CONMED GOLDLINE 130309ARecalls & reported problems
Conmed Corporation · CONMED GOLDLINE ELECTROSURGICAL HANDPIECE, HAND CONTROLLED WITH BLADE ELECTRODE, BUTTON SWITCH, SAFETY HOLSTER AND 10 FT (3 M) CABLE · Class 2
All models of CONMED GOLDLINE (10)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
Quick summary
This Conmed Goldline electrosurgical handpiece is a surgeon-controlled cutting and coagulating pencil with a blade electrode and button switch. The reports mostly describe units arriving at distributors with unsterile or poorly sealed packaging, discovered during incoming inspection and rejected before any patient contact, plus a few that switched on by themselves. Of 33 reports,5 involve injuries and there are no deaths. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 33
- Deaths reported
- 0
- Injuries reported
- 5
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 33 adverse event reports that list this exact device (DI 30653405056151), received between Mar 27, 2017 and Jun 9, 2026 — classified as 5 injuries, 28 malfunctions.
The most frequently reported specific device problem is “Delivered as Unsterile Product” (23 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Delivered as Unsterile Product23
- Insufficient Device Problem Information4
- Self-Activation or Keying3
- Device Packaging Compromised1
- Packaging Problem1
- Unsealed Device Packaging1
Patient problems
- No Clinical Signs, Symptoms or Conditions25
- Burn(s)5
- Not Applicable2
- No Patient Involvement1
Recent adverse event reports
The 25 most recent of 33 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 9, 2026Delivered as Unsterile Product
DURING INCOMING INSPECTION, THE DISTRIBUTOR REJECTED THIS DEVICE, 130309A, FOR AN INSUFFICIENT HEATSEAL. THERE WAS NO CONTACT WITH THE PATIENT AS THIS WAS FOUND DURING INCOMING INSPECTION. THIS WILL BE REPORTED AS A MALFUNCTION WITH POTENTIAL FOR INJURY UPON REOCCURRENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25442015 ↗MalfunctionDec 11, 2025Delivered as Unsterile Product; Delivered as Unsterile Product
DURING INCOMING INSPECTION, THE DISTRIBUTOR REJECTED THIS DEVICE, 130309A, FOR AN INSUFFICIENT HEATSEAL. THERE WAS NO CONTACT WITH THE PATIENT AS THIS WAS FOUND DURING INCOMING INSPECTION. THIS WILL BE REPORTED AS A MALFUNCTION WITH POTENTIAL FOR INJURY UPON REOCCURRENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23779802 ↗MalfunctionDec 11, 2025Delivered as Unsterile Product; Delivered as Unsterile Product
THE DISTRIBUTOR REPORTED THAT THE 130309A DEVICE HAD AN INSUFFICIENT HEATSEAL FOUND DURING INCOMING INSPECTION. NO PATIENT INVOLVEMENT OCCURRED. THIS REPORT IS BEING RAISED DUE TO THE REPORTED MALFUNCTION WITH POTENTIAL FOR INJURY UPON REOCCURRENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23777147 ↗MalfunctionSep 17, 2025Delivered as Unsterile Product; Delivered as Unsterile Product
DURING INCOMING INSPECTION, THE DISTRIBUTOR REJECTED THIS DEVICE, 130309A FOR AN INSUFFICIENT HEATSEAL. THERE WAS NO CONTACT WITH THE PATIENT AS THIS WAS FOUND DURING INCOMING INSPECTION. THIS REPORT IS BEING RAISED DUE TO THE REPORTED MALFUNCTION WITH POTENTIAL FOR INJURY UPON REOCCURRENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23071579 ↗MalfunctionJun 9, 2025Delivered as Unsterile Product; Delivered as Unsterile Product
