Edwards Lifesciences Swan Ganz Thermodilution Catheter, REF… recallZ-0764-2019 · Edwards Lifesciences, LLC · initiated Dec 21, 2018
Class I Terminated All recalls posted in January 2019
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Edwards Lifesciences, LLC or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
These Swan Ganz catheters have incorrect lumen assembly, which causes reversal of the lumens. If the lumens are reversed the clinician may note reverse PA and CVP pressure values and waveforms.
FDA root cause category: Nonconforming Material/Component
What the company is doing
An Urgent Product Recall letter was sent via Fed-Ex to affected customers on 12/21/2018. A second communication was sent via Fed-Ex on 01/10/19. The letter identified the affected product, problem and actions to be taken. Customers were instructed to the return any unused units that are currently in their inventory. Once returned, replacement product will be shipped at no charge. Customers were asked to complete the attached acknowledgement form. For questions contact Technical Support at 1-800-822-8837 option 1.
Recall details
- Recall number
- Z-0764-2019
- Recall class
- Class I (most serious)
- Status
- Terminated
- Initiated by the company
- Dec 21, 2018
- Posted by the FDA
- Jan 31, 2019
- Recalling firm
- Edwards Lifesciences, LLC
- Quantity
- 9 units
- Product code
- DYG
- FDA submissions
- K160084
- Distribution
- Worldwide Distribution
Product description
Edwards Lifesciences Swan Ganz Thermodilution Catheter, REF 131VF7P Product Usage: The Swan-Ganz Thermodilution Catheter provides a diagnostic tool for physicians to rapidly determine hemodynamic pressures and cardiac output when used with a compatible cardiac output computer. Swan-Ganz thermodilution catheters are indicated for the assessment of a patient s hemodynamic condition through direct intra-cardiac and pulmonary artery pressure monitoring, cardiac output determination, and for infusing solutions. The distal (pulmonary artery) port also allows sampling of mixed venous blood for the assessment of oxygen transport balance and the calculation of derived parameters such as oxygen consumption, oxygen utilization coefficient, and intrapulmonary shunt fraction.
Code information (models, lots, serial numbers)
Lot Number: 61176314
Other devices under the same FDA submission
20 devices cleared under K160084 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- SWAN-GANZ CCOMBO V 774F75Edwards Lifesciences LLC · CATHETER, FLOW DIRECTED2 recalls2 deaths41 reports
- SWAN-GANZ VIP 831F75Edwards Lifesciences LLC · CATHETER, FLOW DIRECTED1 recall23 reports
- SWAN-GANZ CONTROLCATH C146F7Edwards Lifesciences LLC · CATHETER, INTRAVASCULAR, DIAGNOSTIC1 recall16 reports
- SWAN-GANZ 131F7Edwards Lifesciences LLC · CATHETER, INTRAVASCULAR, DIAGNOSTIC1 recall13 reports
- SWAN-GANZ VIP+ 834F75Edwards Lifesciences LLC · CATHETER, FLOW DIRECTED1 recall12 reports
- SWAN-GANZ VIP 831F75Edwards Lifesciences LLC · CATHETER, FLOW DIRECTED1 recall6 reports
- SWAN-GANZ VIP 782F75MEdwards Lifesciences LLC · Catheter, continuous flush1 recall4 reports
- SWAN-GANZ HI-SHORE TRUE SIZE 141F7Edwards Lifesciences LLC · CATHETER, INTRAVASCULAR, DIAGNOSTIC1 recall3 reports
- SWAN-GANZ 131F7Edwards Lifesciences LLC · Catheter, Intravascular, Diagnostic1 recall2 reports
- SWAN-GANZ TRUE SIZE 151F7Edwards Lifesciences LLC · CATHETER, INTRAVASCULAR, DIAGNOSTIC1 recall2 reports
- SWAN-GANZ TRUE SIZE TS105F5Edwards Lifesciences LLC · CATHETER, FLOW DIRECTED1 recall2 reports
- SWAN-GANZ CONTROLCATH TRUE SIZE C144F7Edwards Lifesciences LLC · CATHETER, INTRAVASCULAR, DIAGNOSTIC1 recall1 report
Other products in the same recall event
The FDA groups related products into one recall event (event 81878).
- Z-0762-2019 · TerminatedEdwards Lifesciences Swan Ganz Thermodilution Catheter, REF131F 7J Product Usage: The Swan-Ganz Thermodilution Catheter provides a diagnostic tool for physicians to rapidly det…
- Z-0763-2019 · TerminatedEdwards Lifesciences Swan Ganz Thermodilution Catheter, REF 131F7P (5 pack of 131F7) Product Usage: The Swan-Ganz Thermodilution Catheter provides a diagnostic tool for physicians…
- Z-0765-2019 · TerminatedEdwards Lifesciences Swan Ganz Thermodilution Catheter, REF 151F7 Product Usage: The Swan-Ganz Thermodilution Catheter provides a diagnostic tool for physicians to rapidly dete…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page