Medline medical procedure kits, containing Medline Neuro… recallZ-2025-2026 · Medline Industries, LP · initiated Mar 19, 2026
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Medline Industries, LP or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
During an internal review, Medline identified higher-than-expected endotoxin levels on a representative Medline neuro sponge product, indicating that in-market product may be out-of-specification for endotoxin.
FDA root cause category: Under Investigation by firm
What the company is doing
Medline industries issued an URGENT MEDICAL DEVICE RECALL notice to its consignees on 3/19/2026 via USPS first class mail and email. The notice explained the issue, potential health risk, and requested the following actions be taken on the affected sponges: Identify and quarantine affected product, complete response form using website link: https://recalls.medline.com, destroy affected product. Credit will be issued once response form is received. If the product has been further distributed, sold, or transferred, notify the customer to destroy the product. Distributors must notify their customers and should include customer quantities on your response form. For consignees of Medline medical convenience kits containing the affected sponges the actions to be taken are as follows: Identify and quarantine all affected kits, complete response form using website link: https://recalls.medline.com. Upon receipt of your submitted response form, your account will receive over-labels to place on affected inventory, with instructions for your staff to remove and discard the affected component prior to using the kit. These labels will be provided via FedEx overnight, if applicable. Distributors must notify their customers and should include customer quantities on your response form. Upon receipt of your submitted response form, you will receive over-labels to provide to your customers to affix to their affected product. If you have any questions, contact the Recall Department at 866-359-1704 or [email protected].
Recall details
- Recall number
- Z-2025-2026
- Status
- Open, Classified
- Initiated by the company
- Mar 19, 2026
- Posted by the FDA
- Apr 30, 2026
- Recalling firm
- Medline Industries, LP
- Quantity
- 22116 kits
- Product code
- FTN
- Distribution
- Worldwide distribution.
Product description
Medline medical procedure kits, containing Medline Neuro Sponges (102 in total)
Code information (models, lots, serial numbers)
Show the full code information (85k characters)
Medline Kit Number/SKU CDS980375L: UDI/DI each 10193489922882, UDI/DI case 40193489922883, Lot Number: 23KBN711; Medline Kit Number/SKU CDS980375L: UDI/DI each 10193489922882, UDI/DI case 40193489922883, Lot Number: 23GBD593; Medline Kit Number/SKU CDS980375L: UDI/DI each 10193489922882, UDI/DI case 40193489922883, Lot Number: 23GBD781; Medline Kit Number/SKU CDS980375L: UDI/DI each 10193489922882, UDI/DI case 40193489922883, Lot Number: 23DBF906; Medline Kit Number/SKU CDS980375L: UDI/DI each 10193489922882, UDI/DI case 40193489922883, Lot Number: 22LBI985; Medline Kit Number/SKU CDS980375L: UDI/DI each 10193489922882, UDI/DI case 40193489922883, Lot Number: 22KBC504; Medline Kit Number/SKU CDS980375L: UDI/DI each 10193489922882, UDI/DI case 40193489922883, Lot Number: 22IBF741; Medline Kit Number/SKU CDS980375L: UDI/DI each 10193489922882, UDI/DI case 40193489922883, Lot Number: 22FBX069; Medline Kit Number/SKU CDS980375L: UDI/DI each 10193489922882, UDI/DI case 40193489922883, Lot Number: 22DBS948; Medline Kit Number/SKU CDS980375L: UDI/DI each 10193489922882, UDI/DI case 40193489922883, Lot Number: 22CBX667; Medline Kit Number/SKU CDS980375L: UDI/DI each 10193489922882, UDI/DI case 40193489922883, Lot Number: 21IBT137; Medline Kit Number/SKU CDS980375L: UDI/DI each 10193489922882, UDI/DI case 40193489922883, Lot Number: 21HBG608; Medline Kit Number/SKU CDS980375L: UDI/DI each 10193489922882, UDI/DI case 40193489922883, Lot Number: 25GBV215; Medline Kit Number/SKU CDS980375L: UDI/DI each 10193489922882, UDI/DI case 40193489922883, Lot Number: 25GBA852; Medline Kit Number/SKU CDS980375L: UDI/DI each 10193489922882, UDI/DI case 40193489922883, Lot Number: 24KBJ988; Medline Kit Number/SKU CDS980375L: UDI/DI each 10193489922882, UDI/DI case 40193489922883, Lot Number: 24JBC528; Medline Kit Number/SKU CDS980375L: UDI/DI each 10193489922882, UDI/DI case 40193489922883, Lot Number: 24IBV131; Medline Kit Number/SKU CDS980375L: UDI/DI each 10193489922882, UDI/DI case 40193489922883, Lot Number: 24IBP386; Medline Kit Number/SKU CDS980375L: UDI/DI each 10193489922882, UDI/DI case 40193489922883, Lot Number: 24GBA976; Medline Kit Number/SKU CDS980375L: UDI/DI each 10193489922882, UDI/DI case 40193489922883, Lot Number: 24CBU492; Medline Kit Number/SKU CDS980375L: UDI/DI each 10193489922882, UDI/DI case 40193489922883, Lot Number: 23KBW165; Medline Kit Number/SKU CDS980375M: UDI/DI each 10198459447327, UDI/DI case 40198459447328, Lot Number: 26ABK337; Medline Kit Number/SKU CDS980375M: UDI/DI each 10198459447327, UDI/DI case 40198459447328, Lot Number: 25JBE079; Medline Kit Number/SKU CDS980375M: UDI/DI each 10198459447327, UDI/DI case 40198459447328, Lot Number: 25HBV443; Medline Kit Number/SKU CDS982939U: UDI/DI each 10198459333330, UDI/DI case 40198459333331, Lot Number: 25IBR604; Medline Kit Number/SKU CDS982939U: UDI/DI each 10198459333330, UDI/DI case 40198459333331, Lot Number: 25HBR025; Medline Kit Number/SKU CDS982939U: UDI/DI each 10198459333330, UDI/DI case 40198459333331, Lot Number: 25GBB389; Medline Kit Number/SKU CDS982939U: UDI/DI each 10198459333330, UDI/DI case 40198459333331, Lot Number: 25FBC324; Medline Kit Number/SKU CDS983100D: UDI/DI each 10198459007934, UDI/DI case 40198459007935, Lot Number: 24KBH657; Medline Kit Number/SKU CDS983100D: UDI/DI each 10198459007934, UDI/DI case 40198459007935, Lot Number: 24GBN523; Medline Kit Number/SKU CDS983173I: UDI/DI each 10195327170981, UDI/DI case 40195327170982, Lot Number: 23IBT936; Medline Kit Number/SKU CDS983173I: UDI/DI each 10195327170981, UDI/DI case 40195327170982, Lot Number: 23HBA798; Medline Kit Number/SKU CDS983173I: UDI/DI each 10195327170981, UDI/DI case 40195327170982, Lot Number: 23FBF767; Medline Kit Number/SKU CDS983173I: UDI/DI each 10195327170981, UDI/DI case 40195327170982, Lot Number: 23DBD282; Medline Kit Number/SKU CDS983173I: UDI/DI each 10195327170981, UDI/DI case 40195327170982, Lot Number: 22KBJ790; Medline Kit Number/SKU CDS983173I: UDI/DI each 10195327170981, UDI/DI case 40195327170982, Lot Number: 22HBD313; Medline Kit Number/SKU CDS983173I: UDI/DI each 10195327170981, UDI/DI case 40195327170982, Lot Number: 24DBC806; Medline Kit Number/SKU CDS983173I: UDI/DI each 10195327170981, UDI/DI case 40195327170982, Lot Number: 24BBN518; Medline Kit Number/SKU CDS983173I: UDI/DI each 10195327170981, UDI/DI case 40195327170982, Lot Number: 23LBL541; Medline Kit Number/SKU CDS983173I: UDI/DI each 10195327170981, UDI/DI case 40195327170982, Lot Number: 23LBD588; Medline Kit Number/SKU CDS983493L: UDI/DI each 10193489275643, UDI/DI case 40193489275644, Lot Number: 23HDB332; Medline Kit Number/SKU CDS983493L: UDI/DI each 10193489275643, UDI/DI case 40193489275644, Lot Number: 23GDA495; Medline Kit Number/SKU CDS983493L: UDI/DI each 10193489275643, UDI/DI case 40193489275644, Lot Number: 23FDB617; Medline Kit Number/SKU CDS983493L: UDI/DI each 10193489275643, UDI/DI case 40193489275644, Lot Number: 23EDB945; Medline Kit Number/SKU CDS983493L: UDI/DI each 10193489275643, UDI/DI case 40193489275644, Lot Number: 22JDA727; Medline Kit Number/SKU CDS983493L: UDI/DI each 10193489275643, UDI/DI case 40193489275644, Lot Number: 22FDB173; Medline Kit Number/SKU CDS983493L: UDI/DI each 10193489275643, UDI/DI case 40193489275644, Lot Number: 21JDB599; Medline Kit Number/SKU CDS983493L: UDI/DI each 10193489275643, UDI/DI case 40193489275644, Lot Number: 21JDB403; Medline Kit Number/SKU CDS983493L: UDI/DI each 10193489275643, UDI/DI case 40193489275644, Lot Number: 21IDC321; Medline Kit Number/SKU CDS983493L: UDI/DI each 10193489275643, UDI/DI case 40193489275644, Lot Number: 21IDB329; Medline Kit Number/SKU CDS983493L: UDI/DI each 10193489275643, UDI/DI case 40193489275644, Lot Number: 21HDA478; Medline Kit Number/SKU CDS983493L: UDI/DI each 10193489275643, UDI/DI case 40193489275644, Lot Number: 21GDC463; Medline Kit Number/SKU CDS983493L: UDI/DI each 10193489275643, UDI/DI case 40193489275644, Lot Number: 21FDA896; Medline Kit Number/SKU CDS983493L: UDI/DI each 10193489275643, UDI/DI case 40193489275644, Lot Number: 21EDB291; Medline Kit Number/SKU CDS983493L: UDI/DI each 10193489275643, UDI/DI case 40193489275644, Lot Number: 21CDC134; Medline Kit Number/SKU CDS983493L: UDI/DI each 10193489275643, UDI/DI case 40193489275644, Lot Number: 21CDB258; Medline Kit Number/SKU CDS983493M: UDI/DI each 10195327497415, UDI/DI case 40195327497416, Lot Number: 24ADA328; Medline Kit Number/SKU CDS983493M: UDI/DI each 10195327497415, UDI/DI case 40195327497416, Lot Number: 23LDA144; Medline Kit Number/SKU CDS983493O: UDI/DI each 10198459335549, UDI/DI case 40198459335540, Lot Number: 26CDA415; Medline Kit Number/SKU CDS983493O: UDI/DI each 10198459335549, UDI/DI case 40198459335540, Lot Number: 26BDA065; Medline Kit Number/SKU CDS983493O: UDI/DI each 10198459335549, UDI/DI case 40198459335540, Lot Number: 25LDA842; Medline Kit Number/SKU CDS983493O: UDI/DI each 10198459335549, UDI/DI case 40198459335540, Lot Number: 25IDB325; Medline Kit Number/SKU DYNJ0395085P: UDI/DI each 10193489432718, UDI/DI case 40193489432719, Lot Number: 21DKA483; Medline Kit Number/SKU DYNJ0660040R: UDI/DI each 10889942220083, UDI/DI case 40889942220084, Lot Number: 23IMD698; Medline Kit Number/SKU DYNJ0660040R: UDI/DI each 10889942220083, UDI/DI case 40889942220084, Lot Number: 23GMB527; Medline Kit Number/SKU DYNJ0660040R: UDI/DI each 10889942220083, UDI/DI case 40889942220084, Lot Number: 23FMF102; Medline Kit Number/SKU DYNJ0660040R: UDI/DI each 10889942220083, UDI/DI case 40889942220084, Lot Number: 23EMH628; Medline Kit Number/SKU DYNJ0660040R: UDI/DI each 10889942220083, UDI/DI case 40889942220084, Lot Number: 23CMA614; Medline Kit Number/SKU DYNJ0660040R: UDI/DI each 10889942220083, UDI/DI case 40889942220084, Lot Number: 23BME548; Medline Kit Number/SKU DYNJ0660040R: UDI/DI each 10889942220083, UDI/DI case 40889942220084, Lot Number: 22JMB623; Medline Kit Number/SKU DYNJ0660040R: UDI/DI each 10889942220083, UDI/DI case 40889942220084, Lot Number: 22GMG727; Medline Kit Number/SKU DYNJ0660040R: UDI/DI each 10889942220083, UDI/DI case 40889942220084, Lot Number: 22FME591; Medline Kit Number/SKU DYNJ0660040R: UDI/DI each 10889942220083, UDI/DI case 40889942220084, Lot Number: 22DMG455; Medline Kit Number/SKU DYNJ0660040R: UDI/DI each 10889942220083, UDI/DI case 40889942220084, Lot Number: 26AMA936; Medline Kit Number/SKU DYNJ0660040R: UDI/DI each 10889942220083, UDI/DI case 40889942220084, Lot Number: 25KMG652; Medline Kit Number/SKU DYNJ0660040R: UDI/DI each 10889942220083, UDI/DI case 40889942220084, Lot Number: 25JME572; Medline Kit Number/SKU DYNJ0660040R: UDI/DI each 10889942220083, UDI/DI case 40889942220084, Lot Number: 25GMF834; Medline Kit Number/SKU DYNJ0660040R: UDI/DI each 10889942220083, UDI/DI case 40889942220084, Lot Number: 25EMA958; Medline Kit Number/SKU DYNJ0660040R: UDI/DI each 10889942220083, UDI/DI case 40889942220084, Lot Number: 25DME904; Medline Kit Number/SKU DYNJ0660040R: UDI/DI each 10889942220083, UDI/DI case 40889942220084, Lot Number: 25CME519; Medline Kit Number/SKU DYNJ0660040R: UDI/DI each 10889942220083, UDI/DI case 40889942220084, Lot Number: 24LMF057; Medline Kit Number/SKU DYNJ0660040R: UDI/DI each 10889942220083, UDI/DI case 40889942220084, Lot Number: 24JMH001; Medline Kit Number/SKU DYNJ0660040R: UDI/DI each 10889942220083, UDI/DI case 40889942220084, Lot Number: 24IMC603; Medline Kit Number/SKU DYNJ0660040R: UDI/DI each 10889942220083, UDI/DI case 40889942220084, Lot Number: 24HMA307; Medline Kit Number/SKU DYNJ0660040R: UDI/DI each 10889942220083, UDI/DI case 40889942220084, Lot Number: 24EMH324; Medline Kit Number/SKU DYNJ0660040R: UDI/DI each 10889942220083, UDI/DI case 40889942220084, Lot Number: 24EMA549; Medline