Medline medical procedure kits, containing Medline Neuro… recallZ-2022-2026 · Medline Industries, LP · initiated Mar 19, 2026

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Medline Industries, LP or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

During an internal review, Medline identified higher-than-expected endotoxin levels on a representative Medline neuro sponge product, indicating that in-market product may be out-of-specification for endotoxin.

FDA root cause category: Under Investigation by firm

What the company is doing

Medline industries issued an URGENT MEDICAL DEVICE RECALL notice to its consignees on 3/19/2026 via USPS first class mail and email. The notice explained the issue, potential health risk, and requested the following actions be taken on the affected sponges: Identify and quarantine affected product, complete response form using website link: https://recalls.medline.com, destroy affected product. Credit will be issued once response form is received. If the product has been further distributed, sold, or transferred, notify the customer to destroy the product. Distributors must notify their customers and should include customer quantities on your response form. For consignees of Medline medical convenience kits containing the affected sponges the actions to be taken are as follows: Identify and quarantine all affected kits, complete response form using website link: https://recalls.medline.com. Upon receipt of your submitted response form, your account will receive over-labels to place on affected inventory, with instructions for your staff to remove and discard the affected component prior to using the kit. These labels will be provided via FedEx overnight, if applicable. Distributors must notify their customers and should include customer quantities on your response form. Upon receipt of your submitted response form, you will receive over-labels to provide to your customers to affix to their affected product. If you have any questions, contact the Recall Department at 866-359-1704 or [email protected].

Recall details

Recall number
Z-2022-2026
Status
Open, Classified
Initiated by the company
Mar 19, 2026
Posted by the FDA
Apr 30, 2026
Recalling firm
Medline Industries, LP
Quantity
21288 kits
Product code
FDE
Distribution
Worldwide distribution.

Product description

Medline medical procedure kits, containing Medline Neuro Sponges, labeled as follows: 1. PLASTICS FREE FLAP, Medline Kit Number/SKU CDS983493N; 2. TLIF PACK, Medline Kit Number/SKU DYNJ02476O; 3. DONOR FREE FLAP PACK-LF, Medline Kit Number/SKU DYNJ0426069N; 4. DONOR FREE FLAP PACK-LF, Medline Kit Number/SKU DYNJ0426069Q; 5. MINOR MB PACK-LF, Medline Kit Number/SKU DYNJ0429303L; 6. ENDOSCOPIC SKULL BASE PACK-LF, Medline Kit Number/SKU DYNJ0753948J; 7. RR-H&N FREEFLAP BLCKWELL PK-LF, Medline Kit Number/SKU DYNJ0785741K; 8. MINIMALLY INVASIVE PACK-LF, Medline Kit Number/SKU DYNJ0843063K; 9. FREE FLAP BASIC SET-UP PK, Medline Kit Number/SKU DYNJ0991320K; 10. FREE FLAP BASIC SET-UP PK, Medline Kit Number/SKU DYNJ0991320M; 11. MLD PACK, Medline Kit Number/SKU DYNJ19521Q; 12. SINUS ENDOSCOPY PACK, Medline Kit Number/SKU DYNJ34623C; 13. PMD PACK SURG CNTR-LF, Medline Kit Number/SKU DYNJ35305G; 14. LARYNGOSCOPY PACK, Medline Kit Number/SKU DYNJ37547D; 15. FUNCTIONAL SINUS ENDOSCOPY, Medline Kit Number/SKU DYNJ40406A; 16. ENDOSCOPY PACK, Medline Kit Number/SKU DYNJ40645C; 17. FREE FLAP B PACK, Medline Kit Number/SKU DYNJ40648G; 18. MINOR ENT PACK, Medline Kit Number/SKU DYNJ50808A; 19. SINUS ENDOSCOPY PACK-LF, Medline Kit Number/SKU DYNJ52062; 20. LAP PACK, Medline Kit Number/SKU DYNJ52116B; 21. LAP PACK, Medline Kit Number/SKU DYNJ52116C; 22. ENDOSCOPY PK, Medline Kit Number/SKU DYNJ54719C; 23. ENDOSCOPY PK, Medline Kit Number/SKU DYNJ54719D; 24. FREE FLAP PACK, Medline Kit Number/SKU DYNJ56917I; 25. FREE FLAP PACK, Medline Kit Number/SKU DYNJ56917J; 26. FREE FLAP PACK, Medline Kit Number/SKU DYNJ56917K; 27. FREE FLAP PACK, Medline Kit Number/SKU DYNJ56917L; 28. FREE FLAP PACK, Medline Kit Number/SKU DYNJ56917N; 29. LARYNGOSCOPY WEST PACK, Medline Kit Number/SKU DYNJ58324A; 30. CCH DECOMPRESSION PACK, Medline Kit Number/SKU DYNJ67558A; 31. CCH DECOMPRESSION PACK, Medline Kit Number/SKU DYNJ67558B; 32. FREE FLAP PACK, Medline Kit Number/SKU DYNJ68213; 33. FREE FLAP PACK, Medline Kit Number/SKU DYNJ68213D; 34. FREE FLAP PACK, Medline Kit Number/SKU DYNJ68213G; 35. EENT PACK, Medline Kit Number/SKU DYNJ68366B; 36. FREE FLAP NO DRAPE PACK NTX, Medline Kit Number/SKU DYNJ68447; 37. FREE FLAP NO DRAPE PACK NTX, Medline Kit Number/SKU DYNJ68447A; 38. FREE FLAP NO DRAPE PACK NTX, Medline Kit Number/SKU DYNJ68447B; 39. UNIVERSITY DISC PACK, Medline Kit Number/SKU DYNJ69824; 40. MAJOR PACK, Medline Kit Number/SKU DYNJ81325; 41. MAJOR PACK, Medline Kit Number/SKU DYNJ81325A; 42. MAJOR PACK, Medline Kit Number/SKU DYNJ81325B; 43. PLANO DIEP PACK, Medline Kit Number/SKU DYNJ81751; 44. PLANO DIEP PACK, Medline Kit Number/SKU DYNJ81751A; 45. FLAP PACK, Medline Kit Number/SKU DYNJ85359; 46. ENDO PACK, Medline Kit Number/SKU DYNJ86125; 47. ENDO PACK, Medline Kit Number/SKU DYNJ86125A; 48. ENT PACK, Medline Kit Number/SKU DYNJ88448; 49. MAJOR BACK PACK, Medline Kit Number/SKU DYNJ88470; 50. UROLOGY DAVINCI, Medline Kit Number/SKU DYNJ901572O; 51. LARYNGOSCOPY/BRONCH, Medline Kit Number/SKU DYNJ902985F; 52. KIT FREE FLAP MOSES, Medline Kit Number/SKU DYNJ905366B; 53. FLAP, Medline Kit Number/SKU DYNJ907076C; 54. TRANSPHENOIDAL, Medline Kit Number/SKU DYNJ908165; 55. TRANSPHENOIDAL, Medline Kit Number/SKU DYNJ908165A; 56. ENDOSCOPY, Medline Kit Number/SKU DYNJ909512; 57. DIEP FLAP, Medline Kit Number/SKU DYNJ910120A; 58. DIEP BREAST, Medline Kit Number/SKU DYNJ911086; 59. WILLIAMS GENERAL, Medline Kit Number/SKU DYNJ940735F; 60. ROBOTIC LAP PROSTATE PACK CHS, Medline Kit Number/SKU DYNJT3101; 61. CUH FLAP ACCESSORY PACK, Medline Kit Number/SKU DYNJT3243.

Code information (models, lots, serial numbers)

Show the full code information (54k characters)

