Insulin Pump/Model recallZ-0193-2023 · Medtronic MiniMed · initiated Sep 20, 2022
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Medtronic MiniMed or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Medtronic identified a cybersecurity vulnerability in the MiniMed 600 series Insulin Pump Systems associated with the communication protocol that could allow unauthorized access to the pump system. This unauthorized access could be used to deliver too much or too little insulin through delivery of an unintended insulin bolus or because insulin delivery is slowed or stopped, which could lead to hypoglycemia or hyperglycemia. The Remote Bolus feature on insulin pumps should be turned off to prevent the unauthorized access.
FDA root cause category: Component change control
What the company is doing
On 09/20/2022, correction notices were distributed to customers, healthcare professionals, distributors, and service providers. They were informed to take the following actions: 1. Turn off the "Remote Bolus" feature on your pump if it is turned on. Note that the "Remote Bolus" capability is on by default, so take this action even if this feature has never been used. See Appendix: "How to Turn Off Remote Bolus Settings" or by visiting our website at www.medtronicdiabetes.com/remote-bolus 2. Conduct any connection linking of devices in a non-public place. 3. Complete and return the customer confirmation form. RECOMMENDED PRECAUTIONS: 1. Keep your pump and connected system components within your control at all times. 2. Be attentive to pump notifications, alarms, and alerts. 3. Immediately cancel any boluses you or your care partner did not initiate, monitor blood glucose levels closely and reach out to Medtronic 24-Hour Technical Support to report the bolus. NOTE: Turning off remote bolus feature will ensure no remote bolus is possible. 4. Disconnect the USB device from your computer when you're not using it to download pump data. 5. DO NOT confirm remote connection requests or any other remote action on the pump screen unless it is initiated by you or your care partner. 6. DO NOT share your pump's or devices' serial numbers with anyone other than your healthcare provider, distributors, and Medtronic. 7. DO NOT accept, calibrate, or bolus using a blood glucose reading you didn't initiate. 8. DO NOT connect to or allow any third-party devices to be connected to your pump 9. DO NOT use any software which has not been authorized by Medtronic as being safe for use with your pump. 10. Get medical help immediately when experiencing symptoms of severe hypoglycemia or diabetic ketoacidosis. 11. Call the recalling firm's 24-Hour Technical Support, 1-800-646-4633, option 1, if you suspect a pump setting or insulin delivery has changed unexpectedly, without your knowledge.
Recall details
- Recall number
- Z-0193-2023
- Status
- Open, Classified
- Initiated by the company
- Sep 20, 2022
- Posted by the FDA
- Nov 8, 2022
- Recalling firm
- Medtronic MiniMed
- Quantity
- 399651
- Product code
- OZP
- FDA submissions
- P160017, P150001
- Distribution
- US: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, GU, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VI, VT, WA, WI, WV, WY. OUS: Andorra, Australia, Austria, Barbados, Belgium, Bermuda, Bolivia, Botswana, Canada, Canary Islands, Cayman Islands, Chile, Czech Republic, Denmark, Dominican Rep., El Salvador, Faroe Islands, Finland, France, Germany, Gibraltar, Greece, Guatemala, Hungary, Ireland, Israel, Italy, Kenya, Liechtenstein, Luxembourg, Mauritius, Mexico, Namibia, Netherlands, New Zealand, Norway, Paraguay, Peru, Poland, Portugal, Romania, Slovakia, South Africa, Spain, Sweden, Switzerland, Trinidad and Tobago, Turkey, Uganda, United Kingdom, Venezuela
Product description
Insulin Pump/Model: MiniMed 630G/ MMT-1715, MMT-1755, MMT-1754; MiniMed 670G/ MMT-1780, MMT-1781, MMT-1782, MMT-1760, MMT-1761, MMT-1762, MMT-1740, MMT-1741, MMT-1742
Code information (models, lots, serial numbers)
Show the full code information (1k characters)
Model/UDI-GTIN(All Serial Numbers): MMT-1715/643000000000, 763000072520, 643169873834, 00643169656840; MMT-1755/643169752726, 643169873841, 643169878648, 763000179632, 00643169752726; MMT-1754/643169976344; MMT-1780/643169949706, 643169939202, 763000072537, 643000000000, 00643169939219; MMT-1781/763000187415, 763000283568, 00763000140601, 00763000225575, 00763000235765, 00763000187415, 00763000283551, 00763000240707; MMT-1782/763000187422, 763000283599, 00763000140618, 00763000225612, 00763000235772, 00763000187422, 00643169957640, 00763000283582; MMT-1760/643169946040, 763000278427, 763000179649, 643169946057, 763000367084, 00643169946057; MMT-1761/00763000367657, 00763000165413; MMT-1762/00763000239404, 00763000203313, 00763000216481, 00763000187408, 00763000165475; MMT-1740/Investigational; MMT-1741/00763000303556, 00763000408916; MMT-1742/00763000250232
Devices named in this recall
15 devices in the FDA device registry whose identifier appears in the recall's code information.