Kit Number/SKU DYNJ0660040R: UDI/DI each 10889942220083, UDI/DI case 40889942220084, Lot Number: 24AMJ677; Medline Kit Number/SKU DYNJ0660040R: UDI/DI each 10889942220083, UDI/DI case 40889942220084, Lot Number: 23LMC624; Medline Kit Number/SKU DYNJ0785741L: UDI/DI each 10193489935080, UDI/DI case 40193489935081, Lot Number: 23IMG288; Medline Kit Number/SKU DYNJ0785741L: UDI/DI each 10193489935080, UDI/DI case 40193489935081, Lot Number: 23GMF398; Medline Kit Number/SKU DYNJ0785741L: UDI/DI each 10193489935080, UDI/DI case 40193489935081, Lot Number: 23FMH043; Medline Kit Number/SKU DYNJ0785741L: UDI/DI each 10193489935080, UDI/DI case 40193489935081, Lot Number: 23FMD596; Medline Kit Number/SKU DYNJ0785741L: UDI/DI each 10193489935080, UDI/DI case 40193489935081, Lot Number: 23EMF866; Medline Kit Number/SKU DYNJ0785741L: UDI/DI each 10193489935080, UDI/DI case 40193489935081, Lot Number: 23DMH874; Medline Kit Number/SKU DYNJ0785741L: UDI/DI each 10193489935080, UDI/DI case 40193489935081, Lot Number: 23CMH139; Medline Kit Number/SKU DYNJ0785741L: UDI/DI each 10193489935080, UDI/DI case 40193489935081, Lot Number: 23CME193; Medline Kit Number/SKU DYNJ0785741L: UDI/DI each 10193489935080, UDI/DI case 40193489935081, Lot Number: 23BMD471; Medline Kit Number/SKU DYNJ0785741L: UDI/DI each 10193489935080, UDI/DI case 40193489935081, Lot Number: 22LMD522; Medline Kit Number/SKU DYNJ0785741L: UDI/DI each 10193489935080, UDI/DI case 40193489935081, Lot Number: 22HMG591; Medline Kit Number/SKU DYNJ0785741L: UDI/DI each 10193489935080, UDI/DI case 40193489935081, Lot Number: 22GME991; Medline Kit Number/SKU DYNJ0785741L: UDI/DI each 10193489935080, UDI/DI case 40193489935081, Lot Number: 22GME992; Medline Kit Number/SKU DYNJ0785741L: UDI/DI each 10193489935080, UDI/DI case 40193489935081, Lot Number: 22DMC069; Medline Kit Number/SKU DYNJ0785741L: UDI/DI each 10193489935080, UDI/DI case 40193489935081, Lot Number: 22OBD206; Medline Kit Number/SKU DYNJ0785741L: UDI/DI each 10193489935080, UDI/DI case 40193489935081, Lot Number: 22CBF109; Medline Kit Number/SKU DYNJ0785741L: UDI/DI each 10193489935080, UDI/DI case 40193489935081, Lot Number: 21IBM538; Medline Kit Number/SKU DYNJ0785741L: UDI/DI each 10193489935080, UDI/DI case 40193489935081, Lot Number: 21IBM610; Medline Kit Number/SKU DYNJ0785741L: UDI/DI each 10193489935080, UDI/DI case 40193489935081, Lot Number: 21FBP997; Medline Kit Number/SKU DYNJ0785741L: UDI/DI each 10193489935080, UDI/DI case 40193489935081, Lot Number: 24JMH447; Medline Kit Number/SKU DYNJ0785741L: UDI/DI each 10193489935080, UDI/DI case 40193489935081, Lot Number: 24IMF995; Medline Kit Number/SKU DYNJ0785741L: UDI/DI each 10193489935080, UDI/DI case 40193489935081, Lot Number: 24FMI942; Medline Kit Number/SKU DYNJ0785741L: UDI/DI each 10193489935080, UDI/DI case 40193489935081, Lot Number: 24EME920; Medline Kit Number/SKU DYNJ0785741L: UDI/DI each 10193489935080, UDI/DI case 40193489935081, Lot Number: 24DMJ136; Medline Kit Number/SKU DYNJ0785741L: UDI/DI each 10193489935080, UDI/DI case 40193489935081, Lot Number: 23LMH895; Medline Kit Number/SKU DYNJ0991320L: UDI/DI each 10195327028541, UDI/DI case 40195327028542, Lot Number: 23KMC601; Medline Kit Number/SKU DYNJ0991320L: UDI/DI each 10195327028541, UDI/DI case 40195327028542, Lot Number: 23GMC715; Medline Kit Number/SKU DYNJ0991320L: UDI/DI each 10195327028541, UDI/DI case 40195327028542, Lot Number: 23DME468; Medline Kit Number/SKU DYNJ0991320L: UDI/DI each 10195327028541, UDI/DI case 40195327028542, Lot Number: 23BMC887; Medline Kit Number/SKU DYNJ0991320L: UDI/DI each 10195327028541, UDI/DI case 40195327028542, Lot Number: 21LLA690; Medline Kit Number/SKU DYNJ0991320L: UDI/DI each 10195327028541, UDI/DI case 40195327028542, Lot Number: 24BMA691; Medline Kit Number/SKU DYNJ0991320L: UDI/DI each 10195327028541, UDI/DI case 40195327028542, Lot Number: 24AMH656; Medline Kit Number/SKU DYNJ0991320L: UDI/DI each 10195327028541, UDI/DI case 40195327028542, Lot Number: 23LMB786; Medline Kit Number/SKU DYNJ39211F: UDI/DI each 10195327313654, UDI/DI case 40195327313655, Lot Number: 23BMD216; Medline Kit Number/SKU DYNJ40648F: UDI/DI each 10889942611041, UDI/DI case 40889942611042, Lot Number: 22AMG375; Medline Kit Number/SKU DYNJ40648F: UDI/DI each 10889942611041, UDI/DI case 40889942611042, Lot Number: 21KMG254; Medline Kit Number/SKU DYNJ40648F: UDI/DI each 10889942611041, UDI/DI case 40889942611042, Lot Number: 21HMB060; Medline Kit Number/SKU DYNJ40648F: UDI/DI each 10889942611041, UDI/DI case 40889942611042, Lot Number: 21FMD643; Medline Kit Number/SKU DYNJ40648F: UDI/DI each 10889942611041, UDI/DI case 40889942611042, Lot Number: 21DME180; Medline Kit Number/SKU DYNJ41922C: UDI/DI each 10193489677355, UDI/DI case 40193489677356, Lot Number: 23IBB468; Medline Kit Number/SKU DYNJ41922C: UDI/DI each 10193489677355, UDI/DI case 40193489677356, Lot Number: 23FBO732; Medline Kit Number/SKU DYNJ41922C: UDI/DI each 10193489677355, UDI/DI case 40193489677356, Lot Number: 23EBV753; Medline Kit Number/SKU DYNJ41922C: UDI/DI each 10193489677355, UDI/DI case 40193489677356, Lot Number: 23EBQ610; Medline Kit Number/SKU DYNJ41922C: UDI/DI each 10193489677355, UDI/DI case 40193489677356, Lot Number: 23DBM685; Medline Kit Number/SKU DYNJ41922C: UDI/DI each 10193489677355, UDI/DI case 40193489677356, Lot Number: 23ABH385; Medline Kit Number/SKU DYNJ41922C: UDI/DI each 10193489677355, UDI/DI case 40193489677356, Lot Number: 22GBK343; Medline Kit Number/SKU DYNJ41922C: UDI/DI each 10193489677355, UDI/DI case 40193489677356, Lot Number: 22EBB123; Medline Kit Number/SKU DYNJ41922C: UDI/DI each 10193489677355, UDI/DI case 40193489677356, Lot Number: 22ABE215; Medline Kit Number/SKU DYNJ41922C: UDI/DI each 10193489677355, UDI/DI case 40193489677356, Lot Number: 21ABM833; Medline Kit Number/SKU DYNJ41922C: UDI/DI each 10193489677355, UDI/DI case 40193489677356, Lot Number: 24JBE450; Medline Kit Number/SKU DYNJ41922C: UDI/DI each 10193489677355, UDI/DI case 40193489677356, Lot Number: 24JBF281; Medline Kit Number/SKU DYNJ41922C: UDI/DI each 10193489677355, UDI/DI case 40193489677356, Lot Number: 24JBA800; Medline Kit Number/SKU DYNJ41922C: UDI/DI each 10193489677355, UDI/DI case 40193489677356, Lot Number: 24IBT612; Medline Kit Number/SKU DYNJ41922C: UDI/DI each 10193489677355, UDI/DI case 40193489677356, Lot Number: 24FBS381; Medline Kit Number/SKU DYNJ41922C: UDI/DI each 10193489677355, UDI/DI case 40193489677356, Lot Number: 23LBN005; Medline Kit Number/SKU DYNJ42833B: UDI/DI each 10193489880335, UDI/DI case 40193489880336, Lot Number: 26BBK430; Medline Kit Number/SKU DYNJ42833B: UDI/DI each 10193489880335, UDI/DI case 40193489880336, Lot Number: 25JBU219; Medline Kit Number/SKU DYNJ42833B: UDI/DI each 10193489880335, UDI/DI case 40193489880336, Lot Number: 25GBW762; Medline Kit Number/SKU DYNJ42833B: UDI/DI each 10193489880335, UDI/DI case 40193489880336, Lot Number: 25DBU105; Medline Kit Number/SKU DYNJ43859J: UDI/DI each 10193489893700, UDI/DI case 40193489893701, Lot Number: 23EME342; Medline Kit Number/SKU DYNJ44759F: UDI/DI each 10193489789935, UDI/DI case 40193489789936, Lot Number: 22IBI991; Medline Kit Number/SKU DYNJ44759F: UDI/DI each 10193489789935, UDI/DI case 40193489789936, Lot Number: 22EBP134; Medline Kit Number/SKU DYNJ44759F: UDI/DI each 10193489789935, UDI/DI case 40193489789936, Lot Number: 22ABX616; Medline Kit Number/SKU DYNJ44759F: UDI/DI each 10193489789935, UDI/DI case 40193489789936, Lot Number: 22ABA063; Medline Kit Number/SKU DYNJ44759F: UDI/DI each 10193489789935, UDI/DI case 40193489789936, Lot Number: 21JBD678; Medline Kit Number/SKU DYNJ44759F: UDI/DI each 10193489789935, UDI/DI case 40193489789936, Lot Number: 21HBQ587; Medline Kit Number/SKU DYNJ44759F: UDI/DI each 10193489789935, UDI/DI case 40193489789936, Lot Number: 21DBC693; Medline Kit Number/SKU DYNJ46504C: UDI/DI each 10195327005221, UDI/DI case 40195327005222, Lot Number: 23GME117; Medline Kit Number/SKU DYNJ46504C: UDI/DI each 10195327005221, UDI/DI case 40195327005222, Lot Number: 23EMI284; Medline Kit Number/SKU DYNJ46504C: UDI/DI each 10195327005221, UDI/DI case 40195327005222, Lot Number: 23AMJ001; Medline Kit Number/SKU DYNJ46504C: UDI/DI each 10195327005221, UDI/DI case 40195327005222, Lot Number: 22KMD020; Medline Kit Number/SKU DYNJ46504C: UDI/DI each 10195327005221, UDI/DI case 40195327005222, Lot Number: 22GBW975; Medline Kit Number/SKU DYNJ46504C: UDI/DI each 10195327005221, UDI/DI case 40195327005222, Lot Number: 22FBW091; Medline Kit Number/SKU DYNJ46504C: UDI/DI each 10195327005221, UDI/DI case 40195327005222, Lot Number: 22OBJ071; Medline Kit Number/SKU DYNJ46504C: UDI/DI each 10195327005221, UDI/DI case 40195327005222, Lot Number: 22CBC147; Medline Kit Number/SKU DYNJ46504C: UDI/DI each 10195327005221, UDI/DI case 40195327005222, Lot Number: 21VBD092; Medline Kit Number/SKU DYNJ46504D: UDI/DI each 10195327477585, UDI/DI case 40195327477586, Lot Number: 23HMG668; Medline Kit Number/SKU DYNJ46504D: UDI/DI each 10195327477585, UDI/DI case 40195327477586, Lot Number: 24JMC633; Medline Kit Number/SKU DYNJ46504D: UDI/DI each 10195327477585, UDI/DI case 40195327477586, Lot Number: 24IMI654; Medline Kit Number/SKU DYNJ46504D: UDI/DI each 10195327477585, UDI/DI case 40195327477586, Lot Number: 24IMB838; Medline Kit Number/SKU DYNJ46504D: UDI/DI each 10195327477585, UDI/DI case 40195327477586, Lot Number: 24HMB162; Medline Kit Number/SKU DYNJ46504D: UDI/DI each 10195327477585, UDI/DI case 40195327477586, Lot Number: 24FMG045; Medline Kit Number/SKU DYNJ46504D: UDI/DI each 10195327477585, UDI/DI case 40195327477586, Lot Number: 24EMD442; Medline Kit Number/SKU DYNJ46504D: UDI/DI each 10195327477585, UDI/DI case 40195327477586, Lot Number: 24DME855; Medline Kit Number/SKU DYNJ46504D: UDI/DI each 10195327477585, UDI/DI case 40195327477586, Lot Number: 24CMH418; Medline Kit Number/SKU DYNJ46504D: UDI/DI each 10195327477585, UDI/DI case 40195327477586, Lot Number: 24BMK020; Medline Kit Number/SKU DYNJ46504D: UDI/DI each 10195327477585, UDI/DI case 40195327477586, Lot Number: 24AMG855; Medline Kit Number/SKU DYNJ46504D: UDI/DI each 10195327477585, UDI/DI case 40195327477586, Lot Number: 24AMB928; Medline Kit Number/SKU DYNJ46504F: UDI/DI each 10198459140761, UDI/DI case 40198459140762, Lot Number: 25HMI601; Medline Kit Number/SKU DYNJ46504F: UDI/DI each 10198459140761, UDI/DI case 40198459140762, Lot Number: 25GMG101; Medline Kit Number/SKU DYNJ46504F: UDI/DI each 10198459140761, UDI/DI case 40198459140762, Lot Number: 25EMF342; Medline Kit Number/SKU DYNJ46781B: UDI/DI each 10193489825916, UDI/DI case 40193489825917, Lot Number: 22JBA979; Medline Kit Number/SKU DYNJ46781B: UDI/DI each 10193489825916, UDI/DI case 40193489825917, Lot Number: 26BBR125; Medline Kit Number/SKU DYNJ46781B: UDI/DI each 10193489825916, UDI/DI case 40193489825917, Lot Number: 25KBQ493; Medline Kit Number/SKU DYNJ46781B: UDI/DI each 10193489825916, UDI/DI case 40193489825917, Lot Number: 24JBD583; Medline Kit Number/SKU DYNJ46781B: UDI/DI each 10193489825916, UDI/DI case 40193489825917, Lot Number: 24HBA712; Medline Kit Number/SKU DYNJ46781B: UDI/DI each 10193489825916, UDI/DI case 40193489825917, Lot Number: 24EBM092; Medline Kit Number/SKU DYNJ46781B: UDI/DI each 10193489825916, UDI/DI case 40193489825917, Lot Number: 24DBK313; Medline Kit Number/SKU DYNJ46781B: UDI/DI each 10193489825916, UDI/DI case 40193489825917, Lot Number: 24BBO353; Medline Kit Number/SKU DYNJ48191G: UDI/DI each 10193489476606, UDI/DI case 40193489476607, Lot Number: 21JMH268; Medline Kit Number/SKU DYNJ48191G: UDI/DI each 10193489476606, UDI/DI case 40193489476607, Lot Number: 21JMF238; Medline Kit Number/SKU DYNJ48191G: UDI/DI each 10193489476606, UDI/DI case 40193489476607, Lot Number: 21DMB033; Medline Kit Number/SKU DYNJ48191I: UDI/DI each 10195327057817, UDI/DI case 40195327057818, Lot Number: 22EME800; Medline Kit Number/SKU DYNJ48191I: UDI/DI each 10195327057817, UDI/DI case 40195327057818, Lot Number: 22CMD031; Medline Kit Number/SKU DYNJ48191J: UDI/DI each 10195327255039, UDI/DI case 40195327255030, Lot Number: 23EMA429; Medline Kit Number/SKU DYNJ48191J: UDI/DI each 10195327255039, UDI/DI case 40195327255030, Lot Number: 23BMD353; Medline Kit Number/SKU DYNJ48191J: UDI/DI each 