Medline Kit Number/SKU CDS983493N: UDI/DI each 10198459000218, UDI/DI case 40198459000219, Lot Number: 25EDA675; Medline Kit Number/SKU CDS983493N: UDI/DI each 10198459000218, UDI/DI case 40198459000219, Lot Number: 25CDA718; Medline Kit Number/SKU CDS983493N: UDI/DI each 10198459000218, UDI/DI case 40198459000219, Lot Number: 25BDA328; Medline Kit Number/SKU CDS983493N: UDI/DI each 10198459000218, UDI/DI case 40198459000219, Lot Number: 24KDB408; Medline Kit Number/SKU CDS983493N: UDI/DI each 10198459000218, UDI/DI case 40198459000219, Lot Number: 24KDA607; Medline Kit Number/SKU DYNJ02476O: UDI/DI each 10195327119478, UDI/DI case 40195327119479, Lot Number: 23HMH658; Medline Kit Number/SKU DYNJ02476O: UDI/DI each 10195327119478, UDI/DI case 40195327119479, Lot Number: 23GMD341; Medline Kit Number/SKU DYNJ02476O: UDI/DI each 10195327119478, UDI/DI case 40195327119479, Lot Number: 23EMB793; Medline Kit Number/SKU DYNJ02476O: UDI/DI each 10195327119478, UDI/DI case 40195327119479, Lot Number: 23CMG372; Medline Kit Number/SKU DYNJ02476O: UDI/DI each 10195327119478, UDI/DI case 40195327119479, Lot Number: 23BMI499; Medline Kit Number/SKU DYNJ02476O: UDI/DI each 10195327119478, UDI/DI case 40195327119479, Lot Number: 22JMF967; Medline Kit Number/SKU DYNJ02476O: UDI/DI each 10195327119478, UDI/DI case 40195327119479, Lot Number: 22EMC648; Medline Kit Number/SKU DYNJ02476O: UDI/DI each 10195327119478, UDI/DI case 40195327119479, Lot Number: 22CMD889; Medline Kit Number/SKU DYNJ02476O: UDI/DI each 10195327119478, UDI/DI case 40195327119479, Lot Number: 24EMH897; Medline Kit Number/SKU DYNJ02476O: UDI/DI each 10195327119478, UDI/DI case 40195327119479, Lot Number: 24DMF264; Medline Kit Number/SKU DYNJ02476O: UDI/DI each 10195327119478, UDI/DI case 40195327119479, Lot Number: 24BMJ904; Medline Kit Number/SKU DYNJ02476O: UDI/DI each 10195327119478, UDI/DI case 40195327119479, Lot Number: 24AMH650; Medline Kit Number/SKU DYNJ0426069N: UDI/DI each 10195327371319, UDI/DI case 40195327371310, Lot Number: 23IME074; Medline Kit Number/SKU DYNJ0426069N: UDI/DI each 10195327371319, UDI/DI case 40195327371310, Lot Number: 23IMA416; Medline Kit Number/SKU DYNJ0426069N: UDI/DI each 10195327371319, UDI/DI case 40195327371310, Lot Number: 23EME302; Medline Kit Number/SKU DYNJ0426069N: UDI/DI each 10195327371319, UDI/DI case 40195327371310, Lot Number: 23DMH071; Medline Kit Number/SKU DYNJ0426069N: UDI/DI each 10195327371319, UDI/DI case 40195327371310, Lot Number: 23CMH752; Medline Kit Number/SKU DYNJ0426069N: UDI/DI each 10195327371319, UDI/DI case 40195327371310, Lot Number: 23LME133; Medline Kit Number/SKU DYNJ0426069Q: UDI/DI each 10198459311321, UDI/DI case 40198459311322, Lot Number: 25LMB574; Medline Kit Number/SKU DYNJ0426069Q: UDI/DI each 10198459311321, UDI/DI case 40198459311322, Lot Number: 25IMG534; Medline Kit Number/SKU DYNJ0426069Q: UDI/DI each 10198459311321, UDI/DI case 40198459311322, Lot Number: 25EMJ878; Medline Kit Number/SKU DYNJ0426069Q: UDI/DI each 10198459311321, UDI/DI case 40198459311322, Lot Number: 25DME997; Medline Kit Number/SKU DYNJ0426069Q: UDI/DI each 10198459311321, UDI/DI case 40198459311322, Lot Number: 25CMI768; Medline Kit Number/SKU DYNJ0429303L: UDI/DI each 10195327587260, UDI/DI case 40195327587261, Lot Number: 24HMH265; Medline Kit Number/SKU DYNJ0753948J: UDI/DI each 10193489895896, UDI/DI case 40193489895897, Lot Number: 21KMD078; Medline Kit Number/SKU DYNJ0753948J: UDI/DI each 10193489895896, UDI/DI case 40193489895897, Lot Number: 21HME198; Medline Kit Number/SKU DYNJ0753948J: UDI/DI each 10193489895896, UDI/DI case 40193489895897, Lot Number: 21FMD014; Medline Kit Number/SKU DYNJ0785741K: UDI/DI each 10193489603576, UDI/DI case 40193489603577, Lot Number: 21CKA430; Medline Kit Number/SKU DYNJ0843063K: UDI/DI each 10195327286460, UDI/DI case 40195327286461, Lot Number: 23JDA179; Medline Kit Number/SKU DYNJ0843063K: UDI/DI each 10195327286460, UDI/DI case 40195327286461, Lot Number: 23HDC666; Medline Kit Number/SKU DYNJ0843063K: UDI/DI each 10195327286460, UDI/DI case 40195327286461, Lot Number: 23FDB421; Medline Kit Number/SKU DYNJ0843063K: UDI/DI each 10195327286460, UDI/DI case 40195327286461, Lot Number: 23EDA191; Medline Kit Number/SKU DYNJ0843063K: UDI/DI each 10195327286460, UDI/DI case 40195327286461, Lot Number: 23CDC020; Medline Kit Number/SKU DYNJ0843063K: UDI/DI each 10195327286460, UDI/DI case 40195327286461, Lot Number: 23BDA963; Medline Kit Number/SKU DYNJ0843063K: UDI/DI each 10195327286460, UDI/DI case 40195327286461, Lot Number: 23ADB394; Medline Kit Number/SKU DYNJ0843063K: UDI/DI each 10195327286460, UDI/DI case 40195327286461, Lot Number: 23LDA280; Medline Kit Number/SKU DYNJ0991320K: UDI/DI each 10193489925388, UDI/DI case 40193489925389, Lot Number: 21ILA211; Medline Kit Number/SKU DYNJ0991320K: UDI/DI each 10193489925388, UDI/DI case 40193489925389, Lot Number: 21HLA920; Medline Kit Number/SKU DYNJ0991320K: UDI/DI each 10193489925388, UDI/DI case 40193489925389, Lot Number: 21HLA656; Medline Kit Number/SKU DYNJ0991320K: UDI/DI each 10193489925388, UDI/DI case 40193489925389, Lot Number: 21GLA243; Medline Kit Number/SKU DYNJ0991320M: UDI/DI each 10195327632458, UDI/DI case 40195327632459, Lot Number: 24IMH942; Medline Kit Number/SKU DYNJ0991320M: UDI/DI each 10195327632458, UDI/DI case 40195327632459, Lot Number: 24IMB140; Medline Kit Number/SKU DYNJ0991320M: UDI/DI each 10195327632458, UDI/DI case 40195327632459, Lot Number: 24GMB047; Medline Kit Number/SKU DYNJ0991320M: UDI/DI each 10195327632458, UDI/DI case 40195327632459, Lot Number: 24EMG280; Medline Kit Number/SKU DYNJ19521Q: UDI/DI each 10195327318079, UDI/DI case 40195327318070, Lot Number: 23BMB103; Medline Kit Number/SKU DYNJ19521Q: UDI/DI each 10195327318079, UDI/DI case 40195327318070, Lot Number: 25BMG556; Medline Kit Number/SKU DYNJ34623C: UDI/DI each 10889942745807, UDI/DI case 40889942745808, Lot Number: 23CBT880; Medline Kit Number/SKU DYNJ34623C: UDI/DI each 10889942745807, UDI/DI case 40889942745808, Lot Number: 22KBW690; Medline Kit Number/SKU DYNJ34623C: UDI/DI each 10889942745807, UDI/DI case 40889942745808, Lot Number: 21LBG909; Medline Kit Number/SKU DYNJ35305G: UDI/DI each 10195327070328, UDI/DI case 40195327070329, Lot Number: 23AMB303; Medline Kit Number/SKU DYNJ37547D: UDI/DI each 10889942945993, UDI/DI case 40889942945994, Lot Number: 21DDB912; Medline Kit Number/SKU DYNJ37547D: UDI/DI each 10889942945993, UDI/DI case 40889942945994, Lot Number: 21CDB910; Medline Kit Number/SKU DYNJ40406A: UDI/DI each 10889942215904, UDI/DI case 40889942215905, Lot Number: 23JBY647; Medline Kit Number/SKU DYNJ40406A: UDI/DI each 10889942215904, UDI/DI case 40889942215905, Lot Number: 23HBX192; Medline Kit Number/SKU DYNJ40406A: UDI/DI each 10889942215904, UDI/DI case 40889942215905, Lot Number: 23GBC541; Medline Kit Number/SKU DYNJ40406A: UDI/DI each 10889942215904, UDI/DI case 40889942215905, Lot Number: 23DBD185; Medline Kit Number/SKU DYNJ40406A: UDI/DI each 10889942215904, UDI/DI case 40889942215905, Lot Number: 23CBP552; Medline Kit Number/SKU DYNJ40406A: UDI/DI each 10889942215904, UDI/DI case 40889942215905, Lot Number: 23ABG640; Medline Kit Number/SKU DYNJ40406A: UDI/DI each 10889942215904, UDI/DI case 40889942215905, Lot Number: 22JBW980; Medline Kit Number/SKU DYNJ40406A: UDI/DI each 10889942215904, UDI/DI case 40889942215905, Lot Number: 22HBY401; Medline Kit Number/SKU DYNJ40406A: UDI/DI each 10889942215904, UDI/DI case 40889942215905, Lot Number: 22GBG635; Medline Kit Number/SKU DYNJ40406A: UDI/DI each 10889942215904, UDI/DI case 40889942215905, Lot Number: 22EBL969; Medline Kit Number/SKU