Azurion recallZ-3165-2024 · Philips Medical Systems Nederland B.V. · initiated Aug 22, 2024

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Philips Medical Systems Nederland B.V. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Potential Detachment of the Cable Hose Carrier (due to broken plastic and/or bolts) which May Result in Parts Falling and/or Part of the Cable Hose Dropping, and result in injury.

FDA root cause category: Device Design

What the company is doing

Philips Issued URGENT Medical Device Correction Letter (Philips Reference # 2023-IGT-BST-006) on 8/22/24. Letter states reason for recall, health risk and action to take: a. Affected systems may continue to be used in accordance with their Instructions for Use (IFU) and the instructions below. b. Avoid that operators or bystanders stand underneath the Cable Hose Carriers, especially when moving or turning the system monitor. c. Avoid having the patient located underneath the Cable Hose Carriers (e.g., during transport). d. If you notice loose or broken bolts and/or broken plastic parts of the Cable Hose Carriers (see Figure 2), please contact your local Philips representative immediately. e. As part of the preventive maintenance cycle, Philips will check the Monitor Ceiling Suspension, including the Cable Hose Carriers, as indicated in the Preventive Maintenance Manual Update attached in Appendix B. o Keep a copy of this Preventative Maintenance Manual Update with your current manual. o If you do not use Philips to perform the preventative maintenance on your system, provide a copy of the Preventive Maintenance Manual Update to your qualified and authorized service provider. f. Circulate this Urgent Medical Device Correction letter to all users of the system so that they are aware of the issue and follow the instructions above. Keep this Medical Device Correction letter with the documentation of the system until Philips corrects your system. g. Complete and return the attached response form (page 4) to Philips promptly and no later than 30 days from receipt. Completing this form confirms receipt of the Urgent Medical Device Correction letter and understanding of the issue and required actions to be taken. Philips is redesigning the Cable Hose Carriers. Once the redesigned cable hose carriers are available, Philips will contact you to schedule a visit to install them in the affected system(s) (reference FCO72200536, FCO72200576, and FCO72200577). As of the date of t

Recall details

Recall number
Z-3165-2024
Status
Open, Classified
Initiated by the company
Aug 22, 2024
Posted by the FDA
Sep 18, 2024
Recalling firm
Philips Medical Systems Nederland B.V.
Quantity
7091 units
Product code
OWB
FDA submissions
K200917
Distribution
Nationwide, Canada and Worldwide: Albania Algeria Argentina Armenia Australia Austria Azerbaijan Bahrain Bangladesh Belarus Belgium Bolivia Bosnia and Herzegovina Brazil Brunei Darussalam Bulgaria Cambodia Cayman Islands Chile China Colombia Costa Rica C¿te D'Ivoire Croatia Cuba Curacao Cyprus Czech Republic Denmark Dominican Republic Ecuador Egypt Estonia Ethiopia Fiji Finland France French Polynesia Georgia Germany Ghana Gibraltar Greece Guadeloupe Guinea Honduras Hong Kong Hungary Iceland India Indonesia Iran Iraq Ireland Israel Italy Jamaica Japan Jordan Kazakhstan Kenya Korea, Republic of Kosovo Kuwait Kyrgyzstan Lao People's Democratic Republic Latvia Lebanon Libya Lithuania Macao Macedonia Malaysia Maldives Malta Martinique Mauritania Mauritius Mexico Mongolia Morocco Myanmar Nepal Netherlands New Caledonia New Zealand Nicaragua Norway Oman Pakistan Palestine, State of Panama Papua New Guinea Paraguay Peru Philippines Poland Portugal Qatar R¿union Romania Russian Federation Saudi Arabia Senegal Serbia Singapore Slovakia Slovenia South Africa Spain Sri Lanka Sudan Sweden Switzerland Syrian Arab Republic Taiwan Tanzania Thailand Trinidad and Tobago Tunisia Turkey Turkmenistan Ukraine United Arab Emirates United Kingdom Uruguay Uzbekistan Venezuela Viet Nam Virgin Islands (U.S.) Yemen

Product description

System Code Name: 722282 Azurion 7 M20 722281 Azurion 5 M20 722236 Azurion 7 B20 722235 Azurion 7 B12 722234 Azurion 7 M20 722233 Azurion 7 M12 722228 Azurion 5 M20 722227 Azurion 5 M12 722226 Azurion 7 B20 722225 Azurion 7 B12 722224 Azurion 7 M20 722223 Azurion 7 M12 722222 Azurion 3 M15 722221 Azurion 3 M12 722079 Azurion 7 M20 722078 Azurion 7 M12 722068 Azurion 7 B20 722067 Azurion 7 B12 722064 Azurion 3 M15 722063 Azurion 3 M12

Code information (models, lots, serial numbers)

Show the full code information (1k characters)

Model Product UDI-DI: 1. 722282 Azurion 7 M20 (01)00884838110571(21); 2. 722281 Azurion 5 M20 (01)00884838110564(21); 3. 722236 Azurion 7 B20 (01)00884838116801(21); 4. 722235 Azurion 7 B12 (01)00884838116788(21); 5. 722234 Azurion 7 M20 (01)00884838116771(21); 6. 722233 Azurion 7 M12 (01)00884838116764(21); 7. 722228 Azurion 5 M20 (01)00884838099234(21); 8. 722227 Azurion 5 M12 (01)00884838099227(21); 9. 722226 Azurion 7 B20 (01)00884838099272(21); 10. 722225 Azurion 7 B12 (01)00884838099265(21); 11. 722224 Azurion 7 M20 (01)00884838099258(21); 12. 722223 Azurion 7 M12 (01)00884838099241(21); 13. 722222 Azurion 3 M15 (01)00884838099210(21); 14. 722221 Azurion 3 M12 (01)00884838099203(21); 15. 722079 Azurion 7 M20 (01)00884838085268(21); 16. 722078 Azurion 7 M12 (01)00884838085251(21); 17. 722068 Azurion 7 B20 (01)00884838085367(21); 18.722067 Azurion 7 B12 (01)00884838085350(21); 19. 722064 Azurion 3 M15 (01)00884838085282(21); 20. 722063 Azurion 3 M12 (01)00884838085275(21) All serial numbers

Devices named in this recall

18 devices in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other products in the same recall event

The FDA groups related products into one recall event (event 95256).