Conmed Corporation · CONMED GOLDLINE ELECTROSURGICAL HANDPIECE, HAND CONTROLLED WITH UltraClean BLADE ELECTRODE, BUTTON SWITCH, SAFETY HOLSTER AND 10' (3 M) CABLE · Class 2

All models of CONMED GOLDLINE (10)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
8
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 30653405056274), received between Aug 29, 2019 and Sep 18, 2025 — classified as 1 injury, 7 malfunctions.

The most frequently reported specific device problem is “Delivered as Unsterile Product” (6 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Delivered as Unsterile Product6
  • Arcing1
  • Self-Activation or Keying1

Patient problems

  • No Clinical Signs, Symptoms or Conditions7
  • Burn(s)1

Recent adverse event reports

The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 18, 2025Delivered as Unsterile Product; Delivered as Unsterile Product

    DURING INCOMING INSPECTION, THE DISTRIBUTOR REJECTED THIS DEVICE, 131309A, FOR AN INSUFFICIENT HEATSEAL. THERE WAS NO CONTACT WITH THE PATIENT AS THIS WAS FOUND DURING INCOMING INSPECTION. THIS WILL BE REPORTED AS A MALFUNCTION WITH POTENTIAL FOR INJURY UPON REOCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23082571 ↗
  • MalfunctionAug 6, 2025Delivered as Unsterile Product; Delivered as Unsterile Product

    DURING INCOMING INSPECTION, THE DISTRIBUTOR REJECTED THIS DEVICE, 131309A, FOR AN INSUFFICIENT HEATSEAL. THERE WAS NO CONTACT WITH THE PATIENT AS THIS WAS FOUND DURING INCOMING INSPECTION. THIS WILL BE REPORTED AS A MALFUNCTION WITH POTENTIAL FOR INJURY UPON REOCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22726186 ↗
  • MalfunctionFeb 26, 2025Delivered as Unsterile Product; Delivered as Unsterile Product

    THE DISTRIBUTOR REPORTED THAT THE 131309A DEVICE HAD AN INSUFFICIENT HEATSEAL FOUND DURING INCOMING INSPECTION. NO PATIENT INVOLVEMENT OCCURRED. THE SHIPMENT ASSOCIATED WITH THE RGA FOR THIS DEVICE HAS NOT YET BEEN RETURNED FOR EVALUATION AND IT HAS NOT BEEN CONFIRMED THAT AN INSUFFICIENT HEATSEAL CAUSING A STERILITY BREACH HAS OCCURRED. BASED ON THE INFORMATION AVAILABLE, AND THE DECISION TREE RESULTS, THIS COMPLAINT DOES NOT MEET THE DEFINITION OF A REPORTABLE EVENT. UPON THE RECEIPT OF THE DEVICE, AN EVALUATION WILL BE CONDUCTED AND IF A BREACH OF STERILITY IS CONFIRMED THE REPORTING DECISION WILL BE REASSESSED AT THAT TIME. UPDATE: 19FEB2025 - THE DEVICE HAS BEEN RETURNED AND AN EVALUATION HAS BEEN COMPLETED. AN INSUFFICIENT HEATSEAL WAS DISCOVERED AFTER TESTING. THIS WILL BE REPORTED AS A MALFUNCTION WITH POTENTIAL FOR INJURY UPON REOCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21470557 ↗
  • MalfunctionApr 17, 2024Self-Activation or Keying; Self-Activation or Keying

    A CUSTOMER REPORTED THAT THE 131309A, GOLDLINE, BTN, HOLST, ULTRACLEAN DEVICE WAS BEING USED DURING AN UNNAMED PROCEDURE ON (B)(6) 2024, AND THE DEVICE ¿STARTED GOING OFF BY ITS SELF¿. THE PROCEDURE WAS COMPLETED WITHOUT THE USE OF AN ALTERNATE SAME DEVICE AND WITH NO DELAY. THERE WAS NO REPORT OF INJURY, MEDICAL/SURGICAL INTERVENTION, OR EXTENDED HOSPITALIZATION FOR THE PATIENT. THIS REPORT IS BEING RAISED ON THE BASIS OF MALFUNCTION WITH POTENTIAL FOR INJURY UPON REOCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19123786 ↗
  • MalfunctionOct 7, 2022Delivered as Unsterile Product; Delivered as Unsterile Product

    DURING INCOMING INSPECTION, THE DISTRIBUTOR REJECTED THIS DEVICE, 131309A, FOR AN INSUFFICIENT HEATSEAL. THERE WAS NO CONTACT WITH THE PATIENT AS THIS WAS FOUND DURING INCOMING INSPECTION. THIS WILL BE REPORTED AS A MALFUNCTION WITH POTENTIAL FOR INJURY UPON REOCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15560969 ↗
  • MalfunctionOct 5, 2022Delivered as Unsterile Product; Delivered as Unsterile Product

    DURING INCOMING INSPECTION, THE DISTRIBUTOR REJECTED THIS DEVICE, 131309A, FOR AN INSUFFICIENT HEATSEAL. THERE WAS NO CONTACT WITH THE PATIENT AS THIS WAS FOUND DURING INCOMING INSPECTION. THIS WILL BE REPORTED AS A MALFUNCTION WITH POTENTIAL FOR INJURY UPON REOCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15543567 ↗
  • MalfunctionNov 4, 2021Delivered as Unsterile Product; Delivered as Unsterile Product

    DURING INCOMING INSPECTION, THE DISTRIBUTOR REJECTED THIS DEVICE, (B)(4), FOR AN INSUFFICIENT HEATSEAL. THERE WAS NO CONTACT WITH THE PATIENT AS THIS WAS FOUND DURING INCOMING INSPECTION. THIS REPORT IS BEING RAISED ON THE BASIS OF MALFUNCTION WITH POTENTIAL FOR INJURY UPON REOCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12751421 ↗
  • InjuryAug 29, 2019Arcing; Arcing

    THE SALES REPRESENTATIVE REPORTED ON BEHALF OF THE CUSTOMER THAT THE 131309A DEVICE WAS BEING USED DURING A MASTOPEXY ON (B)(6) 2019. THE SURGEON STATED THAT BLOOD ACCUMULATED AT THE TIP WHICH COULD HAVE CAUSED AN ARCED/SPARK CAUSING HIS GLOVES TO POP. THE EVENT CAUSED A 5-MINUTE DELAY; HOWEVER, THE PROCEDURE WAS COMPLETED SUCCESSFULLY. THE REPRESENTATIVE WAS LATER SHOWN BY THE SURGEON THAT HE HAD A BLISTER ON THE FIRST KNUCKLE OF THE SECOND DIGIT. THIS REPORT IS BEING RAISED ON THE BASIS OF INJURY DUE TO THE BURN OF UNKNOWN DEGREE.

    Patient problems: Burn(s)

    MAUDE report 8944156 ↗

Other ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES devices