Skip to content

Edwards Lifesciences Swan Ganz Thermodilution Catheter, REF… recallZ-0765-2019 · Edwards Lifesciences, LLC · initiated Dec 21, 2018

Class I Terminated All recalls posted in January 2019

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Edwards Lifesciences, LLC or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

These Swan Ganz catheters have incorrect lumen assembly, which causes reversal of the lumens. If the lumens are reversed the clinician may note reverse PA and CVP pressure values and waveforms.

FDA root cause category: Nonconforming Material/Component

What the company is doing

An Urgent Product Recall letter was sent via Fed-Ex to affected customers on 12/21/2018. A second communication was sent via Fed-Ex on 01/10/19. The letter identified the affected product, problem and actions to be taken. Customers were instructed to the return any unused units that are currently in their inventory. Once returned, replacement product will be shipped at no charge. Customers were asked to complete the attached acknowledgement form. For questions contact Technical Support at 1-800-822-8837 option 1.

Recall details

Recall number
Z-0765-2019
Recall class
Class I (most serious)
Status
Terminated
Initiated by the company
Dec 21, 2018
Posted by the FDA
Jan 31, 2019
Recalling firm
Edwards Lifesciences, LLC
Quantity
459 units
Product code
DYG
FDA submissions
K160084
Distribution
Worldwide Distribution

Product description

Edwards Lifesciences Swan Ganz Thermodilution Catheter, REF 151F7 Product Usage: The Swan-Ganz Thermodilution Catheter provides a diagnostic tool for physicians to rapidly determine hemodynamic pressures and cardiac output when used with a compatible cardiac output computer. Swan-Ganz thermodilution catheters are indicated for the assessment of a patient s hemodynamic condition through direct intra-cardiac and pulmonary artery pressure monitoring, cardiac output determination, and for infusing solutions. The distal (pulmonary artery) port also allows sampling of mixed venous blood for the assessment of oxygen transport balance and the calculation of derived parameters such as oxygen consumption, oxygen utilization coefficient, and intrapulmonary shunt fraction.

Code information (models, lots, serial numbers)

Lot Numbers: 61176367, 61176370

Other devices under the same FDA submission

20 devices cleared under K160084 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 81878).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page