Skip to content

SHOCKPULSE SPL-PR150Recalls & reported problems

Gyrus ACMI, LLC · LITHOTRIPSY PROBE, REUSABLE, ShockPulse-SE 4.5 Fr (1.50 mm) x 564 mm · Class 2

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00821925043787), received between Oct 29, 2025 and Feb 23, 2026 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Break” (2 reports).

1 other recall cover devices cleared under the same FDA submission (K142428); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break2
  • Detachment of Device or Device Component1
  • Output Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • Insufficient Information1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 23, 2026Detachment of Device or Device Component; Output Problem

    IT WAS REPORTED THE SHOCKPULSE LITHOTRIPSY WHEN USING THE PROBE FOR LITHOTRIPSY OF A CONCREMENT OF THE LOWER URETER, AFTER THE THIRD PRESSING AND ACTIVATING THE PEDAL, THE PROBE STOPPED WORKING AND UPON INSPECTION THE DISTAL PART OF THE PROBE SEPARATED FROM THE PART THAT SHOULD BE ON THE INTRODUCER. THE ISSUE OCCURRED DURING A THERAPEUTIC LITHOTRIPSY PROCEDURE THAT WAS COMPLETED WITH A SIMILAR DEVICE. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24413603 ↗
  • MalfunctionDec 23, 2025Break; Break

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE SHOCKPULSE LITHOTRIPSY PROBE EXHIBITED THE THREAD WAS STUCK IN THE HAND PIECE AND RUST WAS DETECTED AT THE PROXIMAL END SIDE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: Insufficient Information

    MAUDE report 23882692 ↗
  • MalfunctionOct 29, 2025Break; Break

    NO ADDITIONAL INFORMATION RECEIVED FROM CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23418523 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K142428. One submission often covers many models, so check the product and lot details in each record.

  • Z-2197-2020 · Apr 23, 2020 · Terminated
    OLYMPUS ShockPulse-SE Lithotripsy System, Model SPL-T Transducer

Other Lithotriptor, Ultrasonic devices

All Lithotriptor, Ultrasonic devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.