Medline medical procedure kits, containing Medline Neuro… recallZ-2024-2026 · Medline Industries, LP · initiated Mar 19, 2026

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Medline Industries, LP or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

During an internal review, Medline identified higher-than-expected endotoxin levels on a representative Medline neuro sponge product, indicating that in-market product may be out-of-specification for endotoxin.

FDA root cause category: Under Investigation by firm

What the company is doing

Medline industries issued an URGENT MEDICAL DEVICE RECALL notice to its consignees on 3/19/2026 via USPS first class mail and email. The notice explained the issue, potential health risk, and requested the following actions be taken on the affected sponges: Identify and quarantine affected product, complete response form using website link: https://recalls.medline.com, destroy affected product. Credit will be issued once response form is received. If the product has been further distributed, sold, or transferred, notify the customer to destroy the product. Distributors must notify their customers and should include customer quantities on your response form. For consignees of Medline medical convenience kits containing the affected sponges the actions to be taken are as follows: Identify and quarantine all affected kits, complete response form using website link: https://recalls.medline.com. Upon receipt of your submitted response form, your account will receive over-labels to place on affected inventory, with instructions for your staff to remove and discard the affected component prior to using the kit. These labels will be provided via FedEx overnight, if applicable. Distributors must notify their customers and should include customer quantities on your response form. Upon receipt of your submitted response form, you will receive over-labels to provide to your customers to affix to their affected product. If you have any questions, contact the Recall Department at 866-359-1704 or [email protected].

Recall details

Recall number
Z-2024-2026
Status
Open, Classified
Initiated by the company
Mar 19, 2026
Posted by the FDA
Apr 30, 2026
Recalling firm
Medline Industries, LP
Quantity
2153 kits
Product code
PPB
Distribution
Worldwide distribution.

Product description

Medline medical procedure kits, containing Medline Neuro Sponges, labeled as follows: 1. ENDOSCOPY-LF, Medline Kit Number/SKU DYNJ22890M; 2. ENDOSCOPY PACK, Medline Kit Number/SKU DYNJ40645B; 3. ENT PK 1008847, Medline Kit Number/SKU DYNJ42319C; 4. ENDO SINUS PACK, Medline Kit Number/SKU DYNJ66449; 5. ENT, Medline Kit Number/SKU DYNJ900282I; 6. SEPTO/ENDO, Medline Kit Number/SKU DYNJ902831D; 7. ENDOSCOPIC SKULL BASE, Medline Kit Number/SKU DYNJ906630.

Code information (models, lots, serial numbers)

Show the full code information (7k characters)