DURING INCOMING INSPECTION, THE DISTRIBUTOR REJECTED THIS DEVICE, 130309A, FOR AN INSUFFICIENT HEATSEAL. THERE WAS NO CONTACT WITH THE PATIENT AS THIS WAS FOUND DURING INCOMING INSPECTION. THIS WILL BE REPORTED AS A MALFUNCTION WITH POTENTIAL FOR INJURY UPON REOCCURRENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22167624 ↗MalfunctionApr 25, 2025Delivered as Unsterile Product; Delivered as Unsterile Product
DURING INCOMING INSPECTION, THE DISTRIBUTOR REJECTED THIS DEVICE, 130309A, FOR AN INSUFFICIENT HEATSEAL. THERE WAS NO CONTACT WITH THE PATIENT AS THIS WAS FOUND DURING INCOMING INSPECTION. THIS REPORT IS BEING RAISED DUE TO THE REPORTED MALFUNCTION WITH POTENTIAL FOR INJURY UPON REOCCURRENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21908718 ↗MalfunctionApr 25, 2025Delivered as Unsterile Product; Delivered as Unsterile Product
DURING INCOMING INSPECTION, THE DISTRIBUTOR REJECTED THIS DEVICE, 130309A, FOR AN INSUFFICIENT HEATSEAL. THERE WAS NO CONTACT WITH THE PATIENT AS THIS WAS FOUND DURING INCOMING INSPECTION. THIS WILL BE REPORTED AS A MALFUNCTION WITH POTENTIAL FOR INJURY UPON REOCCURRENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21908615 ↗MalfunctionFeb 26, 2025Delivered as Unsterile Product; Delivered as Unsterile Product
THE DISTRIBUTOR REPORTED THAT THE 130309A DEVICE HAD AN INSUFFICIENT HEATSEAL FOUND DURING INCOMING INSPECTION. NO PATIENT INVOLVEMENT OCCURRED. THE SHIPMENT ASSOCIATED WITH THE RGA FOR THIS DEVICE HAS NOT YET BEEN RETURNED FOR EVALUATION AND IT HAS NOT BEEN CONFIRMED THAT AN INSUFFICIENT HEATSEAL CAUSING A STERILITY BREACH HAS OCCURRED. BASED ON THE INFORMATION AVAILABLE, AND THE DECISION TREE RESULTS, THIS COMPLAINT DOES NOT MEET THE DEFINITION OF A REPORTABLE EVENT. UPON THE RECEIPT OF THE DEVICE, AN EVALUATION WILL BE CONDUCTED AND IF A BREACH OF STERILITY IS CONFIRMED THE REPORTING DECISION WILL BE REASSESSED AT THAT TIME. UPDATE: 19FEB25 - THE DEVICE HAS BEEN RETURNED AND AN EVALUATION HAS BEEN COMPLETED. AN INSUFFICIENT HEATSEAL WAS DISCOVERED AFTER TESTING. THIS WILL BE REPORTED AS A MALFUNCTION WITH POTENTIAL FOR INJURY UPON REOCCURRENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21469879 ↗MalfunctionSep 4, 2024Delivered as Unsterile Product; Delivered as Unsterile Product
DURING INCOMING INSPECTION, THE DISTRIBUTOR REJECTED THIS DEVICE, 130309A, GOLDLINE, BUTTON, HOLSTER, FOR AN INSUFFICIENT HEATSEAL. THERE WAS NO CONTACT WITH THE PATIENT AS THIS WAS FOUND DURING INCOMING INSPECTION. THIS WILL BE REPORTED AS A MALFUNCTION WITH POTENTIAL FOR INJURY UPON REOCCURRENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20137692 ↗MalfunctionJul 15, 2024Delivered as Unsterile Product; Delivered as Unsterile Product
DURING INCOMING INSPECTION, THE DISTRIBUTOR REJECTED THIS DEVICE, 130309A, GOLDLINE, BUTTON, HOLSTER, FOR AN INSUFFICIENT HEATSEAL. THERE WAS NO CONTACT WITH THE PATIENT AS THIS WAS FOUND DURING INCOMING INSPECTION. THIS REPORT IS BEING RAISED DUE TO THE REPORTED MALFUNCTION WITH POTENTIAL FOR INJURY UPON REOCCURRENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19744358 ↗MalfunctionJul 12, 2024Delivered as Unsterile Product; Delivered as Unsterile Product