10195327255039, UDI/DI case 40195327255030, Lot Number: 22LMH655; Medline Kit Number/SKU DYNJ48191J: UDI/DI each 10195327255039, UDI/DI case 40195327255030, Lot Number: 22JME786; Medline Kit Number/SKU DYNJ48191K: UDI/DI each 10195327437541, UDI/DI case 40195327437542, Lot Number: 23GMG423; Medline Kit Number/SKU DYNJ48191K: UDI/DI each 10195327437541, UDI/DI case 40195327437542, Lot Number: 23LMH905; Medline Kit Number/SKU DYNJ48191L: UDI/DI each 10195327630690, UDI/DI case 40195327630691, Lot Number: 24EMH956; Medline Kit Number/SKU DYNJ48191M: UDI/DI each 10198459078910, UDI/DI case 40198459078911, Lot Number: 25GMD380; Medline Kit Number/SKU DYNJ48191M: UDI/DI each 10198459078910, UDI/DI case 40198459078911, Lot Number: 25EMA899; Medline Kit Number/SKU DYNJ48191M: UDI/DI each 10198459078910, UDI/DI case 40198459078911, Lot Number: 25BMD703; Medline Kit Number/SKU DYNJ48191M: UDI/DI each 10198459078910, UDI/DI case 40198459078911, Lot Number: 25BMB374; Medline Kit Number/SKU DYNJ48191O: UDI/DI each 10198459602641, UDI/DI case 40198459602642, Lot Number: 26AMB007; Medline Kit Number/SKU DYNJ53076D: UDI/DI each 10193489239119, UDI/DI case 40193489239110, Lot Number: 21JMD339; Medline Kit Number/SKU DYNJ53076D: UDI/DI each 10193489239119, UDI/DI case 40193489239110, Lot Number: 21FMF721; Medline Kit Number/SKU DYNJ53076D: UDI/DI each 10193489239119, UDI/DI case 40193489239110, Lot Number: 21FME498; Medline Kit Number/SKU DYNJ53076D: UDI/DI each 10193489239119, UDI/DI case 40193489239110, Lot Number: 21EMA403; Medline Kit Number/SKU DYNJ54761B: UDI/DI each 10193489434286, UDI/DI case 40193489434287, Lot Number: 21HBG397; Medline Kit Number/SKU DYNJ54761B: UDI/DI each 10193489434286, UDI/DI case 40193489434287, Lot Number: 21FBE715; Medline Kit Number/SKU DYNJ54761B: UDI/DI each 10193489434286, UDI/DI case 40193489434287, Lot Number: 21CBW195; Medline Kit Number/SKU DYNJ54761B: UDI/DI each 10193489434286, UDI/DI case 40193489434287, Lot Number: 21ABG033; Medline Kit Number/SKU DYNJ54761B: UDI/DI each 10193489434286, UDI/DI case 40193489434287, Lot Number: 21ABH228; Medline Kit Number/SKU DYNJ54761C: UDI/DI each 10195327020217, UDI/DI case 40195327020218, Lot Number: 23JBY274; Medline Kit Number/SKU DYNJ54761C: UDI/DI each 10195327020217, UDI/DI case 40195327020218, Lot Number: 23FBM738; Medline Kit Number/SKU DYNJ54761C: UDI/DI each 10195327020217, UDI/DI case 40195327020218, Lot Number: 23EBS795; Medline Kit Number/SKU DYNJ54761C: UDI/DI each 10195327020217, UDI/DI case 40195327020218, Lot Number: 23DBQ176; Medline Kit Number/SKU DYNJ54761C: UDI/DI each 10195327020217, UDI/DI case 40195327020218, Lot Number: 23DBB931; Medline Kit Number/SKU DYNJ54761C: UDI/DI each 10195327020217, UDI/DI case 40195327020218, Lot Number: 23CBB075; Medline Kit Number/SKU DYNJ54761C: UDI/DI each 10195327020217, UDI/DI case 40195327020218, Lot Number: 23BBC518; Medline Kit Number/SKU DYNJ54761C: UDI/DI each 10195327020217, UDI/DI case 40195327020218, Lot Number: 22JBD638; Medline Kit Number/SKU DYNJ54761C: UDI/DI each 10195327020217, UDI/DI case 40195327020218, Lot Number: 22FBJ435; Medline Kit Number/SKU DYNJ54761C: UDI/DI each 10195327020217, UDI/DI case 40195327020218, Lot Number: 22EBL603; Medline Kit Number/SKU DYNJ54761C: UDI/DI each 10195327020217, UDI/DI case 40195327020218, Lot Number: 22CBV860; Medline Kit Number/SKU DYNJ54761C: UDI/DI each 10195327020217, UDI/DI case 40195327020218, Lot Number: 22CBP225; Medline Kit Number/SKU DYNJ54761C: UDI/DI each 10195327020217, UDI/DI case 40195327020218, Lot Number: 21LBU279; Medline Kit Number/SKU DYNJ54761C: UDI/DI each 10195327020217, UDI/DI case 40195327020218, Lot Number: 21KBP799; Medline Kit Number/SKU DYNJ54761C: UDI/DI each 10195327020217, UDI/DI case 40195327020218, Lot Number: 26CBB708; Medline Kit Number/SKU DYNJ54761C: UDI/DI each 10195327020217, UDI/DI case 40195327020218, Lot Number: 26ABM764; Medline Kit Number/SKU DYNJ54761C: UDI/DI each 10195327020217, UDI/DI case 40195327020218, Lot Number: 25LBQ922; Medline Kit Number/SKU DYNJ54761C: UDI/DI each 10195327020217, UDI/DI case 40195327020218, Lot Number: 25KBI476; Medline Kit Number/SKU DYNJ54761C: UDI/DI each 10195327020217, UDI/DI case 40195327020218, Lot Number: 25IBS238; Medline Kit Number/SKU DYNJ54761C: UDI/DI each 10195327020217, UDI/DI case 40195327020218, Lot Number: 25IBB148; Medline Kit Number/SKU DYNJ54761C: UDI/DI each 10195327020217, UDI/DI case 40195327020218, Lot Number: 25HBM494; Medline Kit Number/SKU DYNJ54761C: UDI/DI each 10195327020217, UDI/DI case 40195327020218, Lot Number: 25FBI994; Medline Kit Number/SKU DYNJ54761C: UDI/DI each 10195327020217, UDI/DI case 40195327020218, Lot Number: 25DBU225; Medline Kit Number/SKU DYNJ54761C: UDI/DI each 10195327020217, UDI/DI case 40195327020218, Lot Number: 25BBE810; Medline Kit Number/SKU DYNJ54761C: UDI/DI each 10195327020217, UDI/DI case 40195327020218, Lot Number: 25ABK916; Medline Kit Number/SKU DYNJ54761C: UDI/DI each 10195327020217, UDI/DI case 40195327020218, Lot Number: 24LBB828; Medline Kit Number/SKU DYNJ54761C: UDI/DI each 10195327020217, UDI/DI case 40195327020218, Lot Number: 24JBX741; Medline Kit Number/SKU DYNJ54761C: UDI/DI each 10195327020217, UDI/DI case 40195327020218, Lot Number: 24IBR297; Medline Kit Number/SKU DYNJ54761C: UDI/DI each 10195327020217, UDI/DI case 40195327020218, Lot Number: 24HBI405; Medline Kit Number/SKU DYNJ54761C: UDI/DI each 10195327020217, UDI/DI case 40195327020218, Lot Number: 24FBP312; Medline Kit Number/SKU DYNJ54761C: UDI/DI each 10195327020217, UDI/DI case 40195327020218, Lot Number: 24EBN616; Medline Kit Number/SKU DYNJ54761C: UDI/DI each 10195327020217, UDI/DI case 40195327020218, Lot Number: 24CBL687; Medline Kit Number/SKU DYNJ54761C: UDI/DI each 10195327020217, UDI/DI case 40195327020218, Lot Number: 24CBJ095; Medline Kit Number/SKU DYNJ54761C: UDI/DI each 10195327020217, UDI/DI case 40195327020218, Lot Number: 24BBS591; Medline Kit Number/SKU DYNJ57604: UDI/DI each 10889942756902, UDI/DI case 40889942756903, Lot Number: 23IBD079; Medline Kit Number/SKU DYNJ57604: UDI/DI each 10889942756902, UDI/DI case 40889942756903, Lot Number: 23GBL994; Medline Kit Number/SKU DYNJ57604: UDI/DI each 10889942756902, UDI/DI case 40889942756903, Lot Number: 23EBS598; Medline Kit Number/SKU DYNJ57604: UDI/DI each 10889942756902, UDI/DI case 40889942756903, Lot Number: 23EBL984; Medline Kit Number/SKU DYNJ57604: UDI/DI each 10889942756902, UDI/DI case 40889942756903, Lot Number: 23DBH411; Medline Kit Number/SKU DYNJ57604: UDI/DI each 10889942756902, UDI/DI case 40889942756903, Lot Number: 23CBV513; Medline Kit Number/SKU DYNJ57604: UDI/DI each 10889942756902, UDI/DI case 40889942756903, Lot Number: 23BBD492; Medline Kit Number/SKU DYNJ57604: UDI/DI each 10889942756902, UDI/DI case 40889942756903, Lot Number: 22DBQ354; Medline Kit Number/SKU DYNJ57604: UDI/DI each 10889942756902, UDI/DI case 40889942756903, Lot Number: 22BBB405; Medline Kit Number/SKU DYNJ57604: UDI/DI each 10889942756902, UDI/DI case 40889942756903, Lot Number: 21KBW367; Medline Kit Number/SKU DYNJ57604: UDI/DI each 10889942756902, UDI/DI case 40889942756903, Lot Number: 21KBC969; Medline Kit Number/SKU DYNJ57604: UDI/DI each 10889942756902, UDI/DI case 40889942756903, Lot Number: 21FBE526; Medline Kit Number/SKU DYNJ57604: UDI/DI each 10889942756902, UDI/DI case 40889942756903, Lot Number: 21FBE506; Medline Kit Number/SKU DYNJ57604: UDI/DI each 10889942756902, UDI/DI case 40889942756903, Lot Number: 21DBM118; Medline Kit Number/SKU DYNJ57604: UDI/DI each 10889942756902, UDI/DI case 40889942756903, Lot Number: 21DBD151; Medline Kit Number/SKU DYNJ57604: UDI/DI each 10889942756902, UDI/DI case 40889942756903, Lot Number: 21BBA803; Medline Kit Number/SKU DYNJ57604: UDI/DI each 10889942756902, UDI/DI case 40889942756903, Lot Number: 21ABF228; Medline Kit Number/SKU DYNJ57604: UDI/DI each 10889942756902, UDI/DI case 40889942756903, Lot Number: 26BBS083; Medline Kit Number/SKU DYNJ57604: UDI/DI each 10889942756902, UDI/DI case 40889942756903, Lot Number: 25IBE859; Medline Kit Number/SKU DYNJ57604: UDI/DI each 10889942756902, UDI/DI case 40889942756903, Lot Number: 25BBJ593; Medline Kit Number/SKU DYNJ57604: UDI/DI each 10889942756902, UDI/DI case 40889942756903, Lot Number: 25BBJ601; Medline Kit Number/SKU DYNJ57604: UDI/DI each 10889942756902, UDI/DI case 40889942756903, Lot Number: 24HBD973; Medline Kit Number/SKU DYNJ57604: UDI/DI each 10889942756902, UDI/DI case 40889942756903, Lot Number: 24FBB019; Medline Kit Number/SKU DYNJ57604: UDI/DI each 10889942756902, UDI/DI case 40889942756903, Lot Number: 24CBK963; Medline Kit Number/SKU DYNJ57604: UDI/DI each 10889942756902, UDI/DI case 40889942756903, Lot Number: 23KBX604; Medline Kit Number/SKU DYNJ59400K: UDI/DI each 10195327492489, UDI/DI case 40195327492480, Lot Number: 24GMC759; Medline Kit Number/SKU DYNJ59456A: UDI/DI each 10193489362060, UDI/DI case 40193489362061, Lot Number: 23IBC981; Medline Kit Number/SKU DYNJ59456A: UDI/DI each 10193489362060, UDI/DI case 40193489362061, Lot Number: 23CBB054; Medline Kit Number/SKU DYNJ59456A: UDI/DI each 10193489362060, UDI/DI case 40193489362061, Lot Number: 23ABK120; Medline Kit Number/SKU DYNJ59456A: UDI/DI each 10193489362060, UDI/DI case 40193489362061, Lot Number: 22GBU138; Medline Kit Number/SKU DYNJ59456A: UDI/DI each 10193489362060, UDI/DI case 40193489362061, Lot Number: 22BBT750; Medline Kit Number/SKU DYNJ59456A: UDI/DI each 10193489362060, UDI/DI case 40193489362061, Lot Number: 21JBZ859; Medline Kit Number/SKU DYNJ59456A: UDI/DI each 10193489362060, UDI/DI case 40193489362061, Lot Number: 21FBA944; Medline Kit Number/SKU DYNJ59456A: UDI/DI each 10193489362060, UDI/DI case 40193489362061, Lot Number: 21ABM493; Medline Kit Number/SKU DYNJ59456A: UDI/DI each 10193489362060, UDI/DI case 40193489362061, Lot Number: 26BBG778; Medline Kit Number/SKU DYNJ59456A: UDI/DI each 10193489362060, UDI/DI case 40193489362061, Lot Number: 25KBN868; Medline Kit Number/SKU DYNJ59456A: UDI/DI each 10193489362060, UDI/DI case 40193489362061, Lot Number: 25EBF013; Medline Kit Number/SKU DYNJ59456A: UDI/DI each 10193489362060, UDI/DI case 40193489362061, Lot Number: 24LBK254; Medline Kit Number/SKU DYNJ59456A: UDI/DI each 10193489362060, UDI/DI case 40193489362061, Lot Number: 24HBM708; Medline Kit Number/SKU DYNJ59456A: UDI/DI each 10193489362060, UDI/DI case 40193489362061, Lot Number: 24DBB730; Medline Kit Number/SKU DYNJ59456A: UDI/DI each 10193489362060, UDI/DI case 40193489362061, Lot Number: 24ABC264; Medline Kit Number/SKU DYNJ60392C: UDI/DI each 10193489338973, UDI/DI case 40193489338974, Lot Number: 23EMG712; Medline Kit Number/SKU DYNJ60392C: UDI/DI each 10193489338973, UDI/DI case 40193489338974, Lot Number: 23EMA976; Medline Kit Number/SKU DYNJ60392C: UDI/DI each 10193489338973, UDI/DI case 40193489338974, Lot Number: 23CMH213; Medline Kit Number/SKU DYNJ60392C: UDI/DI each 10193489338973, UDI/DI case 40193489338974, Lot Number: 23BMH981; Medline Kit Number/SKU DYNJ60392C: UDI/DI each 10193489338973, UDI/DI case 40193489338974, Lot Number: 23AMI112; Medline Kit Number/SKU DYNJ60392C: UDI/DI each 10193489338973, UDI/DI case 40193489338974, Lot Number: 22JMF099; Medline Kit Number/SKU DYNJ60392C: UDI/DI each 10193489338973, UDI/DI case 40193489338974, Lot Number: 22EMA459; Medline Kit Number/SKU DYNJ60392C: UDI/DI each 10193489338973, UDI/DI case 40193489338974, Lot Number: 22DMB172; Medline Kit Number/SKU DYNJ60392C: UDI/DI each 10193489338973, UDI/DI case 40193489338974, Lot Number: 22BMG965; Medline Kit Number/SKU DYNJ60392C: UDI/DI each 10193489338973, UDI/DI case 40193489338974, Lot Number: 21LMG903; Medline Kit Number/SKU DYNJ60392C: UDI/DI each 10193489338973, UDI/DI case 40193489338974, Lot Number: 21KMH623; Medline Kit Number/SKU DYNJ60392C: UDI/DI each 10193489338973, UDI/DI case 40193489338974, Lot Number: 21JMH398; Medline Kit Number/SKU DYNJ60392C: UDI/DI each 10193489338973, UDI/DI case 40193489338974, Lot Number: 21JMF247; Medline Kit Number/SKU DYNJ60392C: UDI/DI each 10193489338973, UDI/DI case 40193489338974, Lot Number: 21HMD136; Medline Kit Number/SKU DYNJ60392C: UDI/DI each 10193489338973, UDI/DI case 40193489338974, Lot Number: 21GMD160; Medline Kit Number/SKU DYNJ60392C: UDI/DI each 10193489338973, UDI/DI case 