Skip to content

Hudson RCI Sheridan recallZ-1763-2019 · Teleflex Medical · initiated May 24, 2019

Class I Terminated All recalls posted in June 2019

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Teleflex Medical or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

This voluntary recall is due to reported complaints (<0.0025% of all in scope distributed product) indicating that there is an increased incidence of specific lots of the 15 mm Sheridan connector becoming disconnected from the Endotracheal tube.

FDA root cause category: Process control

What the company is doing

Teleflex sent an Urgent Medical Device Recall letter dated May 24, 2019 to customers. The letter identified the affected product, problem and actions to be taken. The letter directed customers to discontinue use, quarantine the product, and contact Teleflex for return of the product. Distributors were directed to notify their customers and request the customer return the recalled products to them for consolidation and subsequent return to Teleflex.. For questions contact your local sales representative or Customer Service at 1-866-396-2111

Recall details

Recall number
Z-1763-2019
Recall class
Class I (most serious)
Status
Terminated
Initiated by the company
May 24, 2019
Posted by the FDA
Jun 19, 2019
Recalling firm
Teleflex Medical
Quantity
14357
Product code
BTR
FDA submissions
K822082
Distribution
Worldwide - US Nationwide Distribution

Product description

Hudson RCI Sheridan: 1) Uncuffed 6.0 mm, Product Code 5-10412 2) Uncuffed 6.5 mm, Product Code 5-10413 3) Uncuffed 7.0 mm, Product Code 5-10414 Product Usage: Tracheal tube/airway management

Code information (models, lots, serial numbers)

1) Uncuffed 6.0 mm, Product Code 5-10412 Lot Numbers: 73J1600363 73K1600581 73L1600347 73L1600552 73L1600850 73M1600138 73A1700264 73B1700300 73C1700688 73E1700098 73E1700712 73G1700132 73H1700327 73J1700290 73K1700065 73K1700415 73L1700169 73L1700320 73M1700250 73A1800060 73B1800348 73B1800487 73C1800349 73C1800482 73D1800464 73E1800389 73F1800171 73G1800923  2) Uncuffed 6.5 mm, Product Code 5-10413 Lot Numbers: 73M1600139 73B1700301 73E1700472 73G1700127 73K1700416 73A1800061 73B1800349 73B1800488 73D1800634 73F1800544 73G1800924  3) Uncuffed 7.0 mm, Product Code 5-10414 Lot Numbers: 73A1700100 73F1700064 73L1800406

Other devices under the same FDA submission

60 devices cleared under K822082 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 82964).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page