DURING INCOMING INSPECTION, THE DISTRIBUTOR REJECTED THIS DEVICE, 130309A, GOLDLINE, BUTTON, HOLSTER, FOR AN INSUFFICIENT HEATSEAL. THERE WAS NO CONTACT WITH THE PATIENT AS THIS WAS FOUND DURING INCOMING INSPECTION. THIS REPORT IS BEING RAISED DUE TO THE REPORTED MALFUNCTION WITH POTENTIAL FOR INJURY UPON REOCCURRENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19730664 ↗MalfunctionJul 9, 2024Delivered as Unsterile Product; Delivered as Unsterile Product
DURING INCOMING INSPECTION, THE DISTRIBUTOR REJECTED TWO 130309A, GOLDLINE, BUTTON, HOLSTER DEVICES FOR AN INSUFFICIENT HEATSEAL. THERE WAS NO CONTACT WITH THE PATIENT AS THIS WAS FOUND DURING INCOMING INSPECTION. THIS WILL BE REPORTED AS A MALFUNCTION WITH POTENTIAL FOR INJURY UPON REOCCURRENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19704967 ↗MalfunctionMay 10, 2024Self-Activation or Keying; Self-Activation or Keying
THE CUSTOMER REPORTED THAT THE 130309A, GOLDLINE, BUTTON, HOLSTER, DEVICE WAS BEING USED DURING AN UNKNOWN PROCEDURE ON (B)(6) 2024 WHEN IT WAS REPORTED ¿THE DEFECTIVE PENCIL CAUTERIZED WITHOUT THE SURGEON HANDLING IT.¿. THERE WAS NO REPORT OF INJURY, MEDICAL INTERVENTION, OR EXTENDED HOSPITALIZATION FOR THE PATIENT OR USER. THIS REPORT IS BEING RAISED ON THE BASIS OF MALFUNCTION WITH POTENTIAL FOR INJURY UPON REOCCURRENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19285747 ↗InjuryMar 12, 2024Self-Activation or Keying; Self-Activation or Keying
THE SALES REPRESENTATIVE REPORTED ON BEHALF OF THE CUSTOMER, THAT THE 130309A, GOLDLINE, BUTTON, HOLSTER WAS BEING USED DURING AN UNKNOWN PROCEDURE ON (B)(6) 2024 DATE WHEN IT WAS REPORTED ¿WHEN THEY PLUGGED THE PEN IT ACTIVATED ITSELF, WITH ESG-400 NOT ABLE TO BE REPRODUCED BY BIOMED. THE HOSPITAL BUT THEY INFORMED ME THERE WAS 3 BURNS TO THE PATIENT THAT THEY DIDN'T MENTION WHEN THEY SENT IT TO ME FURTHER ASSESSMENT QUESTIONS WERE REQUESTED OF THE REPORTER AND A GOOD FAITH EFFORT WAS COMPLETED; HOWEVER, THE REPORTER HAS NOT RESPONDED TO OUR ATTEMPTS FOR INFORMATION. THE PROCEDURE WAS COMPLETED WITH AN UNKNOWN ALTERNATE DEVICE AND A 2-MINUTE DELAY WAS REPORTED. THERE WAS NO REPORT OF MEDICAL INTERVENTION, OR EXTENDED HOSPITALIZATION TO THE PATIENT OR USER. THIS REPORT IS BEING RAISED ON THE REPORTED INJURY DUE TO UNKNOWN DEGREE OF BURN.
Patient problems: Burn(s)
MAUDE report 18887804 ↗MalfunctionJan 29, 2024Delivered as Unsterile Product; Delivered as Unsterile Product
DURING INCOMING INSPECTION, THE DISTRIBUTOR REJECTED THIS DEVICE, 130309A, FOR AN INSUFFICIENT HEATSEAL. THERE WAS NO CONTACT WITH THE PATIENT AS THIS WAS FOUND DURING INCOMING INSPECTION. THIS REPORT IS BEING RAISED DUE TO THE REPORTED MALFUNCTION WITH POTENTIAL FOR INJURY UPON REOCCURRENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18594465 ↗MalfunctionJan 25, 2024Delivered as Unsterile Product; Delivered as Unsterile Product
DURING INCOMING INSPECTION, THE DISTRIBUTOR REJECTED TWO OF THIS DEVICE, 130309A, GOLDLINE, BUTTON, HOLSTER, FOR AN INSUFFICIENT HEATSEAL. THERE WAS NO CONTACT WITH THE PATIENT AS THIS WAS FOUND DURING INCOMING INSPECTION. THIS WILL BE REPORTED AS A MALFUNCTION WITH POTENTIAL FOR INJURY UPON REOCCURRENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18582100 ↗MalfunctionDec 21, 2023Delivered as Unsterile Product; Delivered as Unsterile Product
