Hudson RCI Sheridan LTS, Product Code 5-11112 recallZ-1759-2019 · Teleflex Medical · initiated May 24, 2019

Class I Terminated All recalls posted in June 2019

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Teleflex Medical or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

This voluntary recall is due to reported complaints (<0.0025% of all in scope distributed product) indicating that there is an increased incidence of specific lots of the 15 mm Sheridan connector becoming disconnected from the Endotracheal tube.

FDA root cause category: Process control

What the company is doing

Teleflex sent an Urgent Medical Device Recall letter dated May 24, 2019 to customers. The letter identified the affected product, problem and actions to be taken. The letter directed customers to discontinue use, quarantine the product, and contact Teleflex for return of the product. Distributors were directed to notify their customers and request the customer return the recalled products to them for consolidation and subsequent return to Teleflex.. For questions contact your local sales representative or Customer Service at 1-866-396-2111

Recall details

Recall number
Z-1759-2019
Recall class
Class I (most serious)
Status
Terminated
Initiated by the company
May 24, 2019
Posted by the FDA
Jun 19, 2019
Recalling firm
Teleflex Medical
Quantity
3483 units
Product code
BTR
FDA submissions
K822082
Distribution
Worldwide - US Nationwide Distribution

Product description

Hudson RCI Sheridan LTS, Product Code 5-11112 Product Usage: Tracheal tube/airway management

Code information (models, lots, serial numbers)

Lot Numbers: 73K1600369, 73L1600011, 73A1700646, 73C1700238, 73D1700625, 73E1700475, 73F1700387, 73H1700041, 73H1700483, 73J1700292, 73L1700503, 73M1700453, 73A1800671, 73B1800350, 73D1800138, 73F1800174, 73G1800126, 73G1800806, 73J1800308, 73K1800173, 73L1800225, 73K1800172

Other devices under the same FDA submission

60 devices cleared under K822082 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 82964).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page