40193489338974, Lot Number: 21FMC247; Medline Kit Number/SKU DYNJ60392D: UDI/DI each 10195327434380, UDI/DI case 40195327434381, Lot Number: 23JMG294; Medline Kit Number/SKU DYNJ60392D: UDI/DI each 10195327434380, UDI/DI case 40195327434381, Lot Number: 23IMC541; Medline Kit Number/SKU DYNJ60392D: UDI/DI each 10195327434380, UDI/DI case 40195327434381, Lot Number: 23GMF129; Medline Kit Number/SKU DYNJ60392D: UDI/DI each 10195327434380, UDI/DI case 40195327434381, Lot Number: 23FMI543; Medline Kit Number/SKU DYNJ60392D: UDI/DI each 10195327434380, UDI/DI case 40195327434381, Lot Number: 25CMG672; Medline Kit Number/SKU DYNJ60392D: UDI/DI each 10195327434380, UDI/DI case 40195327434381, Lot Number: 25AMB065; Medline Kit Number/SKU DYNJ60392D: UDI/DI each 10195327434380, UDI/DI case 40195327434381, Lot Number: 24LMA593; Medline Kit Number/SKU DYNJ60392D: UDI/DI each 10195327434380, UDI/DI case 40195327434381, Lot Number: 24JMI177; Medline Kit Number/SKU DYNJ60392D: UDI/DI each 10195327434380, UDI/DI case 40195327434381, Lot Number: 24HMH536; Medline Kit Number/SKU DYNJ60392D: UDI/DI each 10195327434380, UDI/DI case 40195327434381, Lot Number: 24HMD457; Medline Kit Number/SKU DYNJ60392D: UDI/DI each 10195327434380, UDI/DI case 40195327434381, Lot Number: 24GMJ778; Medline Kit Number/SKU DYNJ60392D: UDI/DI each 10195327434380, UDI/DI case 40195327434381, Lot Number: 24FMC742; Medline Kit Number/SKU DYNJ60392D: UDI/DI each 10195327434380, UDI/DI case 40195327434381, Lot Number: 24EME114; Medline Kit Number/SKU DYNJ60392D: UDI/DI each 10195327434380, UDI/DI case 40195327434381, Lot Number: 24CMI272; Medline Kit Number/SKU DYNJ60392D: UDI/DI each 10195327434380, UDI/DI case 40195327434381, Lot Number: 23KMJ516; Medline Kit Number/SKU DYNJ60392F: UDI/DI each 10198459313424, UDI/DI case 40198459313425, Lot Number: 25EMH894; Medline Kit Number/SKU DYNJ60392F: UDI/DI each 10198459313424, UDI/DI case 40198459313425, Lot Number: 25DMJ856; Medline Kit Number/SKU DYNJ62636: UDI/DI each 10193489675108, UDI/DI case 40193489675109, Lot Number: 21AMA163; Medline Kit Number/SKU DYNJ62803: UDI/DI each 10193489688993, UDI/DI case 40193489688994, Lot Number: 23KBT422; Medline Kit Number/SKU DYNJ62803: UDI/DI each 10193489688993, UDI/DI case 40193489688994, Lot Number: 23GBF930; Medline Kit Number/SKU DYNJ62803: UDI/DI each 10193489688993, UDI/DI case 40193489688994, Lot Number: 22HBQ384; Medline Kit Number/SKU DYNJ62803: UDI/DI each 10193489688993, UDI/DI case 40193489688994, Lot Number: 22GBS133; Medline Kit Number/SKU DYNJ62803: UDI/DI each 10193489688993, UDI/DI case 40193489688994, Lot Number: 26ABN799; Medline Kit Number/SKU DYNJ62803: UDI/DI each 10193489688993, UDI/DI case 40193489688994, Lot Number: 25KBN615; Medline Kit Number/SKU DYNJ62803: UDI/DI each 10193489688993, UDI/DI case 40193489688994, Lot Number: 25FBQ548; Medline Kit Number/SKU DYNJ62803: UDI/DI each 10193489688993, UDI/DI case 40193489688994, Lot Number: 25FBO124; Medline Kit Number/SKU DYNJ62803: UDI/DI each 10193489688993, UDI/DI case 40193489688994, Lot Number: 25FBF480; Medline Kit Number/SKU DYNJ62803: UDI/DI each 10193489688993, UDI/DI case 40193489688994, Lot Number: 24LBP070; Medline Kit Number/SKU DYNJ62803: UDI/DI each 10193489688993, UDI/DI case 40193489688994, Lot Number: 23LBN303; Medline Kit Number/SKU DYNJ63332D: UDI/DI each 10195327062675, UDI/DI case 40195327062676, Lot Number: 22IBS344; Medline Kit Number/SKU DYNJ63332D: UDI/DI each 10195327062675, UDI/DI case 40195327062676, Lot Number: 22GBC932; Medline Kit Number/SKU DYNJ63332D: UDI/DI each 10195327062675, UDI/DI case 40195327062676, Lot Number: 22DBL674; Medline Kit Number/SKU DYNJ63332D: UDI/DI each 10195327062675, UDI/DI case 40195327062676, Lot Number: 22DBB480; Medline Kit Number/SKU DYNJ63332D: UDI/DI each 10195327062675, UDI/DI case 40195327062676, Lot Number: 22CBV253; Medline Kit Number/SKU DYNJ63332D: UDI/DI each 10195327062675, UDI/DI case 40195327062676, Lot Number: 21LBO403; Medline Kit Number/SKU DYNJ64658: UDI/DI each 10193489210569, UDI/DI case 40193489210560, Lot Number: 22EBR214; Medline Kit Number/SKU DYNJ64658: UDI/DI each 10193489210569, UDI/DI case 40193489210560, Lot Number: 22BBG688; Medline Kit Number/SKU DYNJ64658: UDI/DI each 10193489210569, UDI/DI case 40193489210560, Lot Number: 22ABW065; Medline Kit Number/SKU DYNJ64658: UDI/DI each 10193489210569, UDI/DI case 40193489210560, Lot Number: 21KBK820; Medline Kit Number/SKU DYNJ64658: UDI/DI each 10193489210569, UDI/DI case 40193489210560, Lot Number: 21JBL906; Medline Kit Number/SKU DYNJ64658: UDI/DI each 10193489210569, UDI/DI case 40193489210560, Lot Number: 21FBR921; Medline Kit Number/SKU DYNJ64658: UDI/DI each 10193489210569, UDI/DI case 40193489210560, Lot Number: 21FBR939; Medline Kit Number/SKU DYNJ64658: UDI/DI each 10193489210569, UDI/DI case 40193489210560, Lot Number: 21EBA643; Medline Kit Number/SKU DYNJ64658: UDI/DI each 10193489210569, UDI/DI case 40193489210560, Lot Number: 21BBQ790; Medline Kit Number/SKU DYNJ64658: UDI/DI each 10193489210569, UDI/DI case 40193489210560, Lot Number: 21BBC028; Medline Kit Number/SKU DYNJ64658A: UDI/DI each 10195327303914, UDI/DI case 40195327303915, Lot Number: 23JBE733; Medline Kit Number/SKU DYNJ64658A: UDI/DI each 10195327303914, UDI/DI case 40195327303915, Lot Number: 23GBK608; Medline Kit Number/SKU DYNJ64658A: UDI/DI each 10195327303914, UDI/DI case 40195327303915, Lot Number: 23DBJ226; Medline Kit Number/SKU DYNJ64658A: UDI/DI each 10195327303914, UDI/DI case 40195327303915, Lot Number: 22LBJ519; Medline Kit Number/SKU DYNJ64658A: UDI/DI each 10195327303914, UDI/DI case 40195327303915, Lot Number: 25GBL417; Medline Kit Number/SKU DYNJ64658A: UDI/DI each 10195327303914, UDI/DI case 40195327303915, Lot Number: 25EBC969; Medline Kit Number/SKU DYNJ64658A: UDI/DI each 10195327303914, UDI/DI case 40195327303915, Lot Number: 24KBQ659; Medline Kit Number/SKU DYNJ64658A: UDI/DI each 10195327303914, UDI/DI case 40195327303915, Lot Number: 24HBH157; Medline Kit Number/SKU DYNJ64658A: UDI/DI each 10195327303914, UDI/DI case 40195327303915, Lot Number: 24FBM160; Medline Kit Number/SKU DYNJ64658A: UDI/DI each 10195327303914, UDI/DI case 40195327303915, Lot Number: 24BBA851; Medline Kit Number/SKU DYNJ64658A: UDI/DI each 10195327303914, UDI/DI case 40195327303915, Lot Number: 23LBL662; Medline Kit Number/SKU DYNJ64658B: UDI/DI each 10198459560675, UDI/DI case 40198459560676, Lot Number: 26CBA976; Medline Kit Number/SKU DYNJ64658B: UDI/DI each 10198459560675, UDI/DI case 40198459560676, Lot Number: 25LBF574; Medline Kit Number/SKU DYNJ65028: UDI/DI each 10193489240894, UDI/DI case 40193489240895, Lot Number: 21DBG884; Medline Kit Number/SKU DYNJ65028: UDI/DI each 10193489240894, UDI/DI case 40193489240895, Lot Number: 21BBQ227; Medline Kit Number/SKU DYNJ65028: UDI/DI each 10193489240894, UDI/DI case 40193489240895, Lot Number: 21ABC103; Medline Kit Number/SKU DYNJ65028A: UDI/DI each 10193489934342, UDI/DI case 40193489934343, Lot Number: 22DMA079; Medline Kit Number/SKU DYNJ65028A: UDI/DI each 10193489934342, UDI/DI case 40193489934343, Lot Number: 22CBA570; Medline Kit Number/SKU DYNJ65028A: UDI/DI each 10193489934342, UDI/DI case 40193489934343, Lot Number: 22BBS875; Medline Kit Number/SKU DYNJ65028A: UDI/DI each 10193489934342, UDI/DI case 40193489934343, Lot Number: 22BBH976; Medline Kit Number/SKU DYNJ65028A: UDI/DI each 10193489934342, UDI/DI case 40193489934343, Lot Number: 21LBB949; Medline Kit Number/SKU DYNJ65028A: UDI/DI each 10193489934342, UDI/DI case 40193489934343, Lot Number: 21JBG683; Medline Kit Number/SKU DYNJ65028A: UDI/DI each 10193489934342, UDI/DI case 40193489934343, Lot Number: 21GBO340; Medline Kit Number/SKU DYNJ65028A: UDI/DI each 10193489934342, UDI/DI case 40193489934343, Lot Number: 21GBL438; Medline Kit Number/SKU DYNJ67968: UDI/DI each 10193489460513, UDI/DI case 40193489460514, Lot Number: 21GLA583; Medline Kit Number/SKU DYNJ67968: UDI/DI each 10193489460513, UDI/DI case 40193489460514, Lot Number: 21ELA300; Medline Kit Number/SKU DYNJ67968: UDI/DI each 10193489460513, UDI/DI case 40193489460514, Lot Number: 21DLA305; Medline Kit Number/SKU DYNJ67968: UDI/DI each 10193489460513, UDI/DI case 40193489460514, Lot Number: 21DLA054; Medline Kit Number/SKU DYNJ67968: UDI/DI each 10193489460513, UDI/DI case 40193489460514, Lot Number: 21CLA170; Medline Kit Number/SKU DYNJ67968: UDI/DI each 10193489460513, UDI/DI case 40193489460514, Lot Number: 21CLA152; Medline Kit Number/SKU DYNJ67968A: UDI/DI each 10193489962949, UDI/DI case 40193489962940, Lot Number: 23DMA746; Medline Kit Number/SKU DYNJ67968A: UDI/DI each 10193489962949, UDI/DI case 40193489962940, Lot Number: 23BMC966; Medline Kit Number/SKU DYNJ67968A: UDI/DI each 10193489962949, UDI/DI case 40193489962940, Lot Number: 22IMF531; Medline Kit Number/SKU DYNJ67968A: UDI/DI each 10193489962949, UDI/DI case 40193489962940, Lot Number: 22GMG777; Medline Kit Number/SKU DYNJ67968A: UDI/DI each 10193489962949, UDI/DI case 40193489962940, Lot Number: 22FMA171; Medline Kit Number/SKU DYNJ67968A: UDI/DI each 10193489962949, UDI/DI case 40193489962940, Lot Number: 22CLA200; Medline Kit Number/SKU DYNJ67968A: UDI/DI each 10193489962949, UDI/DI case 40193489962940, Lot Number: 22ALA499; Medline Kit Number/SKU DYNJ67968A: UDI/DI each 10193489962949, UDI/DI case 40193489962940, Lot Number: 21JLB002; Medline Kit Number/SKU DYNJ67968A: UDI/DI each 10193489962949, UDI/DI case 40193489962940, Lot Number: 21HLA839; Medline Kit Number/SKU DYNJ67968B: UDI/DI each 10195327418649, UDI/DI case 40195327418640, Lot Number: 23IMG461; Medline Kit Number/SKU DYNJ67968B: UDI/DI each 10195327418649, UDI/DI case 40195327418640, Lot Number: 23EMF925; Medline Kit Number/SKU DYNJ67968B: UDI/DI each 10195327418649, UDI/DI case 40195327418640, Lot Number: 24JMA680; Medline Kit Number/SKU DYNJ67968B: UDI/DI each 10195327418649, UDI/DI case 40195327418640, Lot Number: 24HMI731; Medline Kit Number/SKU DYNJ67968B: UDI/DI each 10195327418649, UDI/DI case 40195327418640, Lot Number: 24HMG728; Medline Kit Number/SKU DYNJ67968B: UDI/DI each 10195327418649, UDI/DI case 40195327418640, Lot Number: 24FMA661; Medline Kit Number/SKU DYNJ67968B: UDI/DI each 10195327418649, UDI/DI case 40195327418640, Lot Number: 24BMC283; Medline Kit Number/SKU DYNJ67968B: UDI/DI each 10195327418649, UDI/DI case 40195327418640, Lot Number: 24AMG558; Medline Kit Number/SKU DYNJ68844: UDI/DI each 10193489843408, UDI/DI case 40193489843409, Lot Number: 22GMA633; Medline Kit Number/SKU DYNJ68844: UDI/DI each 10193489843408, UDI/DI case 40193489843409, Lot Number: 22EMG944; Medline Kit Number/SKU DYNJ68844: UDI/DI each 10193489843408, UDI/DI case 40193489843409, Lot Number: 22EMA059; Medline Kit Number/SKU DYNJ68844: UDI/DI each 10193489843408, UDI/DI case 40193489843409, Lot Number: 22DMA733; Medline Kit Number/SKU DYNJ68844: UDI/DI each 10193489843408, UDI/DI case 40193489843409, Lot Number: 21KMF575; Medline Kit Number/SKU DYNJ68844: UDI/DI each 10193489843408, UDI/DI case 40193489843409, Lot Number: 21JMD129; Medline Kit Number/SKU DYNJ68844: UDI/DI each 10193489843408, UDI/DI case 40193489843409, Lot Number: 21HMB364; Medline Kit Number/SKU DYNJ68844: UDI/DI each 10193489843408, UDI/DI case 40193489843409, Lot Number: 21GMB819; Medline Kit Number/SKU DYNJ68844: UDI/DI each 10193489843408, UDI/DI case 40193489843409, Lot Number: 21EMD513; Medline Kit Number/SKU DYNJ68844: UDI/DI each 10193489843408, UDI/DI case 40193489843409, Lot Number: 21CMB774; Medline Kit Number/SKU DYNJ68844A: UDI/DI each 10195327216504, UDI/DI case 40195327216505, Lot Number: 23JMJ037; Medline Kit Number/SKU DYNJ68844A: UDI/DI each 10195327216504, UDI/DI case 40195327216505, Lot Number: 23IMC044; Medline Kit Number/SKU DYNJ68844A: UDI/DI each 10195327216504, UDI/DI case 40195327216505, Lot Number: 23DMI656; Medline Kit Number/SKU DYNJ68844A: UDI/DI each 10195327216504, UDI/DI case 40195327216505, Lot Number: 23BMH340; Medline Kit Number/SKU DYNJ68844A: UDI/DI each 10195327216504, UDI/DI case 40195327216505, Lot Number: 24IME987; Medline Kit Number/SKU DYNJ68844A: UDI/DI each 10195327216504, UDI/DI case 40195327216505, Lot Number: 24HMA605; Medline Kit Number/SKU DYNJ68844A: UDI/DI each 10195327216504, UDI/DI case 40195327216505, Lot Number: 24GMA632; Medline