Gyrus ACMI Reusable Active Cord for Bovie/ValleyLab… recallZ-1952-2018 · Gyrus Acmi, Incorporated · initiated Apr 17, 2018
Terminated
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Gyrus Acmi, Incorporated or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
There is a potential for the cord to spark and cause a fire.
FDA root cause category: Device Design
What the company is doing
Notification letters were distributed on April 17, 2018. The letters instruct customers to perform the following: Immediately assess any affected product you have in stock and replace the IFU with the revised IFU. Also ensure that any potential users of the devices are aware of the additional warning in the revised IFU. Future versions of the IFU will include the warning. Remove and discard all cords in use greater than one year. If you need additional copies of the IFU, call your Olympus customer service representative at 1-888-524-7266 to arrange for them to be provided. Please note on the enclosed Reply Form that you have received this information. Fax the completed Reply Form to 484-896-7128 regardless of whether you have any affected inventory at your facility. In addition, if you may have further distributed this product, please identify your customers, inform them at once of this action, provide the revised IFU and appropriately document your notification process. Your notification to your customers may be enhanced by including a copy of this letter. For further questions, please call (508) 804-2600.
Recall details
- Recall number
- Z-1952-2018
- Status
- Terminated
- Initiated by the company
- Apr 17, 2018
- Recalling firm
- Gyrus Acmi, Incorporated
- Quantity
- 6447
- Product code
- FAS
- FDA submissions
- K890328
- Distribution
- Worldwide Distribution - USA (nationwide) Distribution to the states of : AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, WA, WI, WV, and WY., and to the countries of : Japan, Canada, Mexico, Australia, and EU.
Product description
Gyrus ACMI Reusable Active Cord for Bovie/ValleyLab Generators, Catalog No C650-129A (00821925008700) The cord consists of a single conductor stranded wire terminated with plugs that connect to a generator on one end and a plug that connects to a Gyrus ACMI working element on the other end. They are provided non-sterile and are reusable. The cord is designed to be used with Gyrus ACMI GOCUS 640, Bovie, and other general purpose monopolar electrosurgical generators and appropriate electrodes and endoscopes.
Code information (models, lots, serial numbers)
All lots manufactured before December 7, 2017 UDI (00821925008700)
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Classic Series C650-129AGyrus ACMI, LLC · Hysteroscope (And Accessories)1 recall4 reports
Other devices under the same FDA submission
39 devices cleared under K890328 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- M3 Gold M3-30AGyrus ACMI, LLC · ELECTRODE, ELECTROSURGICAL, ACTIVE, UROLOGICAL1 recall11 reports
- USA Elite MLE-24-012Gyrus ACMI, LLC · ELECTRODE, ELECTROSURGICAL, ACTIVE, UROLOGICAL1 recall9 reports
- Blue Silicone CS-B612Gyrus ACMI, LLC · Hysteroscope (And Accessories)1 recall2 reports
- USA Elite RBGyrus ACMI, LLC · ELECTRODE, ELECTROSURGICAL, ACTIVE, UROLOGICAL1 recall1 report
- Gyrus ACMI 194Gyrus ACMI, LLC · Hysteroscope (And Accessories)1 recall0 reports
- USA Elite 190-NRSGyrus ACMI, LLC · Hysteroscope (And Accessories)1 recall0 reports
- M3 Gold M3-70AGyrus ACMI, LLC · ELECTRODE, ELECTROSURGICAL, ACTIVE, UROLOGICAL1 recall0 reports
- Gyrus ACMI CS-G7Gyrus ACMI, LLC · Hysteroscope (And Accessories)1 recall0 reports
- USA Elite EIS-HCF25Gyrus ACMI, LLC · Hysteroscope (And Accessories)25 reports
- M3 Gold REXM3-30AGyrus ACMI, LLC · Hysteroscope (And Accessories)14 reports
- USA Elite EMRS-26BGyrus ACMI, LLC · Hysteroscope (And Accessories)9 reports
- Elite 2 System E2OS-CF25Gyrus ACMI, LLC · Hysteroscope (And Accessories)7 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page