Azurion recallZ-1969-2025 · Philips Medical Systems Nederland B.V. · initiated May 28, 2025

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Philips Medical Systems Nederland B.V. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

A software issue was identified in the internal communication process between the system software and the X-ray generator firmware. Loss of imaging (X-ray) functionality caused by the software issue in either of the two situations identified could result in a delay of therapy. The potential delay may result in serious adverse health outcomes, including the possibility of death, especially when the system is used with patients undergoing complex and/or urgent interventions for potentially life-threatening conditions (e.g., acute ischemic stroke, ST-segment elevation myocardial ischemia, life-threatening bleedings).

FDA root cause category: Software design

What the company is doing

On May 28, 2025, URGENT MEDICAL DEVICE CORRECTION letters were sent to customers. Actions that should be taken by the customer / user in order to prevent risks for patients or users - Circulate this Urgent Medical Device Correction letter to all users of the system so that they are aware of the issue. - Avoid rapidly pressing and releasing the footswitch pedals. - In case X-ray functionality is not available following the situations described in this letter, perform a cold system restart of the system, as follows: o On the Review Module, press and hold Power Off . o Release the button when the indicator light begins to flash. o When the indicator light stops flashing, wait for 10 seconds. o On the Review Module, press and hold Power On . NOTE: Do not operate any of the controls while the system is powering on, as this may inhibit the start-up process - In case the affected system has been transferred to another organization, please send a copy of this Urgent Medical Device Correction letter to that organization and inform Philips about this transfer through your local Philips representative. - Keep this Urgent Medical Device Correction letter with the documentation of the system until Philips corrects your system. Ensure that the letter is in a place likely to be seen/viewed. - Complete and return the response form included in this Urgent Medical Device Correction to Philips promptly and no later than 30 days from receipt. Completing this form confirms receipt of the Urgent Medical Device Correction letter and understanding of the issue and required actions to be taken. - If you experience the issue described in this letter, please report it to your local Philips representative. 5. Actions planned by Philips IGT Systems to correct the issue Philips will address the identified issue by implementing a software update in all affected systems. For systems featuring an Interventional Workspot and/or EchoNavigator, Philips will upgrade the Interventional Workspo

Recall details

Recall number
Z-1969-2025
Status
Open, Classified
Initiated by the company
May 28, 2025
Posted by the FDA
Jun 12, 2025
Recalling firm
Philips Medical Systems Nederland B.V.
Quantity
9,707 units (2300 US, 7407 OUS)
Product code
OWB
FDA submissions
K172822, K181830, K200917
Distribution
Worldwide distribution - US Nationwide and the countries of Afghanistan, Albania, Algeria, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Bosnia-Herz., Brazil, Brunei Darussal, Bulgaria, Cambodia, Cayman Islands, Chile, China, Colombia, Costa Rica, C¿te D'Ivoire, Croatia, Cyprus, Czech Republic, Denmark, Dominican Rep, Ecuador, Egypt, Estonia, Ethiopia, Fiji, Finland, France, French Polynesia, Germany, Ghana, Gibraltar, Greece, Guadeloupe, Hong Kong, Hungary, Iceland, India, Indonesia, IRAN, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Kosovo, Kuwait, Kyrgyzstan, Lao, Latvia, Lebanon, Libya, Lithuania, Macao, Macedonia, Malaysia, Maldives, Martinique, Mauritania, Mauritius, Mexico, Moldova, Mongolia, Morocco, Myanmar, Nepal, Netherlands, New Zealand, Nicaragua, Nigeria, Norway, Oman, Pakistan, Palestine, Panama, Peru, Philippines, Poland, Portugal, Qatar, R¿union, Romania, Russian Fed., Saudi Arabia, Senegal, Serbia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sri Lanka, Sweden, Switzerland, Syria, Taiwan, Tanzania, Thailand, Tunisia, T¿rkiye, Turkmenistan, Ukraine, United Kingdom, Uruguay, Utd.Arab.Emir., Uzbekistan, Vietnam, Yemen.

Product description

Azurion R1.x and R2.x systems; Model Numbers: (1) 722063, (2) 722064, (3) 722067, (4) 722068, (5) 722078, (6) 722079, (7) 722221, (8) 722222, (9) 722223, (10) 722224, (11) 722225, (12) 722226, (13) 722227, (14) 722228, (15) 722280, (16) 722281, (17) 722282; Software Version: All software versions;

Code information (models, lots, serial numbers)

Model Numbers: (1) 722063, (2) 722064, (3) 722067, (4) 722068, (5) 722078, (6) 722079, (7) 722221, (8) 722222, (9) 722223, (10) 722224, (11) 722225, (12) 722226, (13) 722227, (14) 722228, (15) 722280, (16) 722281, (17) 722282; UDI-DIs: (1) 00884838085275, (2) 00884838085282, (3) 00884838085350, (4) 00884838085367, (5) 00884838085251, (6) 00884838085268, (7) 00884838099203, (8) 00884838099210, (9) 00884838099241, (10) 00884838099258, (11) 00884838099265, (12) 00884838099272, (13) 00884838099227, (14) 00884838099234, (15) Not commercialized in the US, (16) Not commercialized in the US, (17) Not commercialized in the US;

Devices named in this recall

14 devices in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Sources and updates