Kit Number/SKU DYNJ68844A: UDI/DI each 10195327216504, UDI/DI case 40195327216505, Lot Number: 24DMG133; Medline Kit Number/SKU DYNJ68844A: UDI/DI each 10195327216504, UDI/DI case 40195327216505, Lot Number: 24CMK124; Medline Kit Number/SKU DYNJ68844A: UDI/DI each 10195327216504, UDI/DI case 40195327216505, Lot Number: 24BMI337; Medline Kit Number/SKU DYNJ68844A: UDI/DI each 10195327216504, UDI/DI case 40195327216505, Lot Number: 23LMD063; Medline Kit Number/SKU DYNJ81727: UDI/DI each 10195327137113, UDI/DI case 40195327137114, Lot Number: 22GMI084; Medline Kit Number/SKU DYNJ81727: UDI/DI each 10195327137113, UDI/DI case 40195327137114, Lot Number: 22FMH411; Medline Kit Number/SKU DYNJ81727A: UDI/DI each 10195327277895, UDI/DI case 40195327277896, Lot Number: 23KME276; Medline Kit Number/SKU DYNJ81727A: UDI/DI each 10195327277895, UDI/DI case 40195327277896, Lot Number: 23IMB927; Medline Kit Number/SKU DYNJ81727A: UDI/DI each 10195327277895, UDI/DI case 40195327277896, Lot Number: 23BMG594; Medline Kit Number/SKU DYNJ81727A: UDI/DI each 10195327277895, UDI/DI case 40195327277896, Lot Number: 23BMG926; Medline Kit Number/SKU DYNJ81727A: UDI/DI each 10195327277895, UDI/DI case 40195327277896, Lot Number: 23BMG593; Medline Kit Number/SKU DYNJ81727F: UDI/DI each 10198459260940, UDI/DI case 40198459260941, Lot Number: 25CMA943; Medline Kit Number/SKU DYNJ81727F: UDI/DI each 10198459260940, UDI/DI case 40198459260941, Lot Number: 25BMH565; Medline Kit Number/SKU DYNJ81772: UDI/DI each 10195327141417, UDI/DI case 40195327141418, Lot Number: 23JDA488; Medline Kit Number/SKU DYNJ81772: UDI/DI each 10195327141417, UDI/DI case 40195327141418, Lot Number: 23FDC241; Medline Kit Number/SKU DYNJ81772: UDI/DI each 10195327141417, UDI/DI case 40195327141418, Lot Number: 23FDA551; Medline Kit Number/SKU DYNJ81772: UDI/DI each 10195327141417, UDI/DI case 40195327141418, Lot Number: 23DDB518; Medline Kit Number/SKU DYNJ81772: UDI/DI each 10195327141417, UDI/DI case 40195327141418, Lot Number: 22GDB463; Medline Kit Number/SKU DYNJ81772: UDI/DI each 10195327141417, UDI/DI case 40195327141418, Lot Number: 23LDA469; Medline Kit Number/SKU DYNJ81772C: UDI/DI each 10198459349362, UDI/DI case 40198459349363, Lot Number: 26BME027; Medline Kit Number/SKU DYNJ81772C: UDI/DI each 10198459349362, UDI/DI case 40198459349363, Lot Number: 25LMI396; Medline Kit Number/SKU DYNJ81772C: UDI/DI each 10198459349362, UDI/DI case 40198459349363, Lot Number: 25LMH594; Medline Kit Number/SKU DYNJ81772C: UDI/DI each 10198459349362, UDI/DI case 40198459349363, Lot Number: 25IMB428; Medline Kit Number/SKU DYNJ81772C: UDI/DI each 10198459349362, UDI/DI case 40198459349363, Lot Number: 25EMG601; Medline Kit Number/SKU DYNJ82124: UDI/DI each 10195327174804, UDI/DI case 40195327174805, Lot Number: 23KMH124; Medline Kit Number/SKU DYNJ82124: UDI/DI each 10195327174804, UDI/DI case 40195327174805, Lot Number: 23FMJ566; Medline Kit Number/SKU DYNJ82124: UDI/DI each 10195327174804, UDI/DI case 40195327174805, Lot Number: 23FMF306; Medline Kit Number/SKU DYNJ82124: UDI/DI each 10195327174804, UDI/DI case 40195327174805, Lot Number: 23EME060; Medline Kit Number/SKU DYNJ82124: UDI/DI each 10195327174804, UDI/DI case 40195327174805, Lot Number: 23CMA243; Medline Kit Number/SKU DYNJ82124: UDI/DI each 10195327174804, UDI/DI case 40195327174805, Lot Number: 23AMC472; Medline Kit Number/SKU DYNJ82124: UDI/DI each 10195327174804, UDI/DI case 40195327174805, Lot Number: 22IMH894; Medline Kit Number/SKU DYNJ82124: UDI/DI each 10195327174804, UDI/DI case 40195327174805, Lot Number: 22IMG830; Medline Kit Number/SKU DYNJ82124: UDI/DI each 10195327174804, UDI/DI case 40195327174805, Lot Number: 26BMA823; Medline Kit Number/SKU DYNJ82124: UDI/DI each 10195327174804, UDI/DI case 40195327174805, Lot Number: 25JMF655; Medline Kit Number/SKU DYNJ82124: UDI/DI each 10195327174804, UDI/DI case 40195327174805, Lot Number: 25HMC540; Medline Kit Number/SKU DYNJ82124: UDI/DI each 10195327174804, UDI/DI case 40195327174805, Lot Number: 25EMG428; Medline Kit Number/SKU DYNJ82124: UDI/DI each 10195327174804, UDI/DI case 40195327174805, Lot Number: 25BMG415; Medline Kit Number/SKU DYNJ82124: UDI/DI each 10195327174804, UDI/DI case 40195327174805, Lot Number: 24LME660; Medline Kit Number/SKU DYNJ82124: UDI/DI each 10195327174804, UDI/DI case 40195327174805, Lot Number: 24KMB725; Medline Kit Number/SKU DYNJ82124: UDI/DI each 10195327174804, UDI/DI case 40195327174805, Lot Number: 24HMJ214; Medline Kit Number/SKU DYNJ82124: UDI/DI each 10195327174804, UDI/DI case 40195327174805, Lot Number: 24EMA834; Medline Kit Number/SKU DYNJ82124: UDI/DI each 10195327174804, UDI/DI case 40195327174805, Lot Number: 24BMI629; Medline Kit Number/SKU DYNJ82124: UDI/DI each 10195327174804, UDI/DI case 40195327174805, Lot Number: 23LMD692; Medline Kit Number/SKU DYNJ83417: UDI/DI each 10195327265038, UDI/DI case 40195327265039, Lot Number: 22LBO285; Medline Kit Number/SKU DYNJ83417A: UDI/DI each 10195327382339, UDI/DI case 40195327382330, Lot Number: 23FBJ724; Medline Kit Number/SKU DYNJ83417A: UDI/DI each 10195327382339, UDI/DI case 40195327382330, Lot Number: 23EBI932; Medline Kit Number/SKU DYNJ83417A: UDI/DI each 10195327382339, UDI/DI case 40195327382330, Lot Number: 23EBI500; Medline Kit Number/SKU DYNJ83417A: UDI/DI each 10195327382339, UDI/DI case 40195327382330, Lot Number: 25DBJ446; Medline Kit Number/SKU DYNJ83417A: UDI/DI each 10195327382339, UDI/DI case 40195327382330, Lot Number: 25CBM860; Medline Kit Number/SKU DYNJ83417A: UDI/DI each 10195327382339, UDI/DI case 40195327382330, Lot Number: 25BBC119; Medline Kit Number/SKU DYNJ83417A: UDI/DI each 10195327382339, UDI/DI case 40195327382330, Lot Number: 24JBF417; Medline Kit Number/SKU DYNJ83417A: UDI/DI each 10195327382339, UDI/DI case 40195327382330, Lot Number: 24FBQ165; Medline Kit Number/SKU DYNJ83417A: UDI/DI each 10195327382339, UDI/DI case 40195327382330, Lot Number: 24EBS240; Medline Kit Number/SKU DYNJ83417A: UDI/DI each 10195327382339, UDI/DI case 40195327382330, Lot Number: 24DBK432; Medline Kit Number/SKU DYNJ83417A: UDI/DI each 10195327382339, UDI/DI case 40195327382330, Lot Number: 24BBR722; Medline Kit Number/SKU DYNJ86166: UDI/DI each 10195327599331, UDI/DI case 40195327599332, Lot Number: 26BBI958; Medline Kit Number/SKU DYNJ86166: UDI/DI each 10195327599331, UDI/DI case 40195327599332, Lot Number: 25JBO782; Medline Kit Number/SKU DYNJ86166: UDI/DI each 10195327599331, UDI/DI case 40195327599332, Lot Number: 25IBA640; Medline Kit Number/SKU DYNJ86166: UDI/DI each 10195327599331, UDI/DI case 40195327599332, Lot Number: 25GBD814; Medline Kit Number/SKU DYNJ86166: UDI/DI each 10195327599331, UDI/DI case 40195327599332, Lot Number: 25DBG181; Medline Kit Number/SKU DYNJ86166: UDI/DI each 10195327599331, UDI/DI case 40195327599332, Lot Number: 25CBG994; Medline Kit Number/SKU DYNJ86166: UDI/DI each 10195327599331, UDI/DI case 40195327599332, Lot Number: 24LBS022; Medline Kit Number/SKU DYNJ86166: UDI/DI each 10195327599331, UDI/DI case 40195327599332, Lot Number: 24CBQ298; Medline Kit Number/SKU DYNJ87503: UDI/DI each 10198459041822, UDI/DI case 40198459041823, Lot Number: 24LBJ556; Medline Kit Number/SKU DYNJ87503A: UDI/DI each 10198459337680, UDI/DI case 40198459337681, Lot Number: 25FBT430; Medline Kit Number/SKU DYNJ87503B: UDI/DI each 10198459563836, UDI/DI case 40198459563837, Lot Number: 25KBA102; Medline Kit Number/SKU DYNJ88201: UDI/DI each 10198459121524, UDI/DI case 40198459121525, Lot Number: 24JBF518; Medline Kit Number/SKU DYNJ88201A: UDI/DI each 10198459356360, UDI/DI case 40198459356361, Lot Number: 25IBJ505; Medline Kit Number/SKU DYNJ88201A: UDI/DI each 10198459356360, UDI/DI case 40198459356361, Lot Number: 25HBK860; Medline Kit Number/SKU DYNJ88201A: UDI/DI each 10198459356360, UDI/DI case 40198459356361, Lot Number: 25FBV160; Medline Kit Number/SKU DYNJ89375A: UDI/DI each 10198459720147, UDI/DI case 40198459720148, Lot Number: 26CMC035; Medline Kit Number/SKU DYNJ900845B: UDI/DI each 10889942634422, UDI/DI case 40889942634423, Lot Number: 23IBH113; Medline Kit Number/SKU DYNJ900845B: UDI/DI each 10889942634422, UDI/DI case 40889942634423, Lot Number: 23GBH719; Medline Kit Number/SKU DYNJ900845B: UDI/DI each 10889942634422, UDI/DI case 40889942634423, Lot Number: 23GBH824; Medline Kit Number/SKU DYNJ900845B: UDI/DI each 10889942634422, UDI/DI case 40889942634423, Lot Number: 23FBL900; Medline Kit Number/SKU DYNJ900845B: UDI/DI each 10889942634422, UDI/DI case 40889942634423, Lot Number: 23EBW826; Medline Kit Number/SKU DYNJ900845B: UDI/DI each 10889942634422, UDI/DI case 40889942634423, Lot Number: 23DBT401; Medline Kit Number/SKU DYNJ900845B: UDI/DI each 10889942634422, UDI/DI case 40889942634423, Lot Number: 23CBQ543; Medline Kit Number/SKU DYNJ900845B: UDI/DI each 10889942634422, UDI/DI case 40889942634423, Lot Number: 23CBE800; Medline Kit Number/SKU DYNJ900845B: UDI/DI each 10889942634422, UDI/DI case 40889942634423, Lot Number: 23CBE801; Medline Kit Number/SKU DYNJ900845B: UDI/DI each 10889942634422, UDI/DI case 40889942634423, Lot Number: 22LBS875; Medline Kit Number/SKU DYNJ900845B: UDI/DI each 10889942634422, UDI/DI case 40889942634423, Lot Number: 22LBS876; Medline Kit Number/SKU DYNJ900845B: UDI/DI each 10889942634422, UDI/DI case 40889942634423, Lot Number: 22HBV144; Medline Kit Number/SKU DYNJ900845B: UDI/DI each 10889942634422, UDI/DI case 40889942634423, Lot Number: 22HBV582; Medline Kit Number/SKU DYNJ900845B: UDI/DI each 10889942634422, UDI/DI case 40889942634423, Lot Number: 22GBZ303; Medline Kit Number/SKU DYNJ900845B: UDI/DI each 10889942634422, UDI/DI case 40889942634423, Lot Number: 22GBS213; Medline Kit Number/SKU DYNJ900845B: UDI/DI each 10889942634422, UDI/DI case 40889942634423, Lot Number: 22DBM741; Medline Kit Number/SKU DYNJ900845B: UDI/DI each 10889942634422, UDI/DI case 40889942634423, Lot Number: 22DBC328; Medline Kit Number/SKU DYNJ900845B: UDI/DI each 10889942634422, UDI/DI case 40889942634423, Lot Number: 21LBW483; Medline Kit Number/SKU DYNJ900845B: UDI/DI each 10889942634422, UDI/DI case 40889942634423, Lot Number: 21KBK354; Medline Kit Number/SKU DYNJ900845B: UDI/DI each 10889942634422, UDI/DI case 40889942634423, Lot Number: 21JBX680; Medline Kit Number/SKU DYNJ900845B: UDI/DI each 10889942634422, UDI/DI case 40889942634423, Lot Number: 21IBR084; Medline Kit Number/SKU DYNJ900845B: UDI/DI each 10889942634422, UDI/DI case 40889942634423, Lot Number: 21HBI219; Medline Kit Number/SKU DYNJ900845B: UDI/DI each 10889942634422, UDI/DI case 40889942634423, Lot Number: 21HBB375; Medline Kit Number/SKU DYNJ901296A: UDI/DI each 10889942009992, UDI/DI case 40889942009993, Lot Number: 21DBE128; Medline Kit Number/SKU DYNJ901753J: UDI/DI each 10193489604429, UDI/DI case 40193489604420, Lot Number: 23FDB702; Medline Kit Number/SKU DYNJ901753J: UDI/DI each 10193489604429, UDI/DI case 40193489604420, Lot Number: 23FDA114; Medline Kit Number/SKU DYNJ901753J: UDI/DI each 10193489604429, UDI/DI case 40193489604420, Lot Number: 22GDA782; Medline Kit Number/SKU DYNJ901753J: UDI/DI each 10193489604429, UDI/DI case 40193489604420, Lot Number: 22FDB058; Medline Kit Number/SKU DYNJ901753J: UDI/DI each 10193489604429, UDI/DI case 40193489604420, Lot Number: 21HDB858; Medline Kit Number/SKU DYNJ901753J: UDI/DI each 10193489604429, UDI/DI case 40193489604420, Lot Number: 21HDA339; Medline Kit Number/SKU DYNJ901753J: UDI/DI each 10193489604429, UDI/DI case 40193489604420, Lot Number: 21GDB427; Medline Kit Number/SKU DYNJ901753J: UDI/DI each 10193489604429, UDI/DI case 40193489604420, Lot Number: 21GDB224; Medline Kit Number/SKU DYNJ901753J: UDI/DI each 10193489604429, UDI/DI case 40193489604420, Lot Number: 21FDB291; Medline Kit Number/SKU DYNJ901753J: UDI/DI each 10193489604429, UDI/DI case 40193489604420, Lot Number: 21DDB137; Medline Kit Number/SKU DYNJ901753J: UDI/DI each 10193489604429, UDI/DI case 40193489604420, Lot Number: 21CDB779; Medline Kit Number/SKU DYNJ901753J: UDI/DI each 10193489604429, UDI/DI case 40193489604420, Lot Number: 21BDA755; Medline Kit Number/SKU DYNJ901753K: UDI/DI each 10195327431211, UDI/DI case 40195327431212, Lot Number: 23IDA281; Medline Kit Number/SKU DYNJ901753L: UDI/DI each 10195327496180, UDI/DI case 40195327496181, Lot Number: 26ADB691; Medline