Medline medical procedure kits, containing Medline Neuro… recallZ-2014-2026 · Medline Industries, LP · initiated Mar 19, 2026
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Medline Industries, LP or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
During an internal review, Medline identified higher-than-expected endotoxin levels on a representative Medline neuro sponge product, indicating that in-market product may be out-of-specification for endotoxin.
FDA root cause category: Under Investigation by firm
What the company is doing
Medline industries issued an URGENT MEDICAL DEVICE RECALL notice to its consignees on 3/19/2026 via USPS first class mail and email. The notice explained the issue, potential health risk, and requested the following actions be taken on the affected sponges: Identify and quarantine affected product, complete response form using website link: https://recalls.medline.com, destroy affected product. Credit will be issued once response form is received. If the product has been further distributed, sold, or transferred, notify the customer to destroy the product. Distributors must notify their customers and should include customer quantities on your response form. For consignees of Medline medical convenience kits containing the affected sponges the actions to be taken are as follows: Identify and quarantine all affected kits, complete response form using website link: https://recalls.medline.com. Upon receipt of your submitted response form, your account will receive over-labels to place on affected inventory, with instructions for your staff to remove and discard the affected component prior to using the kit. These labels will be provided via FedEx overnight, if applicable. Distributors must notify their customers and should include customer quantities on your response form. Upon receipt of your submitted response form, you will receive over-labels to provide to your customers to affix to their affected product. If you have any questions, contact the Recall Department at 866-359-1704 or [email protected].
Recall details
- Recall number
- Z-2014-2026
- Status
- Open, Classified
- Initiated by the company
- Mar 19, 2026
- Posted by the FDA
- Apr 30, 2026
- Recalling firm
- Medline Industries, LP
- Quantity
- 160 kits
- Product code
- OEQ
- Distribution
- Worldwide distribution.
Product description
Medline medical procedure kits, containing Medline Neuro Sponges, labeled as follows: ANGIO PACK-LF, Medline Kit Number/SKU DYNJ0774765U
Code information (models, lots, serial numbers)
UDI/DI each 10193489634594, UDI/DI case 40193489634595, Lot Number: 24ILA931
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Medline Industries, Inc. DYNJ0774765UMedline Industries, Inc. · Angiography/angioplasty kit2 recalls0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 98664).
- Z-2008-2026 · Open, ClassifiedMedline Neuro Sponges, various dimensions, labeled as: 1. Neuro Sponges, Eponges neuro, Neuroesponjas, size 0.25" X 0.25", XR, 10/PK, Medline SKU NEUROSPNG02; 2. Neuro Sponge…
- Z-2009-2026 · Open, ClassifiedMedline medical procedure kits, containing Medline Neuro Sponges, labeled as follows: PEDS AIRWAY PACK, Medline Kit Number SKU DYNJ86657A
- Z-2010-2026 · Open, ClassifiedMedline medical procedure kits, containing Medline Neuro Sponges, labeled as follows: EH BRONCHOSCOPY, Medline Kit Number/SKU DYNJT3270
- Z-2011-2026 · Open, ClassifiedMedline medical procedure kits, containing Medline Neuro Sponges, labeled as follows: SPINAL FUSION, Medline Kit Number/SKU CDS860017AG
- Z-2012-2026 · Open, ClassifiedMedline medical procedure kits, containing Medline Neuro Sponges, labeled as follows: SPINE PACK-LF, Medline Kit Number/SKU PHS719414B
- Z-2013-2026 · Open, ClassifiedMedline medical procedure kits, containing Medline Neuro Sponges, labeled as follows: 1. NEURO PACK, Medline kit number/SKU DYNJ64479; 2. CRANIO/NEURO/SPINE, Medline kit number/…
- Z-2015-2026 · Open, ClassifiedMedline medical procedure kits, containing Medline Neuro Sponges, labeled as follows: 1. MAJOR NEURO PACK-LF, Medline Kit Number/SKU DYNJ0578916X; 2. ANGIOGRAPHY PACK, Medline K…
- Z-2016-2026 · Open, ClassifiedMedline medical procedure kits, containing Medline Neuro Sponges, labeled as follows: 1. OPEN HEART CDS PART A, Medline Kit Number/SKU CDS840402AI; 2. OPEN HEART CDS, Medline Ki…
- Z-2017-2026 · Open, ClassifiedMedline medical procedure kits, containing Medline Neuro Sponges, labeled as follows: 1. VP SHUNT, Medline Kit Number/SKU CDS840193V; 2. SHY VP SHUNT CDS, Medline Kit Number/SKU…
- Z-2018-2026 · Open, ClassifiedMedline medical procedure kits, containing Medline Neuro Sponges, labeled as follows: 1. OSC COLEMAN FAT HARVEST KIT, Medline Kit Number/SKU DYNJ32854B; 2. RR-DR. CHHETRI MDL PA…
- Z-2019-2026 · Open, ClassifiedMedline medical procedure kits, containing Medline Neuro Sponges, labeled as follows: 1. RR-OMFS PACK-LF, Medline Kit Number/SKU DYNJ0045920D; 2. RR-OMFS PACK-LF, Medline Kit Nu…
- Z-2020-2026 · Open, ClassifiedMedline medical procedure kits, containing Medline Neuro Sponges, labeled as follows: 1. ZALE STEREOTACTIC CDS, Medline Kit Number/SKU CDS983173G; 2. EPISTAXIS TRAY, Medline Kit…
- Z-2021-2026 · Open, ClassifiedMedline and Centurion medical procedure kits, containing Medline Neuro Sponges (1229 in total), HEAD & NECK TRAY, SINUS, SMR, etc.
