Medline medical procedure kits, containing Medline Neuro… recallZ-2019-2026 · Medline Industries, LP · initiated Mar 19, 2026

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Medline Industries, LP or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

During an internal review, Medline identified higher-than-expected endotoxin levels on a representative Medline neuro sponge product, indicating that in-market product may be out-of-specification for endotoxin.

FDA root cause category: Under Investigation by firm

What the company is doing

Medline industries issued an URGENT MEDICAL DEVICE RECALL notice to its consignees on 3/19/2026 via USPS first class mail and email. The notice explained the issue, potential health risk, and requested the following actions be taken on the affected sponges: Identify and quarantine affected product, complete response form using website link: https://recalls.medline.com, destroy affected product. Credit will be issued once response form is received. If the product has been further distributed, sold, or transferred, notify the customer to destroy the product. Distributors must notify their customers and should include customer quantities on your response form. For consignees of Medline medical convenience kits containing the affected sponges the actions to be taken are as follows: Identify and quarantine all affected kits, complete response form using website link: https://recalls.medline.com. Upon receipt of your submitted response form, your account will receive over-labels to place on affected inventory, with instructions for your staff to remove and discard the affected component prior to using the kit. These labels will be provided via FedEx overnight, if applicable. Distributors must notify their customers and should include customer quantities on your response form. Upon receipt of your submitted response form, you will receive over-labels to provide to your customers to affix to their affected product. If you have any questions, contact the Recall Department at 866-359-1704 or [email protected].

Recall details

Recall number
Z-2019-2026
Status
Open, Classified
Initiated by the company
Mar 19, 2026
Posted by the FDA
Apr 30, 2026
Recalling firm
Medline Industries, LP
Quantity
2228 kits
Product code
OFX
Distribution
Worldwide distribution.

Product description

Medline medical procedure kits, containing Medline Neuro Sponges, labeled as follows: 1. RR-OMFS PACK-LF, Medline Kit Number/SKU DYNJ0045920D; 2. RR-OMFS PACK-LF, Medline Kit Number/SKU DYNJ0045920F; 3. MEM MAXILLOFACIAL PACK-LF, Medline Kit Number/SKU DYNJ0660040S; 4. ORAL SURGERY PACK, Medline Kit Number/SKU DYNJ66450B; 5. ORAL SURGERY PACK RFID, Medline Kit Number/SKU DYNJ66450C; 6. ORAL SURGERY PACK, Medline Kit Number/SKU DYNJ66450D; 7. ORAL SURGERY PACK RFID-V2, Medline Kit Number/SKU DYNJ66450F; 8. ORAL SURGERY PACK RFID-V2, Medline Kit Number/SKU DYNJ66450G.

Code information (models, lots, serial numbers)

Show the full code information (6k characters)

