Azurion recallZ-2502-2025 · Philips Medical Systems Nederland B.V. · initiated Aug 1, 2025
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Philips Medical Systems Nederland B.V. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Philips Azurion and Allura systems configured with a 1 Phase Uninterruptible Power Supply (UPS) have the potential of shutting down and/or not starting up, if power to the system is lost.
FDA root cause category: Under Investigation by firm
What the company is doing
Philips notified consignees on about 08/01/2025 via letter. Consignees were informed that systems may still be utilized in accordance with their Instructions for Use (IFU), notify all applicable personnel of this recall, maintain the Urgent Medical Device Correction letter with documentation of the system until Philips corrects the system, and establish an emergency protocol prior to all applicable diagnostics, interventional and minimally invasive procedures to manage the situation should you experience the issue during a procedure. If an affected system has been transferred to another organization, please send a copy of this Urgent Medical Device Correction letter to that organization and inform Philips about this transfer through your local Philips representative. Additionally, consignees were requested to complete and return the provided response form and report any issue experienced to Philips.
Recall details
- Recall number
- Z-2502-2025
- Status
- Open, Classified
- Initiated by the company
- Aug 1, 2025
- Posted by the FDA
- Sep 3, 2025
- Recalling firm
- Philips Medical Systems Nederland B.V.
- Quantity
- 1,861 units
- Product code
- OWB
- FDA submissions
- K172822, K181830, K200917
- Distribution
- Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Austria, Belarus, Belgium, Bolivia, Bosnia and Herzegovina, Brazil, Bulgaria, Canada, Chile, China, Colombia, Cyprus, Czech Republic, Dominican Republic, Ecuador, Egypt, Finland, France, Germany, Greece, Hong Kong, Hungary, India, Iran, Ireland, Israel, Italy, Jamaica, Japan, Jordan, Kazakhstan, Korea, Republic of, Kosovo, Latvia, Lithuania, Macedonia, Maldives, Malta, Mexico, Moldova, Netherlands, New Caledonia, New Zealand, Nicaragua, Norway, Panama, Paraguay, Peru, Philippines, Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Tunisia, United Arab Emirates, United Kingdom, Uruguay, Uzbekistan, Viet Nam.
Product description
Philips Azurion System configured with a 1 Phase UPS. Azurion family (R1.x, R2.x). Labeled as the following with corresponding model numbers: 1. Azurion 3 M12, Model Number 722063. 2. Azurion 3 M15, Model Number 722064. 3. Azurion 7 B12, Model Number 722067. 4. Azurion 7 B20, Model Number 722068. 5. Azurion 7 M12, Model Number 722078. 6. Azurion 7 M20, Model Number 722079. 7. Azurion 3 M12, Model Number 722221. 8. Azurion 3 M15, Model Number 722222. 9. Azurion 7 M12, Model Number 722223. 10. Azurion 7 M20, Model Number 722224. 11. Azurion 7 B12, Model Number 722225. 12. Azurion 7 B20, Model Number 722226. 13. Azurion 5 M12, Model Number 722227. 14. Azurion 5 M20, Model Number 722228. 15. Azurion 7 M20, Model Number 722282.
Code information (models, lots, serial numbers)
Show the full code information (11k characters)
