SUREFIT 410-2000Recalls & reported problems
Conmed Corporation · SUREFIT, DUAL DISPERSIVE ELECTRODE, CONTACT QUALITY MONITOR · Class 2
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 30653405033831), received between Jan 5, 2016 and Apr 19, 2023 — classified as 2 injuries, 1 malfunction.
The most frequently reported specific device problem is “Insufficient Device Problem Information” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Insufficient Device Problem Information2
- Disconnection1
Patient problems
- Burn(s)1
- No Clinical Signs, Symptoms or Conditions1
- Partial thickness (Second Degree) Burn1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 19, 2023Disconnection
BOVIE PAD FULLY ADHERED TO PATIENT'S SKIN, HOWEVER, BOVIE MACHINE ALARMING DISCONNECTED PAD. BOVIE MACHINE CLEARED VIA MEDICAL ENGINEERING AS WORKING APPROPRIATELY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16768362 ↗InjuryJan 20, 2017Insufficient Device Problem Information; Insufficient Device Problem Information
AS REPORTED ON MAUDE ADVERSE EVENT REPORT NUMBER (B)(4): "A (B)(6) MALE ADMITTED FOR AN ELECTIVE NEUROSURGICAL PROCEDURE ON (B)(6) 2016. AT THE START OF THE PROCEDURE, THE NEUROSURGEON NOTED THE BOVIE WAS NOT DELIVERING THE DESIRED POWER. THE PROCEDURE WAS STOPPED AS THE CIRCULATING RN TRACED THE ELECTROSURGICAL EQUIPMENT FROM PEN TO GENERATOR AND NOTED "GREEN" LIGHT PRESENT INDICATING GOOD GROUNDING AND CAN PROCEED. THEN CONTINUED TRACING TO PT AT BOVIE PAD. THE BOVIE PAD WAS REMOVED WHICH REVEALED A SMALL BURN AT THE EDGE OF THE PAD ON THE PT'S RIGHT LATERAL THIGH. THE NEUROSURGEON DECIDED TO ABORT THE CASE DUE TO THE BURN. A GENERAL SURGEON ASSESSED THE SITE AND EXCISED THE BURNED TISSUE. THIS RESULTED IN A 2CM X 6 CM INCISION WOUND THAT WAS CLOSED WITH NYLON SUTURES AND DRESSING APPLIED. SUTURES TO BE REMOVED BY GENERAL SURGEON IN ONE WEEK. ORIGINAL NEUROSURGICAL PROCEDURE TO BE RESCHEDULED IN 2-3 WEEKS." NO CONTACT INFORMATION FOR REPORTER OR USER FACILITY IS LISTED ON MAUDE REPORT AND THEREFORE, TO DATE, THERE HAS BEEN NO ADDITIONAL INFORMATION RECEIVED REGARDING THE PATIENT'S LATEST CONDITION OR ANY INDICATION OF LONG TERM ADVERSE EFFECT HAS OCCURRED.
Patient problems: Burn(s)
MAUDE report 6266723 ↗InjuryJan 5, 2016Insufficient Device Problem Information
THE END-USER FACILITY REPORTED THAT DURING A ROTATOR CUFF REPAIR PROCEDURE, A CONMED SUREFIT DISPERSIVE ELECTRODE WAS UTILIZED AND PLACED ON THE PATIENT'S RIGHT THIGH. REPORTEDLY, UPON COMPLETION OF THE SURGERY, AND WHEN THE DISPERSIVE PAD WAS REMOVED, A SECOND DEGREE BURN TO THE PATIENT'S RIGHT THIGH WAS IDENTIFIED AT THE SITE OF THE DISPERSIVE ELECTRODE. THE "BURN" WAS APPROXIMATELY 1 X 1.5 INCHES AND OVAL SHAPED. TREATMENT TO THE REPORTED "BURN" WAS SILVADENE OINTMENT AND A NON-ADHERENT BANDAGE. THERE WAS NO REPORT OF ANY LONG TERM CONSEQUENCES THAT RESULTED FROM THE REPORTED INCIDENT.
Patient problems: Partial thickness (Second Degree) Burn
MAUDE report 5343595 ↗
Other ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES devices
- Vasoview Hemopro 2 VH-4000Maquet Cardiovascular LLC · Electrosurgical, Cutting & Coagulation & Accessories6 recalls1 death3,232 reports
- LigaSure LF1937Covidien LP · Electrosurgical, cutting & coagulation & accessories3 deaths2,090 reports
- Olympus WB91051WOLYMPUS Winter & Ibe GmbH · Electrosurgical, Cutting & Coagulation & Accessories1 death1,951 reports
- Valleylab FT10 VLFT10GENCovidien LP · Electrosurgical, cutting & coagulation & accessories1 recall5 deaths1,062 reports
- LigaSure LF1837Covidien LP · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES1 recall967 reports
- LigaSure LF2019Covidien LP · Electrosurgical, cutting & coagulation & accessories948 reports
- VAPR 225028Depuy Mitek, LLC · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES895 reports
- LigaSure LF1923Covidien LP · Electrosurgical, cutting & coagulation & accessories855 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.