INFINITY Occipitocervical Upper Thoracic System recallZ-1409-2023 · Medtronic Sofamor Danek USA Inc · initiated Feb 16, 2023
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Medtronic Sofamor Danek USA Inc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Incorrect type of titanium was used which reduces the gripping strength of the screw head of 5 batches of Infinity Multi-Axial Screws
FDA root cause category: Under Investigation by firm
What the company is doing
The firm issued an URGENT MEDICAL DEVICE RECALL NOTICE to its consignees on 02/16/2023 by UPS 2nd day mail. The letter explained the issue and the risk and requested the following actions be taken: Immediately identify, segregate, and quarantine affected products within your inventory. Pass this notice to all those who need to be aware within your organization or to any organization where the potentially affected devices have been transferred. Contact Medtronic to return affected product: 1-800-933-2635 to receive an return authorization (RGA) and coordinate inventory replacement/credit. Reference FA1315. The letter also includes specific patient recommendations.
Recall details
- Recall number
- Z-1409-2023
- Status
- Open, Classified
- Initiated by the company
- Feb 16, 2023
- Posted by the FDA
- Apr 19, 2023
- Recalling firm
- Medtronic Sofamor Danek USA Inc
- Quantity
- 45 units
- Product code
- KWP
- FDA submissions
- K163375
- Distribution
- Worldwide distribution - US Nationwide and the country of Canada.
Product description
INFINITY Occipitocervical Upper Thoracic System, MULTI AXIAL SCREW, Size: 4.0mm x 24mm, REF 3604024; for immobilization and stabilization of spinal segment as an adjunct to fusion for the following acute and chronic instabilities of the craniocervical unction, the cervical spine (C1 to C7), and the thoracic spine from T1-T3.
Code information (models, lots, serial numbers)
GTIN 00643169554689, Lot/Batch numbers: H5802983
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- INFINITY™ Occipitocervical Upper Thoracic System 3604024Medtronic Sofamor Danek, Inc. · Orthosis, cervical pedicle screw spinal fixation1 recall0 reports
Product lines
Other devices under the same FDA submission
34 devices cleared under K163375 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- INFINITY™ Occipitocervical Upper Thoracic System 3604026Medtronic Sofamor Danek, Inc. · Orthosis, cervical pedicle screw spinal fixation2 recalls1 report
- INFINITY™ Occipitocervical Upper Thoracic System 3604532Medtronic Sofamor Danek, Inc. · Orthosis, cervical pedicle screw spinal fixation1 recall0 reports
- INFINITY™ Occipitocervical Upper Thoracic System 360PT4022Medtronic Sofamor Danek, Inc. · Orthosis, cervical pedicle screw spinal fixation1 recall0 reports
- INFINITY™ Occipitocervical Upper Thoracic System 3600315Medtronic Sofamor Danek, Inc. · APPLIANCE, FIXATION, SPINAL INTERLAMINAL8 reports
- INFINITY™ Occipitocervical Upper Thoracic System G3605525Medtronic Sofamor Danek, Inc. · Orthosis, cervical pedicle screw spinal fixation6 reports
- INFINITY™ Occipitocervical Upper Thoracic System 3604012Medtronic Sofamor Danek, Inc. · Orthosis, cervical pedicle screw spinal fixation6 reports
- INFINITY™ Occipitocervical Upper Thoracic System 3600215Medtronic Sofamor Danek, Inc. · APPLIANCE, FIXATION, SPINAL INTERLAMINAL4 reports
- INFINITY™ Occipitocervical Upper Thoracic System G3600215Medtronic Sofamor Danek, Inc. · APPLIANCE, FIXATION, SPINAL INTERLAMINAL4 reports
- INFINITY™ Occipitocervical Upper Thoracic System G3603514Medtronic Sofamor Danek, Inc. · APPLIANCE, FIXATION, SPINAL INTERLAMINAL3 reports
- INFINITY™ Occipitocervical Upper Thoracic System 3603514Medtronic Sofamor Danek, Inc. · APPLIANCE, FIXATION, SPINAL INTERLAMINAL3 reports
- INFINITY™ Occipitocervical Upper Thoracic System 3602536Medtronic Sofamor Danek, Inc. · APPLIANCE, FIXATION, SPINAL INTERLAMINAL2 reports
- INFINITY™ Occipitocervical Upper Thoracic System G3603522Medtronic Sofamor Danek, Inc. · Orthosis, cervical pedicle screw spinal fixation2 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 91790).
- Z-1410-2023 · Open, ClassifiedINFINITY Occipitocervical Upper Thoracic System, MULTI AXIAL SCREW, Size: 4.0mm x 32mm, REF 3604532; for immobilization and stabilization of spinal segment as an adjunct to fusion…
- Z-1411-2023 · Open, ClassifiedINFINITY Occipitocervical Upper Thoracic System, PARTIALLY THREADED MULTI AXIAL SCREW, Size: 4.0mm x 22mm, REF 360PT4022; for immobilization and stabilization of spinal segment as…
- Z-1412-2023 · Open, ClassifiedINFINITY Occipitocervical Upper Thoracic System, MULTI AXIAL SCREW, Size: 4.0mm x 26mm, REF 3604026; for immobilization and stabilization of spinal segment as an adjunct to fusion…
- Z-1413-2023 · Open, ClassifiedINFINITY Occipitocervical Upper Thoracic System, MULTI AXIAL SCREW, Size: 4.0mm x 34mm, REF 3604034; for immobilization and stabilization of spinal segment as an adjunct to fusion…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page