Optima Coil System recallZ-1680-2022 · Balt USA, LLC · initiated Jul 26, 2022
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Balt USA, LLC or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Products were released to customers prior to the approval and release of a test report authorizing the use of an existing material from an alternate supplier.
FDA root cause category: Under Investigation by firm
What the company is doing
On or about 07/15/2022, Balt USA LLC communicated an "URGENT FIELD SAFETY NOTICE (FSN)/PRODUCT RECALL" Letter informing customers that Balt became aware that finished goods were released and distributed to customers prior to the approval and release of a test report authorizing the use of existing material from an alternate supplier. On 08/08/2022 and updated/revised "URGENT MEDICAL DEVICE RECALL" letter was sent via certified mail to customers re-emphasizing the recall and providing updated instructions Customer are instructed to: -Inform within their hospital safety officers, pharmacists, head of neuroradiology and the neuroradiology department staff, as well as any other person. -Return to Balt USA any unused product from lot #F220600470: +Collect and put in quarantine the Optima products concerned by this recall and then return them to Balt USA through the usual "RMA" (Return Material Authorization) procedure by contacting Balt USA's Customer Service Department +Keepj informed Balt USA about the status of every unit of Optima products concerned by this recall; +Fulfill the "Notice of receipt" then return to Balt USA via the indicated contact. -If product from lot #F220600470 has been used, no further action is required. For question contact - Quality Department at [email protected] or telephone number 949-788-1443 or Fax Number: 949-788-1444
Recall details
- Recall number
- Z-1680-2022
- Status
- Open, Classified
- Initiated by the company
- Jul 26, 2022
- Recalling firm
- Balt USA, LLC
- Quantity
- 4 devices
- Product code
- HCG
- FDA submissions
- K172390, K200030
- Distribution
- U.S.: AZ and FL O.U.S.: None
Product description
Optima Coil System Model OPTI0520CSF10. For endovascular embolization.
Code information (models, lots, serial numbers)
Model Number: OPTI0520CSF10 UDI-DI Code: 00818053026164 Lot Number: F220600470
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Optima Coil System OPTI0520CSF10Balt USA LLC · Device, Vascular, For Promoting Embolization2 recalls1 report
Product lines
Other devices under the same FDA submission
399 devices cleared under K172390, K200030 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- Optima Coil System OPTI0208CSS10Balt USA LLC · Device, Neurovascular Embolization3 recalls3 reports
- Optima Coil System OPTI0308CSS10Balt USA LLC · Device, Neurovascular Embolization2 recalls6 reports
- Optima Coil System OPTI0720CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls4 reports
- Optima Coil System OPTI0930CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls4 reports
- Optima Coil System OPTI0413CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls3 reports
- Optima Coil System OPTI0203CSS10Balt USA LLC · Device, Neurovascular Embolization2 recalls3 reports
- Optima Coil System OPTI0256CSS10Balt USA LLC · Device, Neurovascular Embolization2 recalls3 reports
- Optima Coil System OPTI0410CSS10Balt USA LLC · Device, Neurovascular Embolization2 recalls3 reports
- Optima Coil System OPTI0413CSF10Balt USA LLC · Device, Neurovascular Embolization2 recalls3 reports
- Prestige Coil System PRES0520HPKPLBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls3 reports
- Optima Coil System OPTI0407CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls2 reports
- Optima Coil System OPTI4510CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls2 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page