Hudson RCI Sheridan recallZ-1761-2019 · Teleflex Medical · initiated May 24, 2019
Class I Terminated All recalls posted in June 2019
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Teleflex Medical or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
This voluntary recall is due to reported complaints (<0.0025% of all in scope distributed product) indicating that there is an increased incidence of specific lots of the 15 mm Sheridan connector becoming disconnected from the Endotracheal tube.
FDA root cause category: Process control
What the company is doing
Teleflex sent an Urgent Medical Device Recall letter dated May 24, 2019 to customers. The letter identified the affected product, problem and actions to be taken. The letter directed customers to discontinue use, quarantine the product, and contact Teleflex for return of the product. Distributors were directed to notify their customers and request the customer return the recalled products to them for consolidation and subsequent return to Teleflex.. For questions contact your local sales representative or Customer Service at 1-866-396-2111
Recall details
- Recall number
- Z-1761-2019
- Recall class
- Class I (most serious)
- Status
- Terminated
- Initiated by the company
- May 24, 2019
- Posted by the FDA
- Jun 19, 2019
- Recalling firm
- Teleflex Medical
- Quantity
- 55312 units
- Product code
- BTR
- FDA submissions
- K822082
- Distribution
- Worldwide - US Nationwide Distribution
Product description
Hudson RCI Sheridan: 1) EZ-ENDO 6.0 mm, Product Code 5-22512 2) EZ-ENDO 6.5 mm, Product Code 5-22513 3) EZ-ENDO 7.0 mm, Product Code 5-22514 4) EZ-ENDO 7.5 mm, Product Code 5-22515 5) EZ-ENDO 8.0 mm, Product Code 5-22516 6) EZ-ENDO 8.5 mm, Product Code 5-22517 Product Usage: Tracheal tube/airway management
Code information (models, lots, serial numbers)
Show the full code information (1k characters)
1) EZ-ENDO 6.0 mm, Product Code 5-22512 Lot Numbers: 73K1600152 73M1600163 73A1700280 73C1700567 73E1700496 73F1700403 73H1700347 73M1700272 73E1800401 73E1800575 73F1800354 2) EZ-ENDO 6.5 mm, Product Code 5-22513 Lot Numbers: 73J1600766 73B1700155 73C1700568 73F1700404 73G1700432 73H1700348 73E1800576 73F1800355 3) EZ-ENDO 7.0 mm, Product Code 5-22514 Lot Numbers: 73A1700118 73C1700569 73E1700722 73F1700175 73F1700405 73G1700433 73H1700349 73H1700494 73H1700711 73J1700143 73K1700081 73K1700610 73L1700181 73M1700273 73D1800653 73F1800562 73F1800720 73H1800450 73J1800531 73K1800328 73K1800786 4) EZ-ENDO 7.5 mm, Product Code 5-22515 Lot Numbers: 73B1700316 73C1700262 73C1700418 73E1700723 73E1700104 73F1700176 73F1700406 73G1700434 73H1700350 73J1700302 73J1700497 73K1700747 73M1700274 73A1800068 73A1800683 73B1800502 73C1800201 73D1800154 73E1800402 73F1800565 73G1800591 73H1800451 73H1800686 73J1800113 73K1800736 5) EZ-ENDO 8.0 mm, Product Code 5-22516 Lot Numbers: 73K1600594 73A1700282 73B1700158 73B1700317 73C1700570 73F1700036 73F1700177 73F1700407 73G1700614 73H1700495 73J1700487 73K1700082 73A1800684 73B1800503 73C1800371 73E1800577 73G1800232 73J1800532 73K1800327 6) EZ-ENDO 8.5 mm, Product Code 5-22517 Lot Numbers: 73J1600767 73C1800202
Other devices under the same FDA submission
60 devices cleared under K822082 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- HUDSON RCI IPN044731Teleflex Incorporated · Tube, tracheal (w/wo connector)42 reports
- HUDSON RCI IPN048896Teleflex Incorporated · Tube, tracheal (w/wo connector)2 deaths32 reports
- HUDSON RCI IPN044767Teleflex Incorporated · Tube, tracheal (w/wo connector)31 reports
- HUDSON RCI IPN048895Teleflex Incorporated · Tube, tracheal (w/wo connector)1 death27 reports
- HUDSON RCI IPN046756Teleflex Incorporated · Tube, tracheal (w/wo connector)25 reports
- HUDSON RCI IPN048873Teleflex Incorporated · Tube, tracheal (w/wo connector)25 reports
- HUDSON RCI IPN044732Teleflex Incorporated · Tube, tracheal (w/wo connector)21 reports
- HUDSON RCI IPN044730Teleflex Incorporated · Tube, tracheal (w/wo connector)20 reports
- HUDSON RCI IPN048874Teleflex Incorporated · Tube, tracheal (w/wo connector)20 reports
- HUDSON RCI IPN046759Teleflex Incorporated · Tube, tracheal (w/wo connector)19 reports
- HUDSON RCI IPN044768Teleflex Incorporated · Tube, tracheal (w/wo connector)17 reports
- HUDSON RCI IPN048897Teleflex Incorporated · Tube, tracheal (w/wo connector)15 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 82964).
- Z-1759-2019 · TerminatedHudson RCI Sheridan LTS, Product Code 5-11112 Product Usage: Tracheal tube/airway management
- Z-1760-2019 · TerminatedHudson RCI Sheridan: 1) CF 6.0 mm, Product Codes: 5-10112, 5-10212 2) CF 6.5 mm, Product Codes: 5-10113, 5-10213 3) CF 7.0 mm, Product Codes: 5-10114, 5-10214 4) CF 7.5 mm, …
- Z-1762-2019 · TerminatedHudson RCI Sheridan: 1) Preformed 6.0 mm, Product Codes: a) 5-22212, b) 5-22312, c) 5-22112, d) 5-22012, 2) Preformed 6.5 mm, Product Codes: a) 5-22313, b) 5-222…
- Z-1763-2019 · TerminatedHudson RCI Sheridan: 1) Uncuffed 6.0 mm, Product Code 5-10412 2) Uncuffed 6.5 mm, Product Code 5-10413 3) Uncuffed 7.0 mm, Product Code 5-10414 Product Usage: Tracheal tube/…
- Z-1764-2019 · TerminatedHudson RCI Sheridan: 1) HVT 6.0 mm, Product Code 5-10312 2) HVT 6.5 mm, Product Code 5-10313 3) HVT 7.0 mm, Product Code 5-10314 4) HVT 7.5 mm, Product Code 5-10315 5) HVT 8.0…
- Z-1765-2019 · TerminatedSheridan/CF Novaplus 7.0 mm, Product Code V5-10114 Product Usage: Tracheal tube/airway management
- Z-1766-2019 · TerminatedSheridan/HVT: 1) Novaplus 7.0 mm, Product Code V5-10314 2) Novaplus 7.5 mm, Product Code V5-10315 3) Novaplus 8.0 mm, Product Code V5-10316 4) Novaplus 8.5 mm, Product Code V5-…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page