DYNJ40406A: UDI/DI each 10889942215904, UDI/DI case 40889942215905, Lot Number: 22CBX216; Medline Kit Number/SKU DYNJ40406A: UDI/DI each 10889942215904, UDI/DI case 40889942215905, Lot Number: 22ABL872; Medline Kit Number/SKU DYNJ40406A: UDI/DI each 10889942215904, UDI/DI case 40889942215905, Lot Number: 21KBK659; Medline Kit Number/SKU DYNJ40406A: UDI/DI each 10889942215904, UDI/DI case 40889942215905, Lot Number: 21IBR941; Medline Kit Number/SKU DYNJ40406A: UDI/DI each 10889942215904, UDI/DI case 40889942215905, Lot Number: 21FBG655; Medline Kit Number/SKU DYNJ40406A: UDI/DI each 10889942215904, UDI/DI case 40889942215905, Lot Number: 21DBG029; Medline Kit Number/SKU DYNJ40406A: UDI/DI each 10889942215904, UDI/DI case 40889942215905, Lot Number: 21CBJ871; Medline Kit Number/SKU DYNJ40406A: UDI/DI each 10889942215904, UDI/DI case 40889942215905, Lot Number: 26CBE134; Medline Kit Number/SKU DYNJ40406A: UDI/DI each 10889942215904, UDI/DI case 40889942215905, Lot Number: 26ABM247; Medline Kit Number/SKU DYNJ40406A: UDI/DI each 10889942215904, UDI/DI case 40889942215905, Lot Number: 25KBF571; Medline Kit Number/SKU DYNJ40406A: UDI/DI each 10889942215904, UDI/DI case 40889942215905, Lot Number: 25IBS828; Medline Kit Number/SKU DYNJ40406A: UDI/DI each 10889942215904, UDI/DI case 40889942215905, Lot Number: 25HBF128; Medline Kit Number/SKU DYNJ40406A: UDI/DI each 10889942215904, UDI/DI case 40889942215905, Lot Number: 25GBD444; Medline Kit Number/SKU DYNJ40406A: UDI/DI each 10889942215904, UDI/DI case 40889942215905, Lot Number: 25EBN746; Medline Kit Number/SKU DYNJ40406A: UDI/DI each 10889942215904, UDI/DI case 40889942215905, Lot Number: 25DBU598; Medline Kit Number/SKU DYNJ40406A: UDI/DI each 10889942215904, UDI/DI case 40889942215905, Lot Number: 25BBP703; Medline Kit Number/SKU DYNJ40406A: UDI/DI each 10889942215904, UDI/DI case 40889942215905, Lot Number: 24LBL146; Medline Kit Number/SKU DYNJ40406A: UDI/DI each 10889942215904, UDI/DI case 40889942215905, Lot Number: 24HBR568; Medline Kit Number/SKU DYNJ40406A: UDI/DI each 10889942215904, UDI/DI case 40889942215905, Lot Number: 24GBW861; Medline Kit Number/SKU DYNJ40406A: UDI/DI each 10889942215904, UDI/DI case 40889942215905, Lot Number: 24FBL120; Medline Kit Number/SKU DYNJ40406A: UDI/DI each 10889942215904, UDI/DI case 40889942215905, Lot Number: 24EBD749; Medline Kit Number/SKU DYNJ40406A: UDI/DI each 10889942215904, UDI/DI case 40889942215905, Lot Number: 23LBI152; Medline Kit Number/SKU DYNJ40645C: UDI/DI each 10195327107802, UDI/DI case 40195327107803, Lot Number: 23JMA313; Medline Kit Number/SKU DYNJ40645C: UDI/DI each 10195327107802, UDI/DI case 40195327107803, Lot Number: 23HMH325; Medline Kit Number/SKU DYNJ40645C: UDI/DI each 10195327107802, UDI/DI case 40195327107803, Lot Number: 23GMA570; Medline Kit Number/SKU DYNJ40645C: UDI/DI each 10195327107802, UDI/DI case 40195327107803, Lot Number: 23EMI670; Medline Kit Number/SKU DYNJ40645C: UDI/DI each 10195327107802, UDI/DI case 40195327107803, Lot Number: 23DMC678; Medline Kit Number/SKU DYNJ40645C: UDI/DI each 10195327107802, UDI/DI case 40195327107803, Lot Number: 23CMB630; Medline Kit Number/SKU DYNJ40645C: UDI/DI each 10195327107802, UDI/DI case 40195327107803, Lot Number: 22LMF935; Medline Kit Number/SKU DYNJ40645C: UDI/DI each 10195327107802, UDI/DI case 40195327107803, Lot Number: 22HMF705; Medline Kit Number/SKU DYNJ40645C: UDI/DI each 10195327107802, UDI/DI case 40195327107803, Lot Number: 22FMD171; Medline Kit Number/SKU DYNJ40645C: UDI/DI each 10195327107802, UDI/DI case 40195327107803, Lot Number: 22FMD415; Medline Kit Number/SKU DYNJ40645C: UDI/DI each 10195327107802, UDI/DI case 40195327107803, Lot Number: 22EMB584; Medline Kit Number/SKU DYNJ40645C: UDI/DI each 10195327107802, UDI/DI case 40195327107803, Lot Number: 22DMF559; Medline Kit Number/SKU DYNJ40648G: UDI/DI each 10195327106997, UDI/DI case 40195327106998, Lot Number: 23DMH298; Medline Kit Number/SKU DYNJ40648G: UDI/DI each 10195327106997, UDI/DI case 40195327106998, Lot Number: 22KMH672; Medline Kit Number/SKU DYNJ40648G: UDI/DI each 10195327106997, UDI/DI case 40195327106998, Lot Number: 22GMD348; Medline Kit Number/SKU DYNJ40648G: UDI/DI each 10195327106997, UDI/DI case 40195327106998, Lot Number: 22FMD240; Medline Kit Number/SKU DYNJ40648G: UDI/DI each 10195327106997, UDI/DI case 40195327106998, Lot Number: 22EME295; Medline Kit Number/SKU DYNJ40648G: UDI/DI each 10195327106997, UDI/DI case 40195327106998, Lot Number: 22DMG577; Medline Kit Number/SKU DYNJ50808A: UDI/DI each 10889942368105, UDI/DI case 40889942368106, Lot Number: 23HBR726; Medline Kit Number/SKU DYNJ50808A: UDI/DI each 10889942368105, UDI/DI case 40889942368106, Lot Number: 23FBG834; Medline Kit Number/SKU DYNJ50808A: UDI/DI each 10889942368105, UDI/DI case 40889942368106, Lot Number: 23CBL006; Medline Kit Number/SKU DYNJ50808A: UDI/DI each 10889942368105, UDI/DI case 40889942368106, Lot Number: 22IBL408; Medline Kit Number/SKU DYNJ50808A: UDI/DI each 10889942368105, UDI/DI case 40889942368106, Lot Number: 22HBY173; Medline Kit Number/SKU DYNJ50808A: UDI/DI each 10889942368105, UDI/DI case 40889942368106, Lot Number: 22EBF344; Medline Kit Number/SKU DYNJ50808A: UDI/DI each 10889942368105, UDI/DI case 40889942368106, Lot Number: 22ABB710; Medline Kit Number/SKU DYNJ50808A: UDI/DI each 10889942368105, UDI/DI case 40889942368106, Lot Number: 21LBR187; Medline Kit Number/SKU DYNJ50808A: UDI/DI each 10889942368105, UDI/DI case 40889942368106, Lot Number: 21KBE554; Medline Kit Number/SKU DYNJ50808A: UDI/DI each 10889942368105, UDI/DI case 40889942368106, Lot Number: 21HBQ602; Medline Kit Number/SKU DYNJ50808A: UDI/DI each 10889942368105, UDI/DI case 40889942368106, Lot Number: 21HBQ631; Medline Kit Number/SKU DYNJ50808A: UDI/DI each 10889942368105, UDI/DI case 40889942368106, Lot Number: 21GBJ642; Medline Kit Number/SKU DYNJ50808A: UDI/DI each 10889942368105, UDI/DI case 40889942368106, Lot Number: 21CBS316; Medline Kit Number/SKU DYNJ50808A: UDI/DI each 10889942368105, UDI/DI case 40889942368106, Lot Number: 21BBN842; Medline Kit Number/SKU DYNJ50808A: UDI/DI each 10889942368105, UDI/DI case 40889942368106, Lot Number: 25EBR511; Medline Kit Number/SKU DYNJ50808A: UDI/DI each 10889942368105, UDI/DI case 40889942368106, Lot Number: 25BBU313; Medline Kit Number/SKU DYNJ50808A: UDI/DI each 10889942368105, UDI/DI case 40889942368106, Lot Number: 24IBF115; Medline Kit Number/SKU DYNJ50808A: UDI/DI each 10889942368105, UDI/DI case 40889942368106, Lot Number: 24GBX974; Medline Kit Number/SKU DYNJ50808A: UDI/DI each 10889942368105, UDI/DI case 40889942368106, Lot Number: 24FBR318; Medline Kit Number/SKU DYNJ50808A: UDI/DI each 10889942368105, UDI/DI case 40889942368106, Lot Number: 24ABP408; Medline Kit Number/SKU DYNJ52062: UDI/DI each 10889942235742, UDI/DI case 40889942235743, Lot Number: 21LBU391; Medline Kit Number/SKU DYNJ52062: UDI/DI each 10889942235742, UDI/DI case 40889942235743, Lot Number: 21VBD187; Medline Kit Number/SKU DYNJ52116B: UDI/DI each 10193489755589, UDI/DI case 40193489755580, Lot Number: 23JMB108; Medline Kit Number/SKU DYNJ52116B: UDI/DI each 10193489755589, UDI/DI case 40193489755580, Lot Number: 23EMJ664; Medline Kit Number/SKU DYNJ52116B: UDI/DI each 10193489755589, UDI/DI case 40193489755580, Lot Number: 23CMC992; Medline Kit Number/SKU DYNJ52116B: UDI/DI each 10193489755589, UDI/DI case 40193489755580, Lot Number: 22LMF460; Medline Kit Number/SKU DYNJ52116B: UDI/DI each 10193489755589, UDI/DI case 40193489755580, Lot Number: 22FMD641; Medline Kit Number/SKU DYNJ52116B: UDI/DI each 10193489755589, UDI/DI case 40193489755580, Lot Number: 22ABG291; Medline