UNIVERSAL PLUS 60-5272-132Recalls & reported problems

Conmed Corporation · UNIVERSAL PLUS LAPAROSCOPIC ELECTRODE WITH SUCTION/IRRIGATION, L-HOOK 5 MM X 32 CM · Class 2

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 30653405010627), received between May 17, 2024 and May 17, 2024 — classified as 1 malfunction.

The most frequently reported specific device problem is “Energy Output Problem” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Energy Output Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 17, 2024Energy Output Problem; Energy Output Problem

    UPDATE: FURTHER ASSESSMENT INFORMATION FOUND THAT "NO INTERVENTION NECESSARY FOR THE USER. THE PATIENT WAS PLACED FIBRIN GLUE IN THE COAGULATED AREA. NO SUBSEQUENT CONSEQUENCES FOR THE PATIENT. PATIENT AND USER ARE WELL.". THE PROCEDURE WAS COMPLETED. THE CUSTOMER REPORTED THAT THE 60-5272-132, LAP ELECT,REMOVABLE,32,L-HOOK, DEVICE WAS BEING USED DURING A BILATERAL SYMPATHECTOMY PROCEDURE ON (B)(6) 2024 WHEN IT WAS REPORTED ¿DURING THORACOSCOPIC PROCEDURE OF BILATERAL SYMPATHECTOMY, A J ELECTRODE WAS USED. NORMALLY THE DISTAL PART OF THE LAPAROSCOPIC ELECTRODE ¿J¿ DELIVERS COAGULATING CURRENT TO THE TISSUE, BUT DURING THIS PROCEDURE THE CURRENT WAS DELIVERED THROUGH THE SHEATH, TRANSFERRING UNDESIRED COAGULATING CURRENT TO THE PART OF THE LUNG THAT WAS IN CONTACT. LUCKILY, NO VESSEL WAS NEARBY AND TOUCHED FROM THE CURRENT, OTHERWISE THERE COULD HAPPEN SEVERE CONSEQUENCES.¿. FURTHER ASSESSMENT INFORMATION WAS REQUESTED OF THE REPORTER BUT TO DATE NO FURTHER INFORMATION IS AVAILABLE. THERE WAS NO REPORT OF INJURY, MEDICAL INTERVENTION, OR EXTENDED HOSPITALIZATION FOR THE PATIENT OR USER. THIS REPORT IS BEING RAISED ON THE BASIS OF MALFUNCTION WITH POTENTIAL FOR INJURY UPON REOCCURREN

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19343053 ↗

Other ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES devices

All ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES devices →

Sources and updates