1. Model Number/System Code (UDI); Serial Numbers: 722063 (00884838085275); 25, 151, 91, 85, 90, 86, 107, 89, 88, 87, 120, 153, 96, 108. 2. Model Number/System Code (UDI); Serial Numbers: 722064 (00884838085282); 64, 260, 261, 335, 218, 267, 270, 258, 35, 10, 111, 82, 125, 94, 278, 36, 55, 12, 57, 69, 484, 451, 497, 155, 272, 490, 132, 70, 251, 226, 290, 186, 48, 373. 3. Model Number/System Code (UDI); Serial Numbers: 722067 (00884838085350); 150, 69, 221, 148, 42, 98, 123, 277, 92, 103, 41, 214, 173, 172, 163, 128, 151, 22, 23, 268, 112, 97, 243, 45, 108, 113, 127, 176, 165, 33, 30, 206, 203, 48, 149, 251, 262. 4. Model Number/System Code (UDI); Serial Numbers: 722068 (00884838085367); 213, 111, 85, 89, 320, 73, 308, 172, 295, 406, 285, 413, 322, 324, 184, 442, 279, 361, 490, 42, 103, 520, 470, 472, 457, 400, 547, 510, 239, 275, 265, 365, 481, 62, 552, 44, 158, 36, 276, 68, 551, 516, 372, 46, 185, 515, 347, 416, 242, 219, 299, 126, 211, 468, 238, 568, 583, 69, 60, 426, 427, 93, 303, 302, 141. 5. Model Number/System Code (UDI); Serial Numbers: 722078 (00884838085251); 465, 143, 278, 402, 290, 524, 412, 166, 539, 532, 291, 415, 753, 441, 305, 289, 803, 399, 536, 537, 478, 479, 224, 163, 719, 364, 239, 383, 829, 513, 661, 215, 677, 709, 410, 269, 185, 902, 360, 546, 294, 906, 603, 870, 523, 598, 317, 535, 359, 436, 165, 733, 218, 204, 205, 206, 888, 923, 197, 195, 797, 444, 222, 489, 361, 362, 30, 223, 584, 962, 893, 27, 977, 156, 909, 953, 182, 958, 847, 959, 610, 490, 103, 184, 506, 23, 112, 199, 1047, 1074, 850, 643, 319, 547, 879, 572, 38, 865, 881, 261, 472, 368, 369, 243, 244, 731, 202, 442, 351, 636, 777, 40, 477, 134, 341, 563, 633, 286, 336, 512, 158, 357, 397, 896, 281, 52, 366, 190, 617, 496, 461, 463, 270, 528, 638, 529, 431, 712, 714, 219, 203, 575, 498, 577, 574, 139, 178, 179, 265, 264, 266, 511, 526, 746, 1082, 426, 36, 242, 300, 301, 795, 32, 241, 49, 51, 50, 250, 145, 774, 785, 474, 475, 452, 789, 552, 157, 818, 460, 459, 352, 46, 39, 311, 312, 277, 455, 608, 878, 220, 221, 321, 533, 699, 877. 6. Model Number/System Code (UDI); Serial Numbers: 722079 (00884838085268); 942, 948, 702, 703465, 1316, 1042, 465, 674, 703413, 1142, 1090, 526, 439, 637, 455, 334, 875, 737, 599, 29, 704281, 801, 671, 1387, 981, 1211, 1213, 499, 1230, 708, 496, 903, 887, 912, 1041, 703324, 528, 703584, 738, 351, 820, 32, 1411, 537, 445, 428, 587, 1290, 251, 1402, 318, 273, 877, 854, 897, 896, 836, 891, 940, 297, 483, 689, 717, 396, 860, 840, 790, 862, 774, 703344, 1436, 629, 1049, 1355, 1414, 703404, 703387, 703317, 703540, 58, 810, 939, 634, 53, 57, 841, 1282, 703310, 1100, 703314, 54, 703408, 246, 890, 703409, 658, 776, 703495, 18, 314, 828, 1323, 645, 646, 325, 731, 262, 1252, 703418, 474, 481, 800, 1359, 1358, 1360, 809, 1258, 1006, 602, 335, 837, 472, 703559, 342, 344, 345, 343, 1232, 1226, 1237, 1236, 34, 1400, 863, 987, 703389, 1391, 1043, 905, 577, 714, 573, 805, 228, 30, 366, 543, 921, 1200, 1222, 353, 703379, 703402, 350, 673, 669, 1212, 1285, 1216, 416, 290, 272, 703306, 35, 43, 40, 1270, 274, 466, 1155, 1033, 703356, 284, 703685, 703550, 703380, 703381, 332, 331, 330, 679, 845, 846, 328, 329, 783, 1016, 1017, 74, 1077, 37, 293, 952, 312, 703351, 703384, 759, 557, 1169, 756, 1306, 703597, 703611, 1129, 234, 75, 316, 740, 1104, 703653, 803, 1388, 67, 