Kit Number/SKU DYNJ901753L: UDI/DI each 10195327496180, UDI/DI case 40195327496181, Lot Number: 25GDB518; Medline Kit Number/SKU DYNJ901753L: UDI/DI each 10195327496180, UDI/DI case 40195327496181, Lot Number: 25GDA588; Medline Kit Number/SKU DYNJ901753L: UDI/DI each 10195327496180, UDI/DI case 40195327496181, Lot Number: 25EDB366; Medline Kit Number/SKU DYNJ901753L: UDI/DI each 10195327496180, UDI/DI case 40195327496181, Lot Number: 25CDB331; Medline Kit Number/SKU DYNJ901753L: UDI/DI each 10195327496180, UDI/DI case 40195327496181, Lot Number: 24IDB178; Medline Kit Number/SKU DYNJ901753L: UDI/DI each 10195327496180, UDI/DI case 40195327496181, Lot Number: 24GDA863; Medline Kit Number/SKU DYNJ901753L: UDI/DI each 10195327496180, UDI/DI case 40195327496181, Lot Number: 24BDA185; Medline Kit Number/SKU DYNJ901753L: UDI/DI each 10195327496180, UDI/DI case 40195327496181, Lot Number: 23LDB476; Medline Kit Number/SKU DYNJ903515B: UDI/DI each 10193489618938, UDI/DI case 40193489618939, Lot Number: 23JBL937; Medline Kit Number/SKU DYNJ903515B: UDI/DI each 10193489618938, UDI/DI case 40193489618939, Lot Number: 23FBV736; Medline Kit Number/SKU DYNJ903515B: UDI/DI each 10193489618938, UDI/DI case 40193489618939, Lot Number: 23CBR362; Medline Kit Number/SKU DYNJ903515B: UDI/DI each 10193489618938, UDI/DI case 40193489618939, Lot Number: 23ABC423; Medline Kit Number/SKU DYNJ903515B: UDI/DI each 10193489618938, UDI/DI case 40193489618939, Lot Number: 22LBF244; Medline Kit Number/SKU DYNJ903515B: UDI/DI each 10193489618938, UDI/DI case 40193489618939, Lot Number: 22HBS342; Medline Kit Number/SKU DYNJ903515B: UDI/DI each 10193489618938, UDI/DI case 40193489618939, Lot Number: 22FBX915; Medline Kit Number/SKU DYNJ903515B: UDI/DI each 10193489618938, UDI/DI case 40193489618939, Lot Number: 22EBR566; Medline Kit Number/SKU DYNJ903515B: UDI/DI each 10193489618938, UDI/DI case 40193489618939, Lot Number: 22CBM758; Medline Kit Number/SKU DYNJ903515B: UDI/DI each 10193489618938, UDI/DI case 40193489618939, Lot Number: 22ABF168; Medline Kit Number/SKU DYNJ903515B: UDI/DI each 10193489618938, UDI/DI case 40193489618939, Lot Number: 21LBM300; Medline Kit Number/SKU DYNJ903515B: UDI/DI each 10193489618938, UDI/DI case 40193489618939, Lot Number: 21KBZ611; Medline Kit Number/SKU DYNJ903515B: UDI/DI each 10193489618938, UDI/DI case 40193489618939, Lot Number: 21IBJ211; Medline Kit Number/SKU DYNJ903515B: UDI/DI each 10193489618938, UDI/DI case 40193489618939, Lot Number: 21IBJ212; Medline Kit Number/SKU DYNJ903515B: UDI/DI each 10193489618938, UDI/DI case 40193489618939, Lot Number: 21EBH572; Medline Kit Number/SKU DYNJ903515B: UDI/DI each 10193489618938, UDI/DI case 40193489618939, Lot Number: 21EBC781; Medline Kit Number/SKU DYNJ903515B: UDI/DI each 10193489618938, UDI/DI case 40193489618939, Lot Number: 24KBA482; Medline Kit Number/SKU DYNJ903515B: UDI/DI each 10193489618938, UDI/DI case 40193489618939, Lot Number: 24IBN252; Medline Kit Number/SKU DYNJ903515B: UDI/DI each 10193489618938, UDI/DI case 40193489618939, Lot Number: 24GBN507; Medline Kit Number/SKU DYNJ903515B: UDI/DI each 10193489618938, UDI/DI case 40193489618939, Lot Number: 24FBG214; Medline Kit Number/SKU DYNJ903515B: UDI/DI each 10193489618938, UDI/DI case 40193489618939, Lot Number: 24FBE668; Medline Kit Number/SKU DYNJ903515B: UDI/DI each 10193489618938, UDI/DI case 40193489618939, Lot Number: 24BBF142; Medline Kit Number/SKU DYNJ903515B: UDI/DI each 10193489618938, UDI/DI case 40193489618939, Lot Number: 24ABC727; Medline Kit Number/SKU DYNJ903515B: UDI/DI each 10193489618938, UDI/DI case 40193489618939, Lot Number: 23KBX358; Medline Kit Number/SKU DYNJ904275F: UDI/DI each 10198459096112, UDI/DI case 40198459096113, Lot Number: 25LMH954; Medline Kit Number/SKU DYNJ904275F: UDI/DI each 10198459096112, UDI/DI case 40198459096113, Lot Number: 25LMC207; Medline Kit Number/SKU DYNJ904275F: UDI/DI each 10198459096112, UDI/DI case 40198459096113, Lot Number: 25HMG134; Medline Kit Number/SKU DYNJ904275F: UDI/DI each 10198459096112, UDI/DI case 40198459096113, Lot Number: 25EMH523; Medline Kit Number/SKU DYNJ904275F: UDI/DI each 10198459096112, UDI/DI case 40198459096113, Lot Number: 25DMA957; Medline Kit Number/SKU DYNJ904275F: UDI/DI each 10198459096112, UDI/DI case 40198459096113, Lot Number: 25CMH640; Medline Kit Number/SKU DYNJ904275F: UDI/DI each 10198459096112, UDI/DI case 40198459096113, Lot Number: 25BMD941; Medline Kit Number/SKU DYNJ904275F: UDI/DI each 10198459096112, UDI/DI case 40198459096113, Lot Number: 25AMG740; Medline Kit Number/SKU DYNJ904275F: UDI/DI each 10198459096112, UDI/DI case 40198459096113, Lot Number: 25AMA942; Medline Kit Number/SKU DYNJ904275F: UDI/DI each 10198459096112, UDI/DI case 40198459096113, Lot Number: 24LMC564; Medline Kit Number/SKU DYNJ904275F: UDI/DI each 10198459096112, UDI/DI case 40198459096113, Lot Number: 24IMG084; Medline Kit Number/SKU DYNJ904565D: UDI/DI each 10193489379150, UDI/DI case 40193489379151, Lot Number: 21FBD283; Medline Kit Number/SKU DYNJ904565D: UDI/DI each 10193489379150, UDI/DI case 40193489379151, Lot Number: 21DBG641; Medline Kit Number/SKU DYNJ904565D: UDI/DI each 10193489379150, UDI/DI case 40193489379151, Lot Number: 21CBK121; Medline Kit Number/SKU DYNJ904565F: UDI/DI each 10193489916805, UDI/DI case 40193489916806, Lot Number: 23DBF757; Medline Kit Number/SKU DYNJ904565F: UDI/DI each 10193489916805, UDI/DI case 40193489916806, Lot Number: 23CBG746; Medline Kit Number/SKU DYNJ904565F: UDI/DI each 10193489916805, UDI/DI case 40193489916806, Lot Number: 23ABS372; Medline Kit Number/SKU DYNJ904565F: UDI/DI each 10193489916805, UDI/DI case 40193489916806, Lot Number: 23ABI406; Medline Kit Number/SKU DYNJ904565F: UDI/DI each 10193489916805, UDI/DI case 40193489916806, Lot Number: 22IBO075; Medline Kit Number/SKU DYNJ904565F: UDI/DI each 10193489916805, UDI/DI case 40193489916806, Lot Number: 22IBN313; Medline Kit Number/SKU DYNJ904565F: UDI/DI each 10193489916805, UDI/DI case 40193489916806, Lot Number: 22IBN315; Medline Kit Number/SKU DYNJ904565F: UDI/DI each 10193489916805, UDI/DI case 40193489916806, Lot Number: 22GBE006; Medline Kit Number/SKU DYNJ904565F: UDI/DI each 10193489916805, UDI/DI case 40193489916806, Lot Number: 22GBE793; Medline Kit Number/SKU DYNJ904565F: UDI/DI each 10193489916805, UDI/DI case 40193489916806, Lot Number: 22EBQ475; Medline Kit Number/SKU DYNJ904565F: UDI/DI each 10193489916805, UDI/DI case 40193489916806, Lot Number: 22DBG440; Medline Kit Number/SKU DYNJ904565F: UDI/DI each 10193489916805, UDI/DI case 40193489916806, Lot Number: 22CBP145; Medline Kit Number/SKU DYNJ904565F: UDI/DI each 10193489916805, UDI/DI case 40193489916806, Lot Number: 21LBE727; Medline Kit Number/SKU DYNJ904565F: UDI/DI each 10193489916805, UDI/DI case 40193489916806, Lot Number: 21HBM868; Medline Kit Number/SKU DYNJ904565G: UDI/DI each 10195327371784, UDI/DI case 40195327371785, Lot Number: 23DBP675; Medline Kit Number/SKU DYNJ904565I: UDI/DI each 10195327468422, UDI/DI case 40195327468423, Lot Number: 23JBY083; Medline Kit Number/SKU DYNJ904565I: UDI/DI each 10195327468422, UDI/DI case 40195327468423, Lot Number: 23JBC192; Medline Kit Number/SKU DYNJ904565I: UDI/DI each 10195327468422, UDI/DI case 40195327468423, Lot Number: 23IBH848; Medline Kit Number/SKU DYNJ904565I: UDI/DI each 10195327468422, UDI/DI case 40195327468423, Lot Number: 25FBM817; Medline Kit Number/SKU DYNJ904565I: UDI/DI each 10195327468422, UDI/DI case 40195327468423, Lot Number: 25CBH988; Medline Kit Number/SKU DYNJ904565I: UDI/DI each 10195327468422, UDI/DI case 40195327468423, Lot Number: 25ABW505; Medline Kit Number/SKU DYNJ904565I: UDI/DI each 10195327468422, UDI/DI case 40195327468423, Lot Number: 25ABC626; Medline Kit Number/SKU DYNJ904565I: UDI/DI each 10195327468422, UDI/DI case 40195327468423, Lot Number: 24KBG651; Medline Kit Number/SKU DYNJ904565I: UDI/DI each 10195327468422, UDI/DI case 40195327468423, Lot Number: 24IBP291; Medline Kit Number/SKU DYNJ904565I: UDI/DI each 10195327468422, UDI/DI case 40195327468423, Lot Number: 24HBP497; Medline Kit Number/SKU DYNJ904565I: UDI/DI each 10195327468422, UDI/DI case 40195327468423, Lot Number: 24GBR801; Medline Kit Number/SKU DYNJ904565I: UDI/DI each 10195327468422, UDI/DI case 40195327468423, Lot Number: 24FBS752; Medline Kit Number/SKU DYNJ904565I: UDI/DI each 10195327468422, UDI/DI case 40195327468423, Lot Number: 24FBC560; Medline Kit Number/SKU DYNJ904565I: UDI/DI each 10195327468422, UDI/DI case 40195327468423, Lot Number: 24CBA256; Medline Kit Number/SKU DYNJ904565I: UDI/DI each 10195327468422, UDI/DI case 40195327468423, Lot Number: 23LBN257; Medline Kit Number/SKU DYNJ904565J: UDI/DI each 10198459397356, UDI/DI case 40198459397357, Lot Number: 25IBW100; Medline Kit Number/SKU DYNJ904565J: UDI/DI each 10198459397356, UDI/DI case 40198459397357, Lot Number: 25IBH103; Medline Kit Number/SKU DYNJ904565J: UDI/DI each 10198459397356, UDI/DI case 40198459397357, Lot Number: 25HBN347; Medline Kit Number/SKU DYNJ904565J: UDI/DI each 10198459397356, UDI/DI case 40198459397357, Lot Number: 25GBS779; Medline Kit Number/SKU DYNJ905365B: UDI/DI each 10195327048099, UDI/DI case 40195327048090, Lot Number: 23KBO199; Medline Kit Number/SKU DYNJ905365B: UDI/DI each 10195327048099, UDI/DI case 40195327048090, Lot Number: 23KBN420; Medline Kit Number/SKU DYNJ905365B: UDI/DI each 10195327048099, UDI/DI case 40195327048090, Lot Number: 23KBF177; Medline Kit Number/SKU DYNJ905365B: UDI/DI each 10195327048099, UDI/DI case 40195327048090, Lot Number: 23FBU813; Medline Kit Number/SKU DYNJ905365B: UDI/DI each 10195327048099, UDI/DI case 40195327048090, Lot Number: 23FBP021; Medline Kit Number/SKU DYNJ905365B: UDI/DI each 10195327048099, UDI/DI case 40195327048090, Lot Number: 23BBJ851; Medline Kit Number/SKU DYNJ905365B: UDI/DI each 10195327048099, UDI/DI case 40195327048090, Lot Number: 23BBJ884; Medline Kit Number/SKU DYNJ905365B: UDI/DI each 10195327048099, UDI/DI case 40195327048090, Lot Number: 22LBB596; Medline Kit Number/SKU DYNJ905365B: UDI/DI each 10195327048099, UDI/DI case 40195327048090, Lot Number: 22HBR849; Medline Kit Number/SKU DYNJ905365B: UDI/DI each 10195327048099, UDI/DI case 40195327048090, Lot Number: 22GBQ034; Medline Kit Number/SKU DYNJ905365B: UDI/DI each 10195327048099, UDI/DI case 40195327048090, Lot Number: 22GBQ036; Medline Kit Number/SKU DYNJ905365B: UDI/DI each 10195327048099, UDI/DI case 40195327048090, Lot Number: 22EBA005; Medline Kit Number/SKU DYNJ905365B: UDI/DI each 10195327048099, UDI/DI case 40195327048090, Lot Number: 22DBU733; Medline Kit Number/SKU DYNJ905365B: UDI/DI each 10195327048099, UDI/DI case 40195327048090, Lot Number: 22EBA141; Medline Kit Number/SKU DYNJ905365B: UDI/DI each 10195327048099, UDI/DI case 40195327048090, Lot Number: 25EBN366; Medline Kit Number/SKU DYNJ905365B: UDI/DI each 10195327048099, UDI/DI case 40195327048090, Lot Number: 25DBT886; Medline Kit Number/SKU DYNJ905365B: UDI/DI each 10195327048099, UDI/DI case 40195327048090, Lot Number: 25CBV172; Medline Kit Number/SKU DYNJ905365B: UDI/DI each 10195327048099, UDI/DI case 40195327048090, Lot Number: 25ABF384; Medline Kit Number/SKU DYNJ905365B: UDI/DI each 10195327048099, UDI/DI case 40195327048090, Lot Number: 24LBM934; Medline Kit Number/SKU DYNJ905365B: UDI/DI each 10195327048099, UDI/DI case 40195327048090, Lot Number: 24JBL515; Medline Kit Number/SKU DYNJ905365B: UDI/DI each 10195327048099, UDI/DI case 40195327048090, Lot Number: 24HBM973; Medline Kit Number/SKU DYNJ905365B: UDI/DI each 10195327048099, UDI/DI case 40195327048090, Lot Number: 24HBG658; Medline Kit Number/SKU DYNJ905365B: UDI/DI each 10195327048099, UDI/DI case 40195327048090, Lot Number: 24CBH571; Medline Kit Number/SKU DYNJ905365B: UDI/DI each 10195327048099, UDI/DI case 40195327048090, Lot Number: 24CBH400; Medline Kit Number/SKU DYNJ905365B: UDI/DI each 10195327048099, UDI/DI case 40195327048090, Lot Number: 24CBH416; Medline Kit Number/SKU DYNJ905365C: UDI/DI each 10198459346279, UDI/DI case 40198459346270, Lot Number: 26BBQ345; Medline Kit Number/SKU DYNJ905365C: UDI/DI each 10198459346279, UDI/DI case 40198459346270, Lot Number: 25LBS099; Medline Kit Number/SKU DYNJ905365C: UDI/DI each 10198459346279, UDI/DI case 40198459346270, Lot Number: 25JBU372; Medline Kit