- Z-2022-2026 · Open, ClassifiedMedline medical procedure kits, containing Medline Neuro Sponges, labeled as follows: 1. PLASTICS FREE FLAP, Medline Kit Number/SKU CDS983493N; 2. TLIF PACK, Medline Kit Number/…
- Z-2023-2026 · Open, ClassifiedMedline medical procedure kits, containing Medline Neuro Sponges, labeled as follows: 1. ENDOSCOPIC SKULL BASE, Medline Kit Number/SKU DYNJ58648J; 2. ENT ENDO SINUS WILMINGTON PA…
- Z-2024-2026 · Open, ClassifiedMedline medical procedure kits, containing Medline Neuro Sponges, labeled as follows: 1. ENDOSCOPY-LF, Medline Kit Number/SKU DYNJ22890M; 2. ENDOSCOPY PACK, Medline Kit Number/…
- Z-2025-2026 · Open, ClassifiedMedline medical procedure kits, containing Medline Neuro Sponges (102 in total)
- Z-2026-2026 · Open, ClassifiedMedline medical procedure kits, containing Medline Neuro Sponges, labeled as follows: 1. NEURO BASIN PACK-LF, Medline Kit Number/SKU DYNJ0532910L; 2. NEURO BASIN PACK-LF, Medlin…
- Z-2027-2026 · Open, ClassifiedMedline medical procedure kits, containing Medline Neuro Sponges (1050 in total), BIOBURDEN TEST KIT, NEURO, ROBOTIC, ACDF, ENT, etc.
- Z-2028-2026 · Open, ClassifiedCenturion medical procedure kits, containing Medline Neuro Sponges, labeled as follows: SURGICAL ENT PREP KIT, Medline Kit Number/SKU MNS11715
- Z-2029-2026 · Open, ClassifiedMedline medical procedure kits, containing Medline Neuro Sponges, labeled as follows: CERVICAL PACK, Medline Kit Number/SKU DYNJ37902F
- Z-2030-2026 · Open, ClassifiedMedline medical procedure kits, containing Medline Neuro Sponges, labeled as follows: 1. MICRODISECTOMY PACK, Medline Kit Number/SKU DYNJ80585B; 2. MICRODISECTOMY PACK, Medline …
- Z-2031-2026 · Open, ClassifiedMedline medical procedure kits, containing Medline Neuro Sponges, labeled as follows: GENERAL KIT, Medline Kit Number/SKU DYNJ905664I
- Z-2033-2026 · Open, ClassifiedMedline medical procedure kits, containing Medline Neuro Sponges (961 in total), NEURO,
- Z-2034-2026 · Open, ClassifiedMedline medical procedure kits, containing Medline Neuro Sponges, labeled as follows: 1. SPINE PROCEDURES PACK-LF, Medline Kit Number/SKU DYNJ0751220P; 2. ENSEMBLE RACHIS-LF, Medl…
- Z-2035-2026 · Open, ClassifiedMedline medical procedure kits, containing Medline Neuro Sponges, labeled as follows: 1. ROBOTIC HYSTERECTOMY, Medline Kit Number/SKU DYNJ904269B; 2. ROBOTIC HYSTERECTOMY, Medli…
- Z-2036-2026 · Open, ClassifiedMedline medical procedure kits, containing Medline Neuro Sponges, labeled as follows: 1. ANTERIOR CERVICAL FUSION PACK, Medline Kit Number/SKU DYNJ40566C; 2. OB PACK (OBSJD)642-L…
- Z-2037-2026 · Open, ClassifiedMedline medical procedure kits, containing Medline Neuro Sponges, labeled as follows: 1. CATARACT PACK, Medline Kit Number/SKU DYNJ48725A; 2. OCULAR PACK, Medline Kit Number/SKU …
- Z-2038-2026 · Open, ClassifiedMedline medical procedure kits, containing Medline Neuro Sponges (4077 in total), LAMINECTOMY, CRANI, etc.
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page