1133, 319, 76, 760, 966, 632, 1082, 421, 704131, 704252, 704009, 704215, 1181, 703815, 703330, 782, 1372, 703713, 725, 907, 901, 960, 542, 703470, 1428, 1037, 215, 1035, 703463, 1261, 703321, 831, 832, 410, 224, 495, 340, 839, 728, 339, 282, 444, 1338, 529, 1121, 682, 703560, 727, 576, 630, 703424, 381, 1123, 703385, 71, 583, 703386, 1154, 748, 278, 876, 478, 787, 477, 283, 997, 784, 1116, 703753, 1278, 703313, 703872, 1217, 258, 260, 259, 395, 433, 451, 1405, 703355, 703449, 703654, 703699, 703870, 256, 498, 703638, 703841, 704101, 704171, 1347, 1264, 703834, 704277, 704284, 704150, 703740, 703935, 703995, 703966, 703824, 703961, 1317, 704258, 430, 934, 492, 36, 1202, 833, 834, 1223, 364, 720, 490, 1008, 703845, 1165, 1089, 1330, 743, 271, 791, 370, 1175, 703438, 1150, 72, 844, 786, 1239, 230, 796, 257, 522, 523, 338, 1272, 764. 7. Model Number/System Code (UDI); Serial Numbers: 722221 (00884838099203); 108, 19, 153, 49, 115, 171, 6, 8, 127, 59, 56, 126, 106, 166, 155. 8. Model Number/System Code (UDI); Serial Numbers: 722222 (00884838099210); 55, 23, 44, 45, 187, 223, 99, 100, 101, 332, 51, 65, 11, 268, 188, 77, 80, 245, 109, 83, 50, 244, 177, 280, 13, 300, 5, 9. 9. Model Number/System Code (UDI); Serial Numbers: 722223 (00884838099241); 580, 669, 164, 127, 570, 131, 816, 365, 390, 352, 340, 242, 205, 627, 229, 262, 214, 458, 256, 728, 937, 944, 945, 951, 180, 439, 631, 681, 47, 126, 210, 336, 206, 367, 357, 572, 801, 197, 192, 87, 282, 177, 882, 208, 388, 581, 189, 479, 211, 186, 187, 566, 37, 370, 387, 116, 154, 99, 100, 598, 619, 854, 325, 787, 553, 185, 38, 859, 313, 314, 201, 1083, 860, 382, 645, 647, 646, 648, 30, 296, 297, 41, 914, 461, 808, 165, 33, 559, 563, 564, 558, 114, 176, 321, 661, 913, 246, 245, 422, 2, 639, 91, 333, 1056, 761, 21, 95, 531, 1017, 1037, 392, 358, 618, 590, 65, 295, 840, 12, 68, 465, 547, 702, 34, 450, 376, 451, 792, 876, 1077, 140, 988, 49, 442, 157, 539, 726, 342, 232, 492, 322, 591, 587, 812, 875, 904, 1167, 971, 532, 82, 109, 170, 4, 641, 640, 542, 543, 805, 121, 122, 374, 118, 13, 428, 983, 623, 1009, 515, 660, 712, 305, 179, 1007, 769, 754, 720, 721, 753, 755, 674, 344, 620, 768, 475, 474, 378, 294, 163, 815, 497, 362, 544, 707, 535, 536. 10. Model Number/System Code (UDI); Serial Numbers: 722224 (00884838099258); 606, 1828, 1829, 1830, 1201, 612, 503, 277, 261, 75, 182, 415, 876, 1488, 2957, 266, 512, 1160, 766, 2156, 2138, 250, 226, 360, 790, 452, 1147, 1278, 88, 1142, 792, 595, 892, 814, 813, 2995, 3120, 189, 2139, 2140, 108, 100, 591, 385, 260, 1827, 850, 338, 113, 438, 45, 444, 39, 554, 434, 3207, 696, 1499, 42, 841, 2656, 56, 2331, 421, 589, 2520, 793, 129, 84, 1810, 362, 327, 124, 125, 2280, 239, 1324, 950, 946, 381, 443, 252, 441, 621, 622, 435, 433, 99, 291, 332, 422, 1917, 2248, 2247, 2229, 2246, 1376, 1377, 1514, 351, 1960, 2045, 2403, 80, 183, 366, 101, 564, 396, 120, 454, 300, 1507, 1510, 329, 328, 1336, 307, 953, 1252, 1869, 1458, 904, 940, 1486, 165, 216, 527, 2376, 220, 176, 188, 466, 1112, 2913, 2426, 190, 490, 593, 469, 273, 333, 306, 432, 439, 244, 417, 414, 770, 516, 645, 1409, 565, 1625, 78, 267, 268, 789, 788, 478, 346, 298, 1299, 289, 1873, 2417, 1484, 1335, 1611, 657, 476, 251, 517, 507, 2018, 2239, 1555, 163, 240, 467, 468, 539, 128, 1126, 1246, 1259, 1255, 