Skip to content

CS 100i Intra-Aortic Balloon Pump recallZ-2736-2017 · Maquet Datascope Corp - Cardiac Assist Division · initiated Jun 16, 2017

Class I Terminated All recalls posted in June 2017

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Maquet Datascope Corp - Cardiac Assist Division or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The device failed to pump due to an electrical test failure code #58 (power up vent test fail), maintenance code #3, and an autofill failure which has been associated to a patient death due to the failure of the device to initiate therapy.

FDA root cause category: Component design/selection

What the company is doing

Datascope Corp/MAQUET issued a press release on June 16, 2017. Datascope Corp/MAQUET sent an Urgent Product Recall Letter on June 19, 2017, to their customers via FedEx next day delivery service. Pursuant to the WARNINGS section of our CS100i, CS100 or CS300 IABP Operating/User Instructions, clinicians are instructed not to leave the patient unattended during IABP therapy. Until the service is performed, we recommend the IABP be powered on prior to initiation of therapy to allow the IABP to successfully complete its self-test. This action will take less than 60 seconds to perform. In the event the IABP fails to successfully complete the self-test and exhibits electrical test failure code 58, please remove the IABP from service and contact your local Maquet/Getinge Sales & Service Office. Customers will be instructed to complete and return the attached Medical Device Field Correction Response Form via fax to 1-973-807-9217 or send via email to [email protected]@getinge.com. If you are a distributor who has shipped any affected products to customers, please forward this document to their attention for appropriate action. For U.S. customers with technical questions, please contact our Technical Support Department (at 1-888-627-8383 and press 3), Monday through Friday, between the hours of 8:00 a.m. and 6:00 p.m. EST

Recall details

Recall number
Z-2736-2017
Recall class
Class I (most serious)
Status
Terminated
Initiated by the company
Jun 16, 2017
Posted by the FDA
Jun 23, 2017
Recalling firm
Maquet Datascope Corp - Cardiac Assist Division
Quantity
12,319 units total
Product code
DSP
FDA submissions
K031636
Distribution
Worldwide Distribution - (in over 100 countries) US (Nationwide)

Product description

CS 100i Intra-Aortic Balloon Pump

Code information (models, lots, serial numbers)

0998-UC-0446HXX 0998-UC-0479HXX

Other devices under the same FDA submission

21 devices cleared under K031636 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 77566).

Sources and updates