CONMED GOLDLINE 131307ARecalls & reported problems
Conmed Corporation · CONMED GOLDLINE ELECTROSURGICAL HANDPIECE, HAND CONTROLLED WITH UltraClean BLADE ELECTRODE, ROCKER SWITCH, SAFETY HOLSTER AND 10' (3 M) CABLE · Class 2
All models of CONMED GOLDLINE (10)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 30653405056632), received between Nov 10, 2016 and Feb 26, 2025 — classified as 6 malfunctions.
The most frequently reported specific device problem is “Delivered as Unsterile Product” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Delivered as Unsterile Product4
- Unsealed Device Packaging2
Patient problems
- No Clinical Signs, Symptoms or Conditions4
- Not Applicable2
- No Patient Involvement1
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 26, 2025Delivered as Unsterile Product; Delivered as Unsterile Product
THE DISTRIBUTOR REPORTED THAT THE 131307A DEVICE HAD AN INSUFFICIENT HEATSEAL FOUND DURING INCOMING INSPECTION. NO PATIENT INVOLVEMENT OCCURRED. THE SHIPMENT ASSOCIATED WITH THE RGA FOR THIS DEVICE HAS NOT YET BEEN RETURNED FOR EVALUATION AND IT HAS NOT BEEN CONFIRMED THAT AN INSUFFICIENT HEATSEAL CAUSING A STERILITY BREACH HAS OCCURRED. BASED ON THE INFORMATION AVAILABLE, AND THE DECISION TREE RESULTS, THIS COMPLAINT DOES NOT MEET THE DEFINITION OF A REPORTABLE EVENT. UPON THE RECEIPT OF THE DEVICE, AN EVALUATION WILL BE CONDUCTED AND IF A BREACH OF STERILITY IS CONFIRMED THE REPORTING DECISION WILL BE REASSESSED AT THAT TIME. UPDATE: THE DEVICE WAS RETURNED AND EVALUATED. THE COMPLETION OF THE DEVICE EVALUATION CONFIRMED AN INSUFFICIENT HEATSEAL; THEREFORE, THIS COMPLAINT HAS BEEN REASSESSED AND MEETS THE CRITERIA FOR A REPORTABLE EVENT. THIS REPORT IS BEING RAISED DUE TO THE REPORTED MALFUNCTION WITH POTENTIAL FOR INJURY UPON REOCCURRENCE
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21469644 ↗MalfunctionFeb 18, 2022Delivered as Unsterile Product; Delivered as Unsterile Product
DURING INCOMING INSPECTION, THE DISTRIBUTOR REJECTED THIS DEVICE, 131307A, FOR AN INSUFFICIENT HEATSEAL. THERE WAS NO CONTACT WITH THE PATIENT AS THIS WAS FOUND DURING INCOMING INSPECTION. THIS WILL BE REPORTED AS A MALFUNCTION WITH POTENTIAL FOR INJURY UPON REOCCURRENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13558177 ↗MalfunctionJan 10, 2022Delivered as Unsterile Product; Delivered as Unsterile Product
DURING INCOMING INSPECTION, THE DISTRIBUTOR REJECTED THIS DEVICE, 131307A, FOR AN INSUFFICIENT HEATSEAL. THERE WAS NO CONTACT WITH THE PATIENT AS THIS WAS FOUND DURING INCOMING INSPECTION. THIS REPORT IS BEING RAISED ON THE BASIS OF MALFUNCTION WITH POTENTIAL FOR INJURY UPON REOCCURRENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13205273 ↗MalfunctionAug 26, 2021Delivered as Unsterile Product; Delivered as Unsterile Product
DURING INCOMING INSPECTION, THE DISTRIBUTOR REJECTED THIS DEVICE,131307A, FOR AN INSUFFICIENT HEATSEAL. THERE WAS NO CONTACT WITH THE PATIENT AS THIS WAS FOUND DURING INCOMING INSPECTION. THIS REPORT IS BEING RAISED ON THE BASIS OF MALFUNCTION WITH POTENTIAL FOR INJURY UPON REOCCURRENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12370407 ↗MalfunctionMar 17, 2017Unsealed Device Packaging; Unsealed Device Packaging
THE DISTRIBUTOR IN (B)(6). REPORTED THAT DURING RECEIVING AND INSPECTION OF INCOMING PRODUCTS, ONE (1) PACKAGE CONTAINING AN ULTRA CLEAN PENCIL 131307A WAS DISCOVERED WITH "INSUFFICIENT HEAT SEAL." IN THIS INSTANCE, THERE WAS NO PATIENT INVOLVEMENT WITH THIS REPORTED PROBLEM, AS THE PACKAGING ANOMALY WAS DISCOVERED DURING INSPECTION AT THE DISTRIBUTOR FACILITY PRIOR TO DISTRIBUTION TO AN END-USER.