DURING INCOMING INSPECTION, THE DISTRIBUTOR REJECTED THIS DEVICE, 130309A, FOR AN INSUFFICIENT HEATSEAL. THERE WAS NO CONTACT WITH THE PATIENT AS THIS WAS FOUND DURING INCOMING INSPECTION. PERFORMED A FUNCTIONAL INSPECTION, THE DEVICES WERE DYE LEAK TESTED WHICH INDICATED THAT THE PACKAGING HAD AN INSUFFICIENT HEAT SEAL. THIS WILL BE REPORTED AS A MALFUNCTION WITH POTENTIAL FOR INJURY UPON REOCCURRENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18371056 ↗MalfunctionNov 13, 2023Self-Activation or Keying; Self-Activation or Keying
THE CUSTOMER REPORTED THAT THE 130309A, GOLDLINE, BUTTON, HOLSTER WAS BEING USED ON (B)(6) 2023. TERY SOUND ON ESU GOES OFF EVEN WHEN GOLDLINE PENCIL BUTTONS ARE NOT BEING PUSHED. CUSTOMER SAYS SOME OF THE PENCILS IN THE BATCH HAVE DONE IT BUT DON'T BELIEVE THE BOVIE WAS ACTUALLY FIRING/HEATING UP DURING MISFIRE TONE.¿. THERE WAS NO REPORT OF IMPACT/INJURY, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION TO THE PATIENT. THIS REPORT IS BEING RAISED DUE TO THE REPORTED MALFUNCTION WITH POTENTIAL FOR INJURY UPON REOCCURRENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18120799 ↗InjuryOct 20, 2023Insufficient Device Problem Information; Insufficient Device Problem Information
THE CUSTOMER REPORTED THAT THE 130309A, GOLDLINE, BUTTON, HOLSTER DEVICE WAS BEING USED DURING AN UNKNOWN PROCEDURE ON (B)(6) 2023 WHEN IT WAS REPORTED ¿PATIENT WAS BURNED.¿. FURTHER ASSESSMENT QUESTIONS WERE REQUESTED OF THE REPORTER AND A GOOD FAITH EFFORT WAS COMPLETED; HOWEVER, THE REPORTER HAS NOT RESPONDED TO OUR ATTEMPTS FOR INFORMATION. THE CURRENT STATUS OF THE PATIENT WAS REPORTED AS ¿THE PATIENT WILL BE FINE¿. THERE WAS NO REPORT OF MEDICAL INTERVENTION OR EXTENDED HOSPITALIZATION FOR THE PATIENT OR USER. THIS REPORT IS BEING RAISED ON THE REPORTED INJURY DUE TO PATIENT BEING DIAGNOSED WITH UNKNOWN DEGREE OF BURN.
Patient problems: Burn(s)
MAUDE report 17974511 ↗InjuryOct 18, 2023Insufficient Device Problem Information; Insufficient Device Problem Information
THE CUSTOMER REPORTED THAT THE 130309A, GOLDLINE, BUTTON, HOLSTER DEVICE WAS BEING USED DURING AN UNKNOWN PROCEDURE ON (B)(6) 2023 WHEN IT WAS REPORTED ¿PATIENT WAS BURNED SECOND CASE.¿. FURTHER ASSESSMENT ANSWERS WERE REQUESTED OF THE REPORTER AND A GOOD FAITH EFFORT WAS COMPLETED; HOWEVER, THE REPORTER HAS NOT RESPONDED TO OUR ATTEMPTS FOR INFORMATION. THERE WAS NO REPORT OF MEDICAL INTERVENTION, OR EXTENDED HOSPITALIZATION FOR THE PATIENT OR USER. THIS REPORT IS BEING RAISED ON THE REPORTED INJURY DUE TO PATIENT BEING DIAGNOSED WITH UNKNOWN DEGREE OF BURN.
Patient problems: Burn(s)
MAUDE report 17956067 ↗InjuryOct 18, 2023Insufficient Device Problem Information; Insufficient Device Problem Information
THE CUSTOMER REPORTED THAT THE 130309A, GOLDLINE, BUTTON, HOLSTER DEVICE WAS BEING USED DURING AN UNKNOWN PROCEDURE ON (B)(6) 2023 WHEN IT WAS REPORTED ¿PATIENT WAS BURNED.¿. FURTHER ASSESSMENT ANSWERS WERE REQUESTED OF THE REPORTER AND A GOOD FAITH EFFORT WAS COMPLETED; HOWEVER, THE REPORTER HAS NOT RESPONDED TO OUR ATTEMPTS FOR INFORMATION. THERE WAS NO REPORT OF MEDICAL INTERVENTION, OR EXTENDED HOSPITALIZATION FOR THE PATIENT OR USER. THIS REPORT IS BEING RAISED ON THE REPORTED INJURY DUE TO PATIENT BEING DIAGNOSED WITH UNKNOWN DEGREE OF BURN.