Kit Number/SKU DYNJ52116B: UDI/DI each 10193489755589, UDI/DI case 40193489755580, Lot Number: 21KBF791; Medline Kit Number/SKU DYNJ52116B: UDI/DI each 10193489755589, UDI/DI case 40193489755580, Lot Number: 21HBT589; Medline Kit Number/SKU DYNJ52116B: UDI/DI each 10193489755589, UDI/DI case 40193489755580, Lot Number: 21DBJ867; Medline Kit Number/SKU DYNJ52116B: UDI/DI each 10193489755589, UDI/DI case 40193489755580, Lot Number: 21CBA789; Medline Kit Number/SKU DYNJ52116B: UDI/DI each 10193489755589, UDI/DI case 40193489755580, Lot Number: 24FMF952; Medline Kit Number/SKU DYNJ52116B: UDI/DI each 10193489755589, UDI/DI case 40193489755580, Lot Number: 24BMJ713; Medline Kit Number/SKU DYNJ52116C: UDI/DI each 10198459270352, UDI/DI case 40198459270353, Lot Number: 25LMD153; Medline Kit Number/SKU DYNJ52116C: UDI/DI each 10198459270352, UDI/DI case 40198459270353, Lot Number: 25KMH925; Medline Kit Number/SKU DYNJ52116C: UDI/DI each 10198459270352, UDI/DI case 40198459270353, Lot Number: 25BMD068; Medline Kit Number/SKU DYNJ54719C: UDI/DI each 10193489433449, UDI/DI case 40193489433440, Lot Number: 21CBH091; Medline Kit Number/SKU DYNJ54719C: UDI/DI each 10193489433449, UDI/DI case 40193489433440, Lot Number: 21ABM862; Medline Kit Number/SKU DYNJ54719D: UDI/DI each 10193489885293, UDI/DI case 40193489885294, Lot Number: 21HBE730; Medline Kit Number/SKU DYNJ54719D: UDI/DI each 10193489885293, UDI/DI case 40193489885294, Lot Number: 21FBO645; Medline Kit Number/SKU DYNJ56917I: UDI/DI each 10193489375206, UDI/DI case 40193489375207, Lot Number: 21GDB034; Medline Kit Number/SKU DYNJ56917I: UDI/DI each 10193489375206, UDI/DI case 40193489375207, Lot Number: 21FDC469; Medline Kit Number/SKU DYNJ56917I: UDI/DI each 10193489375206, UDI/DI case 40193489375207, Lot Number: 21EDC231; Medline Kit Number/SKU DYNJ56917I: UDI/DI each 10193489375206, UDI/DI case 40193489375207, Lot Number: 21ADA830; Medline Kit Number/SKU DYNJ56917J: UDI/DI each 10193489991994, UDI/DI case 40193489991995, Lot Number: 22KDA276; Medline Kit Number/SKU DYNJ56917J: UDI/DI each 10193489991994, UDI/DI case 40193489991995, Lot Number: 22JDA493; Medline Kit Number/SKU DYNJ56917J: UDI/DI each 10193489991994, UDI/DI case 40193489991995, Lot Number: 22FDB151; Medline Kit Number/SKU DYNJ56917J: UDI/DI each 10193489991994, UDI/DI case 40193489991995, Lot Number: 22EDB139; Medline Kit Number/SKU DYNJ56917J: UDI/DI each 10193489991994, UDI/DI case 40193489991995, Lot Number: 22DDA699; Medline Kit Number/SKU DYNJ56917J: UDI/DI each 10193489991994, UDI/DI case 40193489991995, Lot Number: 22BDA826; Medline Kit Number/SKU DYNJ56917J: UDI/DI each 10193489991994, UDI/DI case 40193489991995, Lot Number: 21KDB492; Medline Kit Number/SKU DYNJ56917J: UDI/DI each 10193489991994, UDI/DI case 40193489991995, Lot Number: 21IDB626; Medline Kit Number/SKU DYNJ56917K: UDI/DI each 10195327322571, UDI/DI case 40195327322572, Lot Number: 23EDC310; Medline Kit Number/SKU DYNJ56917K: UDI/DI each 10195327322571, UDI/DI case 40195327322572, Lot Number: 23DDA529; Medline Kit Number/SKU DYNJ56917K: UDI/DI each 10195327322571, UDI/DI case 40195327322572, Lot Number: 23BDC220; Medline Kit Number/SKU DYNJ56917K: UDI/DI each 10195327322571, UDI/DI case 40195327322572, Lot Number: 23ADB232; Medline Kit Number/SKU DYNJ56917L: UDI/DI each 10195327466855, UDI/DI case 40195327466856, Lot Number: 23KDB950; Medline Kit Number/SKU DYNJ56917L: UDI/DI each 10195327466855, UDI/DI case 40195327466856, Lot Number: 23JDB895; Medline Kit Number/SKU DYNJ56917L: UDI/DI each 10195327466855, UDI/DI case 40195327466856, Lot Number: 24BDB140; Medline Kit Number/SKU DYNJ56917N: UDI/DI each 10198459140488, UDI/DI case 40198459140489, Lot Number: 25IMB406; Medline Kit Number/SKU DYNJ56917N: UDI/DI each 10198459140488, UDI/DI case 40198459140489, Lot Number: 25EMA512; Medline Kit Number/SKU DYNJ56917N: UDI/DI each 10198459140488, UDI/DI case 40198459140489, Lot Number: 25DMG834; Medline Kit Number/SKU DYNJ56917N: UDI/DI each 10198459140488, UDI/DI case 40198459140489, Lot Number: 25CMC310; Medline Kit Number/SKU DYNJ56917N: UDI/DI each 10198459140488, UDI/DI case 40198459140489, Lot Number: 25BMC774; Medline Kit Number/SKU DYNJ56917N: UDI/DI each 10198459140488, UDI/DI case 40198459140489, Lot Number: 25BMC034; Medline Kit Number/SKU DYNJ56917N: UDI/DI each 10198459140488, UDI/DI case 40198459140489, Lot Number: 24JDB128; Medline Kit Number/SKU DYNJ58324A: UDI/DI each 10193489485639, UDI/DI case 40193489485630, Lot Number: 23FBS890; Medline Kit Number/SKU DYNJ58324A: UDI/DI each 10193489485639, UDI/DI case 40193489485630, Lot Number: 23DBO019; Medline Kit Number/SKU DYNJ58324A: UDI/DI each 10193489485639, UDI/DI case 40193489485630, Lot Number: 22IBG087; Medline Kit Number/SKU DYNJ58324A: UDI/DI each 10193489485639, UDI/DI case 40193489485630, Lot Number: 22BBZ926; Medline Kit Number/SKU DYNJ58324A: UDI/DI each 10193489485639, UDI/DI case 40193489485630, Lot Number: 21CBO062; Medline Kit Number/SKU DYNJ58324A: UDI/DI each 10193489485639, UDI/DI case 40193489485630, Lot Number: 21CBT626; Medline Kit Number/SKU DYNJ58324A: UDI/DI each 10193489485639, UDI/DI case 40193489485630, Lot Number: 21ABA621; Medline Kit Number/SKU DYNJ58324A: UDI/DI each 10193489485639, UDI/DI case 40193489485630, Lot Number: 25FBB275; Medline Kit Number/SKU DYNJ58324A: UDI/DI each 10193489485639, UDI/DI case 40193489485630, Lot Number: 24HBQ403; Medline Kit Number/SKU DYNJ58324A: UDI/DI each 10193489485639, UDI/DI case 40193489485630, Lot Number: 24HBC985; Medline Kit Number/SKU DYNJ58324A: UDI/DI each 10193489485639, UDI/DI case 40193489485630, Lot Number: 24FBB168; Medline Kit Number/SKU DYNJ58324A: UDI/DI each 10193489485639, UDI/DI case 40193489485630, Lot Number: 24EBK104; Medline Kit Number/SKU DYNJ58324A: UDI/DI each 10193489485639, UDI/DI case 40193489485630, Lot Number: 24CBG181; Medline Kit Number/SKU DYNJ67558A: UDI/DI each 10193489917437, UDI/DI case 40193489917438, Lot Number: 21FBF038; Medline Kit Number/SKU DYNJ67558B: UDI/DI each 10195327099053, UDI/DI case 40195327099054, Lot Number: 22OBG478; Medline Kit Number/SKU DYNJ68213: UDI/DI each 10193489816129, UDI/DI case 40193489816120, Lot Number: 22EMA482; Medline Kit Number/SKU DYNJ68213: UDI/DI each 10193489816129, UDI/DI case 40193489816120, Lot Number: 22CMI811; Medline Kit Number/SKU DYNJ68213: UDI/DI each 10193489816129, UDI/DI case 40193489816120, Lot Number: 22CMG430; Medline Kit Number/SKU DYNJ68213: UDI/DI each 10193489816129, UDI/DI case 40193489816120, Lot Number: 22AMB156; Medline Kit Number/SKU DYNJ68213: UDI/DI each 10193489816129, UDI/DI case 40193489816120, Lot Number: 21KMB789; Medline Kit Number/SKU DYNJ68213: UDI/DI each 10193489816129, UDI/DI case 40193489816120, Lot Number: 21IMG069; Medline Kit Number/SKU DYNJ68213: UDI/DI each 10193489816129, UDI/DI case 40193489816120, Lot Number: 21IMA320; Medline Kit Number/SKU DYNJ68213: UDI/DI each 10193489816129, UDI/DI case 40193489816120, Lot Number: 21GMF680; Medline Kit Number/SKU DYNJ68213: UDI/DI each 10193489816129, UDI/DI case 40193489816120, Lot Number: 21GMC469; Medline Kit Number/SKU DYNJ68213: UDI/DI each 10193489816129, UDI/DI case 40193489816120, Lot Number: 21FMB885; Medline Kit Number/SKU DYNJ68213: UDI/DI each 10193489816129, UDI/DI case 40193489816120, Lot Number: 21EME095; Medline Kit Number/SKU DYNJ68213: UDI/DI each 10193489816129, UDI/DI case 40193489816120, Lot Number: 21DMB893; Medline Kit Number/SKU DYNJ68213: UDI/DI each 10193489816129, UDI/DI case 40193489816120, Lot Number: 21BMC546; Medline Kit Number/SKU DYNJ68213D: UDI/DI each 10198459093395, UDI/DI case 40198459093396, Lot Number: 25FMB744; Medline