Why it was recalled

During an internal review, Medline identified higher-than-expected endotoxin levels on a representative Medline neuro sponge product, indicating that in-market product may be out-of-specification for endotoxin.

FDA root cause category: Under Investigation by firm

What the company is doing

Medline industries issued an URGENT MEDICAL DEVICE RECALL notice to its consignees on 3/19/2026 via USPS first class mail and email. The notice explained the issue, potential health risk, and requested the following actions be taken on the affected sponges: Identify and quarantine affected product, complete response form using website link: https://recalls.medline.com, destroy affected product. Credit will be issued once response form is received. If the product has been further distributed, sold, or transferred, notify the customer to destroy the product. Distributors must notify their customers and should include customer quantities on your response form. For consignees of Medline medical convenience kits containing the affected sponges the actions to be taken are as follows: Identify and quarantine all affected kits, complete response form using website link: https://recalls.medline.com. Upon receipt of your submitted response form, your account will receive over-labels to place on affected inventory, with instructions for your staff to remove and discard the affected component prior to using the kit. These labels will be provided via FedEx overnight, if applicable. Distributors must notify their customers and should include customer quantities on your response form. Upon receipt of your submitted response form, you will receive over-labels to provide to your customers to affix to their affected product. If you have any questions, contact the Recall Department at 866-359-1704 or [email protected].

Recall details

Recall number
Z-2028-2026
Status
Open, Classified
Initiated by the company
Mar 19, 2026
Posted by the FDA
Apr 30, 2026
Recalling firm
Medline Industries, LP
Quantity
3000 kits
Product code
LRP
Distribution
Worldwide distribution.

Product description

Centurion medical procedure kits, containing Medline Neuro Sponges, labeled as follows: SURGICAL ENT PREP KIT, Medline Kit Number/SKU MNS11715

Code information (models, lots, serial numbers)

UDI/DI EACH 10653160323068, UDI/DI case 40653160323069. Lot Numbers: 23JBY390, 23IBS518, 23FBW067, 23CBO045, 22HBC645, 22GBL079, 22DBP674, 22ABU694, 21IBM298, 21IBM495, 21GBB532, 21EBR576, 21CBL756, 21ABI942, 26ABG606, 25LBS129, 25IBK286, 25GBB222, 25DBH934, 25BBD132, 24LBO080, 24JBW176, 24HBT740, 24EBU727, 24ABW463.