System Name: 722059 Allura Xper FD20/15 OR Table 722058… recallZ-3166-2024 · Philips Medical Systems Nederland B.V. · initiated Aug 22, 2024
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Philips Medical Systems Nederland B.V. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Potential Detachment of the Cable Hose Carrier (due to broken plastic and/or bolts) which May Result in Parts Falling and/or Part of the Cable Hose Dropping, and result in injury.
FDA root cause category: Device Design
What the company is doing
Philips Issued URGENT Medical Device Correction Letter (Philips Reference # 2023-IGT-BST-006) on 8/22/24. Letter states reason for recall, health risk and action to take: a. Affected systems may continue to be used in accordance with their Instructions for Use (IFU) and the instructions below. b. Avoid that operators or bystanders stand underneath the Cable Hose Carriers, especially when moving or turning the system monitor. c. Avoid having the patient located underneath the Cable Hose Carriers (e.g., during transport). d. If you notice loose or broken bolts and/or broken plastic parts of the Cable Hose Carriers (see Figure 2), please contact your local Philips representative immediately. e. As part of the preventive maintenance cycle, Philips will check the Monitor Ceiling Suspension, including the Cable Hose Carriers, as indicated in the Preventive Maintenance Manual Update attached in Appendix B. o Keep a copy of this Preventative Maintenance Manual Update with your current manual. o If you do not use Philips to perform the preventative maintenance on your system, provide a copy of the Preventive Maintenance Manual Update to your qualified and authorized service provider. f. Circulate this Urgent Medical Device Correction letter to all users of the system so that they are aware of the issue and follow the instructions above. Keep this Medical Device Correction letter with the documentation of the system until Philips corrects your system. g. Complete and return the attached response form (page 4) to Philips promptly and no later than 30 days from receipt. Completing this form confirms receipt of the Urgent Medical Device Correction letter and understanding of the issue and required actions to be taken. Philips is redesigning the Cable Hose Carriers. Once the redesigned cable hose carriers are available, Philips will contact you to schedule a visit to install them in the affected system(s) (reference FCO72200536, FCO72200576, and FCO72200577). As of the date of t
Recall details
- Recall number
- Z-3166-2024
- Status
- Open, Classified
- Initiated by the company
- Aug 22, 2024
- Posted by the FDA
- Sep 18, 2024
- Recalling firm
- Philips Medical Systems Nederland B.V.
- Quantity
- 9318 units
- Product code
- OWB
- FDA submissions
- K141979
- Distribution
- Nationwide, Canada and Worldwide: Albania Algeria Argentina Armenia Australia Austria Azerbaijan Bahrain Bangladesh Belarus Belgium Bolivia Bosnia and Herzegovina Brazil Brunei Darussalam Bulgaria Cambodia Cayman Islands Chile China Colombia Costa Rica C¿te D'Ivoire Croatia Cuba Curacao Cyprus Czech Republic Denmark Dominican Republic Ecuador Egypt Estonia Ethiopia Fiji Finland France French Polynesia Georgia Germany Ghana Gibraltar Greece Guadeloupe Guinea Honduras Hong Kong Hungary Iceland India Indonesia Iran Iraq Ireland Israel Italy Jamaica Japan Jordan Kazakhstan Kenya Korea, Republic of Kosovo Kuwait Kyrgyzstan Lao People's Democratic Republic Latvia Lebanon Libya Lithuania Macao Macedonia Malaysia Maldives Malta Martinique Mauritania Mauritius Mexico Mongolia Morocco Myanmar Nepal Netherlands New Caledonia New Zealand Nicaragua Norway Oman Pakistan Palestine, State of Panama Papua New Guinea Paraguay Peru Philippines Poland Portugal Qatar R¿union Romania Russian Federation Saudi Arabia Senegal Serbia Singapore Slovakia Slovenia South Africa Spain Sri Lanka Sudan Sweden Switzerland Syrian Arab Republic Taiwan Tanzania Thailand Trinidad and Tobago Tunisia Turkey Turkmenistan Ukraine United Arab Emirates United Kingdom Uruguay Uzbekistan Venezuela Viet Nam Virgin Islands (U.S.) Yemen
Product description