Kit Number/SKU DYNJ68213D: UDI/DI each 10198459093395, UDI/DI case 40198459093396, Lot Number: 25DMG846; Medline Kit Number/SKU DYNJ68213D: UDI/DI each 10198459093395, UDI/DI case 40198459093396, Lot Number: 25BMB216; Medline Kit Number/SKU DYNJ68213D: UDI/DI each 10198459093395, UDI/DI case 40198459093396, Lot Number: 24KMI278; Medline Kit Number/SKU DYNJ68213D: UDI/DI each 10198459093395, UDI/DI case 40198459093396, Lot Number: 24JMG883; Medline Kit Number/SKU DYNJ68213D: UDI/DI each 10198459093395, UDI/DI case 40198459093396, Lot Number: 24JMC380; Medline Kit Number/SKU DYNJ68213G: UDI/DI each 10198459648175, UDI/DI case 40198459648176, Lot Number: 26AMA296; Medline Kit Number/SKU DYNJ68366B: UDI/DI each 10195327666330, UDI/DI case 40195327666331, Lot Number: 24EMB363; Medline Kit Number/SKU DYNJ68447: UDI/DI each 10193489447095, UDI/DI case 40193489447096, Lot Number: 21CBL485; Medline Kit Number/SKU DYNJ68447A: UDI/DI each 10193489913804, UDI/DI case 40193489913805, Lot Number: 21VBB288; Medline Kit Number/SKU DYNJ68447A: UDI/DI each 10193489913804, UDI/DI case 40193489913805, Lot Number: 21IBV467; Medline Kit Number/SKU DYNJ68447A: UDI/DI each 10193489913804, UDI/DI case 40193489913805, Lot Number: 21IBG628; Medline Kit Number/SKU DYNJ68447A: UDI/DI each 10193489913804, UDI/DI case 40193489913805, Lot Number: 21GBN769; Medline Kit Number/SKU DYNJ68447B: UDI/DI each 10195327034160, UDI/DI case 40195327034161, Lot Number: 22NBF961; Medline Kit Number/SKU DYNJ68447B: UDI/DI each 10195327034160, UDI/DI case 40195327034161, Lot Number: 22ABV162; Medline Kit Number/SKU DYNJ68447B: UDI/DI each 10195327034160, UDI/DI case 40195327034161, Lot Number: 21LBR924; Medline Kit Number/SKU DYNJ69824: UDI/DI each 10193489981698, UDI/DI case 40193489981699, Lot Number: 23JBU091; Medline Kit Number/SKU DYNJ69824: UDI/DI each 10193489981698, UDI/DI case 40193489981699, Lot Number: 23JBC614; Medline Kit Number/SKU DYNJ69824: UDI/DI each 10193489981698, UDI/DI case 40193489981699, Lot Number: 23EBR266; Medline Kit Number/SKU DYNJ69824: UDI/DI each 10193489981698, UDI/DI case 40193489981699, Lot Number: 23EBB687; Medline Kit Number/SKU DYNJ69824: UDI/DI each 10193489981698, UDI/DI case 40193489981699, Lot Number: 23CBV202; Medline Kit Number/SKU DYNJ69824: UDI/DI each 10193489981698, UDI/DI case 40193489981699, Lot Number: 23BBP560; Medline Kit Number/SKU DYNJ69824: UDI/DI each 10193489981698, UDI/DI case 40193489981699, Lot Number: 23ABL869; Medline Kit Number/SKU DYNJ69824: UDI/DI each 10193489981698, UDI/DI case 40193489981699, Lot Number: 23ABI200; Medline Kit Number/SKU DYNJ69824: UDI/DI each 10193489981698, UDI/DI case 40193489981699, Lot Number: 22KBJ128; Medline Kit Number/SKU DYNJ69824: UDI/DI each 10193489981698, UDI/DI case 40193489981699, Lot Number: 22IBU121; Medline Kit Number/SKU DYNJ69824: UDI/DI each 10193489981698, UDI/DI case 40193489981699, Lot Number: 22HBY433; Medline Kit Number/SKU DYNJ69824: UDI/DI each 10193489981698, UDI/DI case 40193489981699, Lot Number: 22GBR086; Medline Kit Number/SKU DYNJ69824: UDI/DI each 10193489981698, UDI/DI case 40193489981699, Lot Number: 22GBQ113; Medline Kit Number/SKU DYNJ69824: UDI/DI each 10193489981698, UDI/DI case 40193489981699, Lot Number: 22FBP609; Medline Kit Number/SKU DYNJ69824: UDI/DI each 10193489981698, UDI/DI case 40193489981699, Lot Number: 22BBU834; Medline Kit Number/SKU DYNJ69824: UDI/DI each 10193489981698, UDI/DI case 40193489981699, Lot Number: 25FBB994; Medline Kit Number/SKU DYNJ69824: UDI/DI each 10193489981698, UDI/DI case 40193489981699, Lot Number: 25DBQ190; Medline Kit Number/SKU DYNJ69824: UDI/DI each 10193489981698, UDI/DI case 40193489981699, Lot Number: 25CBN006; Medline Kit Number/SKU DYNJ69824: UDI/DI each 10193489981698, UDI/DI case 40193489981699, Lot Number: 25ABQ389; Medline Kit Number/SKU DYNJ69824: UDI/DI each 10193489981698, UDI/DI case 40193489981699, Lot Number: 24KBO095; Medline Kit Number/SKU DYNJ69824: UDI/DI each 10193489981698, UDI/DI case 40193489981699, Lot Number: 24JBQ827; Medline Kit Number/SKU DYNJ69824: UDI/DI each 10193489981698, UDI/DI case 40193489981699, Lot Number: 24IBH741; Medline Kit Number/SKU DYNJ69824: UDI/DI each 10193489981698, UDI/DI case 40193489981699, Lot Number: 24HBO248; Medline Kit Number/SKU DYNJ69824: UDI/DI each 10193489981698, UDI/DI case 40193489981699, Lot Number: 24GBK688; Medline Kit Number/SKU DYNJ69824: UDI/DI each 10193489981698, UDI/DI case 40193489981699, Lot Number: 24GBE944; Medline Kit Number/SKU DYNJ69824: UDI/DI each 10193489981698, UDI/DI case 40193489981699, Lot Number: 24FBQ277; Medline Kit Number/SKU DYNJ69824: UDI/DI each 10193489981698, UDI/DI case 40193489981699, Lot Number: 24EBO557; Medline Kit Number/SKU DYNJ69824: UDI/DI each 10193489981698, UDI/DI case 40193489981699, Lot Number: 24EBB354; Medline Kit Number/SKU DYNJ69824: UDI/DI each 10193489981698, UDI/DI case 40193489981699, Lot Number: 24BBU070; Medline Kit Number/SKU DYNJ69824: UDI/DI each 10193489981698, UDI/DI case 40193489981699, Lot Number: 23LBT175; Medline Kit Number/SKU DYNJ81325: UDI/DI each 10195327111014, UDI/DI case 40195327111015, Lot Number: 22JBC490; Medline Kit Number/SKU DYNJ81325: UDI/DI each 10195327111014, UDI/DI case 40195327111015, Lot Number: 22GBW885; Medline Kit Number/SKU DYNJ81325: UDI/DI each 10195327111014, UDI/DI case 40195327111015, Lot Number: 22EBA478; Medline Kit Number/SKU DYNJ81325A: UDI/DI each 10195327270360, UDI/DI case 40195327270361, Lot Number: 23KBD697; Medline Kit Number/SKU DYNJ81325A: UDI/DI each 10195327270360, UDI/DI case 40195327270361, Lot Number: 23JBC513; Medline Kit Number/SKU DYNJ81325A: UDI/DI each 10195327270360, UDI/DI case 40195327270361, Lot Number: 23IBD507; Medline Kit Number/SKU DYNJ81325A: UDI/DI each 10195327270360, UDI/DI case 40195327270361, Lot Number: 23FBS833; Medline Kit Number/SKU DYNJ81325A: UDI/DI each 10195327270360, UDI/DI case 40195327270361, Lot Number: 23DBS310; Medline Kit Number/SKU DYNJ81325A: UDI/DI each 10195327270360, UDI/DI case 40195327270361, Lot Number: 23CBW627; Medline Kit Number/SKU DYNJ81325A: UDI/DI each 10195327270360, UDI/DI case 40195327270361, Lot Number: 23BBO729; Medline Kit Number/SKU DYNJ81325A: UDI/DI each 10195327270360, UDI/DI case 40195327270361, Lot Number: 22KBT451; Medline Kit Number/SKU DYNJ81325A: UDI/DI each 10195327270360, UDI/DI case 40195327270361, Lot Number: 22KBD178; Medline Kit Number/SKU DYNJ81325A: UDI/DI each 10195327270360, UDI/DI case 40195327270361, Lot Number: 24KBA572; Medline Kit Number/SKU DYNJ81325A: UDI/DI each 10195327270360, UDI/DI case 40195327270361, Lot Number: 24JBH042; Medline Kit Number/SKU DYNJ81325A: UDI/DI each 10195327270360, UDI/DI case 40195327270361, Lot Number: 24IBT346; Medline Kit Number/SKU DYNJ81325A: UDI/DI each 10195327270360, UDI/DI case 40195327270361, Lot Number: 24FBJ124; Medline Kit Number/SKU DYNJ81325A: UDI/DI each 10195327270360, UDI/DI case 40195327270361, Lot Number: 24EBN022; Medline Kit Number/SKU DYNJ81325A: UDI/DI each 10195327270360, UDI/DI case 40195327270361, Lot Number: 24DBL535; Medline Kit Number/SKU DYNJ81325A: UDI/DI each 10195327270360, UDI/DI case 40195327270361, Lot Number: 24BBS543; Medline Kit Number/SKU DYNJ81325A: UDI/DI each 10195327270360, UDI/DI case 40195327270361, Lot Number: 24ABN491; Medline Kit Number/SKU DYNJ81325A: UDI/DI each 10195327270360, UDI/DI case 40195327270361, Lot Number: 23LBF176; Medline Kit Number/SKU DYNJ81325B: UDI/DI each 10198459217029, UDI/DI case 40198459217020, Lot Number: 25FBD987; Medline Kit Number/SKU DYNJ81325B: UDI/DI each 10198459217029, UDI/DI case 40198459217020, Lot Number: 25EBD835; Medline Kit