Medacta Quadra Trial Broach size 0 part # 01.10.10.045, for… recallZ-0791-2013 · Medacta Usa Inc · initiated Dec 17, 2012

Class II Terminated All recalls posted in February 2013

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Medacta Usa Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The tip of the Quadra Trial Broach size 0, broke during a surgery.

FDA root cause category: Device Design

What the company is doing

Medacta sent a Urgent Safety Notification letter dated December 13, 2012 to all affected customers. The letter identified the recalled product, problem and actions to be taken. Customers were instructed to check inventory for affected product and contact [email protected] for an RMA# to return items to Medacta USA. A new version of the recalled product will be provided to all affected customers. For questions call 805-437-7085 x26.

Recall details

Recall number
Z-0791-2013
Recall class
Class II
Status
Terminated
Initiated by the company
Dec 17, 2012
Posted by the FDA
Feb 7, 2013
Recalling firm
Medacta Usa Inc
Quantity
26 pieces U.S.A. (226 pieces Worldwide)
Product code
MEH
FDA submissions
K072857, K082792, K083558, K093944, K103189
Distribution
US Nationwide Distribution including the states of Arizona, California, Idaho, Illinois, North Carolina, Ohio, Utah, Washington and US Virgin Islands

Product description

Medacta Quadra Trial Broach size 0 part # 01.10.10.045, for use with the Medacta Total Hip Prosthesis System. Product Usage - used to broach the intramedullary canal of the femur in order to create space for the femoral stem during a hip replacement surgery.

Code information (models, lots, serial numbers)

Lot/Unit Codes: 075023, 075064, 084000, 084001, 084326, 084743, 084815, 085202, 085455

Other devices under the same FDA submission

98 devices cleared under K072857, K082792, K083558, K093944, K103189 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.