Valleylab E0510Recalls & reported problems
Covidien LP · Laparoscopic Monopolar Cord · Class 2
All models of Valleylab (97)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 20884524001774), received between Sep 22, 2017 and Feb 15, 2026 — classified as 1 injury, 4 malfunctions.
The most frequently reported specific device problem is “Fire” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Fire2
- Material Integrity Problem2
- Adverse Event Without Identified Device or Use Problem1
- Connection Problem1
- Overheating of Device1
- Peeled/Delaminated1
- Thermal Decomposition of Device1
Patient problems
- No Consequences Or Impact To Patient2
- Burn(s)1
- No Known Impact Or Consequence To Patient1
- Superficial (First Degree) Burn1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 15, 2026Connection Problem; Material Integrity Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT DURING THE PROCEDURE, THE REUSABLE NON-MEDTRONIC SUCTION DIATHERMY WAS NOT WORKING PROPERLY. THE DIATHERMY MACHINE USED WAS THE DEVICE THEN THEY TRANSFERRED IT TO A NON-MEDTRONIC DIATHERMY MACHINE. THEY TRIED TO DISCONNECT AND CHECKED THE DIATHERMY EXTENSION LEAD AND WHILE DOING SO, THE USER RECEIVED A TINY/SMALL ELECTRICAL BURN IN THE RIGHT-HAND RING FINGER.
Patient problems: Burn(s)
MAUDE report 24341657 ↗MalfunctionOct 29, 2020Thermal Decomposition of Device
A LAPAROSCOPIC MONOPOLAR CORD WAS DISCOVERED TO HAVE BURNED THROUGH THE CONNECTING END WHERE IT PLUGS INTO A ROBOTIC MONOPOLAR SCISSOR. THE RESIDENT AT BEDSIDE NOTED THAT IT WAS HOT TO THE TOUCH, AND UPON INSPECTION THE END OF THE CORD HAD A BLACKENED BURNT END.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10754989 ↗MalfunctionMar 18, 2020Fire; Material Integrity Problem; Fire; Material Integrity Problem
ACCORDING TO THE REPORTER, DURING A ROBOTIC PROSTATE SURGERY, THE CORD SPARKED AND CAUGHT FIRE. PICTURES WERE SUBMITTED WHICH SHOWED THE BURNT CORD. THE FIRE WAS PUT OUT BY THE NURSE PUTTING THE CORD IN THE PLASTIC HOLDER. THE PROCEDURE WAS COMPLETED USING A NEW OPENED CORD. THERE WAS NO PATIENT INJURY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9847150 ↗MalfunctionNov 20, 2017Peeled/Delaminated; Peeled/Delaminated
THE CUSTOMER REPORTED THAT THERE WAS A HOLE IN THE CORD INSULATION, EXPOSING THE INNER WIRES. THERE WAS NO PATIENT INJURY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7047756 ↗InjurySep 22, 2017Fire; Overheating of Device
ON (B)(6) 2017 DURING AN UNEVENTFUL LAPAROSCOPIC CHOLECYSTECTOMY IN THE MAIN OPERATING ROOM, A FLAME WAS NOTED WHILE USING A COVIDIEN VALLEYLAB LAPAROSCOPIC SINGLE-USE/DISPOSABLE MONOPOLAR CAUTERY CORD ((B)(4)). THE FLAME OCCURRED NEAR THE JUNCTION WHERE THE CORD CONNECTS TO THE CAUTERY INSTRUMENT. THE RESIDENT SURGEON'S GLOVE WAS BURNED AND HE SUSTAINED A FIRST DEGREE BURN TO THE HAND. THE PT WAS NOT HARMED. THE EQUIPMENT WAS IMMEDIATELY SEQUESTERED AND SURGERY WAS COMPLETED WITHOUT FURTHER INCIDENT. WE ARE AWARE OF AT LEAST ONE OTHER MEDICAL CENTER THAT HAS HAD OVERHEATING ISSUES WITH THIS CAUTERY CORD. WE HAVE NOW DISCONTINUED THE USE OF THIS CORD AND HAVE SWITCHED TO A DIFFERENT CORD FROM A DIFFERENT MFR.
Patient problems: Superficial (First Degree) Burn
MAUDE report 6886534 ↗
Other ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES devices
- Vasoview Hemopro 2 VH-4000Maquet Cardiovascular LLC · Electrosurgical, Cutting & Coagulation & Accessories6 recalls1 death3,232 reports
- LigaSure LF1937Covidien LP · Electrosurgical, cutting & coagulation & accessories3 deaths2,090 reports
- Olympus WB91051WOLYMPUS Winter & Ibe GmbH · Electrosurgical, Cutting & Coagulation & Accessories1 death1,951 reports
- Valleylab FT10 VLFT10GENCovidien LP · Electrosurgical, cutting & coagulation & accessories1 recall5 deaths1,062 reports
- LigaSure LF1837Covidien LP · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES1 recall967 reports
- LigaSure LF2019Covidien LP · Electrosurgical, cutting & coagulation & accessories948 reports
- VAPR 225028Depuy Mitek, LLC · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES895 reports
- LigaSure LF1923Covidien LP · Electrosurgical, cutting & coagulation & accessories855 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page