Medline Kit Number/SKU DYNJ22890M: UDI/DI each 10193489766486, UDI/DI case 40193489766487, Lot Number: 23DBM600; Medline Kit Number/SKU DYNJ22890M: UDI/DI each 10193489766486, UDI/DI case 40193489766487, Lot Number: 22JBM081; Medline Kit Number/SKU DYNJ22890M: UDI/DI each 10193489766486, UDI/DI case 40193489766487, Lot Number: 22ABJ346; Medline Kit Number/SKU DYNJ22890M: UDI/DI each 10193489766486, UDI/DI case 40193489766487, Lot Number: 21KBO873; Medline Kit Number/SKU DYNJ22890M: UDI/DI each 10193489766486, UDI/DI case 40193489766487, Lot Number: 21KBO877; Medline Kit Number/SKU DYNJ22890M: UDI/DI each 10193489766486, UDI/DI case 40193489766487, Lot Number: 21JBX107; Medline Kit Number/SKU DYNJ22890M: UDI/DI each 10193489766486, UDI/DI case 40193489766487, Lot Number: 21BBE213; Medline Kit Number/SKU DYNJ22890M: UDI/DI each 10193489766486, UDI/DI case 40193489766487, Lot Number: 21BBB024; Medline Kit Number/SKU DYNJ40645B: UDI/DI each 10193489292183, UDI/DI case 40193489292184, Lot Number: 22AMD386; Medline Kit Number/SKU DYNJ40645B: UDI/DI each 10193489292183, UDI/DI case 40193489292184, Lot Number: 21LMC822; Medline Kit Number/SKU DYNJ40645B: UDI/DI each 10193489292183, UDI/DI case 40193489292184, Lot Number: 21JMF409; Medline Kit Number/SKU DYNJ40645B: UDI/DI each 10193489292183, UDI/DI case 40193489292184, Lot Number: 21IMC184; Medline Kit Number/SKU DYNJ40645B: UDI/DI each 10193489292183, UDI/DI case 40193489292184, Lot Number: 21GMB555; Medline Kit Number/SKU DYNJ40645B: UDI/DI each 10193489292183, UDI/DI case 40193489292184, Lot Number: 21EMD626; Medline Kit Number/SKU DYNJ40645B: UDI/DI each 10193489292183, UDI/DI case 40193489292184, Lot Number: 21CMC668; Medline Kit Number/SKU DYNJ40645B: UDI/DI each 10193489292183, UDI/DI case 40193489292184, Lot Number: 21AMB081; Medline Kit Number/SKU DYNJ42319C: UDI/DI each 10193489219722, UDI/DI case 40193489219723, Lot Number: 23FMD067; Medline Kit Number/SKU DYNJ42319C: UDI/DI each 10193489219722, UDI/DI case 40193489219723, Lot Number: 23EMB306; Medline Kit Number/SKU DYNJ42319C: UDI/DI each 10193489219722, UDI/DI case 40193489219723, Lot Number: 23CMD967; Medline Kit Number/SKU DYNJ42319C: UDI/DI each 10193489219722, UDI/DI case 40193489219723, Lot Number: 22KMB384; Medline Kit Number/SKU DYNJ42319C: UDI/DI each 10193489219722, UDI/DI case 40193489219723, Lot Number: 22OBB494; Medline Kit Number/SKU DYNJ42319C: UDI/DI each 10193489219722, UDI/DI case 40193489219723, Lot Number: 22BBT747; Medline Kit Number/SKU DYNJ42319C: UDI/DI each 10193489219722, UDI/DI case 40193489219723, Lot Number: 22ABA711; Medline Kit Number/SKU DYNJ42319C: UDI/DI each 10193489219722, UDI/DI case 40193489219723, Lot Number: 21LBG900; Medline Kit Number/SKU DYNJ42319C: UDI/DI each 10193489219722, UDI/DI case 40193489219723, Lot Number: 21KBD492; Medline Kit Number/SKU DYNJ42319C: UDI/DI each 10193489219722, UDI/DI case 40193489219723, Lot Number: 21EBS624; Medline Kit Number/SKU DYNJ42319C: UDI/DI each 10193489219722, UDI/DI case 40193489219723, Lot Number: 21CBY739; Medline Kit Number/SKU DYNJ42319C: UDI/DI each 10193489219722, UDI/DI case 40193489219723, Lot Number: 21ABL492; Medline Kit Number/SKU DYNJ66449: UDI/DI each 10193489356113, UDI/DI case 40193489356114, Lot Number: 22KBD211; Medline Kit Number/SKU DYNJ66449: UDI/DI each 10193489356113, UDI/DI