Number/SKU DYNJ905365C: UDI/DI each 10198459346279, UDI/DI case 40198459346270, Lot Number: 25IBT895; Medline Kit Number/SKU DYNJ905365C: UDI/DI each 10198459346279, UDI/DI case 40198459346270, Lot Number: 25HBP096; Medline Kit Number/SKU DYNJ905366C: UDI/DI each 10198459346286, UDI/DI case 40198459346287, Lot Number: 26BBP163; Medline Kit Number/SKU DYNJ905366C: UDI/DI each 10198459346286, UDI/DI case 40198459346287, Lot Number: 26BBG647; Medline Kit Number/SKU DYNJ905366C: UDI/DI each 10198459346286, UDI/DI case 40198459346287, Lot Number: 25KBR622; Medline Kit Number/SKU DYNJ905366C: UDI/DI each 10198459346286, UDI/DI case 40198459346287, Lot Number: 25JBM109; Medline Kit Number/SKU DYNJ905366C: UDI/DI each 10198459346286, UDI/DI case 40198459346287, Lot Number: 25FBO089; Medline Kit Number/SKU DYNJ905366C: UDI/DI each 10198459346286, UDI/DI case 40198459346287, Lot Number: 25FBG065; Medline Kit Number/SKU DYNJ906037A: UDI/DI each 10193489766202, UDI/DI case 40193489766203, Lot Number: 21BBF015; Medline Kit Number/SKU DYNJ907076D: UDI/DI each 10198459045103, UDI/DI case 40198459045104, Lot Number: 24KBQ125; Medline Kit Number/SKU DYNJ907076D: UDI/DI each 10198459045103, UDI/DI case 40198459045104, Lot Number: 24JBT924; Medline Kit Number/SKU DYNJ907076F: UDI/DI each 10198459201417, UDI/DI case 40198459201418, Lot Number: 25HBT028; Medline Kit Number/SKU DYNJ907076F: UDI/DI each 10198459201417, UDI/DI case 40198459201418, Lot Number: 25GBG354; Medline Kit Number/SKU DYNJ907076F: UDI/DI each 10198459201417, UDI/DI case 40198459201418, Lot Number: 25FBJ135; Medline Kit Number/SKU DYNJ907076F: UDI/DI each 10198459201417, UDI/DI case 40198459201418, Lot Number: 25BBD094; Medline Kit Number/SKU DYNJ907446: UDI/DI each 10193489419726, UDI/DI case 40193489419727, Lot Number: 22HBS670; Medline Kit Number/SKU DYNJ907863: UDI/DI each 10193489913187, UDI/DI case 40193489913188, Lot Number: 22EBO903; Medline Kit Number/SKU DYNJ908264: UDI/DI each 10195327020118, UDI/DI case 40195327020119, Lot Number: 23IBH949; Medline Kit Number/SKU DYNJ908264: UDI/DI each 10195327020118, UDI/DI case 40195327020119, Lot Number: 23EBS103; Medline Kit Number/SKU DYNJ908264: UDI/DI each 10195327020118, UDI/DI case 40195327020119, Lot Number: 24ABT534; Medline Kit Number/SKU DYNJ908264: UDI/DI each 10195327020118, UDI/DI case 40195327020119, Lot Number: 24ABC072; Medline Kit Number/SKU DYNJ908264A: UDI/DI each 10195327593988, UDI/DI case 40195327593989, Lot Number: 26CDA182; Medline Kit Number/SKU DYNJ908264A: UDI/DI each 10195327593988, UDI/DI case 40195327593989, Lot Number: 26BDA043; Medline Kit Number/SKU DYNJ908264A: UDI/DI each 10195327593988, UDI/DI case 40195327593989, Lot Number: 26ADB372; Medline Kit Number/SKU DYNJ908264A: UDI/DI each 10195327593988, UDI/DI case 40195327593989, Lot Number: 26ADA380; Medline Kit Number/SKU DYNJ908264A: UDI/DI each 10195327593988, UDI/DI case 40195327593989, Lot Number: 25LDA636; Medline Kit Number/SKU DYNJ908264A: UDI/DI each 10195327593988, UDI/DI case 40195327593989, Lot Number: 25LDA379; Medline Kit Number/SKU DYNJ908264A: UDI/DI each 10195327593988, UDI/DI case 40195327593989, Lot Number: 25JDB762; Medline Kit Number/SKU DYNJ908264A: UDI/DI each 10195327593988, UDI/DI case 40195327593989, Lot Number: 25GDA732; Medline Kit Number/SKU DYNJ908264A: UDI/DI each 10195327593988, UDI/DI case 40195327593989, Lot Number: 25EDA402; Medline Kit Number/SKU DYNJ908264A: UDI/DI each 10195327593988, UDI/DI case 40195327593989, Lot Number: 25BDA442; Medline Kit Number/SKU DYNJ908264A: UDI/DI each 10195327593988, UDI/DI case 40195327593989, Lot Number: 25ADB889; Medline Kit Number/SKU DYNJ908264A: UDI/DI each 10195327593988, UDI/DI case 40195327593989, Lot Number: 24KDB984; Medline Kit Number/SKU DYNJ908264A: UDI/DI each 10195327593988, UDI/DI case 40195327593989, Lot Number: 24KDA219; Medline Kit Number/SKU DYNJ908264A: UDI/DI each 10195327593988, UDI/DI case 40195327593989, Lot Number: 24JDC398; Medline Kit Number/SKU DYNJ908264A: UDI/DI each 10195327593988, UDI/DI case 40195327593989, Lot Number: 24HDC065; Medline Kit Number/SKU DYNJ908264A: UDI/DI each 10195327593988, UDI/DI case 40195327593989, Lot Number: 24GDB441; Medline Kit Number/SKU DYNJ908264A: UDI/DI each 10195327593988, UDI/DI case 40195327593989, Lot Number: 24GDA311; Medline Kit Number/SKU DYNJ908264A: UDI/DI each 10195327593988, UDI/DI case 40195327593989, Lot Number: 24ADB838; Medline Kit Number/SKU DYNJ909439: UDI/DI each 10195327354824, UDI/DI case 40195327354825, Lot Number: 23JBD316; Medline Kit Number/SKU DYNJ909439: UDI/DI each 10195327354824, UDI/DI case 40195327354825, Lot Number: 23DBF809; Medline Kit Number/SKU DYNJ909439: UDI/DI each 10195327354824, UDI/DI case 40195327354825, Lot Number: 24KBU152; Medline Kit Number/SKU DYNJ909439: UDI/DI each 10195327354824, UDI/DI case 40195327354825, Lot Number: 24IBH163; Medline Kit Number/SKU DYNJ909439: UDI/DI each 10195327354824, UDI/DI case 40195327354825, Lot Number: 24HBV805; Medline Kit Number/SKU DYNJ909439: UDI/DI each 10195327354824, UDI/DI case 40195327354825, Lot Number: 24GBC814; Medline Kit Number/SKU DYNJ909439: UDI/DI each 10195327354824, UDI/DI case 40195327354825, Lot Number: 24EBR806; Medline Kit Number/SKU DYNJ909439A: UDI/DI each 10198459228001, UDI/DI case 40198459228002, Lot Number: 26BBP159; Medline Kit Number/SKU DYNJ909439A: UDI/DI each 10198459228001, UDI/DI case 40198459228002, Lot Number: 26ABH446; Medline Kit Number/SKU DYNJ909439A: UDI/DI each 10198459228001, UDI/DI case 40198459228002, Lot Number: 25KBN794; Medline Kit Number/SKU DYNJ909439A: UDI/DI each 10198459228001, UDI/DI case 40198459228002, Lot Number: 25IBR922; Medline Kit Number/SKU DYNJ909439A: UDI/DI each 10198459228001, UDI/DI case 40198459228002, Lot Number: 25EBA894; Medline Kit Number/SKU DYNJ909439A: UDI/DI each 10198459228001, UDI/DI case 40198459228002, Lot Number: 25ABL737; Medline Kit Number/SKU DYNJ909472A: UDI/DI each 10198459119101, UDI/DI case 40198459119102, Lot Number: 25BMI762; Medline Kit Number/SKU DYNJ909472A: UDI/DI each 10198459119101, UDI/DI case 40198459119102, Lot Number: 25BMD422; Medline Kit Number/SKU DYNJ909472A: UDI/DI each 10198459119101, UDI/DI case 40198459119102, Lot Number: 24KMG059; Medline Kit Number/SKU DYNJ909472A: UDI/DI each 10198459119101, UDI/DI case 40198459119102, Lot Number: 24KMB410; Medline Kit Number/SKU DYNJ909472C: UDI/DI each 10198459454226, UDI/DI case 40198459454227, Lot Number: 25IMH655; Medline Kit Number/SKU DYNJ909472C: UDI/DI each 10198459454226, UDI/DI case 40198459454227, Lot Number: 25IMC642; Medline Kit Number/SKU DYNJ909926: UDI/DI each 10195327506827, UDI/DI case 40195327506828, Lot Number: 23LBF552; Medline Kit Number/SKU DYNJ909926B: UDI/DI each 10198459188596, UDI/DI case 40198459188597, Lot Number: 25FBV690; Medline Kit Number/SKU DYNJ909926B: UDI/DI each 10198459188596, UDI/DI case 40198459188597, Lot Number: 25EBK630; Medline Kit Number/SKU DYNJ909926B: UDI/DI each 10198459188596, UDI/DI case 40198459188597, Lot Number: 25ABV651; Medline Kit Number/SKU DYNJ909926B: UDI/DI each 10198459188596, UDI/DI case 40198459188597, Lot Number: 25ABH588; Medline Kit Number/SKU DYNJ909926C: UDI/DI each 10198459423109, UDI/DI case 40198459423100, Lot Number: 26ABF779; Medline Kit Number/SKU DYNJ909926C: UDI/DI each 10198459423109, UDI/DI case 40198459423100, Lot Number: 25LBR659; Medline Kit Number/SKU DYNJ909926C: UDI/DI each 10198459423109, UDI/DI case 40198459423100, Lot Number: 25LBC067; Medline Kit Number/SKU DYNJ909926C: UDI/DI each 10198459423109, UDI/DI case 40198459423100, Lot Number: 25JBN795; Medline Kit Number/SKU DYNJ910902B: UDI/DI each 10198459387142, UDI/DI case 40198459387143, Lot Number: 25GMB865; Medline Kit Number/SKU DYNJ910987A: UDI/DI each 10198459418532, UDI/DI case 40198459418533, Lot Number: 26AMB048; Medline Kit Number/SKU DYNJ910987A: UDI/DI each 10198459418532, UDI/DI case 40198459418533, Lot Number: 25HMG101; Medline Kit Number/SKU DYNJ911469: UDI/DI each 10198459358753, UDI/DI case 40198459358754, Lot Number: 25JBD450; Medline Kit Number/SKU DYNJ911469: UDI/DI each 10198459358753, UDI/DI case 40198459358754, Lot Number: 25FBS379; Medline Kit Number/SKU DYNJT4005: UDI/DI each 10198459381584, UDI/DI case 40198459381585, Lot Number: 26ABH219; Medline Kit Number/SKU DYNJT4005: UDI/DI each 10198459381584, UDI/DI case 40198459381585, Lot Number: 25GBC350; Medline Kit Number/SKU DYNJT6012: UDI/DI each 10198459531316, UDI/DI case 40198459531317, Lot Number: 25LBN117; Medline Kit Number/SKU DYNJVB91096A: UDI/DI each 10193489995268, UDI/DI case 40193489995269, Lot Number: 22JDA197; Medline Kit Number/SKU DYNJVB91096A: UDI/DI each 10193489995268, UDI/DI case 40193489995269, Lot Number: 22HDC165; Medline Kit Number/SKU DYNJVB91096A: UDI/DI each 10193489995268, UDI/DI case 40193489995269, Lot Number: 22HDB422; Medline Kit Number/SKU DYNJVB91096A: UDI/DI each 10193489995268, UDI/DI case 40193489995269, Lot Number: 22EDA722; Medline Kit Number/SKU DYNJVB91096A: UDI/DI each 10193489995268, UDI/DI case 40193489995269, Lot Number: 21JDA283; Medline Kit Number/SKU DYNJVB91096B: UDI/DI each 10195327257156, UDI/DI case 40195327257157, Lot Number: 23KDA098; Medline Kit Number/SKU DYNJVB91096B: UDI/DI each 10195327257156, UDI/DI case 40195327257157, Lot Number: 23IDB014; Medline Kit Number/SKU DYNJVB91096B: UDI/DI each 10195327257156, UDI/DI case 40195327257157, Lot Number: 23GDA586; Medline Kit Number/SKU DYNJVB91096B: UDI/DI each 10195327257156, UDI/DI case 40195327257157, Lot Number: 23FDB544; Medline Kit Number/SKU DYNJVB91096B: UDI/DI each 10195327257156, UDI/DI case 40195327257157, Lot Number: 23DDB294; Medline Kit Number/SKU DYNJVB91096B: UDI/DI each 10195327257156, UDI/DI case 40195327257157, Lot Number: 23DDB035; Medline Kit Number/SKU DYNJVB91096B: UDI/DI each 10195327257156, UDI/DI case 40195327257157, Lot Number: 23CDA388; Medline Kit Number/SKU DYNJVB91096B: UDI/DI each 10195327257156, UDI/DI case 40195327257157, Lot Number: 22JDB847; Medline Kit Number/SKU DYNJVB91096B: UDI/DI each 10195327257156, UDI/DI case 40195327257157, Lot Number: 26ADB209; Medline Kit Number/SKU DYNJVB91096B: UDI/DI each 10195327257156, UDI/DI case 40195327257157, Lot Number: 25KDB765; Medline Kit Number/SKU DYNJVB91096B: UDI/DI each 10195327257156, UDI/DI case 40195327257157, Lot Number: 25JDA968; Medline Kit Number/SKU DYNJVB91096B: UDI/DI each 10195327257156, UDI/DI case 40195327257157, Lot Number: 25IDA858; Medline Kit Number/SKU DYNJVB91096B: UDI/DI each 10195327257156, UDI/DI case 40195327257157, Lot Number: 25HDA887; Medline Kit Number/SKU DYNJVB91096B: UDI/DI each 10195327257156, UDI/DI case 40195327257157, Lot Number: 25HDA137; Medline Kit Number/SKU DYNJVB91096B: UDI/DI each 10195327257156, UDI/DI case 40195327257157, Lot Number: 25GDB525; Medline Kit Number/SKU DYNJVB91096B: UDI/DI each 10195327257156, UDI/DI case 40195327257157, Lot Number: 25GDB387; Medline Kit Number/SKU DYNJVB91096B: UDI/DI each 10195327257156, UDI/DI case 40195327257157, Lot Number: 25GDB122; Medline Kit Number/SKU DYNJVB91096B: UDI/DI each 10195327257156, UDI/DI case 40195327257157, Lot Number: 25GDA061; Medline Kit Number/SKU DYNJVB91096B: UDI/DI each 10195327257156, UDI/DI case 40195327257157, Lot Number: 25DDB338; Medline Kit Number/SKU DYNJVB91096B: UDI/DI each 10195327257156, UDI/DI case 40195327257157, Lot Number: 25CDB735; Medline Kit Number/SKU DYNJVB91096B: UDI/DI each 10195327257156, UDI/DI case 40195327257157, Lot Number: 25CDA205; Medline Kit Number/SKU DYNJVB91096B: UDI/DI each 10195327257156, UDI/DI case 40195327257157, Lot Number: 24JDB378; Medline Kit Number/SKU DYNJVB91096B: UDI/DI each 10195327257156, UDI/DI case 40195327257157, Lot Number: 24IDA577; Medline Kit Number/SKU DYNJVB91096B: UDI/DI each 10195327257156, UDI/DI case 40195327257157, Lot Number: 24GDA172; Medline Kit Number/SKU DYNJVB91096B: UDI/DI each 10195327257156, UDI/DI case 40195327257157, Lot Number: 24FDA640; Medline Kit Number/SKU DYNJVB91096B: UDI/DI each 10195327257156, UDI/DI case 40195327257157, Lot Number: 24BDB046; Medline Kit Number/SKU DYNJVB91096B: UDI/DI each 10195327257156, UDI/DI case 40195327257157, Lot Number: 24ADA265.
Devices named in this recall
86 devices in the FDA device registry whose identifier appears in the recall's code information.
