B)(6) 2025 AND THE "CAUTERY PENCIL WAS BEING USED INSIDE OPEN CHEST. CAUTERY STOPPED WORKING FOR A MOMENT AND WHEN EXAMINED A SMALL PIECE OF PLASTIC WHERE CAUTERY TIP SITS BROKE OFF AND FELL INTO OPEN CAVITY. SMALL PIECE WAS FOUND.¿ THERE WAS NO IMPACT OR INJURY TO THE PATIENT OR USER. THIS REPORT IS BEING RAISED DUE TO THE REPORTED MALFUNCTION WITH POTENTIAL FOR INJURY UPON REOCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23050377 ↗
  • MalfunctionJul 25, 2025Material Fragmentation; Material Fragmentation

    A CUSTOMER REPORTED THAT THE PLPUL2520, ULTRA SURGICAL SMOKE EVACUATION PENCIL DEVICE WAS USED IN AN OGD & TOTAL GASTRECTOMY PROCEDURE ON (B)(6) 2025, AND ¿THE PLASTIC PART OF THE DIATHERMY TIP EXTENSION WAS BROKEN WHILST USING FOR SURGERY¿. FURTHER ASSESSMENT FOUND THAT, "ON THE PLUME-PEN, THERE IS A PLASTIC ATTACHMENT CONTINUING FROM THE DIATHERMY TIP TO THE PEN, THIS PLASTIC CONNECTION WAS DISPLACED/ BROKEN/ SEPARATED/ NOT IN CONTINUITY. WHEN NOTICED, THE BROKEN PIECE WAS STILL INSIDE THE PEN. THE SITE WAS THOROUGH INSPECTED AND NOTHING FOUND. MALFUNCTIONED INSTRUMENT HAD NO EFFECT ON PATIENT¿S SURGICAL OUTCOME". THE PROCEDURE WAS COMPLETED AS NORMAL WITH AN ALTERNATE SAME DEVICE AND A DELAY OF TWO MINUTES. THERE WAS NO REPORT OF INJURY, MEDICAL/SURGICAL INTERVENTION, OR EXTENDED HOSPITALIZATION FOR THE PATIENT. THIS REPORT IS BEING RAISED ON THE BASIS OF MALFUNCTION WITH POTENTIAL FOR INJURY UPON REOCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22608454 ↗
  • InjuryJun 11, 2025Material Fragmentation; Material Fragmentation

    THE SALES REPRESENTATIVE REPORTED ON BEHALF OF THE CUSTOMER THAT THE DEVICE, PLPUL2520, ULTRA SURGICAL SMOKE EVACUATION PENCIL, WAS BEING USED DURING AN OPEN-HEART PROCEDURE ON (B)(6) 2025 WHEN IT WAS REPORTED, ¿BLADE AND HOUSING BROKE OFF AND INTO THE SMOKE CAPTURE PORT.¿. AN ALTERNATE SAME DEVICE WAS USED TO COMPLETE THE PROCEDURE CAUSING A DELAY OF 5-MINUTES. THERE WAS NO REPORT OF INJURY, MEDICAL INTERVENTION, OR HOSPITALIZATION FOR THE PATIENT. THERE WAS NO REPORT OF DEVICE BEING IN CONTACT WITH THE PATIENT AT THE TIME OF THE MALFUNCTION. A GOOD FAITH EFFORT WAS CONDUCTED BY CONMED TO ASSURE NO FRAGMENTATION CAME INTO CONTACT WITH THE PATIENT; HOWEVER, TO DATE WE HAVE NOT RECEIVED A RESPONSE. THIS REPORT IS BEING RAISED DUE TO THE REPORTED INJURY OF UNKNOWN FRAGMENTATION CONTACT TO THE PATIENT.

    Patient problems: Foreign Body In Patient

    MAUDE report 22182858 ↗
  • MalfunctionOct 26, 2022Material Fragmentation; Material Fragmentation

    THIS COMPLAINT WAS CREATED DUE TO THE RECEIPT OF A MEDWATCH REPORT (MW5112276) ON 12OCT22. THE CURRENT COMPLAINT DATABASE HAS BEEN RESEARCHED FOR THIS EVENT AND THERE WERE NO FINDINGS. THE REPORT WAS FOUND TO BE WRITTEN AGAINST PLPUL2520, PLUMEPEN ULTRA SURGICAL SMOKE EVACUATION PENCIL. IT WAS STATED THAT THE DEVICE WAS BEING USED DURING AN UNKNOWN PROCEDURE THAT OCCURRED ON (B)(6) 2022. THE REPORT STATED, ¿WHILE USING THE PLUMEPEN SMOKE EVACUATION PENCIL BOVIE, DURING SURGICAL PROCEDURE, THE BLADE FELL OUT WHILE IN USE AT THE SURGICAL SITE. ALSO NOTED THAT A PLASTIC PIECE FROM INSIDE THE PLASTIC SHEATH ON THE DISTAL END OF THE BOVIE FELL INTO THE PATIENT. THE PIECE WAS RETRIEVED. THE PATIENT WAS NOT INJURED AS A RESULT OF THIS EVENT.¿. THE COMPONENT WAS RETRIEVED FROM THE PATIENT. THERE WAS NO REPORT OF INJURY, MEDICAL INTERVENTION, OR HOSPITALIZATION FOR THE PATIENT. THIS REPORT IS BEING RAISED ON THE BASIS OF MALFUNCTION WITH POTENTIAL FOR INJURY UPON REOCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15674267 ↗

Other Electrosurgical, cutting & coagulation & accessories devices