Patient problems: Burn(s)
MAUDE report 17956066 ↗InjuryOct 18, 2023Insufficient Device Problem Information; Insufficient Device Problem Information
THE CUSTOMER REPORTED THAT THE 130309A, GOLDLINE, BUTTON, HOLSTER DEVICE WAS BEING USED DURING AN UNKNOWN PROCEDURE ON (B)(6) 2023 WHEN IT WAS REPORTED ¿PATIENT WAS BURNED FIRST CASE.¿ FURTHER ASSESSMENT ANSWERS WERE REQUESTED OF THE REPORTER AND A GOOD FAITH EFFORT WAS COMPLETED; HOWEVER, THE REPORTER HAS NOT RESPONDED TO OUR ATTEMPTS FOR INFORMATION. THERE WAS NO REPORT OF MEDICAL INTERVENTION, OR EXTENDED HOSPITALIZATION FOR THE PATIENT OR USER. THIS REPORT IS BEING RAISED ON THE REPORTED INJURY DUE TO PATIENT BEING DIAGNOSED WITH UNKNOWN DEGREE OF BURN.
Patient problems: Burn(s)
MAUDE report 17956064 ↗MalfunctionSep 12, 2023Delivered as Unsterile Product; Delivered as Unsterile Product
DURING INCOMING INSPECTION, THE DISTRIBUTOR REJECTED THIS DEVICE, 130309A, FOR AN INSUFFICIENT HEATSEAL. THERE WAS NO CONTACT WITH THE PATIENT AS THIS WAS FOUND DURING INCOMING INSPECTION. THIS WILL BE REPORTED AS A MALFUNCTION WITH POTENTIAL FOR INJURY UPON REOCCURRENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17730587 ↗MalfunctionJan 30, 2023Delivered as Unsterile Product; Delivered as Unsterile Product
DURING INCOMING INSPECTION, THE DISTRIBUTOR REJECTED THIS DEVICE, 130309A, FOR AN INSUFFICIENT HEATSEAL. THERE WAS NO CONTACT WITH THE PATIENT AS THIS WAS FOUND DURING INCOMING INSPECTION. THIS WILL BE REPORTED AS A MALFUNCTION WITH POTENTIAL FOR INJURY UPON REOCCURRENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16263059 ↗MalfunctionOct 5, 2022Delivered as Unsterile Product; Delivered as Unsterile Product
DURING INCOMING INSPECTION, THE DISTRIBUTOR REJECTED THIS DEVICE, 130309A, FOR AN INSUFFICIENT HEATSEAL. THERE WAS NO CONTACT WITH THE PATIENT AS THIS WAS FOUND DURING INCOMING INSPECTION. THIS WILL BE REPORTED AS A MALFUNCTION WITH POTENTIAL FOR INJURY UPON REOCCURRENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15543615 ↗
Other ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES devices
- Vasoview Hemopro 2 VH-4000Maquet Cardiovascular LLC · Electrosurgical, Cutting & Coagulation & Accessories6 recalls1 death3,232 reports
- LigaSure LF1937Covidien LP · Electrosurgical, cutting & coagulation & accessories3 deaths2,090 reports
- Olympus WB91051WOLYMPUS Winter & Ibe GmbH · Electrosurgical, Cutting & Coagulation & Accessories1 death1,951 reports
- Valleylab FT10 VLFT10GENCovidien LP · Electrosurgical, cutting & coagulation & accessories1 recall5 deaths1,062 reports
- LigaSure LF1837Covidien LP · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES1 recall967 reports
- LigaSure LF2019Covidien LP · Electrosurgical, cutting & coagulation & accessories948 reports
- VAPR 225028Depuy Mitek, LLC · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES895 reports
- LigaSure LF1923Covidien LP · Electrosurgical, cutting & coagulation & accessories855 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page