Patient problems: Not Applicable
MAUDE report 6414442 ↗MalfunctionNov 10, 2016Unsealed Device Packaging
THE DISTRIBUTOR IN (B)(6) REPORTED THAT DURING RECEIVING AND INSPECTION OF INCOMING PRODUCTS, 5 PACKAGES CONTAINING GOLDLINE ROCKER, HOLSTER, AND ULTRACLEAN WAS DISCOVERED WITH "INSUFFICIENT HEAT SEAL." THERE WAS NO PATIENT INVOLVEMENT IN THIS REPORTED PROBLEM, AS THE PACKAGING ANOMALY WAS DISCOVERED DURING INSPECTION AT THE DISTRIBUTOR FACILITY PRIOR TO DISTRIBUTION TO AN END-USER. THESE DEVICES WERE MANUFACTURED 18-MAY-2016. OF THE LOT CONTAINING (B)(4) UNITS, THERE WERE NO OTHER SIMILAR COMPLAINTS RECEIVED FOR THIS ITEM AND LOT NUMBER COMBINATION. A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS LOT HAS VERIFIED THAT THE DEVICES WERE PRODUCED ACCORDING TO THEIR CURRENT AND APPROVED PROCEDURES AND MATERIAL SPECIFICATIONS. ANY NON-CONFORMANCES REGARDING THE PRODUCT'S IDENTITY, QUALITY, SAFETY AND EFFECTIVENESS WERE NOT IDENTIFIED DURING MANUFACTURE. A 2-YEAR REVIEW OF PRODUCT HISTORY FOR THIS DEVICE FAMILY REVEALED 6 COMPLAINTS INVOLVING 11 DEVICES FOR THIS SIMILAR OCCURRENCE "INSUFFICIENT HEAT SEAL." DURING THIS SAME 2-YEAR TIME FRAME, OVER (B)(4) UNITS WERE SOLD WORLDWIDE, MAKING THE OCCURRENCE RATE FOR THIS FAILURE (B)(4) PERCENT. AS OF THIS FILING, THE DEVICE HAS NOT BEEN RECEIV
Patient problems: No Patient Involvement; Not Applicable
MAUDE report 6093634 ↗
Other ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES devices
- Vasoview Hemopro 2 VH-4000Maquet Cardiovascular LLC · Electrosurgical, Cutting & Coagulation & Accessories6 recalls1 death3,232 reports
- LigaSure LF1937Covidien LP · Electrosurgical, cutting & coagulation & accessories3 deaths2,090 reports
- Olympus WB91051WOLYMPUS Winter & Ibe GmbH · Electrosurgical, Cutting & Coagulation & Accessories1 death1,951 reports
- Valleylab FT10 VLFT10GENCovidien LP · Electrosurgical, cutting & coagulation & accessories1 recall5 deaths1,062 reports
- LigaSure LF1837Covidien LP · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES1 recall967 reports
- LigaSure LF2019Covidien LP · Electrosurgical, cutting & coagulation & accessories948 reports
- VAPR 225028Depuy Mitek, LLC · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES895 reports
- LigaSure LF1923Covidien LP · Electrosurgical, cutting & coagulation & accessories855 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page