case 40193489356114, Lot Number: 22JBX995; Medline Kit Number/SKU DYNJ66449: UDI/DI each 10193489356113, UDI/DI case 40193489356114, Lot Number: 22JBL500; Medline Kit Number/SKU DYNJ66449: UDI/DI each 10193489356113, UDI/DI case 40193489356114, Lot Number: 24CBB115; Medline Kit Number/SKU DYNJ900282I: UDI/DI each 10193489798616, UDI/DI case 40193489798617, Lot Number: 22ABB463; Medline Kit Number/SKU DYNJ900282I: UDI/DI each 10193489798616, UDI/DI case 40193489798617, Lot Number: 22ABB461; Medline Kit Number/SKU DYNJ900282I: UDI/DI each 10193489798616, UDI/DI case 40193489798617, Lot Number: 21KBT464; Medline Kit Number/SKU DYNJ900282I: UDI/DI each 10193489798616, UDI/DI case 40193489798617, Lot Number: 21KBP591; Medline Kit Number/SKU DYNJ902831D: UDI/DI each 10193489231670, UDI/DI case 40193489231671, Lot Number: 23EBS895; Medline Kit Number/SKU DYNJ902831D: UDI/DI each 10193489231670, UDI/DI case 40193489231671, Lot Number: 23CBU007; Medline Kit Number/SKU DYNJ902831D: UDI/DI each 10193489231670, UDI/DI case 40193489231671, Lot Number: 23ABE854; Medline Kit Number/SKU DYNJ902831D: UDI/DI each 10193489231670, UDI/DI case 40193489231671, Lot Number: 22LBE910; Medline Kit Number/SKU DYNJ902831D: UDI/DI each 10193489231670, UDI/DI case 40193489231671, Lot Number: 22JBP701; Medline Kit Number/SKU DYNJ902831D: UDI/DI each 10193489231670, UDI/DI case 40193489231671, Lot Number: 22FBV579; Medline Kit Number/SKU DYNJ902831D: UDI/DI each 10193489231670, UDI/DI case 40193489231671, Lot Number: 22BBZ713; Medline Kit Number/SKU DYNJ902831D: UDI/DI each 10193489231670, UDI/DI case 40193489231671, Lot Number: 22BBZ714; Medline Kit Number/SKU DYNJ902831D: UDI/DI each 10193489231670, UDI/DI case 40193489231671, Lot Number: 21LBD683; Medline Kit Number/SKU DYNJ902831D: UDI/DI each 10193489231670, UDI/DI case 40193489231671, Lot Number: 21IBM004; Medline Kit Number/SKU DYNJ902831D: UDI/DI each 10193489231670, UDI/DI case 40193489231671, Lot Number: 21FBJ190; Medline Kit Number/SKU DYNJ902831D: UDI/DI each 10193489231670, UDI/DI case 40193489231671, Lot Number: 21EBE932; Medline Kit Number/SKU DYNJ902831D: UDI/DI each 10193489231670, UDI/DI case 40193489231671, Lot Number: 21DBA613; Medline Kit Number/SKU DYNJ902831D: UDI/DI each 10193489231670, UDI/DI case 40193489231671, Lot Number: 21CBI323; Medline Kit Number/SKU DYNJ906630: UDI/DI each 10193489798852, UDI/DI case 40193489798853, Lot Number: 22FBM378; Medline Kit Number/SKU DYNJ906630: UDI/DI each 10193489798852, UDI/DI case 40193489798853, Lot Number: 22EBE987; Medline Kit Number/SKU DYNJ906630: UDI/DI each 10193489798852, UDI/DI case 40193489798853, Lot Number: 22OBI162; Medline Kit Number/SKU DYNJ906630: UDI/DI each 10193489798852, UDI/DI case 40193489798853, Lot Number: 21KBF463; Medline Kit Number/SKU DYNJ906630: UDI/DI each 10193489798852, UDI/DI case 40193489798853, Lot Number: 21IBU411; Medline Kit Number/SKU DYNJ906630: UDI/DI each 10193489798852, UDI/DI case 40193489798853, Lot Number: 21GBJ251; Medline Kit Number/SKU DYNJ906630: UDI/DI each 10193489798852, UDI/DI case 40193489798853, Lot Number: 21DBU812.

Devices named in this recall

7 devices in the FDA device registry whose identifier appears in the recall's code information.

Other products in the same recall event

The FDA groups related products into one recall event (event 98664).