Number/SKU DYNJ81325B: UDI/DI each 10198459217029, UDI/DI case 40198459217020, Lot Number: 25BBE525; Medline Kit Number/SKU DYNJ81751: UDI/DI each 10195327139315, UDI/DI case 40195327139316, Lot Number: 23ABA516; Medline Kit Number/SKU DYNJ81751: UDI/DI each 10195327139315, UDI/DI case 40195327139316, Lot Number: 22JBE441; Medline Kit Number/SKU DYNJ81751: UDI/DI each 10195327139315, UDI/DI case 40195327139316, Lot Number: 22JBD573; Medline Kit Number/SKU DYNJ81751: UDI/DI each 10195327139315, UDI/DI case 40195327139316, Lot Number: 22GBX010; Medline Kit Number/SKU DYNJ81751A: UDI/DI each 10195327345112, UDI/DI case 40195327345113, Lot Number: 23JBW345; Medline Kit Number/SKU DYNJ81751A: UDI/DI each 10195327345112, UDI/DI case 40195327345113, Lot Number: 23JBG162; Medline Kit Number/SKU DYNJ81751A: UDI/DI each 10195327345112, UDI/DI case 40195327345113, Lot Number: 23FBA908; Medline Kit Number/SKU DYNJ81751A: UDI/DI each 10195327345112, UDI/DI case 40195327345113, Lot Number: 23EBW504; Medline Kit Number/SKU DYNJ81751A: UDI/DI each 10195327345112, UDI/DI case 40195327345113, Lot Number: 26ABU708; Medline Kit Number/SKU DYNJ81751A: UDI/DI each 10195327345112, UDI/DI case 40195327345113, Lot Number: 25LBM684; Medline Kit Number/SKU DYNJ81751A: UDI/DI each 10195327345112, UDI/DI case 40195327345113, Lot Number: 25KBQ995; Medline Kit Number/SKU DYNJ81751A: UDI/DI each 10195327345112, UDI/DI case 40195327345113, Lot Number: 25IBV639; Medline Kit Number/SKU DYNJ81751A: UDI/DI each 10195327345112, UDI/DI case 40195327345113, Lot Number: 25IBU643; Medline Kit Number/SKU DYNJ81751A: UDI/DI each 10195327345112, UDI/DI case 40195327345113, Lot Number: 25HBT909; Medline Kit Number/SKU DYNJ81751A: UDI/DI each 10195327345112, UDI/DI case 40195327345113, Lot Number: 25FBG513; Medline Kit Number/SKU DYNJ81751A: UDI/DI each 10195327345112, UDI/DI case 40195327345113, Lot Number: 25CBQ697; Medline Kit Number/SKU DYNJ81751A: UDI/DI each 10195327345112, UDI/DI case 40195327345113, Lot Number: 25ABE335; Medline Kit Number/SKU DYNJ81751A: UDI/DI each 10195327345112, UDI/DI case 40195327345113, Lot Number: 24KBE398; Medline Kit Number/SKU DYNJ81751A: UDI/DI each 10195327345112, UDI/DI case 40195327345113, Lot Number: 24JBD820; Medline Kit Number/SKU DYNJ81751A: UDI/DI each 10195327345112, UDI/DI case 40195327345113, Lot Number: 24EBU640; Medline Kit Number/SKU DYNJ81751A: UDI/DI each 10195327345112, UDI/DI case 40195327345113, Lot Number: 24DBT114; Medline Kit Number/SKU DYNJ81751A: UDI/DI each 10195327345112, UDI/DI case 40195327345113, Lot Number: 24ABO861; Medline Kit Number/SKU DYNJ81751A: UDI/DI each 10195327345112, UDI/DI case 40195327345113, Lot Number: 23LBJ179; Medline Kit Number/SKU DYNJ85359: UDI/DI each 10195327493158, UDI/DI case 40195327493159, Lot Number: 23KBM252; Medline Kit Number/SKU DYNJ85359: UDI/DI each 10195327493158, UDI/DI case 40195327493159, Lot Number: 23HBX929; Medline Kit Number/SKU DYNJ85359: UDI/DI each 10195327493158, UDI/DI case 40195327493159, Lot Number: 26CBD685; Medline Kit Number/SKU DYNJ85359: UDI/DI each 10195327493158, UDI/DI case 40195327493159, Lot Number: 26ABE085; Medline Kit Number/SKU DYNJ85359: UDI/DI each 10195327493158, UDI/DI case 40195327493159, Lot Number: 25KBN912; Medline Kit Number/SKU DYNJ85359: UDI/DI each 10195327493158, UDI/DI case 40195327493159, Lot Number: 25JBI743; Medline Kit Number/SKU DYNJ85359: UDI/DI each 10195327493158, UDI/DI case 40195327493159, Lot Number: 25GBP180; Medline Kit Number/SKU DYNJ85359: UDI/DI each 10195327493158, UDI/DI case 40195327493159, Lot Number: 25FBL455; Medline Kit Number/SKU DYNJ85359: UDI/DI each 10195327493158, UDI/DI case 40195327493159, Lot Number: 25FBA096; Medline Kit Number/SKU DYNJ85359: UDI/DI each 10195327493158, UDI/DI case 40195327493159, Lot Number: 25DBR251; Medline Kit Number/SKU DYNJ85359: UDI/DI each 10195327493158, UDI/DI case 40195327493159, Lot Number: 25BBT099; Medline Kit Number/SKU DYNJ85359: UDI/DI each 10195327493158, UDI/DI case 40195327493159, Lot Number: 24LBE877; Medline Kit Number/SKU DYNJ85359: UDI/DI each 10195327493158, UDI/DI case 40195327493159, Lot Number: 24KBL906; Medline Kit Number/SKU DYNJ85359: UDI/DI each 10195327493158, UDI/DI case 40195327493159, Lot Number: 24IBF438; Medline Kit Number/SKU DYNJ85359: UDI/DI each 10195327493158, UDI/DI case 40195327493159, Lot Number: 24HBQ248; Medline Kit Number/SKU DYNJ85359: UDI/DI each 10195327493158, UDI/DI case 40195327493159, Lot Number: 24GBO711; Medline Kit Number/SKU DYNJ85359: UDI/DI each 10195327493158, UDI/DI case 40195327493159, Lot Number: 24FBS345; Medline Kit Number/SKU DYNJ85359: UDI/DI each 10195327493158, UDI/DI case 40195327493159, Lot Number: 24DBA574; Medline Kit Number/SKU DYNJ85359: UDI/DI each 10195327493158, UDI/DI case 40195327493159, Lot Number: 24ABN402; Medline Kit Number/SKU DYNJ85359: UDI/DI each 10195327493158, UDI/DI case 40195327493159, Lot Number: 23LBK140; Medline Kit Number/SKU DYNJ86125: UDI/DI each 10195327596002, UDI/DI case 40195327596003, Lot Number: 25GBA828; Medline Kit Number/SKU DYNJ86125: UDI/DI each 10195327596002, UDI/DI case 40195327596003, Lot Number: 25EBF285; Medline Kit Number/SKU DYNJ86125: UDI/DI each 10195327596002, UDI/DI case 40195327596003, Lot Number: 25CBU143; Medline Kit Number/SKU DYNJ86125: UDI/DI each 10195327596002, UDI/DI case 40195327596003, Lot Number: 25ABG027; Medline Kit Number/SKU DYNJ86125: UDI/DI each 10195327596002, UDI/DI case 40195327596003, Lot Number: 24KBN578; Medline Kit Number/SKU DYNJ86125: UDI/DI each 10195327596002, UDI/DI case 40195327596003, Lot Number: 24HBP866; Medline Kit Number/SKU DYNJ86125: UDI/DI each 10195327596002, UDI/DI case 40195327596003, Lot Number: 24FBJ095; Medline Kit Number/SKU DYNJ86125: UDI/DI each 10195327596002, UDI/DI case 40195327596003, Lot Number: 24DBG675; Medline Kit Number/SKU DYNJ86125: UDI/DI each 10195327596002, UDI/DI case 40195327596003, Lot Number: 24CBL699; Medline Kit Number/SKU DYNJ86125A: UDI/DI each 10198459473852, UDI/DI case 40198459473853, Lot Number: 26AMC689; Medline Kit Number/SKU DYNJ86125A: UDI/DI each 10198459473852, UDI/DI case 40198459473853, Lot Number: 25LMD206; Medline Kit Number/SKU DYNJ86125A: UDI/DI each 10198459473852, UDI/DI case 40198459473853, Lot Number: 25JME672; Medline Kit Number/SKU DYNJ88448: UDI/DI each 10198459161322, UDI/DI case 40198459161323, Lot Number: 25JBC138; Medline Kit Number/SKU DYNJ88448: UDI/DI each 10198459161322, UDI/DI case 40198459161323, Lot Number: 24JBP409; Medline Kit Number/SKU DYNJ88470: UDI/DI each 10198459161889, UDI/DI case 40198459161880, Lot Number: 25LMJ318; Medline Kit Number/SKU DYNJ88470: UDI/DI each 10198459161889, UDI/DI case 40198459161880, Lot Number: 25KMF373; Medline Kit Number/SKU DYNJ88470: UDI/DI each 10198459161889, UDI/DI case 40198459161880, Lot Number: 25JMC356; Medline Kit Number/SKU DYNJ88470: UDI/DI each 10198459161889, UDI/DI case 40198459161880, Lot Number: 25IMA237; Medline Kit Number/SKU DYNJ88470: UDI/DI each 10198459161889, UDI/DI case 40198459161880, Lot Number: 24KMA470; Medline Kit Number/SKU DYNJ901572O: UDI/DI each 10193489818376, UDI/DI case 40193489818377, Lot Number: 22LMJ185; Medline Kit Number/SKU DYNJ901572O: UDI/DI each 10193489818376, UDI/DI case 40193489818377, Lot Number: 22KMH569; Medline Kit Number/SKU DYNJ901572O: UDI/DI each 10193489818376, UDI/DI case 40193489818377, Lot Number: 22JMH343; Medline Kit Number/SKU DYNJ901572O: UDI/DI each 10193489818376, UDI/DI case 40193489818377, Lot Number: 22JME988; Medline Kit Number/SKU DYNJ901572O: UDI/DI each 10193489818376, UDI/DI case 40193489818377, Lot Number: 22JMA764; Medline Kit Number/SKU