Skip to content

Shockwave IVL System with the Shockwave M5 Peripheral IVL Catheter M5IVL6060 (UDI M732M5IVL60600)Recalls & reported problems

Shockwave Medical, Inc. · The Shockwave M5 Intravascular Lithotripsy (IVL) Catheter consists of an array of miniaturized lithotripsy emitters, located inside a balloon, that generate acoustic pressure waves within the target treatment site, disrupting calcium within the lesion and allowing subsequent dilation of a peripheral artery stenosis using low balloon pressure. Two radiopaque marker bands inside the balloon aid in positioning. The proximal hub has three ports: one for inflation/deflation of the balloon with 50/50 saline/contrast medium, one for the ‘Over-the-wire’ (OTW) guidewire lumen, and one for connection to the IVL Connector Cable (and IVL Generator). · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI M732M5IVL60600), received between May 29, 2021 and Dec 17, 2021 — classified as 1 injury, 1 malfunction.

The most frequently reported specific device problem is “Material Split, Cut or Torn” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Material Split, Cut or Torn2
  • Detachment of Device or Device Component1
  • Difficult to Remove1
  • Material Separation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryDec 17, 2021Material Separation; Detachment of Device or Device Component; Material Split, Cut or Torn; Material Separation; Detachment of Device or Device Component; Material Split, Cut or Torn

    A SHOCKWAVE M5 PERIPHERAL LITHOTRIPSY (IVL) DEVICE WAS USED TO TREAT A SUPERFICIAL FEMORAL ARTERY (SFA). THE IVL CATHETER COMPLETED ALL 300 PULSES WITHOUT INCIDENT. THE PHYSICIAN DECIDED TO GO BACK TO POST-DILATE THE PROXIMAL POPLITEAL TO PROXIMAL SFA WITH THE SAME IVL BALLOON AT 8ATM, WHICH IS OUTSIDE THE INSTRUCTIONS FOR USE FOR THE M5 DEVICE. DURING THE FINAL INFLATION, A BALLOON LOSS OF PRESSURE OCCURRED. UPON ATTEMPTED REMOVAL OF THE IVL CATHETER, THE BALLOON PORTION WAS FOUND TO BE DETACHED FROM THE IVL CATHETER. ANGIOGRAM WAS USED TO LOCATE THE DETACHED BALLOON IN THE PROXIMAL SFA. THE PHYSICIAN ELECTED TO PERFORM A SURGICAL CUTDOWN OF THE SFA TO RETRIEVE THE DETACHED BALLOON, AND ALL DETACHED COMPONENTS WERE SUCCESSFULLY REMOVED FROM THE PATIENT. THE PHYSICIAN BELIEVES THAT A SHARP PIECE OF CALCIUM IN THE PROXIMAL SFA MAY HAVE CAUSED THE BALLOON LOSS OF PRESSURE AND SUBSEQUENT DETACHED COMPONENT. TO DATE, NO PATIENT SEQUELAE HAS BEEN REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13029364 ↗
  • MalfunctionMay 29, 2021Difficult to Remove; Material Split, Cut or Torn; Difficult to Remove; Material Split, Cut or Torn

    A M5 6.0X60 SHOCKWAVE IVL CATHETER WAS USED TO TREAT THE DISTAL SUPERFICIAL FEMORAL ARTERY (SFA) WITH 150 PULSES AT 4 ATM IN THE PRESENCE OF NODULAR CALCIUM. AFTER THAT, THE PRESSURE WAS INCREASED TO 10 ATMS, WHICH IS BEYOND THE RECOMMENDED TREATMENT PRESSURE OF 4 ATM OF THE LABEL (OFF-LABEL), AND PULSED AT WHICH TIME THE BALLOON RUPTURED. THE PHYSICIAN WAS WELL AWARE THAT INCREASING THE PRESSURE CAN LEAD TO UN-PREDICTABILITY WITHIN THE BALLOON, BUT ELECTED TO PROCEED ANYWAY. UPON REMOVAL OF THE DEVICE, THE BALLOON PORTION GOT STUCK AT THE END OF THE DESTINATION SHEATH. THE PHYSICIAN ATTEMPTED MULTIPLE METHODS TO REMOVE THE CATHETER, AND DURING ONE OF THOSE ATTEMPTS, THE BALLOON SHEARED OFF THE CATHETER. A SNARE WAS USED IN AN ATTEMPT TO SNARE AND REMOVE THE BALLOON PORTION, BUT ONLY A PORTION OF THE DETACHED BALLOON COULD BE SNARED. A COVERED STENT WAS IMPLANTED TO STENT THE DETACHED BALLOON AGAINST THE VESSEL WALL AND OUT OF THE LUMEN. THE PHYSICIAN BELIEVE THAT THE COMBINATION OF NODULAR CALCIUM AND PULSING AT 10ATM'S LEAD TO A COMPLETE TEAR IN THE MATERIAL OF THE BALLOON THAT WOULD NOT ALLOW IT TO RE-WRAP PROPERLY TO BE REMOVED FROM BODY. THERE HAVE BEEN NO PATIENT SEQUELAE REP

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11905519 ↗

Other Percutaneous Catheter, Ultrasound devices

All Percutaneous Catheter, Ultrasound devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page