Medline Kit Number/SKU DYNJ0045920D: UDI/DI each 10195327267407, UDI/DI case 40195327267408, Lot Number: 23JMB165; Medline Kit Number/SKU DYNJ0045920D: UDI/DI each 10195327267407, UDI/DI case 40195327267408, Lot Number: 23CMH103; Medline Kit Number/SKU DYNJ0045920D: UDI/DI each 10195327267407, UDI/DI case 40195327267408, Lot Number: 23BMH200; Medline Kit Number/SKU DYNJ0045920D: UDI/DI each 10195327267407, UDI/DI case 40195327267408, Lot Number: 23BMB495; Medline Kit Number/SKU DYNJ0045920D: UDI/DI each 10195327267407, UDI/DI case 40195327267408, Lot Number: 22KMH108; Medline Kit Number/SKU DYNJ0045920D: UDI/DI each 10195327267407, UDI/DI case 40195327267408, Lot Number: 22KME285; Medline Kit Number/SKU DYNJ0045920D: UDI/DI each 10195327267407, UDI/DI case 40195327267408, Lot Number: 23KMJ648; Medline Kit Number/SKU DYNJ0045920F: UDI/DI each 10195327643713, UDI/DI case 40195327643714, Lot Number: 26BME013; Medline Kit Number/SKU DYNJ0045920F: UDI/DI each 10195327643713, UDI/DI case 40195327643714, Lot Number: 25JMC725; Medline Kit Number/SKU DYNJ0045920F: UDI/DI each 10195327643713, UDI/DI case 40195327643714, Lot Number: 25IMA571; Medline Kit Number/SKU DYNJ0045920F: UDI/DI each 10195327643713, UDI/DI case 40195327643714, Lot Number: 25HMC497; Medline Kit Number/SKU DYNJ0045920F: UDI/DI each 10195327643713, UDI/DI case 40195327643714, Lot Number: 25FMG199; Medline Kit Number/SKU DYNJ0045920F: UDI/DI each 10195327643713, UDI/DI case 40195327643714, Lot Number: 25EMB523; Medline Kit Number/SKU DYNJ0045920F: UDI/DI each 10195327643713, UDI/DI case 40195327643714, Lot Number: 25DMF918; Medline Kit Number/SKU DYNJ0045920F: UDI/DI each 10195327643713, UDI/DI case 40195327643714, Lot Number: 25CMH298; Medline Kit Number/SKU DYNJ0045920F: UDI/DI each 10195327643713, UDI/DI case 40195327643714, Lot Number: 24LMA598; Medline Kit Number/SKU DYNJ0045920F: UDI/DI each 10195327643713, UDI/DI case 40195327643714, Lot Number: 24JMH963; Medline Kit Number/SKU DYNJ0045920F: UDI/DI each 10195327643713, UDI/DI case 40195327643714, Lot Number: 24IMC899; Medline Kit Number/SKU DYNJ0045920F: UDI/DI each 10195327643713, UDI/DI case 40195327643714, Lot Number: 24IMD398; Medline Kit Number/SKU DYNJ0045920F: UDI/DI each 10195327643713, UDI/DI case 40195327643714, Lot Number: 24EME289; Medline Kit Number/SKU DYNJ0045920F: UDI/DI each 10195327643713, UDI/DI case 40195327643714, Lot Number: 24CMG724; Medline Kit Number/SKU DYNJ0660040S: UDI/DI each 10198459674907, UDI/DI case 40198459674908, Lot Number: 26BMD088; Medline Kit Number/SKU DYNJ66450B: UDI/DI each 10195327191689, UDI/DI case 40195327191680, Lot Number: 23JBW720; Medline Kit Number/SKU DYNJ66450B: UDI/DI each 10195327191689, UDI/DI case 40195327191680, Lot Number: 23EBR729; Medline Kit Number/SKU DYNJ66450B: UDI/DI each 10195327191689, UDI/DI case 40195327191680, Lot Number: 23DBC424; Medline Kit Number/SKU DYNJ66450B: UDI/DI each 10195327191689, UDI/DI case 40195327191680, Lot Number: 23CBD133; Medline Kit Number/SKU DYNJ66450B: UDI/DI each 10195327191689, UDI/DI case 40195327191680, Lot Number: 23BBA368; Medline Kit Number/SKU DYNJ66450B: UDI/DI each 10195327191689, UDI/DI case 40195327191680, Lot Number: 22LBG835; Medline Kit Number/SKU DYNJ66450B: UDI/DI each 10195327191689, UDI/DI case 40195327191680, Lot Number: 22KBV773; Medline Kit Number/SKU DYNJ66450B: UDI/DI each 10195327191689, UDI/DI case 40195327191680, Lot Number: 22IBB948; Medline Kit Number/SKU DYNJ66450B: UDI/DI each 10195327191689, UDI/DI case 40195327191680, Lot Number: 24BBP259; Medline Kit Number/SKU DYNJ66450B: UDI/DI each 10195327191689, UDI/DI case 40195327191680, Lot Number: 24ABS878; Medline Kit Number/SKU DYNJ66450B: UDI/DI each 10195327191689, UDI/DI case 40195327191680, Lot Number: 23LBH114; Medline Kit Number/SKU DYNJ66450C: UDI/DI each 10195327671136, UDI/DI case 40195327671137, Lot Number: 24GBC544; Medline Kit Number/SKU DYNJ66450C: UDI/DI each 10195327671136, UDI/DI case 40195327671137, Lot Number: 24FBL453; Medline Kit Number/SKU DYNJ66450C: UDI/DI each 10195327671136, UDI/DI case 40195327671137, Lot Number: 24EBH347; Medline Kit Number/SKU DYNJ66450D: UDI/DI each 10198459059216, UDI/DI case 40198459059217, Lot Number: 25DBC120; Medline Kit Number/SKU DYNJ66450D: UDI/DI each 10198459059216, UDI/DI case 40198459059217, Lot Number: 25CBF213; Medline Kit Number/SKU DYNJ66450D: UDI/DI each 10198459059216, UDI/DI case 40198459059217, Lot Number: 24KBU249; Medline Kit Number/SKU DYNJ66450D: UDI/DI each 10198459059216, UDI/DI case 40198459059217, Lot Number: 24IBQ393; Medline Kit Number/SKU DYNJ66450D: UDI/DI each 10198459059216, UDI/DI case 40198459059217, Lot Number: 24HBN580; Medline Kit Number/SKU DYNJ66450F: UDI/DI each 10198459334146, UDI/DI case 40198459334147, Lot Number: 25JBA224; Medline Kit Number/SKU DYNJ66450F: UDI/DI each 10198459334146, UDI/DI case 40198459334147, Lot Number: 25HBQ050; Medline Kit Number/SKU DYNJ66450F: UDI/DI each 10198459334146, UDI/DI case 40198459334147, Lot Number: 25GBN753; Medline Kit Number/SKU DYNJ66450F: UDI/DI each 10198459334146, UDI/DI case 40198459334147, Lot Number: 25GBJ601; Medline Kit Number/SKU DYNJ66450F: UDI/DI each 10198459334146, UDI/DI case 40198459334147, Lot Number: 25FBQ773; Medline Kit Number/SKU DYNJ66450F: UDI/DI each 10198459334146, UDI/DI case 40198459334147, Lot Number: 25FBB129; Medline Kit Number/SKU DYNJ66450F: UDI/DI each 10198459334146, UDI/DI case 40198459334147, Lot Number: 25FBE608; Medline Kit Number/SKU DYNJ66450G: UDI/DI each 10198459568817, UDI/DI case 40198459568818, Lot Number: 25LBS231; Medline Kit Number/SKU DYNJ66450G: UDI/DI each 10198459568817, UDI/DI case 40198459568818, Lot Number: 25LBF336; Medline Kit Number/SKU DYNJ66450G: UDI/DI each 10198459568817, UDI/DI case 40198459568818, Lot Number: 25LBC263; Medline Kit Number/SKU DYNJ66450G: UDI/DI each 10198459568817, UDI/DI case 40198459568818, Lot Number: 25JBL075.

Devices named in this recall

6 devices in the FDA device registry whose identifier appears in the recall's code information.

Other products in the same recall event

The FDA groups related products into one recall event (event 98664).