UltraClean 139112Recalls & reported problems
Conmed Corporation · UltraClean Accessory Electrode 4" (10.16 cm) Coated Blade · Class 2
All models of UltraClean (8)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 40653405007235), received between Feb 17, 2022 and Mar 3, 2026 — classified as 1 injury, 5 malfunctions.
The most frequently reported specific device problem is “Delivered as Unsterile Product” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Delivered as Unsterile Product4
- Fracture1
- Material Fragmentation1
- Melted1
Patient problems
- No Clinical Signs, Symptoms or Conditions5
- Burn(s)1
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 3, 2026Material Fragmentation; Melted
THE SALES REPRESENTATIVE REPORTED ON BEHALF OF THE CUSTOMER THAT THE DEVICE, 139112, ULTRACLEAN ACCESSORY ELECTRODE 4" (10.16 CM) COATED BLADE, WAS BEING USED DURING AN UNKNOWN PROCEDURE ON (B)(6) 2026 WHEN IT WAS REPORTED, ¿DURING A PROCEDURE WITH A LOT OF CAUTERY THE 4 INCH GUARDED TIP SECTION APPEARED TO MELT ENOUGH TO THE POINT WHERE THE GUARDED SECTION AND THE GREY STABILITY HOLDER CAME APART FROM THE ACTUAL BLADE.¿. THERE WAS NO REPORT OF INJURY, MEDICAL INTERVENTION, OR HOSPITALIZATION FOR THE PATIENT. FURTHER ASSESSMENT WAS REQUESTED AND FOUND THAT THERE WAS FRAGMENTATION THAT FELL INTO THE SURGICAL SITE. THE FRAGMENTATION WAS RETRIEVED USING FORCEPS. THIS REPORT IS BEING RAISED DUE TO THE REPORTED MALFUNCTION WITH POTENTIAL FOR INJURY UPON REOCCURRENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24494751 ↗MalfunctionJun 1, 2023Delivered as Unsterile Product; Delivered as Unsterile Product
DURING INCOMING INSPECTION, THE DISTRIBUTOR REJECTED THIS DEVICE, 139112, FOR AN INSUFFICIENT HEATSEAL. THERE WAS NO CONTACT WITH THE PATIENT AS THIS WAS FOUND DURING INCOMING INSPECTION. THIS WILL BE REPORTED AS A MALFUNCTION WITH POTENTIAL FOR INJURY UPON REOCCURRENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17037724 ↗InjuryJun 20, 2022Fracture; Fracture
THE SALES REPRESENTATIVE REPORTED ON BEHALF OF THE CUSTOMER THAT THE DEVICE, 139112, ULTRACLEAN 4 IN. BLADE 50/CASE WAS BEING USED ON (B)(6) 2022 DURING A TRANSFORAMINAL LUMBAR INTERBODY FUSION, POSTERIOR SPINAL INSTRUMENTATION FUSION, BILATERAL SACROILIAC JOINT FUSION, AND HARDWARE REMOVAL AND "WHEN USING PRODUCT SKU 139112 THERE WAS NO VISIBLE INDICATION OF ISSUE WITH DEVICE UPON OPENING AND USING AT BEGINNING OF CASE. SURGEON BENT THE ELECTRODE AND IT WAS AFTER THAT SURGEON REALIZED THE INSULATION BREACH WHERE INSULATION MEETS PLASTIC ON ELECTRODE. SURGICAL TEAM EXCHANGED FOR ANOTHER 139112 AND SURGEON WAS ABLE TO COMPLETE THE PROCEDURE. STRAY BURN FROM INSULATION BREACH ON ELECTRODE. DRESSING APPLIED IN ORDERLY FASHION.¿ FURTHER ASSESSMENT FOUND THAT IT IS NOT BELIEVED THAT BENDING THE ELECTRODE CAUSED THE DEVICE TO BREAK. THE SURGEON HAS BEEN OPERATING AND USING THIS TECHNIQUE WITHOUT PREVIOUS ISSUE. PER SURGEON, ELECTRODE WAS USED WITH MINIMAL BENDING. THE BURN WAS ALONG THE INCISION, AND IT IS NOT KNOWN WHAT DEGREE OF BURN WAS RECEIVED. THE COVIDIEN FORCE FX ELECTROSURGICAL GENERATOR WAS USED AT 45 WATTS AND WILL BE LISTED AS A CONCOMITANT DEVICE. THERE WAS NO SIGNIFICANT DE
Patient problems: Burn(s)
MAUDE report 14742580 ↗MalfunctionMar 31, 2022Delivered as Unsterile Product; Delivered as Unsterile Product
