StarBurst 700-103902Recalls & reported problems
Navilyst Medical, Inc. · RITA StarBurst XL Electrosurgical Device · Class 2
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI H7877001039021), received between Oct 10, 2017 and Jun 10, 2020 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Device Contamination with Chemical or Other Material” (1 report).
3 other recalls cover devices cleared under the same FDA submission (K992693); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Device Contamination with Chemical or Other Material1
- Temperature Problem1
Patient problems
- Burn, Thermal1
- No Patient Involvement1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 10, 2020Temperature Problem; Temperature Problem
A DISTRIBUTOR REPORTED A PATIENT ISSUE WHEN USING A STARBURST XL15CM DEVICE. AFTER SETTING UP THE UNIT AND TAPING DISPERSIVE ELECTRODES TO THE PATIENT, THE PROCEDURE WAS STARTED BY OPENING THE ANTENNA 2CM. AFTER 10 MINUTES OF RADIOFREQUENCY ABLATION (RFA) ENERGY START, THE UNIT TEMPERATURE MONITOR ONLY DISPLAYED 50-60 DEGREES CELSIUS AND TEMPERATURE NOT TO TARGET TEMPERATURE (STABLE AT 50-60 DEGREE CELSIUS FOR ALL TEMPERATURE INDICATOR, WATT DELIVERY EQUAL WATT SETTING) . THE DOCTOR REQUESTED THE MACHINE BE POWERED OFF AND BACK ON. 10 MINUTES AFTER THE MACHINE WAS RESTARTED, THE MONITORED STILL DISPLAYED 50-60 DEGREES CELSIUS FOR RFA ENERGY. THE SCRUB NURSE LOOKED AT THE PATIENT'S SKIN AND NOTICED THERE WAS EVIDENCE OF A BURN, SO THE RFA ENERGY WAS STOPPED AND THE STARBURST XL ELECTROSURGICAL DEVICE WAS REMOVED FROM THE PATIENT. THE CASE WAS ABORTED. IT WAS REPORTED THAT THE DEVICE INVOLVED IN THE INCIDENT IS AVAILABLE TO BE RETURNED TO THE MANUFACTURER FOR EVALUATION.
Patient problems: Burn, Thermal
MAUDE report 10139511 ↗MalfunctionOct 10, 2017Device Contamination with Chemical or Other Material; Device Contamination with Chemical or Other Material
AS REPORTED TO ANGIODYNAMICS ON SEPTEMBER 21, 2017: DURING THE PREPARATION, IT WAS REPORTED THAT THE STARBURST XL PROBE TINES WERE DIFFICULT TO DEPLOY. THE PROBE WAS IRRIGATED/FLUSHED WITH STERILE SALINE PREPROCEDURE, AT WHICH TIME IT WAS OBSERVED THAT A DISCOLORED FLUID CAME FROM THE LINE. THE DOCTOR DETERMINED NOT TO USE ANOTHER XL NEEDLE AS A PRECAUTION DUE TO CONCERNS THE EVENT WOULD RE-OCCUR. IT WAS REPORTED THAT THE DEVICE INVOLVED IN THE INCIDENT IS AVAILABLE TO BE RETURNED TO THE MANUFACTURER FOR EVALUATION.
Patient problems: No Patient Involvement
MAUDE report 6935056 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K992693. One submission often covers many models, so check the product and lot details in each record.
- Z-1143-2015 · Dec 19, 2014 · TerminatedStarburst MRI Semi-Flex Electrode Device, 25cm/Attached Cable/14Ga/6.4F, Product No. H7877001039091, Catalog No. 700-103909; 25cm/14Ga/6.4F, Product No. H7877001026151, Catalog No.
- Z-0752-2007 · Apr 3, 2007 · TerminatedRITA Starburst XL (3-5 cm Diameter/25cm Length), Electro-surgical Device, Manufactured by: RITA Medical Systems, Inc., One Horizon Way, Manchester, GA 31816
- Z-0754-2007 · Apr 3, 2007 · TerminatedRITA Starburst MRI SemiFlex (3-5 cm Diameter/25cm Length), Electro-surgical Device, Manufactured by: RITA Medical Systems, Inc., One Horizon Way, Manchester, GA 31816
Other ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES devices
- Vasoview Hemopro 2 VH-4000Maquet Cardiovascular LLC · Electrosurgical, Cutting & Coagulation & Accessories6 recalls1 death3,232 reports
- LigaSure LF1937Covidien LP · Electrosurgical, cutting & coagulation & accessories3 deaths2,090 reports
- Olympus WB91051WOLYMPUS Winter & Ibe GmbH · Electrosurgical, Cutting & Coagulation & Accessories1 death1,951 reports
- Valleylab FT10 VLFT10GENCovidien LP · Electrosurgical, cutting & coagulation & accessories1 recall5 deaths1,062 reports
- LigaSure LF1837Covidien LP · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES1 recall967 reports
- LigaSure LF2019Covidien LP · Electrosurgical, cutting & coagulation & accessories948 reports
- VAPR 225028Depuy Mitek, LLC · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES895 reports
- LigaSure LF1923Covidien LP · Electrosurgical, cutting & coagulation & accessories855 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.