1256, 900, 191, 64, 1493, 1032, 2222, 1251, 1356, 194, 254, 1403, 440, 246, 2184, 2436, 613, 1565, 229, 286, 1508, 1461, 1512, 1511, 640, 1524, 836, 1692, 141, 574, 109, 2729, 46, 1063, 571, 174, 1490, 2569, 275, 262, 295, 272, 178, 228, 651, 462, 1049, 867, 1954, 1956, 116, 471, 233, 1995, 2185, 2860, 833, 874, 875, 1247, 1478, 330, 649, 173, 172, 379, 1477, 2752, 1672, 35, 1570, 406, 204, 206, 2081, 1042, 1591, 544, 2680, 1298, 513, 2349, 184, 2079, 545, 2080, 2082, 908, 573, 827, 1027, 1026, 1812, 781, 1204, 1184, 221, 208, 633, 3380, 816, 732, 647, 2510, 1388, 145, 2534, 2527, 2124, 1325, 2638, 632, 1766, 2214, 2269, 2575, 1169, 2515, 2337, 2338, 2823, 2825, 1539, 463, 368, 2415, 383, 563, 519, 1686, 2089, 2574, 1367, 988, 2163, 2725, 2724, 93, 1671, 1594, 1595, 2568, 1401, 2851, 1092, 534, 86, 1572, 634, 782, 1406, 3223, 19, 2230, 2480, 1962, 2218, 227, 337, 822, 1981, 1953, 1864, 749, 1952, 382, 4, 37, 1778, 3135, 648, 971, 2983, 861, 2482, 31, 1523, 59, 898, 866, 2730, 52, 29, 853, 810, 1722, 724, 1431, 171, 413, 2310, 2804, 596, 1737, 801, 937, 1326, 1330, 1349, 1350, 1369, 1609, 2244, 2245, 2373, 3099, 731, 1505, 1552, 280, 764, 1065, 910, 142, 1421, 721, 1430, 2068, 654, 397, 557, 1188, 2134, 2424, 1814, 2101, 2103, 2618, 2343, 2260, 2259, 2261, 2267, 2706, 2707, 2714, 2758, 2813, 3288, 2812, 2828, 1800, 1799, 2263, 2300, 2299, 2705, 2745, 2704, 2819, 2905, 2788, 2817, 2889, 3287, 3286, 3295, 1815, 2102, 2105, 2749, 2751, 1930, 2715, 1139, 156, 1562, 803, 671, 1538, 1730, 387, 13, 399, 2151, 424, 16, 2007, 3, 1331, 7, 863, 8, 358, 3259, 658, 1051, 1050, 1052, 1053, 1054, 1080, 1348, 2277, 604, 3271, 542, 729, 203, 1728, 1736, 610, 2344, 48, 635, 2848, 1272, 2402, 799, 800, 85, 2050, 537, 1566, 1588, 1656, 1649, 1447, 2046, 1602, 2013, 18, 1536, 1580, 1387, 1433, 1957, 1678, 2011, 2076, 2099, 1703, 1933, 1837, 2275, 1196, 1391, 1393, 1404, 1402, 2916, 1645, 1854, 1359, 1443, 1587, 1392, 1399, 1422, 1677, 1691, 1710, 2009, 1345, 2014, 1600, 1620, 2028, 2071, 1425, 1340, 1333, 1855, 2153, 1585, 1372, 1408, 1432, 2049, 1586, 2055, 1429, 1535, 1537, 1544, 1543, 1865, 1597, 1398, 1462, 1596, 1603, 1598, 1327, 2056, 2175, 2330, 1968, 1936, 1351, 2016, 1980, 1495, 2002, 2087, 1666, 1584, 2069, 2398, 1806, 1969, 1601, 1621, 1589, 1699, 1734, 1758, 2305, 2339, 1835, 1583, 1868, 1805, 1644, 2021, 2104, 727, 2057, 213, 509, 1932, 2621, 3263, 984, 344, 829, 1590, 2017, 1522. 11. Model Number/System Code (UDI); Serial Numbers: 722225 (00884838099265); 259, 55, 268, 147, 340, 72, 71, 216, 299, 291, 289, 249, 110, 298, 46, 29, 122, 233, 157, 149, 237, 167, 364, 4, 398, 30, 231, 38, 219, 170, 171, 90, 89, 77, 369, 161, 163, 137, 162, 40, 91, 74, 125, 190, 303, 377, 211, 329, 280, 86, 8, 31, 384, 87, 283, 421, 422, 267, 241. 12.Model Number/System Code (UDI); Serial Numbers: 722226 (00884838099272); 219, 120, 272, 105, 144, 319, 204, 255, 153, 350, 442, 183, 78, 467, 108, 143, 360, 555, 114, 209, 174, 687, 281, 554, 181, 147, 379, 468, 173, 271, 182, 419, 472, 122, 123, 124, 125, 126, 127, 403, 294, 194, 249, 432, 892, 87, 790, 261, 514, 766, 941, 228, 539, 950, 489, 622, 628, 637, 491, 345, 625, 93, 238, 299, 166, 364, 5, 506, 557, 227, 188, 827, 453, 883, 171, 708, 738, 655, 382, 106, 56, 196, 236, 770, 73, 302, 402, 210, 304, 916, 904, 40, 525, 653, 157, 657, 583, 540, 216, 456, 357, 596, 733, 439, 118, 263, 24, 840, 88, 224, 231, 233, 407, 598, 21, 762, 882, 505, 542, 507, 497, 485, 490, 494, 643, 644, 543, 636, 849, 633, 634, 504, 550, 551, 484, 631, 316, 639, 640, 642, 563, 547, 591, 911, 277, 278, 207, 773, 832, 501, 1000, 810. 