System Name: 722059 Allura Xper FD20/15 OR Table 722058 Allura Xper FD20/15 722039 Allura Xper FD20/20 OR Table 722038 Allura Xper FD20/20 722035 Allura Xper FD20 OR Table 722029 Allura Xper FD20/10 722028 Allura Xper FD20 722027 Allura Xper FD10/10 722026 Allura Xper FD10 722025 Allura Xper FD20 Biplane OR Table 722023 Allura Xper FD20 OR Table 722020 Allura Xper FD20 Biplane OR Table 722015 Allura Xper FD20 OR Table 722013 Allura Xper FD20 Biplane 722012 Allura Xper FD20 722011 Allura Xper FD10/10 722010 Allura Xper FD10 722008 Allura Xper FD20 Biplane 722006 Allura Xper FD20 722005 Allura Xper FD10/10 722003 Allura Xper FD10 722002 Allura Xper FD10F 722001 Allura Xper FD10C
Code information (models, lots, serial numbers)
Show the full code information (1k characters)
Model Product UDI-DI: 722059 Allura Xper FD20/15 OR Table (01)00884838059122(21) 722058 Allura Xper FD20/15 (01)00884838059115(21) 722039 Allura Xper FD20/20 OR Table - 722038 Allura Xper FD20/20 (01)00884838054226(21) 722035 Allura Xper FD20 OR Table (01)00884838054240(21) 722029 Allura Xper FD20/10 (01)00884838054219(21) 722028 Allura Xper FD20 (01)00884838054202(21) 722027 Allura Xper FD10/10 (01)00884838054196(21) 722026 Allura Xper FD10 (01)00884838054189(21) 722025 Allura Xper FD20 Biplane OR Table - 722023 Allura Xper FD20 OR Table - 722020 Allura Xper FD20 Biplane OR Table - 722015 Allura Xper FD20 OR Table - 722013 Allura Xper FD20 Biplane (01)00884838059061(21) 722012 Allura Xper FD20 (01)00884838059054(21) 722011 Allura Xper FD10/10 - 722010 Allura Xper FD10 (01)00884838059030(21) 722008 Allura Xper FD20 Biplane - 722006 Allura Xper FD20 (01)00884838055513(21) 722005 Allura Xper FD10/10 (01)00884838080805(10) 722003 Allura Xper FD10 (01)00884838059191(10) 722002 Allura Xper FD10F - 722001 Allura Xper FD10C - All Serial Numbers
Devices named in this recall
13 devices in the FDA device registry whose identifier appears in the recall's code information.
- Allura Xper FD Allura Xper FD20Philips Medical Systems Nederland B.V. · Interventional Fluoroscopic X-Ray System25 recalls2 deaths3,984 reports
- Allura Xper FD Allura Xper FD20Philips Medical Systems Nederland B.V. · Interventional Fluoroscopic X-Ray System23 recalls2 deaths3,614 reports
- Allura Xper FD Allura Xper FD10Philips Medical Systems Nederland B.V. · Interventional Fluoroscopic X-Ray System23 recalls2 deaths2,015 reports
- Allura Xper FD Allura Xper FD10Philips Medical Systems Nederland B.V. · Interventional Fluoroscopic X-Ray System20 recalls2 deaths1,837 reports
- Allura Xper FD Allura Xper FD20/15Philips Medical Systems Nederland B.V. · Interventional Fluoroscopic X-Ray System20 recalls2 deaths776 reports
- Allura Xper FD Allura Xper FD20 BiplanePhilips Medical Systems Nederland B.V. · Interventional Fluoroscopic X-Ray System16 recalls3 deaths792 reports
- Allura Xper FD Allura Xper FD10/10Philips Medical Systems Nederland B.V. · Interventional Fluoroscopic X-Ray System19 recalls603 reports
- Allura Xper FD Allura Xper FD20 OR TablePhilips Medical Systems Nederland B.V. · Interventional Fluoroscopic X-Ray System18 recalls411 reports
- Allura Xper FD Allura Xper FD20/20Philips Medical Systems Nederland B.V. · Interventional Fluoroscopic X-Ray System18 recalls250 reports
- Allura Xper FD Allura Xper FD20/10Philips Medical Systems Nederland B.V. · Interventional Fluoroscopic X-Ray System19 recalls172 reports
- Interventional Workspot Interventional Workspot - 1Philips Medical Systems Nederland B.V. · System, Image Processing, Radiological13 recalls24 reports
- Allura Xper FD Allura Xper FD20/15 OR TablePhilips Medical Systems Nederland B.V. · Interventional Fluoroscopic X-Ray System12 recalls16 reports
- Dynamic Coronary Roadmap Dynamic Coronary Roadmap - 1Philips Medical Systems Nederland B.V. · Interventional Fluoroscopic X-Ray System5 recalls0 reports
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 95256).
- Z-3165-2024 · Open, ClassifiedSystem Code Name: 722282 Azurion 7 M20 722281 Azurion 5 M20 722236 Azurion 7 B20 722235 Azurion 7 B12 722234 Azurion 7 M20 722233 Azurion 7 M12 722228 Azurion 5 M20 722227 Azurion …
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