DURING INCOMING INSPECTION, THE DISTRIBUTOR REJECTED THIS DEVICE, 139112, FOR AN INSUFFICIENT HEATSEAL. THERE WAS NO CONTACT WITH THE PATIENT AS THIS WAS FOUND DURING INCOMING INSPECTION. THIS WILL BE REPORTED AS A MALFUNCTION WITH POTENTIAL FOR INJURY UPON REOCCURRENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13969806 ↗MalfunctionMar 31, 2022Delivered as Unsterile Product; Delivered as Unsterile Product
DURING INCOMING INSPECTION, THE DISTRIBUTOR REJECTED THIS DEVICE, 138114A, FOR AN INSUFFICIENT HEATSEAL. THERE WAS NO CONTACT WITH THE PATIENT AS THIS WAS FOUND DURING INCOMING INSPECTION. THIS WILL BE REPORTED AS A MALFUNCTION WITH POTENTIAL FOR INJURY UPON REOCCURRENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13969770 ↗MalfunctionFeb 17, 2022Delivered as Unsterile Product; Delivered as Unsterile Product
DURING INCOMING INSPECTION, THE DISTRIBUTOR REJECTED THIS DEVICE, 139112, FOR AN INSUFFICIENT HEATSEAL. THERE WAS NO CONTACT WITH THE PATIENT AS THIS WAS FOUND DURING INCOMING INSPECTION. THIS WILL BE REPORTED AS A MALFUNCTION WITH POTENTIAL FOR INJURY UPON REOCCURRENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13548124 ↗
FDA clearance and classification
The FDA regulates product code GEI (electrosurgical, cutting & coagulation & accessories) under 21 CFR 878.4400 as a Class II device. 21 CFR 878.4400 ↗
Devices of this type normally need 510(k) clearance before sale.
Common questions about the UltraClean 139112
Has the UltraClean 139112 been recalled?
No FDA recall record in our data names this exact device identifier (DI 40653405007235), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the UltraClean 139112?
The FDA's MAUDE database lists 6 adverse event reports for this device, received between Feb 17, 2022 and Mar 3, 2026. In those reports, the most common specific device problems are “Delivered as Unsterile Product” (4), “Fracture” (1) and “Material Fragmentation” (1); the most common patient problem is “Burn(s)” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the UltraClean 139112?
None of the reports are classified as deaths, 1 as an injury and 5 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Who makes the UltraClean 139112, and is it still sold?
The FDA's device database lists Conmed Corporation as the company responsible for it (brand UltraClean, model 139112). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES devices
- Vasoview Hemopro 2 VH-4000Maquet Cardiovascular LLC · Electrosurgical, Cutting & Coagulation & Accessories6 recalls1 death3,137 reports
- LigaSure LF1937Covidien LP · Electrosurgical, cutting & coagulation & accessories3 deaths2,038 reports
- Olympus WB91051WOLYMPUS Winter & Ibe GmbH · Electrosurgical, Cutting & Coagulation & Accessories1 death1,891 reports
- Valleylab FT10 VLFT10GENCovidien LP · Electrosurgical, cutting & coagulation & accessories1 recall5 deaths1,046 reports
- LigaSure LF1837Covidien LP · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES1 recall957 reports
- LigaSure LF2019Covidien LP · Electrosurgical, cutting & coagulation & accessories929 reports
- VAPR 225028Depuy Mitek, LLC · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES874 reports
- LigaSure LF1923Covidien LP · Electrosurgical, cutting & coagulation & accessories831 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.