NA 130187Recalls & reported problems
Conmed Corporation · ELECTROSURGICAL SUCTION COAGULATOR 10FR · Class 2
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 30653405059930), received between May 17, 2018 and Jun 10, 2025 — classified as 4 malfunctions.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (1 report).
1 FDA recall record name this device's identifier.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Recalls naming this device
- Nov 18, 2022Terminated
Suction ports may potentially be occluded on the affected devices. If this occurs, the affected devices will not be able to perform the suction function during use, which could cause a delay in procedure and could require another device to be used.
Z-0474-2023 · Root cause: Process control · FDA record ↗
Most reported problems
Device problems
- Detachment of Device or Device Component1
- Material Fragmentation1
- Sparking1
Patient problems
- Insufficient Information1
- No Clinical Signs, Symptoms or Conditions1
- No Known Impact Or Consequence To Patient1
- Superficial (First Degree) Burn1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 10, 2025Sparking; Sparking
THE CUSTOMER REPORTED THAT THE 130187, ELECTROSURGICAL SUCTION COAGULATOR 10F, DEVICE WAS BEING USED ON AN UNKNOWN DATE DURING AN ADENOIDECTOMY PROCEDURE WHEN IT WAS REPORTED ¿SURGEON REPORTS A SPARK AND A BURN THAT CAME "THROUGH THE INSULATED PART OF THE SUCTION BOVIE.¿. IT WAS ALSO REPORTED ¿PATIENT HAD A 1X2MM SUPERFICIAL BURN TO RIGHT UPPER LIP.¿. THE PROCEDURE WAS COMPLETED WITHOUT THE USE OF AN ALTERNATE DEVICE. THERE WAS NO REPORT OF MEDICAL INTERVENTION, OR EXTENDED HOSPITALIZATION FOR THE PATIENT OR USER. THIS REPORT IS BEING RAISED DUE TO THE REPORTED MALFUNCTION WITH POTENTIAL FOR INJURY UPON REOCCURRENCE.
Patient problems: Superficial (First Degree) Burn
MAUDE report 22178366 ↗MalfunctionJan 25, 2024Material Fragmentation; Material Fragmentation
THIS COMPLAINT ARISES FROM THE RECEIPT OF MEDWATCH UF/IMPORTER REPORT # (B)(4), IN WHICH A CUSTOMER REPORTED THAT THE 130187, SUCTION COAG, F/S, 10FR, 25/CS DEVICE WAS BEING USED DURING AN ADENOIDECTOMY PROCEDURE ON (B)(6) 2023, AND ¿CONMED 10FR SUCTION BOVIE PIECE OF PLASTIC BROKE OFF COMPLETELY AND WAS TEMPORARILY LOST. REPEAT BRONCH AND CONFIRMED THAT NO PLASTIC WAS LEFT IN THE PATIENT¿S AIRWAY. THE PIECE WAS LATER FOUND IN THE SUCTION LINE.¿. FURTHER ASSESSMENT CONFIRMED THAT THE ¿REPEAT BRONCH¿ OCCURRED DURING THE SAME REPORTED EVENT; THERE WAS NOT A SECOND PROCEDURE. THE PROCEDURE WAS COMPLETED AND THERE WAS NO INJURY, MEDICAL/SURGICAL INTERVENTION FOR PATIENT OR USER, AND NO EXTENDED HOSPITAL STAY FOR THE PATIENT. THIS REPORT IS BEING RAISED ON THE BASIS OF MALFUNCTION WITH POTENTIAL FOR INJURY UPON REOCCURRENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18581043 ↗MalfunctionJan 12, 2024
CONMED 10FR SUCTION BOVIE PIECE OF PLASTIC BROKE OFF COMPLETELY AND WAS TEMPORARILY LOST. REPEAT BRONCH AND CONFIRMED THAT NO PLASTIC WAS LEFT IN THE PATIENT¿S AIRWAY. THE PIECE WAS LATER FOUND IN THE SUCTION LINE.
Patient problems: Insufficient Information
MAUDE report 18501832 ↗MalfunctionMay 17, 2018Detachment of Device or Device Component; Detachment of Device or Device Component
THE USER FACILITY REPORTED THAT DURING A PROCEDURE, THE 130187 SUCTION/COAGULATION INSULATION COATING SPLIT. NO SURGICAL DELAYS OR PATIENT INJURY WERE REPORTED. TO DATE, NO ADDITIONAL INFORMATION REGARDING PATIENT STATUS OR PROCEDURE HAS BEEN MADE AVAILABLE. THIS INCIDENT IS BEING RAISED ON THE BASIS OF A REPORTED MALFUNCTION WITH POTENTIAL FOR INJURY WITH RECURRENCE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7522894 ↗
Other Electrosurgical, cutting & coagulation & accessories devices
- Vasoview Hemopro 2 VH-4000Maquet Cardiovascular LLC · Electrosurgical, Cutting & Coagulation & Accessories6 recalls1 death3,232 reports
- LigaSure LF1937Covidien LP · Electrosurgical, cutting & coagulation & accessories3 deaths2,090 reports
- Olympus WB91051WOLYMPUS Winter & Ibe GmbH · Electrosurgical, Cutting & Coagulation & Accessories1 death1,951 reports
- Valleylab FT10 VLFT10GENCovidien LP · Electrosurgical, cutting & coagulation & accessories1 recall5 deaths1,062 reports
- LigaSure LF1837Covidien LP · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES1 recall967 reports
- LigaSure LF2019Covidien LP · Electrosurgical, cutting & coagulation & accessories948 reports
- VAPR 225028Depuy Mitek, LLC · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES895 reports
- LigaSure LF1923Covidien LP · Electrosurgical, cutting & coagulation & accessories855 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.