DYNJ901572O: UDI/DI each 10193489818376, UDI/DI case 40193489818377, Lot Number: 22IMG572; Medline Kit Number/SKU DYNJ901572O: UDI/DI each 10193489818376, UDI/DI case 40193489818377, Lot Number: 22GBD794; Medline Kit Number/SKU DYNJ901572O: UDI/DI each 10193489818376, UDI/DI case 40193489818377, Lot Number: 22EBB404; Medline Kit Number/SKU DYNJ901572O: UDI/DI each 10193489818376, UDI/DI case 40193489818377, Lot Number: 22EBC306; Medline Kit Number/SKU DYNJ901572O: UDI/DI each 10193489818376, UDI/DI case 40193489818377, Lot Number: 22DBI454; Medline Kit Number/SKU DYNJ901572O: UDI/DI each 10193489818376, UDI/DI case 40193489818377, Lot Number: 22CBQ232; Medline Kit Number/SKU DYNJ901572O: UDI/DI each 10193489818376, UDI/DI case 40193489818377, Lot Number: 22ABG615; Medline Kit Number/SKU DYNJ901572O: UDI/DI each 10193489818376, UDI/DI case 40193489818377, Lot Number: 21LBN116; Medline Kit Number/SKU DYNJ901572O: UDI/DI each 10193489818376, UDI/DI case 40193489818377, Lot Number: 21JBZ349; Medline Kit Number/SKU DYNJ901572O: UDI/DI each 10193489818376, UDI/DI case 40193489818377, Lot Number: 21JBM717; Medline Kit Number/SKU DYNJ901572O: UDI/DI each 10193489818376, UDI/DI case 40193489818377, Lot Number: 21JBH386; Medline Kit Number/SKU DYNJ901572O: UDI/DI each 10193489818376, UDI/DI case 40193489818377, Lot Number: 21JBE452; Medline Kit Number/SKU DYNJ901572O: UDI/DI each 10193489818376, UDI/DI case 40193489818377, Lot Number: 21GBG956; Medline Kit Number/SKU DYNJ901572O: UDI/DI each 10193489818376, UDI/DI case 40193489818377, Lot Number: 21EBI236; Medline Kit Number/SKU DYNJ901572O: UDI/DI each 10193489818376, UDI/DI case 40193489818377, Lot Number: 21DBP367; Medline Kit Number/SKU DYNJ901572O: UDI/DI each 10193489818376, UDI/DI case 40193489818377, Lot Number: 21BBC494; Medline Kit Number/SKU DYNJ902985F: UDI/DI each 10193489490862, UDI/DI case 40193489490863, Lot Number: 23KMB541; Medline Kit Number/SKU DYNJ902985F: UDI/DI each 10193489490862, UDI/DI case 40193489490863, Lot Number: 23IMI202; Medline Kit Number/SKU DYNJ902985F: UDI/DI each 10193489490862, UDI/DI case 40193489490863, Lot Number: 22JMG646; Medline Kit Number/SKU DYNJ902985F: UDI/DI each 10193489490862, UDI/DI case 40193489490863, Lot Number: 22EMF849; Medline Kit Number/SKU DYNJ902985F: UDI/DI each 10193489490862, UDI/DI case 40193489490863, Lot Number: 22EMD560; Medline Kit Number/SKU DYNJ902985F: UDI/DI each 10193489490862, UDI/DI case 40193489490863, Lot Number: 22BME333; Medline Kit Number/SKU DYNJ902985F: UDI/DI each 10193489490862, UDI/DI case 40193489490863, Lot Number: 21JMA467; Medline Kit Number/SKU DYNJ902985F: UDI/DI each 10193489490862, UDI/DI case 40193489490863, Lot Number: 21HME157; Medline Kit Number/SKU DYNJ902985F: UDI/DI each 10193489490862, UDI/DI case 40193489490863, Lot Number: 21FMB959; Medline Kit Number/SKU DYNJ902985F: UDI/DI each 10193489490862, UDI/DI case 40193489490863, Lot Number: 21DKA711; Medline Kit Number/SKU DYNJ902985F: UDI/DI each 10193489490862, UDI/DI case 40193489490863, Lot Number: 21DKA257; Medline Kit Number/SKU DYNJ902985F: UDI/DI each 10193489490862, UDI/DI case 40193489490863, Lot Number: 21AKA915; Medline Kit Number/SKU DYNJ902985F: UDI/DI each 10193489490862, UDI/DI case 40193489490863, Lot Number: 23LMG587; Medline Kit Number/SKU DYNJ905366B: UDI/DI each 10195327048068, UDI/DI case 40195327048069, Lot Number: 23EBM739; Medline Kit Number/SKU DYNJ905366B: UDI/DI each 10195327048068, UDI/DI case 40195327048069, Lot Number: 23DBM835; Medline Kit Number/SKU DYNJ905366B: UDI/DI each 10195327048068, UDI/DI case 40195327048069, Lot Number: 23CBE807; Medline Kit Number/SKU DYNJ905366B: UDI/DI each 10195327048068, UDI/DI case 40195327048069, Lot Number: 22LBO631; Medline Kit Number/SKU DYNJ905366B: UDI/DI each 10195327048068, UDI/DI case 40195327048069, Lot Number: 22HBM673; Medline Kit Number/SKU DYNJ905366B: UDI/DI each 10195327048068, UDI/DI case 40195327048069, Lot Number: 22GBE117; Medline Kit Number/SKU DYNJ905366B: UDI/DI each 10195327048068, UDI/DI case 40195327048069, Lot Number: 22GBA660; Medline Kit Number/SKU DYNJ905366B: UDI/DI each 10195327048068, UDI/DI case 40195327048069, Lot Number: 22DBR510; Medline Kit Number/SKU DYNJ905366B: UDI/DI each 10195327048068, UDI/DI case 40195327048069, Lot Number: 25DBG803; Medline Kit Number/SKU DYNJ905366B: UDI/DI each 10195327048068, UDI/DI case 40195327048069, Lot Number: 25ABJ890; Medline Kit Number/SKU DYNJ905366B: UDI/DI each 10195327048068, UDI/DI case 40195327048069, Lot Number: 25ABF688; Medline Kit Number/SKU DYNJ905366B: UDI/DI each 10195327048068, UDI/DI case 40195327048069, Lot Number: 24JBP878; Medline Kit Number/SKU DYNJ905366B: UDI/DI each 10195327048068, UDI/DI case 40195327048069, Lot Number: 24HBB373; Medline Kit Number/SKU DYNJ905366B: UDI/DI each 10195327048068, UDI/DI case 40195327048069, Lot Number: 24GBH830; Medline Kit Number/SKU DYNJ905366B: UDI/DI each 10195327048068, UDI/DI case 40195327048069, Lot Number: 24EBN295; Medline Kit Number/SKU DYNJ905366B: UDI/DI each 10195327048068, UDI/DI case 40195327048069, Lot Number: 24EBD925; Medline Kit Number/SKU DYNJ905366B: UDI/DI each 10195327048068, UDI/DI case 40195327048069, Lot Number: 24ABT424; Medline Kit Number/SKU DYNJ905366B: UDI/DI each 10195327048068, UDI/DI case 40195327048069, Lot Number: 23LBT823; Medline Kit Number/SKU DYNJ907076C: UDI/DI each 10193489465617, UDI/DI case 40193489465618, Lot Number: 23IBH134; Medline Kit Number/SKU DYNJ907076C: UDI/DI each 10193489465617, UDI/DI case 40193489465618, Lot Number: 23EBW674; Medline Kit Number/SKU DYNJ907076C: UDI/DI each 10193489465617, UDI/DI case 40193489465618, Lot Number: 23DBM339; Medline Kit Number/SKU DYNJ907076C: UDI/DI each 10193489465617, UDI/DI case 40193489465618, Lot Number: 23ABR108; Medline Kit Number/SKU DYNJ907076C: UDI/DI each 10193489465617, UDI/DI case 40193489465618, Lot Number: 23ABM208; Medline Kit Number/SKU DYNJ907076C: UDI/DI each 10193489465617, UDI/DI case 40193489465618, Lot Number: 22HBS616; Medline Kit Number/SKU DYNJ907076C: UDI/DI each 10193489465617, UDI/DI case 40193489465618, Lot Number: 22IBC655; Medline Kit Number/SKU DYNJ907076C: UDI/DI each 10193489465617, UDI/DI case 40193489465618, Lot Number: 22GBG134; Medline Kit Number/SKU DYNJ907076C: UDI/DI each 10193489465617, UDI/DI case 40193489465618, Lot Number: 21KBV892; Medline Kit Number/SKU DYNJ907076C: UDI/DI each 10193489465617, UDI/DI case 40193489465618, Lot Number: 21JBS139; Medline Kit Number/SKU DYNJ907076C: UDI/DI each 10193489465617, UDI/DI case 40193489465618, Lot Number: 24DBS653; Medline Kit Number/SKU DYNJ907076C: UDI/DI each 10193489465617, UDI/DI case 40193489465618, Lot Number: 24CBQ078; Medline Kit Number/SKU DYNJ907076C: UDI/DI each 10193489465617, UDI/DI case 40193489465618, Lot Number: 24ABS092; Medline Kit Number/SKU DYNJ908165: UDI/DI each 10193489992014, UDI/DI case 40193489992015, Lot Number: 23KBH043; Medline Kit Number/SKU DYNJ908165: UDI/DI each 10193489992014, UDI/DI case 40193489992015, Lot Number: 23KBH076; Medline Kit Number/SKU DYNJ908165: UDI/DI each 10193489992014, UDI/DI case 40193489992015, Lot Number: 23IBO315; Medline Kit Number/SKU DYNJ908165: UDI/DI each 10193489992014, UDI/DI case 40193489992015, Lot Number: 23FBA102; Medline Kit Number/SKU DYNJ908165: UDI/DI each 10193489992014, UDI/DI case 40193489992015, Lot Number: 23BBL712; Medline Kit Number/SKU DYNJ908165: UDI/DI each 10193489992014, UDI/DI case 40193489992015, Lot Number: 23BBG445; Medline Kit Number/SKU DYNJ908165: UDI/DI each 10193489992014, UDI/DI case 40193489992015, Lot Number: 23BBF440; Medline Kit Number/SKU DYNJ908165: UDI/DI each 