13. Model Number/System Code (UDI); Serial Numbers: 722227 (00884838099227); 88, 135, 261, 19, 133, 134, 241, 164, 82, 32, 16, 12, 72, 64, 242, 176, 2, 75, 145, 143, 120, 112, 74, 188, 147, 146, 77, 141, 234, 14, 15, 47, 95, 163, 48, 3, 1, 123, 92, 173, 174, 121, 204, 57, 282, 8, 9. 14. Model Number/System Code (UDI); Serial Numbers: 722228 (00884838099234); 187, 328, 113, 205, 204, 150, 159, 23, 127, 24, 10, 36, 6, 213, 64, 20, 26, 47, 149, 62, 502, 18, 125, 391, 395, 229, 12, 408, 309, 310, 557, 98, 261, 179, 536, 143, 46, 70, 7, 13, 191, 259, 218, 11, 223, 370, 389, 14, 71, 39, 465, 552. 15. Model Number/System Code (UDI); Serial Numbers: 722282 ((01)00884838110571(21)8); 8.
Devices named in this recall
14 devices in the FDA device registry whose identifier appears in the recall's code information.
- Azurion Azurion 7 M20Philips Medical Systems Nederland B.V. · Interventional Fluoroscopic X-Ray System32 recalls2 deaths2,308 reports
- Azurion Azurion 7 M20Philips Medical Systems Nederland B.V. · Interventional Fluoroscopic X-Ray System28 recalls1 death2,176 reports
- Azurion Azurion 7 M12Philips Medical Systems Nederland B.V. · Interventional Fluoroscopic X-Ray System24 recalls2 deaths1,087 reports
- Azurion Azurion 7 M12Philips Medical Systems Nederland B.V. · Interventional Fluoroscopic X-Ray System24 recalls2 deaths835 reports
- Azurion Azurion 7 B20Philips Medical Systems Nederland B.V. · Interventional Fluoroscopic X-Ray System24 recalls1 death634 reports
- Azurion Azurion 7 B20Philips Medical Systems Nederland B.V. · Interventional Fluoroscopic X-Ray System24 recalls594 reports
- Azurion Azurion 5 M20Philips Medical Systems Nederland B.V. · Interventional Fluoroscopic X-Ray System24 recalls201 reports
- Azurion Azurion 7 B12Philips Medical Systems Nederland B.V. · Interventional Fluoroscopic X-Ray System23 recalls232 reports
- Azurion Azurion 7 B12Philips Medical Systems Nederland B.V. · Interventional Fluoroscopic X-Ray System21 recalls222 reports
- Azurion Azurion 5 M12Philips Medical Systems Nederland B.V. · Interventional Fluoroscopic X-Ray System21 recalls1 death197 reports
- Azurion Azurion 3 M15Philips Medical Systems Nederland B.V. · Interventional Fluoroscopic X-Ray System20 recalls2 deaths205 reports
- Azurion Azurion 3 M15Philips Medical Systems Nederland B.V. · Interventional Fluoroscopic X-Ray System21 recalls104 reports
- Azurion Azurion 3 M12Philips Medical Systems Nederland B.V. · Interventional Fluoroscopic X-Ray System20 recalls121 reports
- Azurion Azurion 3 M12Philips Medical Systems Nederland B.V. · Interventional Fluoroscopic X-Ray System20 recalls73 reports
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 97249).
- Z-2503-2025 · Open, ClassifiedPhilips Allura System configured with a 1 Phase UPS. Allura family (R1.x, R2.x). Labeled as the following with corresponding model numbers: 1. Allura Xper FD10, Model Number 72…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page