- Medline Industries, Inc. DYNJ68844AMedline Industries, Inc. · Plastic surgery and accessories kit4 recalls0 reports
- Medline Industries, Inc. DYNJ0660040RMedline Industries, Inc. · Plastic surgery and accessories kit4 recalls0 reports
- Medline Industries, Inc. DYNJ62803Medline Industries, Inc. · Plastic surgery and accessories kit3 recalls0 reports
- Medline Industries, Inc. DYNJ44759FMedline Industries, Inc. · Plastic surgery and accessories kit2 recalls0 reports
- Medline Industries, Inc. DYNJ42833BMedline Industries, Inc. · Plastic surgery and accessories kit2 recalls0 reports
- Medline Industries, Inc. DYNJ907863Medline Industries, Inc. · Plastic surgery and accessories kit2 recalls0 reports
- Medline Industries, Inc. DYNJ67968AMedline Industries, Inc. · Plastic surgery and accessories kit2 recalls0 reports
- Medline Industries, Inc. DYNJ908264Medline Industries, Inc. · Plastic surgery and accessories kit2 recalls0 reports
- Medline Industries, Inc. DYNJ63332DMedline Industries, Inc. · Plastic surgery and accessories kit2 recalls0 reports
- Medline Industries, Inc. DYNJVB91096BMedline Industries, Inc. · Plastic surgery and accessories kit2 recalls0 reports
- Medline Industries, Inc. DYNJ67968BMedline Industries, Inc. · Plastic surgery and accessories kit2 recalls0 reports
- Medline Industries, Inc. DYNJ901753LMedline Industries, Inc. · Plastic surgery and accessories kit2 recalls0 reports
- Medline Industries, Inc. CDS983493MMedline Industries, Inc. · Plastic surgery and accessories kit2 recalls0 reports
- MEDLINE INDUSTRIES, INC. DYNJ81772CMedline Industries, Inc. · Plastic surgery and accessories kit2 recalls0 reports
- Medline Industries, Inc. DYNJ64658Medline Industries, Inc. · Plastic surgery and accessories kit1 recall0 reports
- Medline Industries, Inc. DYNJ53076DMedline Industries, Inc. · Plastic surgery and accessories kit1 recall0 reports
- Medline Industries, Inc. DYNJ65028Medline Industries, Inc. · Plastic surgery and accessories kit1 recall0 reports
- Medline Industries, Inc. CDS983493LMedline Industries, Inc. · Plastic surgery and accessories kit1 recall0 reports
- Medline Industries, Inc. DYNJ60392CMedline Industries, Inc. · Plastic surgery and accessories kit1 recall0 reports
- Medline Industries, Inc. DYNJ59456AMedline Industries, Inc. · Plastic surgery and accessories kit1 recall0 reports
- Medline Industries, Inc. DYNJ904565DMedline Industries, Inc. · Plastic surgery and accessories kit1 recall0 reports
- Medline Industries, Inc. DYNJ907446Medline Industries, Inc. · Plastic surgery and accessories kit1 recall0 reports
- Medline Industries, Inc. DYNJ0395085PMedline Industries, Inc. · Plastic surgery and accessories kit1 recall0 reports
- Medline Industries, Inc. DYNJ67968Medline Industries, Inc. · Plastic surgery and accessories kit1 recall0 reports
- Medline Industries, Inc. DYNJ48191GMedline Industries, Inc. · Plastic surgery and accessories kit1 recall0 reports
- Medline Industries, Inc. DYNJ901753JMedline Industries, Inc. · Plastic surgery and accessories kit1 recall0 reports
- Medline Industries, Inc. DYNJ903515BMedline Industries, Inc. · Plastic surgery and accessories kit1 recall0 reports
- Medline Industries, Inc. DYNJ62636Medline Industries, Inc. · Plastic surgery and accessories kit1 recall0 reports
- Medline Industries, Inc. DYNJ41922CMedline Industries, Inc. · Plastic surgery and accessories kit1 recall0 reports
- Medline Industries, Inc. DYNJ906037AMedline Industries, Inc. · Plastic surgery and accessories kit1 recall0 reports
- Medline Industries, Inc. DYNJ46781BMedline Industries, Inc. · Plastic surgery and accessories kit1 recall0 reports
- Medline Industries, Inc. DYNJ68844Medline Industries, Inc. · Plastic surgery and accessories kit1 recall0 reports
- Medline Industries, Inc. DYNJ43859JMedline Industries, Inc. · Plastic surgery and accessories kit1 recall0 reports
- Medline Industries, Inc. DYNJ904565FMedline Industries, Inc. · Plastic surgery and accessories kit1 recall0 reports
- Medline Industries, Inc. CDS980375LMedline Industries, Inc. · Plastic surgery and accessories kit1 recall0 reports
- Medline Industries, Inc. DYNJ65028AMedline Industries, Inc. · Plastic surgery and accessories kit1 recall0 reports
- Medline Industries, Inc. DYNJ0785741LMedline Industries, Inc. · Plastic surgery and accessories kit1 recall0 reports
- Medline Industries, Inc. DYNJVB91096AMedline Industries, Inc. · Plastic surgery and accessories kit1 recall0 reports
- Medline Industries, Inc. DYNJ46504CMedline Industries, Inc. · Plastic surgery and accessories kit1 recall0 reports
- Medline Industries, Inc. DYNJ54761CMedline Industries, Inc. · Plastic surgery and accessories kit1 recall0 reports
- Medline Industries, Inc. DYNJ0991320LMedline Industries, Inc. · Plastic surgery and accessories kit1 recall0 reports
- Medline Industries, Inc. DYNJ905365BMedline Industries, Inc. · Plastic surgery and accessories kit1 recall0 reports
- Medline Industries, Inc. DYNJ48191IMedline Industries, Inc. · Plastic surgery and accessories kit1 recall0 reports
- Medline Industries, Inc. DYNJ81727Medline Industries, Inc. · Plastic surgery and accessories kit1 recall0 reports
- Medline Industries, Inc. DYNJ81772Medline Industries, Inc. · Plastic surgery and accessories kit1 recall0 reports
- Medline Industries, Inc. CDS983173IMedline Industries, Inc. · Plastic surgery and accessories kit1 recall0 reports
- Medline Industries, Inc. DYNJ82124Medline Industries, Inc. · Plastic surgery and accessories kit1 recall0 reports
- Medline Industries, Inc. DYNJ48191JMedline Industries, Inc. · Plastic surgery and accessories kit1 recall0 reports
- Medline Industries, Inc. DYNJ83417Medline Industries, Inc. · Plastic surgery and accessories kit1 recall0 reports
- Medline Industries, Inc. DYNJ81727AMedline Industries, Inc. · Plastic surgery and accessories kit1 recall0 reports
- Medline Industries, Inc. DYNJ64658AMedline Industries, Inc. · Plastic surgery and accessories kit1 recall0 reports
- Medline Industries, Inc. DYNJ39211FMedline Industries, Inc. · Plastic surgery and accessories kit1 recall0 reports
- Medline Industries, Inc. DYNJ909439Medline Industries, Inc. · Plastic surgery and accessories kit1 recall0 reports
- Medline Industries, Inc. DYNJ904565GMedline Industries, Inc. · Plastic surgery and accessories kit1 recall0 reports
- Medline Industries, Inc. DYNJ83417AMedline Industries, Inc. · Plastic surgery and accessories kit1 recall0 reports
- Medline Industries, Inc. DYNJ901753KMedline Industries, Inc. · Plastic surgery and accessories kit1 recall0 reports
- Medline Industries, Inc. DYNJ60392DMedline Industries, Inc. · Plastic surgery and accessories kit1 recall0 reports
- Medline Industries, Inc. DYNJ48191KMedline Industries, Inc. · Plastic surgery and accessories kit1 recall0 reports
- Medline Industries, Inc. DYNJ904565IMedline Industries, Inc. · Plastic surgery and accessories kit1 recall0 reports
- Medline Industries, Inc. DYNJ46504DMedline Industries, Inc. · Plastic surgery and accessories kit1 recall0 reports
- Medline Industries, Inc. DYNJ59400KMedline Industries, Inc. · Plastic surgery and accessories kit1 recall0 reports
- Medline Industries, Inc. DYNJ909926Medline Industries, Inc. · Plastic surgery and accessories kit1 recall0 reports
- Medline Industries, Inc. DYNJ908264AMedline Industries, Inc. · Plastic surgery and accessories kit1 recall0 reports
- Medline Industries, Inc. DYNJ86166Medline Industries, Inc. · Plastic surgery and accessories kit1 recall0 reports
- Medline Industries, Inc. DYNJ48191LMedline Industries, Inc. · Plastic surgery and accessories kit1 recall0 reports
- Medline Industries, Inc. CDS983100DMedline Industries, Inc. · Plastic surgery and accessories kit1 recall0 reports
- Medline Industries, Inc. DYNJ87503Medline Industries, Inc. · Plastic surgery and accessories kit1 recall0 reports
- Medline Industries, Inc. DYNJ907076DMedline Industries, Inc. · Plastic surgery and accessories kit1 recall0 reports
- Medline Industries, Inc. DYNJ48191MMedline Industries, Inc. · Plastic surgery and accessories kit1 recall0 reports
- Medline Industries, Inc. DYNJ904275FMedline Industries, Inc. · Plastic surgery and accessories kit1 recall0 reports
- Medline Industries, Inc. DYNJ909472AMedline Industries, Inc. · Plastic surgery and accessories kit1 recall0 reports
- Medline Industries, Inc. DYNJ88201Medline Industries, Inc. · Plastic surgery and accessories kit1 recall0 reports
- Medline Industries, Inc. DYNJ46504FMedline Industries, Inc. · Plastic surgery and accessories kit1 recall0 reports
- Medline Industries, Inc. DYNJ909926BMedline Industries, Inc. · Plastic surgery and accessories kit1 recall0 reports
- Medline Industries, Inc. DYNJ907076FMedline Industries, Inc. · Plastic surgery and accessories kit1 recall0 reports
- Medline Industries, Inc. DYNJ909439AMedline Industries, Inc. · Plastic surgery and accessories kit1 recall0 reports
- Medline Industries, Inc. DYNJ81727FMedline Industries, Inc. · Plastic surgery and accessories kit1 recall0 reports
- Medline Industries, Inc. DYNJ60392FMedline Industries, Inc. · Plastic surgery and accessories kit1 recall0 reports
- MEDLINE INDUSTRIES, INC. DYNJ87503AMedline Industries, Inc. · Plastic surgery and accessories kit1 recall0 reports
- MEDLINE INDUSTRIES, INC. DYNJ905365CMedline Industries, Inc. · Plastic surgery and accessories kit1 recall0 reports
- MEDLINE INDUSTRIES, INC. DYNJ88201AMedline Industries, Inc. · Plastic surgery and accessories kit1 recall0 reports
- MEDLINE INDUSTRIES, INC. DYNJT4005Medline Industries, Inc. · Plastic surgery and accessories kit1 recall0 reports
- Medline Industries, Inc. DYNJ901296AMedline Industries, Inc. · Plastic surgery and accessories kit1 recall0 reports
- Medline Industries, Inc. DYNJ40648FMedline Industries, Inc. · Plastic surgery and accessories kit1 recall0 reports
- Medline Industries, Inc. DYNJ900845BMedline Industries, Inc. · Plastic surgery and accessories kit1 recall0 reports
- Medline Industries, Inc. DYNJ57604Medline Industries, Inc. · Plastic surgery and accessories kit1 recall0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 98664).
- Z-2008-2026 · Open, ClassifiedMedline Neuro Sponges, various dimensions, labeled as: 1. Neuro Sponges, Eponges neuro, Neuroesponjas, size 0.25" X 0.25", XR, 10/PK, Medline SKU NEUROSPNG02; 2. Neuro Sponge…
- Z-2009-2026 · Open, ClassifiedMedline medical procedure kits, containing Medline Neuro Sponges, labeled as follows: PEDS AIRWAY PACK, Medline Kit Number SKU DYNJ86657A
- Z-2010-2026 · Open, ClassifiedMedline medical procedure kits, containing Medline Neuro Sponges, labeled as follows: EH BRONCHOSCOPY, Medline Kit Number/SKU DYNJT3270
- Z-2011-2026 · Open, ClassifiedMedline medical procedure kits, containing Medline Neuro Sponges, labeled as follows: SPINAL FUSION, Medline Kit Number/SKU CDS860017AG
- Z-2012-2026 · Open, ClassifiedMedline medical procedure kits, containing Medline Neuro Sponges, labeled as follows: SPINE PACK-LF, Medline Kit Number/SKU PHS719414B
- Z-2013-2026 · Open, ClassifiedMedline medical procedure kits, containing Medline Neuro Sponges, labeled as follows: 1. NEURO PACK, Medline kit number/SKU DYNJ64479; 2. CRANIO/NEURO/SPINE, Medline kit number/…
- Z-2014-2026 · Open, ClassifiedMedline medical procedure kits, containing Medline Neuro Sponges, labeled as follows: ANGIO PACK-LF, Medline Kit Number/SKU DYNJ0774765U
- Z-2015-2026 · Open, ClassifiedMedline medical procedure kits, containing Medline Neuro Sponges, labeled as follows: 1. MAJOR NEURO PACK-LF, Medline Kit Number/SKU DYNJ0578916X; 2. ANGIOGRAPHY PACK, Medline K…
- Z-2016-2026 · Open, ClassifiedMedline medical procedure kits, containing Medline Neuro Sponges, labeled as follows: 1. OPEN HEART CDS PART A, Medline Kit Number/SKU CDS840402AI; 2. OPEN HEART CDS, Medline Ki…
- Z-2017-2026 · Open, ClassifiedMedline medical procedure kits, containing Medline Neuro Sponges, labeled as follows: 1. VP SHUNT, Medline Kit Number/SKU CDS840193V; 2. SHY VP SHUNT CDS, Medline Kit Number/SKU…
- Z-2018-2026 · Open, ClassifiedMedline medical procedure kits, containing Medline Neuro Sponges, labeled as follows: 1. OSC COLEMAN FAT HARVEST KIT, Medline Kit Number/SKU DYNJ32854B; 2. RR-DR. CHHETRI MDL PA…
- Z-2019-2026 · Open, ClassifiedMedline medical procedure kits, containing Medline Neuro Sponges, labeled as follows: 1. RR-OMFS PACK-LF, Medline Kit Number/SKU DYNJ0045920D; 2. RR-OMFS PACK-LF, Medline Kit Nu…
- Z-2020-2026 · Open, ClassifiedMedline medical procedure kits, containing Medline Neuro Sponges, labeled as follows: 1. ZALE STEREOTACTIC CDS, Medline Kit Number/SKU CDS983173G; 2. EPISTAXIS TRAY, Medline Kit…
- Z-2021-2026 · Open, ClassifiedMedline and Centurion medical procedure kits, containing Medline Neuro Sponges (1229 in total), HEAD & NECK TRAY, SINUS, SMR, etc.
- Z-2022-2026 · Open, ClassifiedMedline medical procedure kits, containing Medline Neuro Sponges, labeled as follows: 1. PLASTICS FREE FLAP, Medline Kit Number/SKU CDS983493N; 2. TLIF PACK, Medline Kit Number/…
- Z-2023-2026 · Open, ClassifiedMedline medical procedure kits, containing Medline Neuro Sponges, labeled as follows: 1. ENDOSCOPIC SKULL BASE, Medline Kit Number/SKU DYNJ58648J; 2. ENT ENDO SINUS WILMINGTON PA…
- Z-2024-2026 · Open, ClassifiedMedline medical procedure kits, containing Medline Neuro Sponges, labeled as follows: 1. ENDOSCOPY-LF, Medline Kit Number/SKU DYNJ22890M; 2. ENDOSCOPY PACK, Medline Kit Number/…
- Z-2026-2026 · Open, ClassifiedMedline medical procedure kits, containing Medline Neuro Sponges, labeled as follows: 1. NEURO BASIN PACK-LF, Medline Kit Number/SKU DYNJ0532910L; 2. NEURO BASIN PACK-LF, Medlin…
- Z-2027-2026 · Open, ClassifiedMedline medical procedure kits, containing Medline Neuro Sponges (1050 in total), BIOBURDEN TEST KIT, NEURO, ROBOTIC, ACDF, ENT, etc.
- Z-2028-2026 · Open, ClassifiedCenturion medical procedure kits, containing Medline Neuro Sponges, labeled as follows: SURGICAL ENT PREP KIT, Medline Kit Number/SKU MNS11715
- Z-2029-2026 · Open, ClassifiedMedline medical procedure kits, containing Medline Neuro Sponges, labeled as follows: CERVICAL PACK, Medline Kit Number/SKU DYNJ37902F
- Z-2030-2026 · Open, ClassifiedMedline medical procedure kits, containing Medline Neuro Sponges, labeled as follows: 1. MICRODISECTOMY PACK, Medline Kit Number/SKU DYNJ80585B; 2. MICRODISECTOMY PACK, Medline …
- Z-2031-2026 · Open, ClassifiedMedline medical procedure kits, containing Medline Neuro Sponges, labeled as follows: GENERAL KIT, Medline Kit Number/SKU DYNJ905664I
- Z-2033-2026 · Open, ClassifiedMedline medical procedure kits, containing Medline Neuro Sponges (961 in total), NEURO,
- Z-2034-2026 · Open, ClassifiedMedline medical procedure kits, containing Medline Neuro Sponges, labeled as follows: 1. SPINE PROCEDURES PACK-LF, Medline Kit Number/SKU DYNJ0751220P; 2. ENSEMBLE RACHIS-LF, Medl…
- Z-2035-2026 · Open, ClassifiedMedline medical procedure kits, containing Medline Neuro Sponges, labeled as follows: 1. ROBOTIC HYSTERECTOMY, Medline Kit Number/SKU DYNJ904269B; 2. ROBOTIC HYSTERECTOMY, Medli…
- Z-2036-2026 · Open, ClassifiedMedline medical procedure kits, containing Medline Neuro Sponges, labeled as follows: 1. ANTERIOR CERVICAL FUSION PACK, Medline Kit Number/SKU DYNJ40566C; 2. OB PACK (OBSJD)642-L…
- Z-2037-2026 · Open, ClassifiedMedline medical procedure kits, containing Medline Neuro Sponges, labeled as follows: 1. CATARACT PACK, Medline Kit Number/SKU DYNJ48725A; 2. OCULAR PACK, Medline Kit Number/SKU …
- Z-2038-2026 · Open, ClassifiedMedline medical procedure kits, containing Medline Neuro Sponges (4077 in total), LAMINECTOMY, CRANI, etc.
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page