10193489992014, UDI/DI case 40193489992015, Lot Number: 23BBD054; Medline Kit Number/SKU DYNJ908165: UDI/DI each 10193489992014, UDI/DI case 40193489992015, Lot Number: 22JBI497; Medline Kit Number/SKU DYNJ908165: UDI/DI each 10193489992014, UDI/DI case 40193489992015, Lot Number: 22FLA058; Medline Kit Number/SKU DYNJ908165: UDI/DI each 10193489992014, UDI/DI case 40193489992015, Lot Number: 22CLA782; Medline Kit Number/SKU DYNJ908165: UDI/DI each 10193489992014, UDI/DI case 40193489992015, Lot Number: 22CLA236; Medline Kit Number/SKU DYNJ908165: UDI/DI each 10193489992014, UDI/DI case 40193489992015, Lot Number: 23LBG978; Medline Kit Number/SKU DYNJ908165A: UDI/DI each 10195327570958, UDI/DI case 40195327570959, Lot Number: 25LBE981; Medline Kit Number/SKU DYNJ908165A: UDI/DI each 10195327570958, UDI/DI case 40195327570959, Lot Number: 25JBS925; Medline Kit Number/SKU DYNJ908165A: UDI/DI each 10195327570958, UDI/DI case 40195327570959, Lot Number: 25IBS527; Medline Kit Number/SKU DYNJ908165A: UDI/DI each 10195327570958, UDI/DI case 40195327570959, Lot Number: 25IBC316; Medline Kit Number/SKU DYNJ908165A: UDI/DI each 10195327570958, UDI/DI case 40195327570959, Lot Number: 25EBR807; Medline Kit Number/SKU DYNJ908165A: UDI/DI each 10195327570958, UDI/DI case 40195327570959, Lot Number: 25DBE206; Medline Kit Number/SKU DYNJ908165A: UDI/DI each 10195327570958, UDI/DI case 40195327570959, Lot Number: 25ABF294; Medline Kit Number/SKU DYNJ908165A: UDI/DI each 10195327570958, UDI/DI case 40195327570959, Lot Number: 24IBV349; Medline Kit Number/SKU DYNJ908165A: UDI/DI each 10195327570958, UDI/DI case 40195327570959, Lot Number: 24GBQ643; Medline Kit Number/SKU DYNJ908165A: UDI/DI each 10195327570958, UDI/DI case 40195327570959, Lot Number: 24EBT890; Medline Kit Number/SKU DYNJ908165A: UDI/DI each 10195327570958, UDI/DI case 40195327570959, Lot Number: 24EBH261; Medline Kit Number/SKU DYNJ908165A: UDI/DI each 10195327570958, UDI/DI case 40195327570959, Lot Number: 24EBC018; Medline Kit Number/SKU DYNJ908165A: UDI/DI each 10195327570958, UDI/DI case 40195327570959, Lot Number: 24DBG691; Medline Kit Number/SKU DYNJ908165A: UDI/DI each 10195327570958, UDI/DI case 40195327570959, Lot Number: 24ABD445; Medline Kit Number/SKU DYNJ909512: UDI/DI each 10195327390976, UDI/DI case 40195327390977, Lot Number: 23FBA333; Medline Kit Number/SKU DYNJ909512: UDI/DI each 10195327390976, UDI/DI case 40195327390977, Lot Number: 23EBF184; Medline Kit Number/SKU DYNJ909512: UDI/DI each 10195327390976, UDI/DI case 40195327390977, Lot Number: 26CBF295; Medline Kit Number/SKU DYNJ909512: UDI/DI each 10195327390976, UDI/DI case 40195327390977, Lot Number: 25IBT900; Medline Kit Number/SKU DYNJ909512: UDI/DI each 10195327390976, UDI/DI case 40195327390977, Lot Number: 25IBD206; Medline Kit Number/SKU DYNJ909512: UDI/DI each 10195327390976, UDI/DI case 40195327390977, Lot Number: 25GBR222; Medline Kit Number/SKU DYNJ909512: UDI/DI each 10195327390976, UDI/DI case 40195327390977, Lot Number: 25DBS691; Medline Kit Number/SKU DYNJ909512: UDI/DI each 10195327390976, UDI/DI case 40195327390977, Lot Number: 25ABQ137; Medline Kit Number/SKU DYNJ909512: UDI/DI each 10195327390976, UDI/DI case 40195327390977, Lot Number: 25ABJ070; Medline Kit Number/SKU DYNJ909512: UDI/DI each 10195327390976, UDI/DI case 40195327390977, Lot Number: 24IBP825; Medline Kit Number/SKU DYNJ909512: UDI/DI each 10195327390976, UDI/DI case 40195327390977, Lot Number: 24CBT484; Medline Kit Number/SKU DYNJ909512: UDI/DI each 10195327390976, UDI/DI case 40195327390977, Lot Number: 23LBD654; Medline Kit Number/SKU DYNJ910120A: UDI/DI each 10195327689872, UDI/DI case 40195327689873, Lot Number: 25LMG501; Medline Kit Number/SKU DYNJ910120A: UDI/DI each 10195327689872, UDI/DI case 40195327689873, Lot Number: 25JMG389; Medline Kit Number/SKU DYNJ910120A: UDI/DI each 10195327689872, UDI/DI case 40195327689873, Lot Number: 25IMJ182; Medline Kit Number/SKU DYNJ910120A: UDI/DI each 10195327689872, UDI/DI case 40195327689873, Lot Number: 25HMC859; Medline Kit Number/SKU DYNJ910120A: UDI/DI each 10195327689872, UDI/DI case 40195327689873, Lot Number: 25GMD539; Medline Kit Number/SKU DYNJ910120A: UDI/DI each 10195327689872, UDI/DI case 40195327689873, Lot Number: 25EME174; Medline Kit Number/SKU DYNJ910120A: UDI/DI each 10195327689872, UDI/DI case 40195327689873, Lot Number: 25DME221; Medline Kit Number/SKU DYNJ910120A: UDI/DI each 10195327689872, UDI/DI case 40195327689873, Lot Number: 25BME721; Medline Kit Number/SKU DYNJ910120A: UDI/DI each 10195327689872, UDI/DI case 40195327689873, Lot Number: 24LMB897; Medline Kit Number/SKU DYNJ910120A: UDI/DI each 10195327689872, UDI/DI case 40195327689873, Lot Number: 24JMI715; Medline Kit Number/SKU DYNJ911086: UDI/DI each 10198459248757, UDI/DI case 40198459248758, Lot Number: 25EBI830; Medline Kit Number/SKU DYNJ911086: UDI/DI each 10198459248757, UDI/DI case 40198459248758, Lot Number: 25CBO097; Medline Kit Number/SKU DYNJ940735F: UDI/DI each 10195327129866, UDI/DI case 40195327129867, Lot Number: 23DBF803; Medline Kit Number/SKU DYNJ940735F: UDI/DI each 10195327129866, UDI/DI case 40195327129867, Lot Number: 23CBN885; Medline Kit Number/SKU DYNJ940735F: UDI/DI each 10195327129866, UDI/DI case 40195327129867, Lot Number: 23BBT129; Medline Kit Number/SKU DYNJ940735F: UDI/DI each 10195327129866, UDI/DI case 40195327129867, Lot Number: 23ABC272; Medline Kit Number/SKU DYNJ940735F: UDI/DI each 10195327129866, UDI/DI case 40195327129867, Lot Number: 22LBC489; Medline Kit Number/SKU DYNJ940735F: UDI/DI each 10195327129866, UDI/DI case 40195327129867, Lot Number: 22JBM643; Medline Kit Number/SKU DYNJ940735F: UDI/DI each 10195327129866, UDI/DI case 40195327129867, Lot Number: 22IBH997; Medline Kit Number/SKU DYNJ940735F: UDI/DI each 10195327129866, UDI/DI case 40195327129867, Lot Number: 22HBJ672; Medline Kit Number/SKU DYNJ940735F: UDI/DI each 10195327129866, UDI/DI case 40195327129867, Lot Number: 22GBH649; Medline Kit Number/SKU DYNJ940735F: UDI/DI each 10195327129866, UDI/DI case 40195327129867, Lot Number: 22FBT217; Medline Kit Number/SKU DYNJ940735F: UDI/DI each 10195327129866, UDI/DI case 40195327129867, Lot Number: 22DBK649; Medline Kit Number/SKU DYNJT3101: UDI/DI each 10198459306914, UDI/DI case 40198459306915, Lot Number: 26BBA753; Medline Kit Number/SKU DYNJT3101: UDI/DI each 10198459306914, UDI/DI case 40198459306915, Lot Number: 26ABA848; Medline Kit Number/SKU DYNJT3101: UDI/DI each 10198459306914, UDI/DI case 40198459306915, Lot Number: 25JBE538; Medline Kit Number/SKU DYNJT3101: UDI/DI each 10198459306914, UDI/DI case 40198459306915, Lot Number: 25HBS777; Medline Kit Number/SKU DYNJT3101: UDI/DI each 10198459306914, UDI/DI case 40198459306915, Lot Number: 25FBO606; Medline Kit Number/SKU DYNJT3101: UDI/DI each 10198459306914, UDI/DI case 40198459306915, Lot Number: 25EBR590; Medline Kit Number/SKU DYNJT3243: UDI/DI each 10198459323072, UDI/DI case 40198459323073, Lot Number: 25EBK899A; Medline Kit Number/SKU DYNJT3243: UDI/DI each 10198459323072, UDI/DI case 40198459323073, Lot Number: 25JBR004; Medline Kit Number/SKU DYNJT3243: UDI/DI each 10198459323072, UDI/DI case 40198459323073, Lot Number: 25JBB805; Medline Kit Number/SKU DYNJT3243: UDI/DI each 10198459323072, UDI/DI case 40198459323073, Lot Number: 25HBL978; Medline Kit Number/SKU DYNJT3243: UDI/DI each 10198459323072, UDI/DI case 40198459323073, Lot Number: 25EBK899.

Devices named in this recall

59 devices in the FDA device registry whose identifier appears in the recall's code information.

Other products in the same recall event

The FDA groups related products into one recall event (event 98664).