- MiniMed® 670G MMT-1780KPKMedtronic Minimed, Inc. · Artificial pancreas device system, single hormonal control3 recalls78 deaths55,723 reports
- MiniMed® 630G MMT-1715KMedtronic Minimed, Inc. · Artificial pancreas device system, threshold suspend4 recalls98 deaths35,170 reports
- MiniMed® 630G MMT-1715KMMedtronic Minimed, Inc. · Artificial pancreas device system, threshold suspend2 recalls50 deaths9,892 reports
- MiniMed® 670G MMT-1780KMedtronic Minimed, Inc. · Artificial pancreas device system, single hormonal control3 recalls7 deaths6,776 reports
- MiniMed® 630G MMT-1715KLMedtronic Minimed, Inc. · Artificial pancreas device system, threshold suspend4 recalls4 deaths1,294 reports
- MiniMed® 670G MMT-1780KLMedtronic Minimed, Inc. · Artificial pancreas device system, single hormonal control3 recalls1 death958 reports
- MiniMed® 630G MMT-1755KRMedtronic Minimed, Inc. · Automated insulin dosing device system, single hormonal control5 recalls13 reports
- MiniMed® 630G MMT-1755KMMedtronic Minimed, Inc. · Artificial pancreas device system, threshold suspend5 recalls0 reports
- MiniMed® 630G MMT-1755KMedtronic Minimed, Inc. · Artificial pancreas device system, threshold suspend4 recalls44 reports
- MiniMed® 630G MMT-1755KMKMedtronic Minimed, Inc. · Artificial pancreas device system, threshold suspend4 recalls0 reports
- MiniMed® 670G MMT-1760KPKMedtronic Minimed, Inc. · Artificial pancreas device system, single hormonal control3 recalls10 reports
- MiniMed® 670G MMT-1760KMedtronic Minimed, Inc. · Artificial pancreas device system, single hormonal control3 recalls3 reports
- MiniMed® 670G MMT-1760KPMedtronic Minimed, Inc. · Automated insulin dosing device system, single hormonal control3 recalls0 reports
- MiniMed® 670G MMT-1760KMMedtronic Minimed, Inc. · Automated insulin dosing device system, single hormonal control3 recalls0 reports
- MiniMed® 670G MMT-1760KRMedtronic Minimed, Inc. · Automated insulin dosing device system, single hormonal control3 recalls0 reports
Product lines
Other devices under the same FDA submission
80 devices cleared under P160017, P150001 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- MiniMed® 670G MMT-1780KPKMedtronic Minimed, Inc. · Automated insulin dosing device system, single hormonal control2 recalls51 deaths33,713 reports
- MiniMed®670G MMT-1780KPKMedtronic Minimed, Inc. · Artificial pancreas device system, single hormonal control3 recalls39 deaths28,789 reports
- MiniMed® 670G MMT-1780KLMedtronic Minimed, Inc. · Automated insulin dosing device system, single hormonal control3 recalls8 deaths8,353 reports
- MiniMed® 630G MMT-1715KMedtronic Minimed, Inc. · Automated insulin dosing , threshold suspend4 recalls36 deaths7,947 reports
- MiniMed®630G MMT-1715KMedtronic Minimed, Inc. · Artificial pancreas device system, threshold suspend2 recalls17 deaths6,645 reports
- MiniMed® 670G MMT-1780KPKMedtronic Minimed, Inc. · Automated insulin dosing device system, single hormonal control3 recalls10 deaths4,325 reports
- MiniMed® 630G MMT-1715KLMedtronic Minimed, Inc. · Automated insulin dosing , threshold suspend2 recalls10 deaths2,492 reports
- Guardian™ Sensor (3) MMT-7020AMedtronic Minimed, Inc. · Automated insulin dosing device system, single hormonal control1,304 reports
- MiniMed® 630G MMT-1715KRMedtronic Minimed, Inc. · Automated insulin dosing , threshold suspend9 deaths1,195 reports
- Enlite ® MMT-7008AMedtronic Minimed, Inc. · Automated insulin dosing , threshold suspend994 reports
- MiniMed® 670G MMT-1780KLMedtronic Minimed, Inc. · Automated insulin dosing device system, single hormonal control3 recalls2 deaths783 reports
- MiniMed® 630G MMT-1715KMedtronic Minimed, Inc. · Automated insulin dosing device system, single hormonal control3 recalls8 deaths781 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.